K181348 · Cartiva, Inc. · QBO · Jun 20, 2018 · Orthopedic
Device Facts
Record ID
K181348
Device Name
Cartiva SCI Disposable Instrumentation Set
Applicant
Cartiva, Inc.
Product Code
QBO · Orthopedic
Decision Date
Jun 20, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.4505
Device Class
Class 2
Indications for Use
Orthopedic surgical instrumentation supplied by Cartiva, Inc. is indicated for use in the press-fit fixation of the implantable medical device products manufactured by Cartiva, Inc.
Device Story
Cartiva SCI Disposable Instrumentation Set consists of hand-held orthopedic surgical instruments; used to manipulate bone and cartilage tissue or implants; facilitates positioning, alignment, defect creation, and placement of press-fit osteochondral implants; utilizes no additional fixation means like sutures or adhesives; intended for single-use; used by surgeons in clinical settings; provides same performance as reusable predicate; benefits include sterility assurance and convenience of disposable format.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility testing per ISO 10993 (cytotoxicity, sensitization, irritation, toxicity, pyrogenicity). Sterility validation per ISO 11137 and ISO 11737 (SAL 10^-6). Shelf-life established via accelerated aging per ASTM F1980. Simulated use testing and ship studies per ASTM D4169 performed.
Technological Characteristics
Hand-held orthopedic surgical instruments; single-use; terminally sterilized via Gamma radiation (ISO 11137/11737); biocompatible materials per ISO 10993; shelf-life 5 years (ASTM F1980); shipping validated per ASTM D4169.
Indications for Use
Indicated for use in the press-fit fixation of Cartiva, Inc. implantable medical devices during orthopedic surgery.
Regulatory Classification
Identification
Orthopedic surgical instruments designed for osteochondral implants with press-fit fixation are hand-held devices intended to manipulate bone and cartilage tissue or the implant for the positioning, alignment, defect creation, and placement of press-fit osteochondral implants that utilize no additional means of fixation (e.g., suture fixation, adhesives). This type of device includes instruments specific to the geometry of the implant.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 888.9. The special controls for this device are:(1) Technical specifications regarding geometry of the instruments must be specified and validated to demonstrate that the instruments can safely position and place the implant.
(2) The patient contacting components of the device must be demonstrated to be biocompatible.
(3) Labeling must include:
(i) Identification of implant(s) and instruments which have been validated for use together; and
(ii) Validated methods and instructions for reprocessing any reusable parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 20, 2018
Cartiva, Inc. Tanya Eberle Senior Director, Regulatory Affairs and Quality Assurance 6120 Windward Parkway Suite 220 Alpharetta, Georgia 30005
Re: K181348
Trade/Device Name: Cartiva® SCI Disposable Instrumentation Set Regulation Number: 21 CFR 888.4505 Regulation Name: Orthopedic surgical instrumentation designed for osteochondral implants with pressfit fixation Regulatory Class: Class II Product Code: QBO Dated: May 18, 2018 Received: May 21, 2018
Dear Ms. Eberle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use Statement
| 510(k) No. (if known): | K181348 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Cartiva® SCI Disposable Instrumentation Set |
| Indications for Use: | <p>Orthopedic surgical instrumentation supplied by Cartiva, Inc. is indicated for use in the press-fit fixation of the implantable medical device products manufactured by Cartiva, Inc.</p> |
| Prescription Use:<br>(Part 21 CFR 801 Subpart D) | AND/OR | Over-the-Counter Use<br>(21 CFR 801 Subpart C) |
|--------------------------------------------------|--------|------------------------------------------------|
| X | | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page __ of ___
{3}------------------------------------------------
# CARTIVA
### 510(k) Summary
#### SUBMITTER'S INFORMATION
| Owner: | Cartiva, Inc. |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 6120 Windward Parkway, Suite 220, Alpharetta, GA 30005 |
| Phone: | 770-754-3800 |
| Fax Numbers: | 770-754-3808 |
| Contact Person: | Tanya Eberle, Sr. Director, Regulatory Affairs |
| Date Summary Prepared: | May 18, 2018 |
| DEVICE INFORMATION | |
| Name of Device: | Cartiva SCI Disposable Instrumentation Set |
| Common/Usual Name: | Orthopedic Surgical Instrumentation |
| Classification Name: | Orthopedic Surgical Instrumentation designed for osteochondral<br>implants with press-fit fixation (Product Code QBO) |
| Predicate Device(s): | Cartiva SCI Instrumentation (Product Code QBO, Regulation Number<br>21 CFR 888.4505) |
| Device Description: | Orthopedic surgical instruments designed for osteochondral implants<br>with press-fit fixation are hand-held devices intended to manipulate<br>bone and cartilage tissue or the implant for the positioning, alignment,<br>defect creation, and placement of press-fit osteochondral implants that<br>utilize no additional means of fixation (e.g., suture fixation, adhesives).<br>This type of device includes instruments specific to the geometry of the<br>implant. |
| Indication for Use: | Orthopedic surgical instrumentation supplied by Cartiva, Inc. is indicated<br>for use in the press-fit fixation of the implantable medical device<br>products manufactured by Cartiva, Inc. |
| Technological<br>Characteristics: | The minor changes to the material of the Cartiva SCI Disposable<br>Instrumentation do not do not raise new questions of safety or<br>effectiveness. All technological aspects of press-fit fixation are<br>preserved. |
| Comparison to Predicate<br>Device: | The Cartiva SCI Disposable Instrumentation Set intended use and<br>performance characteristics are not altered by this modification to allow<br>for a single-use option. |
6120 Windward Parkway, Suite 220 Alpharetta, GA 30005
Tel | 877.336.4616 Fax | 770.754.3808 www.cartiva.net Page 1 of 2
{4}------------------------------------------------
#### Performance Data:
Testing of the Cartiva SCI Disposable Instrumentation Set was carried out. including:
#### Biocompatibility per ISO 10993-1
- Cytotoxicity (ISO 10993-5) ●
- Sensitization (ISO 10993-10)
- Irritation (ISO 10993-10) .
- . Toxicity (ISO 10993-11)
- Pyrogenicity (ISO 10993-11) ●
#### Sterility and Shelf Life
- The Cartiva SCI Disposable Instrumentation Sets are terminally ● sterilized via Gamma radiation.
- . Sterilization validation methodology per ISO 11137 and ISO 11737, Method VDmax25.
- Sterility Assurance Level is 10-6. ●
- Non-pyrogenic claims for the Cartiva SCI Disposable ● Instrumentation Sets were verified through ISO Materials Mediated Rabbit Pyrogen testing.
- . Shelf-Life testing was established as 5 years through accelerated aging per ASTM F 1980. Realtime testing is ongoing to assure stability.
#### Performance/Functional Testing
- Simulated Use Testing of aged product in beyond use ● conditions to demonstrate functionality.
- Ship studies per ASTM D4169. .
All data demonstrated that the safety and performance of the Cartiva SCI Instrumentation is not affected by the modification to allow for disposable, single-use instrumentation.
A Clinical Evaluation was determined not to be required as the device design, intended use and indication for use are all preserved.
This modification falls within the FDA regulations for 510(k) review. The indication for use, intended use, principles of operation, and performance have not been altered. The minor change in technological characteristics (material) of the next-generation disposable instrumentation do not raise any new questions of safety or effectiveness and the single use set has been demonstrated to provide the same level of performance as the predicate reusable device. The Cartiva SCI Disposable Instrumentation Set with next-generation instrumentation is substantially equivalent to the predicate device (Cartiva SCI Instrumentation, Product Code QBO, Regulation Number 21 CFR 888.4505).
The Cartiva SCI next-qeneration disposable instrumentation, as modified by this 510(k), do not raise any new issues regarding safety or effectiveness, and therefore is suitable for commercial sale.
> 6120 Windward Parkway, Suite 220 Alpharetta, GA 30005
Tel | 877.336.4616 Fax | 770.754.3808 www.cartiva.net Page 2 of 2
Rationale for Substantial Equivalence:
Conclusion:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.