The Lunderquist Fixed Core Wire Guides are intended to facilitate placement of devices used during diagnostic and interventional procedures.
Device Story
Lunderquist Fixed Core Wire Guide is a sterile, single-use catheter guide wire. Device consists of a mandril design with a distal spring coil and a proximal cannula. Used by physicians in clinical settings to facilitate the placement of diagnostic and interventional devices. The wire is 0.038 inch in diameter and 120 cm in length with a straight tip. It functions as a mechanical aid for navigation within the vasculature or other anatomical pathways during procedures. No electronic or software components are involved.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-1), corrosion (ISO 11070), flexing, surface examination, fracture, tensile, tip flexibility (FDA 1995 guidance), radiopacity (ASTM F640-12), and dimensional/compatibility testing. All tests met predetermined acceptance criteria.
Technological Characteristics
Class II catheter guide wire. Features mandril design with distal spring coil and proximal cannula. Dimensions: 0.038 inch diameter, 120 cm length. Straight tip configuration. Materials tested per ISO 10993-1. Mechanical performance tested per ISO 11070 and FDA 1995 Coronary and Cerebrovascular Guidewire Guidance. Radiopacity tested per ASTM F640-12. Sterile, single-use device.
Indications for Use
Indicated for patients undergoing diagnostic and interventional procedures requiring device placement.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 15, 2019
Cook Incorporated Ms. Jennifer Allman, RAC Regulatory Affairs Project Manager 750 Daniels Way Bloomington, IN 47402
Re: K181353
Trade/Device Name: Lunderquist® Fixed Core Wire Guide Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: January 15, 2019 Received: January 15, 2019
Dear Ms. Allman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Lydia S. Glaw -S Lydia S. Glaw -S. Date: 2019.02.15 15:30:39 -05'00'
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181353
Device Name Lunderquist® Fixed Core Wire Guide
Indications for Use (Describe)
The Lunderquist Fixed Core Wire Guides are intended to facilitate placement of devices used during diagnostic and interventional procedures.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image features the logo for Cook Medical. The logo consists of the word "COOK" in a bold, sans-serif font, stacked above the word "MEDICAL" which is in a smaller, sans-serif font. The word "COOK" also has the registered trademark symbol to the right of it. The background is a solid red color, while the text is white.
K181353
#### 510(k) SUMMARY
## Lunderquist® Fixed Core Wire Guide 21 CFR §807.92 Date Prepared: May 21, 2018
### Submitted By:
| Applicant: | Cook Incorporated |
|-----------------------|---------------------------------------------------------------|
| Contact: | Jennifer L. Allman |
| Applicant Address: | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 |
| Contact Phone Number: | (812) 335-3575 x104280 |
| Contact Fax Number: | (812) 332-0281 |
#### Device Information:
| Trade Name: | Lunderquist® Fixed Core Wire Guide |
|----------------------|------------------------------------|
| Common Name: | Wire, Guide, Catheter |
| Classification Name: | Catheter guide wire |
| Regulation: | 21 CFR §870.1330 |
| Product Code: | DQX |
#### Predicate Device:
The predicate device is the Mandril Wire Guides (Cook, Inc., K171997) cleared for market on March 30, 2018.
#### Reference Device:
The reference device is the Fixed Core Wire Guides (Cook Inc., K171764) cleared for market on March 9, 2018.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in large, white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font, also on a red background. The red background for "MEDICAL" is a rectangle that is slightly offset from the red background for "COOK".
## Device Description:
The Lunderquist® Fixed Core Wire Guide, subject of this submission, is a Class II device according to 21 CFR §870.1330; product code DQX (Wire, Guide, Catheter). The subject device consists of a mandril design with a distal spring coil anchored over the distal portion of the mandril and a cannula anchored over the proximal portion of the mandril. The Lunderquist Fixed Core Wire Guide is available with an outside diameter of 0.038 inch and a length of 120 centimeters (cm). The tip portion of the subject device is straight. The Lunderquist® Fixed Core Wire Guide is a packaged, sterile device intended for single patient use.
## Intended Use:
The Lunderquist Fixed Core Wire Guides are intended to facilitate placement of devices used during diagnostic and interventional procedures.
## Comparison of Technological Characteristics to Predicate:
The Lunderquist® Fixed Core Wire Guide, subject of this submission, is substantially equivalent to the predicate device, the Mandril Wire Guide (K171997), in that both devices have identical indications for use. The subject device is similar in technological characteristics (i.e., design and materials) to the predicate device with the addition of a 0.038 inch diameter configuration and a proximal cannula design.
### Performance Data:
The following tests were performed to demonstrate that the Lunderquist® Fixed Core Wire Guide met applicable design and performance requirements and support a determination of substantial equivalence.
- . Biocompatibility Testing - Tested in accordance with ISO 10993-1:2009. The predetermined acceptance criteria were met.
- . Corrosion Testing – Tested in accordance with Annex B of ISO 11070:2014. The predetermined acceptance criteria were met.
- . Flexing Test – Tested in accordance with the Annex G of BS EN ISO 11070:2014. The pre-determined acceptance criteria were met.
- . Surface Examination – Tested in accordance with BS EN ISO 11070:2014 Section 4.3. The pre-determined acceptance criteria were met.
- . Fracture Testing – Tested in accordance with Annex F of BS EN ISO 11070:2014. The pre-determined acceptance criteria were met.
- . Tensile Testing – Tested in accordance with the applicable values of BS EN ISO 11070:2014. Annex H. The pre-determined acceptance criteria were met.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, with a registered trademark symbol to the right of the word. Below the word "COOK" is the word "MEDICAL" in white, set against a red background. The background is a red rectangle.
- . Tip Flexibility Testing (Tip Deflection) – Tested in accordance with the FDA Coronary and Cerebrovascular Guidance (1995). The pre-determined acceptance criteria were met.
- . Radiopacity Testing - Tested in accordance with ASTM F640-12. The pre-determined acceptance criteria were met.
- Dimensional Measurement and Catheter Compatibility Testing Tested in accordance . with the FDA Coronary and Cerebrovascular Guidewire Guidance (1995) and with an approved study protocol. The pre-determined acceptance criteria were met.
### Conclusion:
The results of these tests support a conclusion that the Lunderquist® Fixed Core Wire Guide, subject of this submission, met the design input requirements based on the device's intended use and support the conclusion this device does not raise new questions of safety and/or effectiveness. The results of these tests support a determination of substantial equivalence to the predicate device, the Mandril Wire Guides (K171997).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.