K181297 · Hdx Will Corp. · MUH · Nov 21, 2018 · Dental
Device Facts
Record ID
K181297
Device Name
DENTIOIII series (DENTIOIII, DENTIOIII-S)
Applicant
Hdx Will Corp.
Product Code
MUH · Dental
Decision Date
Nov 21, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1800
Device Class
Class 2
Indications for Use
The DENTIOIII series is intended for dental radiographic examination of the teeth and temporomandibular joints, specifically for panoramic and cephalometric examinations. It is to be used only by dental practitioners and/or radiologists
Device Story
Dental X-ray imaging system; captures panoramic and cephalometric images. Operates by rotating X-ray generator and detector around patient; X-rays attenuated by dense structures (bone) and permeate soft tissue; detector measures attenuated X-rays. Data transmitted via Ethernet/RS232 to workstation; software reconstructs images for display and diagnostic analysis. Used in dental clinics; operated by dental practitioners or radiologists. Laser used for patient positioning. Output allows visualization of anatomical structures for diagnosis. Benefits include diagnostic imaging for dental and orthodontic treatment planning.
Clinical Evidence
Bench testing only. Performance assessed via imaging metrics including Modulation Transfer Function (MTF) and Detective Quantum Efficiency (DQE) per IEC 6220-1 and IEC 61223-3-4. Results demonstrate diagnostic image quality is equal or better than predicate devices. Electrical safety and EMC testing performed per IEC 60601 series.
Technological Characteristics
Extraoral X-ray system; CMOS/CCD scanning detectors; Tungsten anode; 0.5 mm focal spot; 60-90 kV voltage range; 4-10 mA current range; DICOM 3.0 compatible. Connectivity via Ethernet/RS232. Class 1 laser for positioning. Software-based image reconstruction. Complies with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-63, and 61223-3-4.
Indications for Use
Indicated for dental radiographic examination of teeth and temporomandibular joints in patients requiring panoramic or cephalometric imaging. For use by dental practitioners and radiologists.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
DENTRIα series (DENTRIα, DENTRI-Cα, DENTRI-Sα) (K160140)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 21, 2018
HDX WILL CORP. Lee Myoung-Joon General Manager #105, 106, 201, 202, 203, 204, 38, Osongsaengmyeong 4-ro Osong-eup, Heungdeok-gu CHEONGJU-SI CHUNGCHEONGBUK-DO 28161 REPUBLIC OF KOREA
Re: K181297
Trade/Device Name: DENTIOIII series (DENTIOIII, DENTIOIII-S) Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral Source X-Ray System Regulatory Class: Class II Product Code: MUH Dated: September 21, 2018 Received: September 25, 2018
Dear Lee Myoung-Joon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Hole 2. Mils
for Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181297
Device Name DENTIOIII series (DENTIOIII, DENTIOIII-S)
Indications for Use (Describe)
The DENTIOIII series is intended for dental radiographic examination of the teeth and temporomandibular joints, specifically for panoramic and cephalome. It is to be used only by dental practitioners and/or radiologists
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
> K181297 Page 1/8
# 510(k) Summary
[As required by 21 CFR 807.92]
# 1. Date Prepared [21 CFR 807.92(a)(a)]
Sept. 21. 2018
### 2. Submitter's Information [21 CFR 807.92(a)(1)]
| - Name of Manufacturer: | HDX WILL CORP. |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| - Address: | #105, 106, 201, 202, 203, 204, 38, Osongsaengmyeong 4-ro,<br>Osong-eup, Heungdeok-gu, Cheongju-si, Chungcheongbuk-do<br>28161, Korea |
| - Contact Name: | Myoung-Joon Lee / General Manager |
| - Telephone No.: | +82-43-710-7318 |
| - Fax No.: | +82-43-710-7312 |
| - Email Address: | mjlee@iwillmed.com |
| - Registration No.: | 3013511605 |
## 3. Trade Name, Regulation Name, Classification [21 CFR 807.92(a)(2)]
| Trade Name | DENTIOIII series (DENTIOIII, DENTIOIII-S) |
|-------------------|-------------------------------------------|
| Regulation Number | 21 CFR 872.1800 |
| Regulation Name | Extraoral source x-ray system |
| Regulation Class | II |
| Product Code | MUH |
| Product Code Name | System, X-Ray, Extraoral Source, Digital |
{4}------------------------------------------------
> K181297 Page 2/8
### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)[
The identified predicate devices within this submission are shown as follow;
Predicate device #1 - 510(k) Number: K142058 - Applicant: RAY Co., Ltd. - Regulation Name: Extraoral source x-ray system - Product Code: MUH - Device Class: Class II - Device Name: RAYSCAN α-Expert
Predicate device #2
| - 510(k) Number: | K992385 |
|--------------------|-------------------------------|
| - Applicant: | INSTRUMENTARIUM CORP. |
| - Regulation Name: | Extraoral source x-ray system |
| - Product Code: | MUH |
| - Device Class: | Class II |
| - Device Name: | Orthopantomograph® OP100 D |
Predicate device #3
| - 510(k) Number: | K160140 |
|--------------------|------------------------------------------------|
| - Applicant: | HDX WILL CORP. |
| - Regulation Name: | Computed tomography x-ray system |
| - Product Code: | OAS |
| - Device Class: | Class II |
| - Device Name: | DENTRIα series (DENTRIα, DENTRI-Cα, DENTRI-Sα) |
There are no significant differences between the DENTIOIII, DENTIOIII, DENTIOIII-S) and the predicate device that would adversely affect the use of the product.
### 5. Description of the Device [21 CFR 807.92(a)(4)]
This equipment is a dental X-ray imaging device used for diagnostic purposes in dental examination. The operating principle of this device is obtaining the panoramic images by rotating the arm to get the recombination data, having the X-ray generator and detector rotate around the patient to irradiate the X-ray, and measuring the penetrated X-ray using the detector. When the X-ray irradiates the teeth area for instance, a large amount of X-ray is attenuated because objects such as bones are highly dense. By contrast, the X-ray permeates more through small molecules with low density such as skin or tissue measuring the X-ray is reconstructed by the software for display and analysis, the anatomical structure can be viewed for the diagnosis purposes. In addition, the Cephalo arm uses the scanning method.
1) The DENTIOIII Series are classified as shown below.
▪ DENTIOIII: PANORAMA Mode
▪ DENTIOIII-S: PANORAMA Mode + CEPHALO Mode (SCAN)
2) Description of the image detectors used.
{5}------------------------------------------------
# K181297 Page 3/8
| Model | DENTIOIII | | DENTIOIII-S | |
|----------------------|-----------------------------|---------------------------------|-----------------------------|------------------------------------|
| Contents | Scanning (Pano) | | Scanning (Pano /Ceph) | |
| Detector model | Xineos-1501 | ARGUS-PAN<br>(DM-20-05K10-00-R) | Xineos-2301 | ARGUS - CEPH<br>(DM-20-08K10-00-R) |
| Manufacturer | Teledyne DALSA | Teledyne DALSA | Teledyne DALSA | Teledyne DALSA |
| Detector type | Scanning detector<br>(CMOS) | Scanning detector<br>(CCD) | Scanning detector<br>(CMOS) | Scanning detector<br>(CMOS) |
| Resolution (pixels) | 1536 x 68 | 5580 x 256 | 2305 x 68 | 8160 x 256 |
| Pixel size (um) | 99.0 | 27.0 | 99.0 | 27.0 |
| MTF | 65% at 1LP/mm | 70% at 1LP/mm | 65% at 1LP/mm | 70% at 1LP/mm |
| DQE | 57% at 1LP/mm | 50% at 0LP/mm | 57% at 1LP/mm | 50% at 0LP/mm |
| Active area (mm) | 152.0 x 7.0 mm | 151.0 x 6.9 mm | 228.0 x 7.0 mm | 221.0 x 6.9 mm |
| A/D Conversion | 14 bits | 16 bits | 14 bits | 16 bits |
| FDA 510(k)<br>Number | Not known | Not known | Not known | Not known* |
* Note. Information for detector own 510(K) number or system in which it was cleared 510(k) number.
| Contents | Model | Own 510(K)<br>number | System in which it was cleared 510(k) number. |
|---------------------|------------------------------------|----------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| CEPHALO<br>detector | ARGUS - CEPH<br>(DM-20-08K10-00-R) | No | 1) System name: DENTRIα series (DENTRIα,<br>DENTRI-Ca, DENTRI-Sa)<br>2) Manufacturer: HDX WILL CORP.<br>3) 510(K) Number: K160140 |
The DENTIOIII series communicates with the Workstation by transmitting data using Ethernet cable and RS232 Cable.
### 3) Laser
The laser is used for patient positioning and is class 1. This technical characteristic is as follows:
| | LM-6501RS (Option1) | P9-655-5L (Option2) |
|---------------------|---------------------|---------------------|
| Optical output | 1 mW or less | 1 mW or less |
| Wavelength | 655 nm ± 5 nm | 655 nm ± 6 nm |
| Line type | Accurate Straight | Accurate Straight |
| Form of beam output | Gaussian | Gaussian |
## 6. Indications for Use [21 CFR 807.92(a)(5)]
The DENTIOIII series is intended for dental radiographic examination of the teeth and temporomandibular joints, specifically for panoramic and cephalometric examinations. It is to be used only by dental practitioners and/or radiologists
{6}------------------------------------------------
# W HDXWILL
#105, 106, 201, 202, 203, 204, 38, Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu, Cheongju-si, Chungcheongbuk-do, 28161, Korea, Tel: +82-43-710-7318/ Fax: +82-43-710-7312
> K181297 Page 4/8
### 7. Determination of Substantial Equivalence
The DENTION series is substantially equivated predicate devices with respect of the companison of technical features. The subject device was fund to be equivalent to predicate devices with regarding to dechnical characteristics. The table below presents comparisons for each origations of technological characteristics of the device compared to the predicate devices. [21 CFR 807.92(a)(6)]
| | Proposed Device | Predicate Device #1 | Predicate Device #2 | Predicate Device #3 |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | K181297 | K142058 | K992385 | K160140 |
| Model | DENTIOIII series<br>· DENTIOIII<br>· DENTIOIII-S | RAYSCAN α- Expert<br>· RAYSCAN α -P<br>· RAYSCAN α -SC<br>· RAYSCAN α -OCL<br>· RAYSCAN α -OCS | Orthopantomograph® OP100 D | DENTRIα series<br>· DENTRIα<br>· DENTRI-Cα<br>· DENTRI-Sα |
| Manufacturer | HDX WILL CORP. | RAY Co., Ltd. | INSTRUMENTARIUM CORP. | HDX WILL CORP. |
| Indications for Use | The DENTIOIII series is intended<br>for dental radiographic examination<br>of the teeth and temporomandibular<br>joints, specifically for panoramic<br>and cephalometric examinations. It<br>is to be used only by dental<br>practitioners and/or radiologists | The RAYSCAN α - Expert Dental<br>X-Ray System is an extraoral<br>source dental panoramic and<br>optional cephalometric X-ray<br>system intended to produce X-rays<br>for dental radiographic examination<br>and diagnosis of diseases of the<br>teeth, jaw, and oral structures. | Orthopantomograph® OP100 D is<br>intended to be used for producing<br>diagnostic X-ray radiograph of<br>dentition, TM-joints and other oral<br>structures. | The DENTRIα series is a Computed<br>Tomography X-Ray imaging device<br>specialized in diagnosing general<br>dental treatments and orthodontic<br>purpose using Panoramic and<br>Cephalometric images respectively.<br>In addition, DENTRIα series is<br>used in the field of Otolaryngology<br>by capturing 360 degree rotation<br>sequence of the head and neck<br>areas, including the ENT and<br>dentomaxillofacial areas for a<br>dental treatment in adult and<br>pediatric dentistry, and obtains x-<br>ray images from different angles<br>and calculate though computer-<br>processed to produce 3D x-ray<br>tomographic images. The |
| | Proposed Device | Predicate Device #1 | Predicate Device #2 | Predicate Device #3 |
| | | | | DENTRIα series is used by physicians, dentists, and x-ray technologists. |
| Operation Mode | 1) Panorama<br>2) Cephalo<br>• Scan type | 1) Panorama<br>2) Cephalo<br>• One shot type<br>• Scan type | 1) Panorama<br>2) Cephalo (Option) | 1) CT<br>2) Panorama<br>3) Cephalo<br>• One shot type<br>• Scan type |
| X-ray tube assembly | | | | |
| X-ray tube | D-054S | Not known | D-051S | OPX/105 (C.E.I. X-ray tube) |
| Focal spot size | 0.5 mm | 0.5 mm | 0.5 mm | 0.5 mm |
| Target angle | Target angle 5° | Target angle 5° | Target angle 5° | Target angle 5° |
| Total filtration of X-ray tube assembly | > 2.8 mmAl | 2.6 mm Al equivalent | > 2.5 mmAl | > 2.5 mmAl |
| Anode material | Tungsten | Tungsten | Tungsten | Tungsten |
| range of X-ray tube voltage settings | 60 - 90 kV | 60 - 90 kV | 57 - 85 kV | 1) CT: 60 - 110 kV<br>2) Panorama: 60 - 90 kV<br>3) Cephalo<br>• Scan type:60 - 90 kV<br>• One shot type: 60 - 110 kV |
| range of X-ray tube current settings | 4 - 10 mA | 4 - 10mA | 2 - 16 mA | 4 - 10mA |
| range of irradiation time settings | 1) Panorama<br>Max 14.2 s<br><br>2) Cephalo<br>• Scan type:<br>Max 8.2 s (Normal) / 4.2 s (Fast) | 1) Panorama:<br>below 14sec<br><br>2) Cephalo<br>• Scan type: below 18sec<br>• One shot type: below 2sec | 1) Panorama<br>Max 17.6 s<br><br>2) Cephalo<br>8 - 20 s | 1) CT(Normal):<br>8 s or 24 s<br>2) Panorama:<br>Max 14.2 s<br>3) Cephalo<br>• Scan type: 8.2 s and less<br>• One shot type:<br>0.5, 1, 1.5, 2 s |
| Detector type | 1) Panorama<br>• Flat panel (CCD or CMOS) | 1) Panorama<br>• Flat panel | 1) Panorama<br>• Flat panel (CCD) | 1) CT: Flat panel<br>2) Panorama: Flat panel |
| | Proposed Device | Predicate Device #1 | Predicate Device #2 | Predicate Device #3 |
| | 2) Cephalo<br>· Scan type:<br>Flat panel (CCD or CMOS) | 2) Cephalo<br>· Scan type:<br>CdTe Direct flat panel sensor<br>· One shot type: Amorphous<br>Silicon | 2) Cephalo<br>· Scan type:<br>Flat panel (CCD) | 3) Cephalo<br>· Scan type: CCD<br>· One shot type: Flat panel |
| Pixel size | 1) Panorama: 99 µm<br>2) Cephalo:<br>99 µm | 1) Panorama: 100 µm<br>2) Cephalo<br>· Scan type: 100 µm<br>· One shot type: 139 µm | 1) Panorama: 96 x 96 µm<br>2) Cephalo<br>· Scan type: 96 x 96 µm | 1) CT: 100.1 or 127 µm<br>2) Panorama: 100.1 or 127 µm<br>3) Cephalo<br>· Scan type: 27 µm<br>· One shot type: 129 µm |
| Active area (mm) | 1) Pano<br>· 152.0 x 7.0 mm or<br>151.0 x 6.9mm<br>2) Pano/Ceph<br>· Scan type:<br>228.0 x 7.0 mm or<br>221.0 x 6.9 mm | 1) Panorama:<br>· 151.2 x 6.0 mm<br>2) Cephalo<br>· Scan type:<br>240.0 x 4.8 mm<br>· One shot type:<br>427.0 x 356.0 mm or<br>302.0 x 249.0 mm | 1) Panorama:<br>· 138.0 x 138.0 mm/ 1440 pixels<br>2) Cephalo<br>· 190.0 x 190.0 mm / 1980 pixels | 1) CT:<br>· 131 x 131 mm or<br>· 130 x 130 mm<br>2) Panorama:<br>· 6 x 131 mm or<br>· 3.94 x 128.78 mm<br>3) Cephalo<br>· Scan type: 6.9 x 221 mm<br>· One shot type:<br>193 x 259 mm |
| MTF | 1) Panorama:<br>· 70 % at 1LP/mm or<br>65% at 1LP/mm<br>2) Cephalo<br>· Scan type:<br>70 % at 1LP/mm or<br>65% at 1LP/mm | 1) Panorama:<br>· 70 % at 1LP/mm<br>2) Cephalo<br>· Scan type:<br>75 % at 1LP/mm<br>· One shot type:<br>54 % at 1LP/mm | Not known | 1) CT: 57% or 55% at 1LP/mm<br>2) Panorama:<br>· 57% or 55% at 1LP/mm<br>3) Cephalo<br>· Scan type:<br>70% at 1LP/mm<br>· One shot type:<br>83.3% at 2 LP/mm |
| DQE | 1) Panorama:<br>· 57% at 1LP/mm or<br>· 50% at 0LP/mm | 1) Panorama:<br>· 50 % at 1LP/mm<br>2) Cephalo | Not known | 1) CT:<br>· 70% at 0 LP/mm or |
| | Proposed Device | Predicate Device #1 | Predicate Device #2 | Predicate Device #3 |
| | 2) Cephalo<br>• Scan type:<br>57% at 1LP/mm or<br>50% at 0LP/mm | • Scan type:<br>88 % at 1LP/mm<br>• One shot type:<br>20 % at 1LP/mm | | • 58% at 1 LP/mm<br>2) Panorama:<br>• 58% at 1 LP/mm or<br>• 70% at 0 LP/mm<br>3) Cephalo<br>• Scan type:<br>50% at 0 LP/mm<br>• One shot type:<br>38.5% at 0 LP/mm |
| Geometry | | | | |
| Source Image Distance<br>(SID) | 1) Panorama: 535mm<br>2) Cephalo<br>• 1735 mm (Single detector type)<br>• 1729mm (Dual detector type) | Not known | 1) Panorama:<br>19.2 inches / 487 mm<br>2) Cephalo<br>68.7 inches / 1745 mm | 1) CT: 600 mm<br>2) Panorama: 560 mm<br>3) Cephalo:<br>• Scan type:1783 mm<br>• One shot type: 1790 mm |
| Format<br>compatible | DICOM 3.0 Format compatible | DICOM 3.0 Format compatible | DICOM 3.0 Format compatible | DICOM 3.0 Format compatible |
[Table of Comparison of Proposed Device to Predicate Device]
{7}------------------------------------------------
# W HDXWILL
#105, 106, 201, 202, 203, 204, 38, Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu, Cheongju-si, Chungcheongbuk-do, 28161, Korea, Tel: +82-43-710-7318/ Fax: +82-43-710-7312
# K181297 Page 5/8
{8}------------------------------------------------
# W HDXWILL
#105, 106, 201, 202, 203, 204, 38, Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu, Cheongju-si, Chungcheongbuk-do, 28161, Korea, Tel: +82-43-710-7318/ Fax: +82-43-710-7312
# K181297 Page 6/8
{9}------------------------------------------------
# K181297 Page 7/8
### Justification to Support Substantial Equivalence
Although the clectors wed differ with predicate including image proposed device is enough to perform each mode with respect to the finctional aspects. The clinical data shows that the DENTIOII series is effective . Clinical inages were provided as further on addition to the combet system works as intended. As licenses or clinical daggoses of the images, it might be proved that the region of interests. Therefore, these technological differences of detectors used do not raise different questions of safety and effectiveness to diagnosis.
{10}------------------------------------------------
> K181297 Page 8/8
### Non-Clinical Test Summary
The DENTIOIII series complies with international standards for electromagnetic compatibility. The following data were provided in support of the substantial equivalence determination:
- 1) Electrical Safety, Electromagnetic Compatibility and Performance:
The DENTIOIII series comply with the electrical safety and electromagnetic compatibility requirements established by the standards.
- · Electrical Basic Safety and Essential Performance requirements in accordance with IEC 60601-1:2005/AMD1:2012
- · Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2007
- · Radiation Protection In Diagnostic X-Ray Equipment requirements of IEC 60601-1-3:2008/AMD1:2013
- " Dental Extra-Oral X-Ray Equipment requirements of IEC 60601-2-63:2012
- · Acceptance Tests Imaging Performance of Dental X-Ray requirements of IEC 61223-3-4:2000
Bench testing was used to assess whether or not the parameter measured required for describing functionalities related to imaging properties of the dental X-ray device and patient dosage satisfies the designated tolerance. Performance (Imaging performance) testing was conducted according to standard of IEC 61223-3-4. All test results were satisfactory. The manufacturing facility is in conformance with the relevant EPRC standards as specified in 21 CFR 1020.30, 31 and the records are available for review.
- 2) Software Validation
The DENTIOIII series use original software. The DENTIOIII series contains MODERATE level of concern software. Software was designed and developed according to a software development process and was verified and validated. Software information is provided in accordance with FDA guidance: "The content of premarket submissions for software contained in medical devices, on May 11, 2005."
- 3) Biocompatibility
A biocompatibility study is not necessary.
### 4) SSXI Report
Non-clinical performance was conducted for imaging performance of the proposed detector in accordance with FDA Guidance "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices". Both of the predicate devices and proposed device are based on Csi scintillator designed for general radiography. MTF (Modulation Transfer Function) and DQE (Detective Quantum Efficiency) were tested and measured by IEC 6220-1. Base on the Non-clinical test results, even though the used detector differs in term of the pixel size and active area, the diagnostic image quality of the detector is equal or better than those of the predicate device and there is no significant difference in efficiency and safety.
## 8. Conclusion [21 CFR 807.92(b)(3)]
Based on the comparison of intended use, technological characteristics and performance specifications, the DENTIOIII series is safe and effective and substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.