K181308 · Super Grow Lasers · OAP · Aug 15, 2018 · Physical Medicine
Device Facts
Record ID
K181308
Device Name
SuperGrow 272, Ultimate 272
Applicant
Super Grow Lasers
Product Code
OAP · Physical Medicine
Decision Date
Aug 15, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SuperGrow 272 laser hat and the Ultimate 272 are intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood – Hamilton Classifications of IIa to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-2, or frontal; both with Fitzpatrick Skin Types I to IV.
Device Story
Non-invasive low-level laser therapy (LLLT) devices; 272 red visible light diode lasers (650 nm, 5 mW max output each); photostimulation of scalp hair follicles; non-thermal energy delivery. SuperGrow 272 is a wearable hat; Ultimate 272 is a stand-mounted version. Used in home setting by patients (over-the-counter); no physician operation required. Output is light energy; intended to promote hair growth. Benefits patient by treating Androgenetic Alopecia.
Clinical Evidence
Usability study conducted on 40 subjects representing the intended user population. Evaluated ability to self-select based on indications/contraindications and comprehension of instructions/warnings. Success rates: 80% for self-selection, 82.5% for instruction comprehension, and 80% for proper device use. Met the 80% success threshold required for OTC distribution.
Technological Characteristics
272 red diode lasers (650 nm, 5 mW each); polyethylene terephthalate (PET) shell/liner; non-thermal energy delivery. Electrical safety/EMC per IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11. Biocompatibility per ISO 10993-1. No software algorithms.
Indications for Use
Indicated for males with Norwood-Hamilton IIa-V and females with Ludwig (Savin) I-1 to I-4, II-1, II-2, or frontal hair loss patterns; Fitzpatrick Skin Types I-IV; for treatment of Androgenetic Alopecia and promotion of hair growth.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 15, 2018
Super Grow Lasers % Ms. Allison Komiyama Principal Consultant AcKnowledge Regulatory Strategies, LLC 2834 Hawthorn Street San Diego, California 92104
Re: K181308
Trade/Device Name: SuperGrow 272, Ultimate 272 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: May 15, 2018 Received: May 17, 2018
Dear Ms. Komiyama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# David Krause -S
- for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181308
Device Name SuperGrow 272, Ultimate 272
#### Indications for Use (Describe)
The SuperGrow 272 laser hat and the Ultimate 272 are intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood – Hamilton Classifications of IIa to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-2, or frontal; both with Fitzpatrick Skin Types I to IV.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Supergrow Lasers. The logo consists of a red icon on the left and the text "SUPERGROW LASERS" on the right. The icon is made up of three red shapes. The text is in a bold, sans-serif font.
# 510(k) Summary K181308
#### DATE PREPARED
August 10, 2018
#### MANUFACTURER AND 510(k) OWNER
Super Grow Lasers 1903 Commonwealth #16, Houston, TX 77006, U.S.A. Telephone: Official Contact: Jeffrey Stanifer, CEO
#### REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, Ph.D., R.A.C. Lucie Dalet, Ph.D. AcKnowledge Regulatory Strategies, LLC Telephone: +1 (619) 208-7888 Email: akomiyama@acknowledge-rs.com
#### DEVICE INFORMATION
| Proprietary Name/Trade Name: | SuperGrow 272 and Ultimate 272 |
|------------------------------|----------------------------------------------------|
| Common Name: | Laser, Comb, Hair |
| Regulation Number: | 21 CFR 890.5500 |
| Class: | II |
| Product Code: | OAP |
| Premarket Review: | ODE/DSD/General Surgery Devices Branch One (GSDB1) |
| Review Panel: | General and Plastic Surgery |
#### PREDICATE DEVICE IDENTIFICATION
The SuperGrow 272 and the Ultimate 272 are substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|---------------|---------------------------------------------------------|-------------------|
| K172982 | SuperGrow 272 and Ultimate 272 / Super Grow Lasers | ✓ |
| K163170 | Capillus272Pro, Capillus 272 OfficePro / Capillus, LLC. | |
The predicate devices have not been subject to a design related recall.
#### DEVICE DESCRIPTION
The SuperGrow 272 and the Ultimate 272 are non-invasive, low level laser therapy (LLLT) devices each containing 272 red, visible light, diode lasers. The devices are designed to deliver non-thermal energy to the hair follicles via photostimulation of the scalp. The SuperGrow 272
{4}------------------------------------------------
asers
consists of 272 red, visible light, diode lasers operating at a 650 nanometer wavelength (maximum output power of each is 5 mW). The diodes are configured within a cap, enclosed between the outer shell and inner liner, both made of polyethylene terephthalate (PET). The Ultimate 272 is exactly the same as the SuperGrow 272 in every way except it is mounted on a stand. The same electrical design (power supply and the on/off switch) has been applied to the Ultimate 272.
## INDICATIONS FOR USE
The SuperGrow 272 laser hat and the Ultimate 272 are intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood – Hamilton Classifications of Ila to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-2, or frontal; both with Fitzpatrick Skin Types I to IV.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
SuperGrow Lasers believes that the SuperGrow 272 and the Ultimate 272 are substantially equivalent to the predicate devices based on the information summarized here:
The subject devices have identical designs and dimensions, and use identical materials as the devices cleared in K172982. The subject devices have intended use and identical or similar technological characteristics (laser wavelength, amount of diodes, energy per diode, output mode) as the devices cleared in K172982 and K163170. The only difference between the subject devices and those cleared in K172982 is that the indications are being expanded to include over-the-counter use (similar to the indications for the devices cleared in K163170).
#### NON-CLINICAL TESTING
No FDA performance standards have been established for the SuperGrow 272 and the Ultimate 272. The following tests were performed to demonstrate safety based on current industry standards:
Electromagnetic Compatibility and Electrical Safety was evaluated per IEC 60601-1, IEC60601-1-2 and IEC 60601-1-11. Biocompatibility was evaluated per ISO 10993-1.
#### CLINICAL TESTING
The capability of the average "retail customer" to self-select and appropriately use the subject device without the counsel of a health care provider was evaluated in a usability study. The study was conducted on 40 subjects of both genders, with a wide range in age, educational background, and disease status, as a sample of the intended use population. The first part of the study evaluated the ability of the subjects to decide if they were an appropriate user based on the indications and contraindications listed on the packaging. The second part assessed the ability of the subjects to understand the instructions for use, warnings, and precautions
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Supergrow Lasers. The logo consists of a red graphic on the left and the text "SUPERGROW LASERS" on the right. The graphic is made up of three red shapes that are arranged in a triangular pattern. The text is in a bold, sans-serif font.
provided inside a standard retail package. An observer evaluated the ability of the subjects to properly use the device according to the instructions provided. To demonstrate equivalence for OTC distribution, a minimum of an 80% success rate was required. For each part of the test, all questions had to be answered correctly in order to pass. The results of each part were considered independently. With a passing rate of 80% for self-selection, 82.5% for the comprehension of the instructions, and 80% for the proper use of the results of the usability testing demonstrate that the SuperGrow 272 and the Ultimate 272 are safe for overthe-counter distribution.
## CONCLUSION
Based on the testing performed, including usability testing, it can be concluded that the subject devices do not raise new issues of safety or effectiveness compared to the predicate devices. The similar or identical indications for use, technological characteristics, and performance characteristics for the proposed SuperGrow 272 and Ultimate 272 are assessed to be substantially equivalent to the predicate device
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.