The Argos Cardiac Output monitor is used for the continuous measurement cardiac output from an intravascular radial arterial blood pressure signal is derived from a blood pressure transducer or from the analog output of a vital signs monitor. The device is intended to be used by clinicians on critically ill patients in an operating room or in an intensive care unit.
Device Story
Argos Monitor; portable hemodynamic monitor; calculates cardiac output (CO), cardiac index (CI), stroke volume (SV), stroke volume index (SVI), systemic vascular resistance (SVR), systemic vascular resistance index (SVRI), mean arterial pressure (MAP), heart rate (HR), and pulse pressure variation (PPV). Inputs: blood pressure waveform from radial arterial catheter or analog output of vital signs monitor; user-entered patient demographics (age, height, weight, gender). Operation: proprietary signal processing algorithm analyzes blood pressure waveform using Windkessel model. Output: touchscreen display of hemodynamic parameters. Used in ICU/OR by clinicians. Benefits: continuous hemodynamic monitoring to guide clinical decision-making in critically ill patients. Includes battery backup for transport.
Clinical Evidence
Clinical performance assessed in 40 patients (20 ICU, 20 OR) comparing Argos and predicate device against pulmonary artery catheter reference. Results showed Argos had same or lower measurement errors than predicate. Usability testing conducted with 15 clinicians (physicians/nurses) confirmed ability to set up, configure, and interpret device data/alarms. No adverse effects reported.
Technological Characteristics
Rigid plastic housing; touchscreen display; pole/table mountable. Inputs: radial arterial catheter or analog BP signal. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-34, IEC 60601-1-8, ISTA 2A, ISO 10993-1. Battery backup included. Algorithm: proprietary pulse contour analysis based on Windkessel model.
Indications for Use
Indicated for patients >18 years old in intensive care units or operating rooms requiring hemodynamic monitoring of cardiac output and derived parameters.
Regulatory Classification
Identification
A single-function, preprogrammed diagnostic computer is a hard-wired computer that calculates a specific physiological or blood-flow parameter based on information obtained from one or more electrodes, transducers, or measuring devices.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
December 13, 2018
Retia Medical, LLC Jerry Korten VP R&D 7 Dana Rd. Suite 111 Valhalla, New York 10595
Re: K181372
Trade/Device Name: Argos Regulation Number: 21 CFR 870.1435 Regulation Name: Single-Function, Preprogrammed Diagnostic Computer Regulatory Class: Class II Product Code: DXG Dated: November 8, 2018 Received: November 9, 2018
Dear Jerry Korten:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shawn W.
Forrest -
JS. o=U.S. Government, ou 42.19200300.100.1.1=1300403 awn W. Forrest -A 2018.12.13 15:41:34 -05'00
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181372
Device Name Retia Medical, LLC : Argos Monitor
#### Indications for Use (Describe)
The Argos Cardiac Output monitoring device is intended for use on patients above the age of 18. It is intended to be used as a hemodynamic monitor for monitoring cardiac output and its derived parameters on patients in the intensive care unit or the operating room.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510(K) SUMMARY
## 5.1. Submitter
Retia Medical 7 Dana Rd Valhalla, NY 10595 914 594 1986
# Contact(s): Jerry Korten
# Date Prepared: May 15, 2018
### 5.2. Device
| Name of Device: | Argos Monitor |
|----------------------|----------------------------------|
| Common or Usual Name | Argos |
| Classification Name: | Cardiovascular (21 CFR 870.1435) |
| Regulatory Class: | Class II |
| Product Code: | DXG |
### 5.3. Predicate Device
## Predicate Name and 510(k) Number:
Primary Predicate: Edwards Lifesciences: Flotrac Vigileo Arterial Pressure Cardiac Output (APCO) Monitor (K043065). Secondary Predicate: LiDCOplus monitor (K023960).
# 5.4. Device Description
The Argos Monitor is a portable hemodynamic monitor that calculates Cardiac Output and other derived parameters, including cardiac index (CI), stroke volume (SV), stroke volume index (SVI), systemic vascular resistance (SVR), systemic vascular resistance index (SVRI), mean arterial pressure (MAP), heart rate (HR), and pulse pressure variation (PPV) based on a proprietary algorithm that analyzes the blood pressure waveform and user-entered patient demographic information (age, height weight and gender). The blood pressure waveform is input into the monitor via a connection with either a radial arterial catheter or the analog blood pressure signal output of a vital signs monitor. The scientific method that underlies the algorithm is based on a novel signal processing technique to determine the parameters of the well-established Windkessel model of the circulation in order to calculate cardiac output.
The Argos Monitor comes with a touchscreen monitor and computer system enclosed in a rigid plastic housing and a power cable. Cables to connect the monitor to a radial blood pressure transducer or to the analog blood pressure signal output of a vital signs monitor are also
{4}------------------------------------------------
provided according to the setup and needs of the individual institution. The Monitor may be attached to a pole or a table stand via a standard screw interface pattern.
# 5.5. Indications for Use
The Argos Cardiac Output monitoring device is intended for use on patients above the age of 18. It is intended to be used as a hemodynamic monitor for cardiac output monitoring and its derived parameters on patients in the intensive care unit or the operating room.
# 5.6. Intended Use
The Argos Cardiac Output monitor is used for the continuous measurement cardiac output from an intravascular radial arterial blood pressure signal is derived from a blood pressure transducer or from the analog output of a vital signs monitor. The device is intended to be used by clinicians on critically ill patients in an operating room or in an intensive care unit.
# 5.7. Comparison of Technological Characteristics with the Predicate Device
The design, materials, basic product function, configuration and parameters displayed for the Retia Argos monitor and the predicate devices share several similar characteristics as follows:
- 1. All three systems house a computer-driven display system in a plastic enclosure with an external power supply source.
- 2. All three systems can be pole mounted.
- 3. All three systems can take input from a blood pressure transducer.
- 4. All three systems use a pre-programmed computer that uses pulse contour analysis to analyse a blood pressure signal along with user-entered patient demographic data (age, height, weight and gender) to calculate cardiac output and other derived hemodynamic parameters (CI, SV, SV, SVI, SVR, SVRI).
- 5. The Retia Argos and the reference predicate device from LiDCO both calculate pulse pressure variation (PPV) from the input blood pressure waveform.
- 6. The Retia Argos and the reference predicate device from LiDCO can both input the analogue blood pressure signal output from an external vital signs monitor.
The main differences between the Retia Argos monitor and the predicate device are as follows:
- 1. All three systems systems use proprietary versions of the pulse contour analysis algorithms to analyse the blood pressure signal. Accordingly, the accuracy of the algorithm used in the Argos monitor was verified via clinical testing using the pulmonary artery catheter as a reference for the CO measurement and shown to have as low or lower errors than the predicate Edwards device.
- 2. The Retia Argos does not provide an oximetry catheter nor compute parameters associated with this measurement method.
{5}------------------------------------------------
- 3. The Retia Argos has a battery backup that automatically switches in if the monitor is disconnected from the AC power input, allowing use during transport. The predicate devices do not have a battery.
Verification and validation testing was conducted to compare the mechanical, electrical, software, usability, and clinical accuracy of the Retia Argos monitor to the predicate devices, considering the technological similarities and differences as described. The Argos monitor passed all verification and validation testing and was shown to be safe, effective and substantially equivalent to the predicate device.
# 5.8. Performance Data
The Argos CO Monitor has successfully passed functional and performance testing, including electrical and mechanical testing, environmental testing, shipping tests, software verification and validation, clinical usability testing, and comparison testing with the predicate device on clinical data.
The monitor was assessed and/or tested for and met the following standard requirements:
- 1. IEC 60601-1
- 2. IEC 60601-1-2
- 3. IEC 60601-2-34
- 4. IEC 60601-1-8
- 5. ISTA 2A
- 6. ISO 10993-1
Clinical usability testing for the monitor was performed on 15 clinical users, comprising physicians and nurses who work in the ICU and OR. All users were successfully able to set up the monitor, connect the appropriate cables, enter the patient demographic information, configure the displayed parameters, and interpret the information provided, including alarms.
The clinical performance of the Argos monitor was assessed in comparison to the predicate device on 40 patients, 20 each from the OR and ICU. The blood pressure waveforms from all patients were connected to both the Argos monitor and the predicate device. No adverse effects or complications were noted during the study. The accuracies of the Argos monitor and the predicate device were assessed using the pulmonary artery catheter as the reference for all measurements. The data demonstrated that the Argos monitor has the same or lower errors in measurement compared to the reference than the errors shown by the predicate device compared to the reference.
In conclusion, the results from the nonclinical and clinical tests demonstrate that the Argos monitor is as safe, as effective and performs as well as or better than the predicate device for its intended use.
{6}------------------------------------------------
# 5.9. Conclusions
The Retia Argos monitor has been shown to be safe and effective as the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.