K181305 · Premier Biotech, Inc. · DJC · Sep 20, 2018 · Clinical Toxicology
Device Facts
Record ID
K181305
Device Name
OralTox Oral fluid Drug Test
Applicant
Premier Biotech, Inc.
Product Code
DJC · Clinical Toxicology
Decision Date
Sep 20, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3610
Device Class
Class 2
Indications for Use
The OralTox® Oral Fluid Drug Test is a competitive binding, lateral flow immunochromatographic assay for the qualitative and simultaneous detection of amphetamine, cocaine, marijuana (THC), methamphetamine, opiates, phencyclidine, oxycodone and methadone in human oral fluid at the cut-off concentrations listed below and their metabolites: Drug (Identifier) Calibrator Cut-off level Amphetamine (AMP) D-Amphetamine 50 ng/mL Cocaine (COC) Benzoylecgonine 20 ng/mL Cannabinoids (THC) Delta-8-Tetrahydrocannabinol 40 ng/mL Methamphetamine (MET) D-Methamphetamine 50 ng/mL Opiates (OPI) Morphine 40 ng/mL Phencyclidine (PCP) Phencyclidine 10 ng/mL Oxycodone (OXY) Oxycodone 20 ng/mL Methadone (MTD) Methadone 30 ng/mL The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Liquid Chromatography/Mass Spectrometry/Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. It is not intended to detect intermittent dosing of oxycodone. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for prescription use.
Device Story
OralTox® is a single-use, lateral flow immunochromatographic test cup for qualitative drug screening in human oral fluid. Device includes integrated collection sponge and saturation indicator strip; indicator turns red when sufficient sample volume is collected for testing and potential laboratory confirmation. Operation involves collecting oral fluid via sponge; fluid migrates via capillary action through test strips. Principle is competitive binding: target drugs below cut-off allow antibody-coated particles to bind to immobilized drug-conjugate, forming a visible test line; drug concentrations above cut-off saturate binding sites, preventing line formation. Control line confirms proper test performance. Used at point-of-care by healthcare professionals. Results are visually read. Preliminary positive results require confirmation via LC-MS/MS. Benefits include rapid, on-site screening for multiple drug analytes to inform clinical decision-making.
Clinical Evidence
No clinical studies were performed. Performance was established via bench testing, including precision/reproducibility studies across three sites and method comparison studies. Method comparison evaluated 932 samples (434 for oxycodone, 498 for methadone) against LC-MS/MS reference method. Results showed high concordance with LC-MS/MS; discordant results were analyzed. Analytical specificity, cross-reactivity, and interference studies (exogenous substances, pH, smoking, hemoglobin) were conducted to validate performance for the new analytes.
Technological Characteristics
Lateral flow immunochromatographic assay. Single-use test cup with integrated collection sponge and saturation indicator. Qualitative, visually read. No energy source required. Materials include monoclonal mouse antibodies. Analyte cut-offs: AMP 50 ng/mL, COC 20 ng/mL, THC 40 ng/mL, MET 50 ng/mL, OPI 40 ng/mL, PCP 10 ng/mL, OXY 20 ng/mL, MTD 30 ng/mL. Stable for 24 months at 4-30°C.
Indications for Use
Indicated for qualitative, simultaneous detection of Amphetamine, Cocaine, Marijuana (THC), Methamphetamine, Opiates, Phencyclidine, Oxycodone, and Methadone in human oral fluid. For prescription use. Not intended to detect intermittent Oxycodone dosing.
Regulatory Classification
Identification
A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.
Special Controls
*Classification.* Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
OralTox® Oral Fluid Drug Test (k171403)
Submission Summary (Full Text)
{0}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATIONS
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k181305
B. Purpose for Submission:
Modification of a previously cleared device to include new test analytes, methadone and oxycodone.
C. Measurands:
Amphetamine, cocaine, cannabinoids, methamphetamine, morphine, phencyclidine, oxycodone and methadone.
D. Type of Test:
Qualitative, lateral flow immunochromatographic
E. Applicant:
Premier Biotech, Inc.
F. Proprietary and Established Names:
OralTox® Oral Fluid Drug Test
G. Regulatory Information:
| Product Code | Classification | Relation Section | Panel |
| --- | --- | --- | --- |
| DKZ | Class II | 21CFR 862.3100, Amphetamine Test System | Toxicology (91) |
| DIO | Class II | 21 CFR 862.3250, Cocaine and metabolites TEST System | Toxicology (91) |
| LDJ | Class II | 21 CFR 862.3870, Cannabinoids Test System | Toxicology (91) |
| DJC | Class II | 21 CFR 862.3610, Methamphetamine Test System | Toxicology (91) |
| LCM | Class II | Unclassified, Enzyme immunoassay Phencyclidine | Toxicology (91) |
| DJG | Class II | 21 CFR 862.3650, Opiate Test System | Toxicology (91) |
1
{1}
| DJR | Class II | 21 CFR, 862.3610 Methadone Test System | Toxicology (91) |
| --- | --- | --- | --- |
### H. Intended Use:
#### 1. Intended use(s):
Refer to Indications for Use below.
#### 2. Indication(s) for use:
The OralTox \( ^{®} \) Oral Fluid Drug Test is a competitive binding, lateral flow immunochromatographic assay for the qualitative and simultaneous detection of amphetamine, cocaine, marijuana (THC), methamphetamine, opiates, phencyclidine, oxycodone and methadone in human oral fluid at the cut-off concentrations listed below and their metabolites:
| Drug (Identifier) | Calibrator | Cut-off level |
| --- | --- | --- |
| Amphetamine (AMP) | D-Amphetamine | 50 ng/mL |
| Cocaine (COC) | Benzoylecgonine | 20 ng/mL |
| Cannabinoids (THC) | Delta-8-Tetrahydrocannabinol | 40 ng/mL |
| Methamphetamine (MET) | D-Methamphetamine | 50 ng/mL |
| Opiates (OPI) | Morphine | 40 ng/mL |
| Phencyclidine (PCP) | Phencyclidine | 10 ng/mL |
| Oxycodone (OXY) | Oxycodone | 20 ng/mL |
| Methadone (MTD) | Methadone | 30 ng/mL |
The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Liquid Chromatography/Mass Spectrometry/Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. It is not intended to detect intermittent dosing of oxycodone. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. The tests are intended for prescription use.
#### 3. Special conditions for use statement(s):
For prescription point-of-care use.
#### 4. Special instrument requirements:
Not applicable; the device is a visually-read single use device.
2
{2}
# I. Device Description:
The OralTox® Oral Fluid Drug Test is an immunochromatographic assay that uses a lateral flow system for the qualitative detection of amphetamine, cocaine, cannabinoids, methamphetamine, morphine, phencyclidine, oxycodone and methadone (target analytes) in human oral fluid.
The products are single-use in vitro diagnostic devices. Each test kit contains a test cup, a package insert and a sample collection sponge. Each test device is sealed with a desiccant in an aluminum pouch.
The device has a built-in sample collection tool, including collection swab and sponge, as well as a saturation indicator strip. The saturation indicator strip turns red when a volume of oral fluid has been collected that is adequate to both run the OralTox® test and to conduct confirmation testing at a laboratory if the result is a presumptive positive.
OralTox® Oral Fluid Drug Tests are the first step in a two-step process. The second step is to send the sample for confirmatory laboratory testing using a more specific method (e.g., LC-MS/MS) if preliminary positive results are obtained.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
OralTox® Oral Fluid Drug Test
2. Predicate 510(k) number(s):
k171403
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Proposed device | Predicate device |
| Intended Use | Same | Preliminary Drug screening test for the qualitative detection of drug analytes in oral fluid (human saliva) For In Vitro Diagnostic Use |
| Drug analytes detected | D-Amphetamine Cocaine Delta-9-Tetrahydrocannabinol D-Methamphetamine Morphine Phencyclidine Methadone | D-Amphetamine Cocaine Delta-9-Tetrahydrocannabinol D-Methamphetamine Morphine Phencyclidine |
3
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Proposed device | Predicate device |
| | Oxycodone | |
| Methodology | Same | Competitive binding, lateral flow immunochromatographic assays based on the principle of antigen antibody immunochemistry |
| Type of Test | Same | Qualitative |
| Specimen Type | Same | Human Oral fluid |
| Cut-offs | AMP 50 ng/mL COC 20 ng/mL THC 40 ng/mL MET 50 ng/mL OPI 40 ng/mL PCP 10 ng/mL OXY 20 ng/mL MTD 30 ng/mL | AMP 50 ng/mL COC 20 ng/mL THC 40 ng/mL MET 50 ng/mL MOP 40 ng/ML PCP 10 ng/mL |
| Differences | | |
| Item | Device | Predicate |
| Drug analytes detected | D-Amphetamine Cocaine Delta-9-Tetrahydrocannabinol D-Methamphetamine Morphine Phencyclidine Methadone Oxycodone | D-Amphetamine Cocaine Delta-9-Tetrahydrocannabinol D-Methamphetamine Morphine Phencyclidine |
| Number of analytes detected | 8 | 6 |
### K. Standard/Guidance Document Referenced (if applicable):
None Referenced.
### L. Test Principle:
The OralTox® Oral Fluid Drug Test is an immunochromatographic assay that uses a lateral flow system for the qualitative detection of amphetamine, cocaine, cannabinoids, methamphetamine, morphine, and phencyclidine, methadone, and oxycodone, (target analytes) in human oral fluid. The tests are lateral flow chromatographic immunoassays. During testing, an oral fluid specimen migrates upward by capillary action. If target drugs present in the oral fluid specimen are below the cut-off concentration, it will not saturate the
4
{4}
binding sites of its specific monoclonal mouse antibody coated on the particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the target drug level exceeds its cut-off-concentration because it will saturate all the binding sites of the antibody coated on the particles. A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample to indicate that the tests have been performed properly.
### M. Performance Characteristics (if/when applicable):
#### 1. Analytical performance:
The analytical performance of the device for the measurement of d-amphetamine (AMP), cocaine (COC), delta-8-tetrahydrocannabinol (THC), d-methamphetamine (MET), morphine (OPI), and phencyclidine (PCP), was reviewed in k171403. Please see the decision summary for k171403 for information regarding performance of these analytes.
##### a. Precision/Reproducibility:
Precision-reproducibility-cut-off studies were carried out for samples with concentrations of -100% cut-off, -75% cut-off, -50% cut-off, -25% cut-off, cut-off, +25% cut-off, +50% cut-off, +75% cut-off and +100% cut-off. Testing was repeated across three sites. These samples were prepared by spiking drug in negative oral fluid samples. The sponsor has stated that each drug concentration was confirmed by LC-MS/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed two runs per day for 10 days per device lot in a randomized order. The sponsor provided the following summaries:
| Methadone | -100% cut-off | -75% cut-off | -50% cut-off | -25% cut-off | cut-off | +25% cut-off | +50% cut-off | +75% cut-off | +100% cut-off |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Lot 1 | 60-/0+ | 60-/0+ | 60-/0+ | 54-/6+ | 49+/11- | 55+/5- | 60+/0- | 60+/0- | 60+/0- |
| Lot 2 | 60-/0+ | 60-/0+ | 60-/0+ | 55-/5+ | 48+/12- | 56+/4- | 60+/0- | 60+/0- | 60+/0- |
| Lot 3 | 60-/0+ | 60-/0+ | 60-/0+ | 55-/5+ | 50+/10- | 55+/5- | 60+/0- | 60+/0- | 60+/0- |
| Oxycodone | -100% cut-off | -75% cut-off | -50% cut-off | -25% cut-off | cut-off | +25% cut-off | +50% cut-off | +75% cut-off | +100% cut-off |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Lot 1 | 60-/0+ | 60-/0+ | 60-/0+ | 55-/5+ | 50+/10- | 55+/5- | 60+/0- | 60+/0- | 60+/0- |
| Lot 2 | 60-/0+ | 60-/0+ | 60-/0+ | 54-/6+ | 49+/11- | 56+/4- | 60+/0- | 60+/0- | 60+/0- |
| Lot 3 | 60-/0+ | 60-/0+ | 60-/0+ | 56-/4+ | 49+/11- | 57+/3- | 60+/0- | 60+/0- | 60+/0- |
5
{5}
The device has been developed with the following cut-offs:
| Calibrator | Cut-off (ng/mL) |
| --- | --- |
| D-methamphetamine | 50 |
| Cocaine | 20 |
| Morphine | 40 |
| D-amphetamine | 50 |
| Phencyclidine | 10 |
| Delta-9-tetrahydrocannabinol | 40 |
| Methadone | 30 |
| Oxycodone | 20 |
Sample Volume:
A sample volume study was conducted in k171403 to confirm the reproducibility of adequate sample volume collection by the device and found to be acceptable.
b. Linearity/assay reportable range:
Not applicable. These devices are intended for qualitative use only.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Accelerated stability and real time studies have been conducted for the device. Protocols and acceptance criteria were described and found to be acceptable. The manufacturer claims that when stored un-opened at 4-30 °C, the device is stable for 24 months.
### Sample Recovery
Volume recovery:
A sample volume recovery study was provided in k171403 and found to be acceptable.
Analyte recovery:
In order to confirm that preliminary positive results can be adequately measured via confirmation testing after being subject to the temperature conditions required for shipping and storage, negative oral fluid samples in glass bottles were spiked with a single analyte/bottle to concentrations approximately -50% and +50% of the cut-off. Samples were spiked using known standards. Each drug concentration was confirmed by LC-MS/MS. The samples were transferred to OralTox® devices using the collection sponges. For each of 3 storage conditions (-20°C, 20-25°C, and 40°C), 12 devices were used (4 devices from each of 3 lots. For each device, drug was measured by LC-MS/MS at time zero and the devices containing the specimens were
6
{6}
stored under the specified condition. Drug in the devices stored at 20-25°C and 40°C was measured by LC-MS/MS at two (2) days, and drug in the devices stored at -20°C was measured by LC-MS/MS at 90 days.
The minimum and maximum recovery from 12 devices per lot per storage condition is shown below. Analyte recovery for d-amphetamine (AMP), cocaine (COC), delta-8-tetrahydrocannabinol (THC), d-methamphetamine (MET), morphine (OPI), and phencyclidine (PCP) was provided in k171403.
Room Temperature 20-25°C (two day storage)
| | OXY +50 | OXY -50 | MTD +50 | MTD -50 |
| --- | --- | --- | --- | --- |
| Max | 99 | 99 | 100 | 99 |
| Min | 91 | 91 | 92 | 92 |
40 °C (two day storage)
| | OXY +50 | OXY -50 | MTD +50 | MTD -50 |
| --- | --- | --- | --- | --- |
| Max | 98 | 99 | 100 | 93 |
| Min | 92 | 93 | 92 | 97 |
-20 °C (90 day storage)
| | OXY +50 | OXY -50 | MTD +50 | MTD -50 |
| --- | --- | --- | --- | --- |
| Max | 100 | 98 | 101 | 105 |
| Min | 93 | 93 | 93 | 91 |
Results indicate that samples may be stored at room temperature for up to two days, elevated temperature for up to do days (40 °C), and for up to 90 days at – 20 °C, prior to confirmatory testing.
# d. Detection limit:
See Precision/Reproducibility section in M.1.a above.
# e. Analytical specificity:
To test cross reactivity, drug metabolites and other components that may be present in oral fluid samples were tested using three lots of the OralTox® device. The following is a summary of the cross-reactivity study. A cross-reactivity study for d-amphetamine (AMP), cocaine (COC), delta-8-tetrahydrocannabinol (THC), d-methamphetamine (MET), morphine (OPI), and phencyclidine (PCP) was provided in k171403.
7
{7}
| Oxycodone (Cut-off=20 ng/mL) | Result Positive at (ng/mL) | % Cross-Reactivity |
| --- | --- | --- |
| Oxycodone | 20 | 100% |
| Hydrocodone | 1000 | 2% |
| Hydromorphone | 6250 | 0.3% |
| Naloxone | 6250 | 0.3% |
| Oxymorphone | 1000 | 2% |
| Dihydrocodeine | Negative at 10000 | <0.2% |
| Buprenorphine | Negative at 10000 | <0.2% |
| 6-AM | Negative at 10000 | <0.2% |
| Codeine | Negative at 10000 | <0.2% |
| Heroin | Negative at 10000 | <0.2% |
| Morphine | Negative at 10000 | <0.2% |
| Morphine -3-β-d- | Negative at 10000 | <0.2% |
| Ethylmorphine | Negative at 10000 | <0.2% |
| Methadone (Cut-off=30 ng/mL) | Result Positive at(ng/ml) | % Cross-Reactivity |
| --- | --- | --- |
| Alpha-methadol | 125 | 24% |
| Doxylamine | 12500 | 0.24% |
| 2-Ethylidene-1,5-dimethyl-3,3-diphenyl pyrrolidine (EDDP) | 10000 | 0.3% |
| Phencyclidine | 12500 | 0.24% |
| 2-Ethyl-5-methyl-3,3-diphenylpyrroline (EMDP) | 100000 | 0.03% |
| LAAM | 10000 | 0.3% |
**Exogenous Interference:** Potential interference from structurally unrelated compounds were tested by spiking the potentially interfering compound at a concentration of 10 ug/mL into drug free oral fluid or fluid containing the target drug with concentrations of 50% below and 50% above cut-off level. The following compounds were found not to interfere with test results at a concentration of 10ug/mL for all samples tested. An exogenous interference study for D-Amphetamine (AMP), Cocaine (COC), Delta-8-Tetrahydrocannabinol (THC), D-Methamphetamine (MET), Morphine (OPI), and Phencyclidine (PCP) was provided in k171403.
| Acetaminophen | Digoxin | Nicotinamide |
| --- | --- | --- |
| Acetylcodeine | Dihydrocodeine | Nicotine |
| Allobarbital | diltiazem HCl | Noscapine |
| Alprazolam | Diphenhydramine HCl | Omeprazole |
| Amobarbital | DL-Propranolol | Papaverine |
8
{8}
| Apomorphine | Doxylamine | Pentazocine |
| --- | --- | --- |
| Atenolol | Ecgonine methylester | Phentermine |
| Atropine | Estradiol | Phenylpropanolamine |
| Baclofen | Estrone | Phenytoin |
| Benzocaine | Fluconazole | Pioglitazone HCl |
| Butabarbital | Furosemide | Prednisolone |
| Caffeine | Hexobarbital | Prednisone |
| Cannabidiol | Hydrochlorothiazide | Procainamide HCl |
| Carbamazepine | Ibuprofen | Procaine HCL |
| Chlordiazepoxide | Imipramine | Promethazine |
| Chlorpromazine | Lamotrigine | Quinine HCl |
| Cimetidine | Levetiracetam | R,r(-)-pseudoephedrine |
| Citalopram HBr | Lidocaine | Salicylic Acid |
| Clobazam | Lormetazepam | Sertraline HCL |
| Clomipramine | L-thyroxine | Simvastin |
| Clonazepam | Metformin HCl | Theophylline |
| Clonidine | Methylphenidate HCl | Thiamine |
| Clopidogrel bisulfate | Metoprolol | Topiramate |
| Cortisol | Metronidazole | Valproic acid |
| Cotinine | Montelukast sodium salt | Verapamil |
| D,l-Salbutamol | Naloxone | Zonisamide |
| Deoxycorticosterone | Naltrexone | |
| Dextromethorphan | Naproxen | |
The following potential interference from substances commonly present in oral fluid were evaluated by spiking into drug free oral fluid or oral fluid containing the target drug with concentrations of 50% below and 50% above cut-off level to a concentration of 5%: alcohol, baking soda, chewing gum, coffee, cola, cough syrup, cranberry juice, food coloring (blue, green, and red), methanol cough drops, milk, mouthwash, MSG, orange juice, salt, sugar, tea, toothpaste, and tomatoes. None of these substances showed any interference with the detection of any analyte (OXY, MTD) using the device. A study testing these substances for interference to detection of d-amphetamine (AMP), cocaine (COC), delta-8-tetrahydrocannabinol (THC), d-methamphetamine (MET), morphine (OPI), and phencyclidine (PCP) was provided in k171403.
Potential interference from cigarette smoking, was evaluated by asking a participant to smoke a cigarette. After 15 minutes of smoking, an oral fluid sample was collected and spiked with each drug at concentrations of cut-off +/- 50%. No interference with the detection of any analyte (OXY, MTD) using the device was observed. A study testing potential interference from cigarette smoking to detection of d-amphetamine (AMP), cocaine (COC), delta-8-tetrahydrocannabinol (THC), d-methamphetamine
9
{9}
(MET), morphine (OPI), and phencyclidine (PCP) was provided in k171403.
Potential interference from hemoglobin was evaluated by adding it to drug free oral fluid or oral fluid containing the target drug with concentrations of 50% above and 50% below cut-off level to a concentration of 100ug/mL. No interference with the detection of any analyte (OXY, MTD) using the device was observed. A study testing for potential interference from hemoglobin to detection of d-amphetamine (AMP), cocaine (COC), delta-8-tetrahydrocannabinol (THC), d-methamphetamine (MET), morphine (OPI), and phencyclidine (PCP) was provided in k171403.
Effect of oral fluid pH: To investigate the effect of oral fluid pH, oral fluid samples with pH 4 to 9 were spiked with target drugs at 50% below and 50% above cut-off levels. These samples were tested using three lots of the device. Results were all positive for samples (OXY, MTD) at and above +50% cut-off and all negative for samples at and below -50% Cut-Off. A study to test the potential interference due to pH for d-amphetamine (AMP), cocaine (COC), delta-8-tetrahydrocannabinol (THC), d-methamphetamine (MET), morphine (OPI), and phencyclidine (PCP) was provided in k171403.
There is the possibility that other substances and/or factors not listed above may interfere with the test and cause false results, e.g., technical or procedural errors.
# f. Assay cut-off:
Characterization of how the device performs analytically around the claimed cut-off concentration appears in the precision section, M.1.a., above.
# 2. Comparison studies:
# a. Method comparison:
Method comparison studies for the OralTox® Oral fluid Drug Test were performed, testing a total of 932 samples. 434 samples were evaluated for oxycodone at four sites, and 498 samples were evaluated for methadone at four sites. Each site utilized three operators. Device results were compared to LC-MS/MS results. The results are presented in the tables below. A method comparison study for d-amphetamine (AMP), cocaine (COC), delta-8-tetrahydrocannabinol (THC), d-methamphetamine (MET), morphine (OPI), and phencyclidine (PCP) was provided in k171403.
10
{10}
### Oxycodone
| Concentration (% of Cut-off) by LC/MS | Number of samples | OralTox Results | | The percentage of correct results (%) |
| --- | --- | --- | --- | --- |
| | | No. of Positive | No. of Negative | |
| Drug free (0%) | 152 | 0 | 152 | 100 |
| <50% of cut-off | 47 | 0 | 47 | 100 |
| 50% of cut-off to cut-off | 32 | 5 | 27 | 84.4 |
| Cut-off to 150% of cut-off | 29 | 25 | 4 | 86.2 |
| >150% of cut-off | 174 | 174 | 0 | 100 |
### Discordant results (oxycodone)
| Sample | LC/MS Result | Device Result |
| --- | --- | --- |
| 1 | 18.0 | Positive |
| 2 | 17.2 | Positive |
| 3 | 19.05 | Positive |
| 4 | 18.23 | Positive |
| 5 | 18.52 | Positive |
| 6 | 23.29 | Negative |
| 7 | 22.04 | Negative |
| 8 | 22.4 | Negative |
| 9 | 22.34 | Negative |
### Methadone
| % of Cut-off | Number of samples | OralTox Results | | The percentage of correct results (%) |
| --- | --- | --- | --- | --- |
| | | No. of Positive | No. of Negative | |
| Drug free (0%) | 277 | 0 | 277 | 100 |
| <50% of cut-off | 13 | 0 | 13 | 100 |
| 50% of cut-off to cut-off | 20 | 4 | 16 | 80 |
| cut-off to 150% of cut-off | 15 | 13 | 2 | 87 |
| >150% of cut-off | 173 | 173 | 0 | 100 |
11
{11}
Discordant results (methadone)
| Sample Number | LC/MS Result | Device Result |
| --- | --- | --- |
| 1 | 26.9 | Positive |
| 2 | 27.85 | Positive |
| 3 | 28.26 | Positive |
| 4 | 28.82 | Positive |
| 5 | 31.07 | Negative |
| 6 | 31.29 | Negative |
b. Matrix comparison:
Not applicable. These devices are for use with oral fluid samples only.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Not applicable.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.