K181358 · Televere Systems, Inc. · MUH · Jun 20, 2018 · Dental
Device Facts
Record ID
K181358
Device Name
Televere Digital Dental Imaging System
Applicant
Televere Systems, Inc.
Product Code
MUH · Dental
Decision Date
Jun 20, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1800
Device Class
Class 2
Indications for Use
The Televere Digital Dental Imaging System is intended for radiographic examination by a dental professional to assist in the diagnosing of diseases of the teeth, jaw and oral structures. Not for mammography. Not for fluorosopy.
Device Story
Televere Digital Dental Imaging System is an extraoral digital x-ray system; consists of an intraoral detector (Hamamatsu CMOS area image sensor) and image management software (TigerView Professional). Detector connects to a PC via USB; captures x-ray images generated by external wall-mounted generators (1-15mA, 50-100kV). Device does not control x-ray generation; acts solely as a receiver. Software allows acquisition, display, editing (resize, contrast, crop), review, storage, printing, and distribution of images within a PACS environment. Used in dental clinics by dental professionals. Output provides high-resolution radiographic images for clinical diagnosis of oral structures. Benefits include digital image management and diagnostic support.
Clinical Evidence
Bench testing and clinical image inspection were conducted. Bench testing included electrical safety and electromagnetic compatibility per IEC standards (IEC 60601-1, IEC 60601-1-2, IEC 62220-1-1:2015, IEC 60601-2-65:2012). Performance testing of the CMOS sensor followed FDA guidance for solid-state digital x-ray panels. Sample clinical images were provided to demonstrate image quality; no formal clinical trial was required to establish substantial equivalence.
Technological Characteristics
Extraoral digital x-ray system. Components: Hamamatsu CMOS area image sensor (S11684-12, S116845-12) with Cesium Iodide (CsI) scintillator. Interface: USB 2.0. Power: 100-240 VAC external supply. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62220-1-1:2015, ANSI 60529-2004, IEC 60601-2-65:2012. Software: TigerView Professional (PACS).
Indications for Use
Indicated for radiographic examination of teeth, jaw, and oral structures by dental professionals to assist in diagnosis. Not for mammography or fluoroscopy.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 20, 2018
Televere Systems, LLC % Robert Bakin Regulatory Consultant Technology and Business Law Advisors, LLC 1244 Capuchino Ave. BURLINGAME CA 94010
Re: K181358
Trade/Device Name: Televere Digital Dental Imaging System Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: MUH Dated: May 20, 2018 Received: May 22, 2018
Dear Robert Bakin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Hole 2. Nils
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
### 510(k) Number (if known) K181358
#### Device Name
Televere Digital Dental Imaging System
#### Indications for Use (Doscribo)
The Televere Digital Dental Imaging System is intended for radiographic examination by a dental professional to assist in the diagnosing of diseases of the teeth, jaw and oral structures. Not for mammography. Not for fluorosopy.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of Information unless It displays a currently valld OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for TELEVERE SYSTEMS. The word "TELEVERE" is in orange, with a stylized font. Below that, in smaller, gray letters, is the word "SYSTEMS". The logo is simple and modern.
COURSE STORE VIEW SHARE
#### 510(k) Summary:
# Televere Systems, LLC 1160 Cavalier Road Arnold, MD 21012 Email: rquaal@tigerview.com Establishment Registration Number: 2954358
CONTACT: Robert E. Bakin Email: rbakin@tblawadvisors.com Technology & Business Law Advisors, LLC 1244 Capuchino Avenue, Burlingame, CA 94010 Phone: 571-215-3507 Prepared May 20, 2018
1. Identification of the Device: Proprietary-Trade Name: Televere Digital Dental Imaging System Classification Name: Extraoral Source X-ray System Product Code: MUH Common/Usual Name: Intraoral Digital X-Ray Sensor Device Class/Regulation Number: Class II per regulation 21 C.F.R. §872.1800
2. Equivalent legally marketed device: I View and Imagen Sensor (K162619). Classification Name: Extraoral Source X-ray System Product Code: MUH Common/Usual Name: Intraoral Digital X-Ray Sensor Device Class/Regulation Number: Class II per regulation 21 C.F.R. §872.1800
3. Associated legally marketed devices: TigerView Professional (PACS Software) K061035. This is the software used with this device.
Classification Name: System, Image Processing, Radiological, Code LLZ Common/Usual Name: Picture Archiving and Communications System Device Class/Regulation Number: Class II per regulation 21 C.F.R. §892.2050
4. Indications for Use: The Televere Digital Dental Imaging System is used for a radiographic examination by a dental professional to assist in the diagnosing of disease of the teeth, jaw and oral structures. Not for mammography. Not for fluoroscopy.
5. Device Description: The Televere Digital Dental Imaging System is an extraoral digital x-ray system comprised of two previously cleared individual components: (1) an intraoral detector component (K162619) which connects to a PC via a USB port and (2) and image management software package (i.e., Tigerview Professional, K061035). The device detector comes in two sizes: 600mm² (Hamamatsu CMOS area image sensor, S11684-12) and 884mm² (Hamamatsu CMOS area image sensor, S116845-12). X-ray generators that can integrate with the Televere Digital Dental Imaging System are wall-mounted x-ray generators (both AC and DC) with a tube current between 1 and 15mA inclusive, and with a tube voltage between 50 and 100kV inclusive, with built in controls to set exposure parameters. Generators allow variable mA/kV to be selected, all will control the exposure time. The device and software cannot act as an x-ray generator controller. All control of xray generation is done by controls built into the generator itself. There is no connection between the device and the x-ray generator. The device is an x-ray receiver and does not control the generator.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for TELEVERE SYSTEMS. The word "TELEVERE" is in orange, and the word "SYSTEMS" is in gray and is located below the word "TELEVERE". The font is sans-serif.
SQUARE STORE SQUARE
The previously cleared Tigerview™ Professional software (K061035) provides basic image adjustment features. An image adjustment system allows the physician to acquire, display, edit (e.g., resize, adjust contrast, crop, etc.), review, store, print, and distribute medical images within a Picture Archiving and Communication System (PACS) environment. Tigerview™ Professional runs on standard PC-compatible computers and is compatible with capture devices which attach to the computer using a Network Adaptor, USB port, PCI slot, parallel port, memory card, S-video port on a video capture card, or SCSI card. NOTE: The Tigerview ™ Professional software incorporated into the Televere Digital Dental Imaging system is unchanged from the Predicate software (i.e., K061035)
6. Safety and Effectiveness, Comparison to predicate device. The results of clinical image inspection, bench, and laboratory test results demonstrate that the new device is as safe and effective as the predicate device. Clinical images highlight equal or better image quality as compared to the predicate.
| | Predicate Device | Proposed Device |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Device Name | I-View and Imagen Sensor K162619 | Televere Digital Dental Imaging System |
| CMOS Sensor | Hamamatsu, S11684/5-12 | SAME |
| Intended Use | The Televere Digital Dental Imaging<br>System is used for a radiographic<br>examination by a dental professional<br>to assist in the diagnosing of disease of<br>the teeth, jaw and oral structures. Not<br>for mammography. Not for<br>fluoroscopy. | SAME |
| Configuration | CMOS sensor and imaging software<br>only. No generator or stand provided. | SAME |
| Digital Panel Pixels | 1000 x 1056 (S11684-12)<br>1300 x 1700 (S11685-12) | SAME |
| Software | Deep-View™ | TIgerview Professional™, K061035 |
| Scintillator | Cesium Iodide (CsI) | SAME |
| Interface | USB 2.0 | SAME |
| Power Source | External Power Supply 100-240 VAC | SAME |
| Standards | IEC 60601-1<br>IEC 60601-1-2<br>IEC 62220-1-1:2015<br>ANSI 60529-2004<br>IEC 60601-2-65:2012 | SAME |
## 7. Substantial Equivalence Chart
Substantial Equivalence Conclusion: The comparison table reveals there are no new technical issues of safety or effectiveness raised by substitution of the imaging software component. Professional evaluation of imaging samples were of excellent quality, high resolution, clinically acceptable and substantially equivalent to the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for TELEVERE SYSTEMS. The word "TELEVERE" is in a large, orange, sans-serif font. Below it, in a smaller, gray, sans-serif font, is the word "SYSTEMS". The logo is simple and modern.
COURSE STORE VIEW SHARE
8. Summary of Bench Testing Conducted: IEC Standards have been employed for: Electrical Safety and Electromagnetic Compatibility. The panel power supply is UL Listed. The CMOS area image sensor manufacturerconductedperformancetestingaccordingtotheFDAguidancedocumentforsolidstate digital x-ray panels. Risk Analysis and System operation verification tests were conducted in accordance with FDAguidancedocuments. Sincenosoftwaremodificationswererequired,confirmationtestingwas performed. Labeling was developed to comply with the FDA solid state panel guidance document.
9. Summary of Clinical Testing: Sample clinical images from the new device were also provided however they were not necessary to demonstrate substantial equivalence.
Conclusion: After analyzing bench, clinical image, and external laboratory testing to applicable 10. standards, it is the conclusion of Televere that the Televere Digital Dental Imaging System is as safe and effective as the predicate device K162619, has few technological differences, and has no new indication(s) for use, thus rendering it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.