← Product Code [DQA](/productcode/DQA) · K181352

# Loop System (K181352)

_Spry Health, Inc. · DQA · Mar 29, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K181352

## Device Facts

- **Applicant:** Spry Health, Inc.
- **Product Code:** [DQA](/productcode/DQA.md)
- **Decision Date:** Mar 29, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** AI/ML

## Indications for Use

The Loop System is intended for adult patients in the home environment for passive, non-invasive, intermittent data collection of physiological parameters that will later be transmitted to a web server for remote review by a clinician. The Loop System measures and records: - arterial oxygen saturation (SpO2) - heart rate (HR) - respiration rate (RR) All of these measurements are made when no motion is detected by the System. The Loop System device does not provide physiological alarms

## Device Story

Loop System collects physiological data via wrist-worn Loop Band; uses reflectance-based photoplethysmogram (PPG) signals from LEDs and photodiodes; incorporates accelerometer to detect resting states; records data only during no-motion periods. Band stores raw data; uploads to Spry Server via charging station. Server processes data to determine respiration rate; displays HR and SpO2 on band. Clinicians access time-stamped CSV files via web interface for remote review. Device intended for adult patients in home environment; prescription-only. Benefits include passive, intermittent monitoring of vital signs without continuous active patient engagement.

## Clinical Evidence

Two clinical studies conducted on 12 healthy adult subjects. Study 1: Respiration rate accuracy compared to gold standard. Study 2: Pulse oximetry accuracy compared to arterial blood gas analysis. Both studies performed in accordance with ISO 80601-2-61 and FDA 2013 Pulse Oximeter guidance. Results demonstrated equivalence to gold standard procedures.

## Technological Characteristics

Reflectance-based PPG sensing using LEDs and photodiodes. Wrist-worn form factor. Internal rechargeable battery. Cellular connectivity via charging station. Compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-1-6, ISO 80601-2-61, and ISO 10993-1. Type BF applied part.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- MightySat RX Fingertip Pulse Oximeter ([K181956](/device/K181956.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

March 29, 2019

Spry Health, Inc. % Craig Coombs Consultant Coombs Medical Device Consulting, Inc. 1193 Sherman St. Alameda, CA 94501

Re: K181352

Trade/Device Name: Loop System Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA, BZQ Dated: March 4, 2019 Received: March 5, 2019

Dear Craig Coombs:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Shawn W. Forrest -S 2019.03.29 10:28:03 -04'00'

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K181352

Device Name Loop System

Indications for Use (Describe)

The Loop System is intended for adult patients in the home environment for passive, intermittent data collection of physiological parameters that will later be transmitted to a web server for remote review by a clinician.

The Loop System measures and records:

- · arterial oxygen saturation (SpO2)
- · heart rate (HR)
- · respiration rate (RR)

All of these measurements are made when no motion is detected by the System.

The Loop System device does not provide physiological alarms

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Spry. The logo consists of four concentric circles, with the outermost circle being blue and the inner circles being shades of green. Below the circles, the word "spry" is written in lowercase letters.

# Section 5: 510(k) Summary

#### A. Device Information

| Category                                 | Comments                                                                                                          |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Sponsor:                                 | Spry Health, Inc<br>235 Alma St<br>Palo Alto, CA 94301<br>Tel: (650) 352-3429                                     |
| Primary Communicant                      | Craig Coombs<br>Coombs Medical Device Consulting<br>1193 Sherman Street<br>Alameda, CA 94501<br>Tel: 510-337-0140 |
| Device Common Name:                      | Oximeter<br>Breathing Frequency Monitor                                                                           |
| Device Classification Number:            | 21 CFR 870.2700<br>21 CFR 868.2375                                                                                |
| Device Classification &<br>Product Code: | Class 2,<br>DQA & BZQ                                                                                             |
| Device Proprietary Name:                 | Loop System                                                                                                       |

#### Predicate Device Information:

| Predicate Device:                          | MightySat RX Fingertip Pulse Oximeter                                        |
|--------------------------------------------|------------------------------------------------------------------------------|
| Predicate Device Manufacturer:             | Masimo                                                                       |
| Predicate Device Common Name:              | Oximeter<br>Breathing Frequency Monitor                                      |
| Predicate Device Premarket Notification #: | K181956                                                                      |
| Predicate Device Classification:           | 21 CFR 870.2700: Oximeter<br>21 CFR 868.2375: Breathing Frequency<br>Monitor |
| Predicate Device Class & Product Code:     | Class 2, DQA & BZQ                                                           |

#### B. Date Summary Prepared

13 March 2019

#### C. Description of Device

The Loop System is a prescription-only medical device indicated for use by adult patients in the home environment for passive, non-invasive, intermittent data collection of resting (i.e., no motion) physiological parameters. The data is derived from reflectance-based photoplethysmogram (PPG) signals from the patient's wrist, collected by Light Emitting Diodes

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Spry. The logo consists of three concentric circles, with the outermost circle being blue, the middle circle being a lighter shade of green, and the innermost circle being a darker shade of green. Below the circles, the word "spry" is written in lowercase letters.

(LEDs) of varying wavelengths and photodiodes sensitive to said wavelengths embedded in the Loop Band. An accelerometer incorporated in the Loop Band determines when the patient is at rest by constantly monitoring the activity level of the patient. The data is recorded during those resting periods. Using filtering technology for removal of noise (including ambient light) and data processing algorithms, the Loop System stores the raw data collected until it is able to send to the Spry Server.

Patients wear the Loop Band on their wrist for periods of up to 24 hours. Patients will then have to charge the Loop Band with a Loop Charging Station provided to them as part of the Loop System. During charging, the data is uploaded to the Spry Server.

The Loop band can independently analyze and display resting heart rate and arterial oxygen saturation (SpO2). The data must be downloaded and analyzed by the Spry Server to determine respiration rate.

All physiological measurements collected by the Loop System are reported on a comma separated value (CSV) file as a time-stamped series. The CSV can then be remotely accessed through a web interface for review and analysis by a clinician.

## D. Indications for Use

The Loop System is intended for adult patients in the home environment for passive, noninvasive, intermittent data collection of physiological parameters that will later be transmitted to a web server for remote review by a clinician.

The Loop System measures and records:

- arterial oxygen saturation (SpO2)
- · heart rate (HR)
- · respiration rate (RR)

All of these measurements are made when no motion is detected by the System.

The Loop System device does not provide physiological alarms

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows a logo for a company called "spry". The logo consists of three concentric circles, with the outermost circle being blue, the middle circle being a lighter shade of blue, and the innermost circle being green. The word "spry" is written in a sans-serif font below the circles.

## E. Comparison to Predicate Device

The application Spry Health Loop System is substantially equivalent to the Masimo MightySat Rx Fingertip Pulse Oximeter (K181956). The devices have a similar Indications for Use, features, technology and accuracy.

|                                                                                         | Predicate Device<br>Masimo MightySat Rx<br>Fingertip Pulse<br>Oximeter<br>K181956                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Application Device<br>Spry Health, Inc.<br>Loop System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Impact on<br>Substantial Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|-----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use Issues                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Indications for Use                                                                     | The Masimo MightySat Rx<br>Fingertip Pulse Oximeter is<br>intended for hospitals,<br>hospital-type facilities, home<br>environments, and transport.<br>The Masimo MightySat Rx<br>Fingertip Pulse Oximeter is<br>indicated for the noninvasive<br>spot checking of functional<br>oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse<br>rate (PR) for adult and<br>pediatric patients during<br>both no motion and motion<br>conditions, and for patients<br>who are well or poorly<br>perfused.<br>The Masimo MightySat Rx<br>Fingertip Pulse Oximeter is<br>indicated for the noninvasive<br>spot checking of respiration<br>rate (RRp) for adult patients. | The Loop System is intended for<br>adult patients in the home<br>environment for passive, non-<br>invasive, intermittent data<br>collection of physiological<br>parameters that will later be<br>transmitted to a web server for<br>remote review by a clinician.<br>The Loop System measures and<br>records:<br>• arterial oxygen saturation<br>(SpO2)<br>• heart rate (HR)<br>• respiration rate (RR)<br>All of these measurements are<br>made when no motion is<br>detected by the System.<br>The Loop System device does not<br>provide physiological alarms. | The application Loop System<br>Indications for Use includes a<br>subset of the patients and use<br>conditions cleared in the<br>predicate device. The Loop<br>System and the predicate are<br>intended for adults in a home<br>environment when the wearers<br>are not moving.<br>Since the application device is<br>intended for a subset of the<br>application conditions of the<br>predicated device, no new<br>types of safety or efficacy<br>questions are raised.<br>Both devices provide SpO2, HR,<br>and RR.<br>Neither device is intended for<br>continuous vital sign<br>monitoring. |
| Product Code<br>Regulation Description                                                  | DQA - Oximeter<br>BZQ -Breathing Frequency<br>Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | DQA - Oximeter<br>BZQ -Breathing Frequency<br>Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Arterial Oxygen saturation and<br>derived HR and RR capabilities<br>all fall under the same product<br>codes for both devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Patient Population                                                                      | HR & SpO2 - Adults,<br>pediatrics<br>Respiratory Rate - Adults<br>Non-critical care                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Adult patients for all parameters<br>Non-critical care                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | The application device<br>intended cohort is a subset of<br>the predicate's; no new types<br>of safety or efficacy questions<br>are raised.                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                                         | Predicate Device<br>Masimo MightySat Rx<br>Fingertip Pulse<br>Oximeter<br>K181956                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Application Device<br>Spry Health, Inc.<br>Loop System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Impact on<br>Substantial Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Use Environment                                                                         | hospitals, hospital-type<br>facilities, home<br>environments, and transport.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | home environment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | The application device use<br>environment is a subset of the<br>predicate's; no new types of<br>safety or efficacy questions are<br>raised.                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Prescribed?                                                                             | Prescription use only                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Prescription use only                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Parameter Sampling<br>Frequency                                                         | Spot check parameters as<br>desired by users. Not<br>designed for continuous<br>wearing.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Designed for nearly continuous<br>measuring and recording of<br>physiological data when the<br>patient is not in motion                                                                                                                                                                                                                                                                                                                                                                                                                                           | Difference in sampling<br>frequency does not raise new<br>safety or efficacy questions                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Technology associated with each parameter                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Respiration Rate<br>Fundamental Scientific<br>Technology                                | Respiration rate (RR)<br>measured by analyzing cyclic<br>variations in the<br>photoplethysmogram. The<br>diodes are mounted in the<br>device such that they are in<br>contact with the skin.                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Technologically equivalent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Respiration Rate Range                                                                  | 4 - 70 respirations per<br>minute (RPM)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 4 - 40 RPM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Application device has a lower<br>upper limit than predicate.<br>Specific RPMs that are >40 are<br>not clinically differentiating in a<br>retrospective monitoring<br>context                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Respiration Rate<br>Accuracy                                                            | 3 RPM ARMS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 3 RPM ARMS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | A clinical study demonstrated<br>that the Loop System could<br>accurately monitor respiration<br>rate as well as the predicate,<br>within the sensor range                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Arterial hemoglobin<br>oxygen saturation (SpO2)<br>Fundamental Scientific<br>Technology | SpO2 measured by analyzing<br>reflectance of certain LED<br>frequencies in a<br>photoplethysmogram design.<br>The diodes are mounted in<br>the device such that they are<br>in contact with the skin.                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Technologically equivalent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                                         | Predicate Device<br>Masimo MightySat Rx<br>Fingertip Pulse<br>Oximeter<br>K181956                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Application Device<br>Spry Health, Inc.<br>Loop System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Impact on<br>Substantial Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| SpO2 Range                                                                              | 70 – 100%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 70 – 100%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Sp02 Accuracy                                                                           | 2% ARMS, no motion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 3% ARMS, no motion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Both comply with ISO 80601-2-<br>61 as well as with FDA<br>Guidance for Pulse Oximeters<br>(2013)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Heart Rate (HR)<br>Fundamental Scientific<br>Technology                                 | HR measured by analyzing<br>cyclic variations in<br>reflectance of certain LED<br>frequencies in a<br>photoplethysmogram design.<br>The diodes are mounted in<br>the device such that they are<br>in contact with the skin.                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Technologically equivalent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Heart Rate Range                                                                        | 25 - 240 beats per minute<br>(BPM)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 25 - 250 beats per minute (BPM)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Clinically equivalent. Both<br>comply with ISO 80601-2-61                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Heart Rate Accuracy                                                                     | 3 BPM ARMS, no motion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 3 BPM ARMS, no motion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Clinically equivalent. Both<br>comply with ISO 80601-2-61                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Standards Compliance                                                                    | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 80601-2-61<br>ISO 10993-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-1-6<br>ISO 80601-2-61<br>ISO 10993-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Both devices meet latest<br>guidelines for safety and<br>efficacy                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Design                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| User Interface                                                                          | Finger clamp                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Wrist band                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Both devices are worn to<br>provide device access to<br>peripheral arteries. Difference<br>in particular user interface does<br>not raise new questions of<br>safety or efficacy.                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                                         | Predicate Device<br>Masimo MightySat Rx<br>Fingertip Pulse<br>Oximeter<br>K181956                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Application Device<br>Spry Health, Inc.<br>Loop System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Impact on<br>Substantial Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Wearing location and<br>frequency                                                       | For intermittent wearing, aka<br>"spot checking"<br>Image: Person using fingertip pulse oximeter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Continuous wearing, nearly<br>continuous recording of all data.<br>Image: Spry Health Loop System on wrist                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Difference in wearing location<br>on the body and frequency<br>does not raise new questions of<br>safety or efficacy. Both<br>locations allow for monitoring<br>peripheral arteries.                                                                                                                                                                                                                                                                                                                                                                                                           |
| Display                                                                                 | OLED color display of HR and<br>SpO2 when device worn<br>Image: Masimo MightySat Rx Fingertip Pulse Oximeter display                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | LED color display of HR and SpO2<br>on demand<br>Image: Spry Health Loop System display                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Similar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Power source                                                                            | 2 AA batteries                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Internal rechargeable batteries                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Both battery-powered                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Data Communication                                                                      | Wireless (Bluetooth pairing)<br>with mobile devices such as<br>smartphones                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Wireless (cellular connection) via<br>charging station to Spry Server.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Both devices are designed to<br>transmit their data to alternate<br>devices or sites.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Type of protection                                                                      | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Type BF - Applied Part per IEC<br>60601-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Application device meets latest<br>guidelines for electrical safety                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Patient contracting<br>materials                                                        | Plastic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Plastic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Similar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Biocompatibility                                                                        | Compliant to ISO 10993-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Compliant to ISO 10993-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Spry. The logo consists of a circular design with multiple concentric rings in shades of blue and green. Below the circular design, the word "spry" is written in lowercase letters.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Spry. The logo consists of a circular design with three concentric rings in shades of blue and green. Below the circular design, the word "spry" is written in lowercase letters.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Spry. The logo consists of three concentric circles, with the outermost circle being blue and the innermost circle being green. Below the circles, the word "spry" is written in lowercase letters.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for Spry. The logo consists of three concentric circles, with the outermost circle being blue, the middle circle being a lighter shade of blue, and the innermost circle being green. Below the circles, the word "spry" is written in lowercase letters.

# F. Summary of Supporting Data

Spry Health has conducted extensive testing to ensure that the Loop System met design specifications, functions as intended, and conforms to internationally recognized standards and FDA Guidelines.

# Bench Testing

All test results demonstrate the performance of Spry Health Loop System met the requirements of its pre-defined acceptance criteria and intended use. The results of the bench (non-clinical) testing demonstrate that the Loop System is as safe and effective as the predicate device.

Bench testing demonstrated the accuracy of the heart rate monitoring with the Loop System. This testing was conducted in accordance with ISO 80601-2-61 Medical electrical equipment -Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment and in accordance with the FDA Guidelines for Pulse Oximeters - Premarket Notification Submissions [510(k)s]: Guidance for Industry and Food and Drug Administration Staff (2013). The Loop System was found to be in compliance with both documents.

Additionally, transit testing and one-year shelf life testing was conducted.

## Electrical Safety and Electromagnetic Compatibility Testing

Electrical safety testing was conducted in accordance with: IEC 60601-1:2005+AM1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

IEC 60601-1-2:2014: Medical Electrical Equipment – Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Disturbances - Requirements and tests

IEC 60601-1-11:2015 Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

IEC 60601-1-6:2013 Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability

The Spry Health Loop System passed all electrical safety and EMC testing.

## Biocompatibility Testing

Biocompatibility testing was conducted in accordance with ISO 10993-1:2009/AC:2010, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process, and FDA's guidance documents, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a rick management process" issued June 16, 2016. This testing demonstrates that the materials in the Loop Band will not cause an adverse biocompatibility reaction when used as intended.

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the logo for Spry. The logo consists of three concentric circles, with the outermost circle being blue, the middle circle being a lighter shade of blue, and the innermost circle being green. Below the circles, the word "spry" is written in lowercase letters.

## Clinical Studies

Two clinical studies were conducted to demonstrate the performance of the Loop System.

One clinical study investigated the accuracy of respiration rate monitoring in 12 healthy adult subjects with a range of skin types and respiratory rates. The Loop System accuracy was shown to be equivalent to a gold standard monitoring procedure.

Another clinical study investigated the accuracy of the pulse oximetry monitoring in 12 healthy adult subjects with a range of skin types. The Loop System accuracy was compared to the gold standard, arterial blood gas analysis.

This testing was conducted in accordance with ISO 80601-2-61 Medical electrical equipment -Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment and in accordance with the FDA Guidelines for Pulse Oximeters - Premarket Notification Submissions [510(k)s]: Guidance for Industry and Food and Drug Administration Staff (2013). The Loop System was found to be in compliance with both documents.

#### G. Conclusion

Spry Health concludes that the application Spry Health Loop System is substantially equivalent to the Masimo MightySat Rx Fingertip Pulse Oximeter (K181956). The devices have a similar Indications for Use, features, technology and accuracy in monitoring respiration rate, heart rate and pulse oximetry saturation.

---

**Source:** [https://fda-staging.innolitics.com/device/K181352](https://fda-staging.innolitics.com/device/K181352)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
