The Catalyst CSR Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint where hemi- or total shoulder arthroplasty is determined by the surgeon to be the preferred method of treatment. The Catalyst CSR Shoulder System is intended for use in patients with the following conditions where the humeral head, neck and glenoid vault are of sufficient bone stock and the rotator cuff is intact or reconstructable. - Osteoarthritis - Avascular Necrosis - Rheumatoid Arthritis - Post-traumatic Arthritis - Correction of functional deformity Both components of the Catalyst CSR Shoulder System are intended for cemented use only.
Device Story
Bone-preserving total shoulder prosthesis; provides alternative to traditional designs requiring extensive bone resection. System components include humeral head implants and glenoid components. Humeral components feature non-spherical polished articulating surfaces and 4-peg design for rotational stability/alignment; back-side geometry matches prepared humeral head surfaces. Used in clinical settings by orthopedic surgeons. Implanted via cemented fixation. Benefits patients by preserving bone stock while restoring joint function. This submission expands the system to include larger, standard humeral component sizes.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including static lever-out testing, endotoxin testing, and size/radial mismatch comparisons on worst-case components.
Technological Characteristics
Humeral components manufactured from cast Co-Cr-Mo alloy (ASTM F75). Features non-spherical polished articulating surface and 4-peg fixation. Designed for cemented use. Bone-preserving geometry with 4-plane back-side interface. Compatible with previously cleared 3-peg glenoid components.
Indications for Use
Indicated for skeletally mature individuals with degenerative glenohumeral joint disease (osteoarthritis, avascular necrosis, rheumatoid arthritis, post-traumatic arthritis, functional deformity) requiring hemi- or total shoulder arthroplasty. Requires sufficient bone stock in humeral head, neck, and glenoid vault, and intact or reconstructable rotator cuff. Cemented use only.
Regulatory Classification
Identification
A shoulder joint metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings,”
(vii) F 1378-97 “Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Catalyst OrthoScience CSR Shoulder System (K152825)
Reference Devices
Biomet Versa-Dial Humeral Head Prosthesis (K060716)
Zimmer Anatomical Shoulder Humeral Head and Anatomical Shoulder Keeled Glenoid (K053274, K051623)
Tornier Aequalis PerFORM+ Shoulder System (K160975)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Catalyst OrthoScience, Inc. Dale Davison Sr. VP of Manufacturing & Product Development 14710 Tamiami Trail North, Suite 102 Naples, Florida 34110
Re: K181287
Trade/Device Name: Catalyst CSR Shoulder System Regulation Number: 21 CFR 888.3650 Regulation Name: Shoulder Joint Metal/Polymer Non-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: KWT, HSD Dated: June 11, 2018 Received: June 12, 2018
July 12, 2018
Dear Dale Davison:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
K181287
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
## K181287 Device Name Catalyst OrthoScience CSR Shoulder System
#### Indications for Use (Describe)
The Catalyst CSR Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint where hemi- or total shoulder arthroplasty is determined by the surgeon to be the preferred method of treatment. The Catalyst CSR Shoulder System is intended for use in patients with the following conditions where the humeral head, neck and glenoid vault are of sufficient bone stock and the rotator cuff is intact or reconstructable.
- Osteoarthritis
- Avascular Necrosis
- Rheumatoid Arthritis ●
- Post-traumatic Arthritis ●
- Correction of functional deformity
Both components of the Catalyst CSR Shoulder System are intended for cemented use only.
Type of Use (Select one or both, as applicable)
区Prescription Use (Part 21 CFR 801 Subpart D)
□Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Form FDA 3881
{3}------------------------------------------------
| K181287<br>7 | |
|--------------|--|
|--------------|--|
| 510(k) Summary | |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prepared: | May 15, 2018 |
| Submitter: | Catalyst OrthoScience, Inc.<br>14710 Tamiami Trail North, Suite 102<br>Naples, FL 34110 |
| Contact: | Dale Davison<br>Sr. VP of Manufacturing & Product Development<br>Catalyst OrthoScience, Inc.<br>1-239-325-9976 ext 102<br>ddavison@catalystortho.com |
| Proprietary Name: | Catalyst CSR Shoulder System |
| Common Name: | Shoulder Prosthesis |
| Classification Names: | 21 CFR 888.3650: Shoulder joint metal/polymer non-<br>constrained cemented prosthesis; Class II<br><br>21 CFR 888.3690: Shoulder joint humeral (hemi-shoulder)<br>metallic uncemented prosthesis; Class II |
| Product Codes: | KWT, HSD |
| Substantially<br>Equivalent Device: | • Catalyst OrthoScience CSR Shoulder System (K152825) |
| Reference Devices: | • Biomet Versa-Dial Humeral Head Prosthesis (K060716)<br>• Zimmer Anatomical Shoulder Humeral Head and<br>Anatomical Shoulder Keeled Glenoid<br>(K053274, K051623)<br>• Tornier Aequalis PerFORM+ Shoulder System<br>(K160975) |
### Device Description:
The Catalyst CSR Shoulder System is a bone preserving total shoulder prosthesis designed for use in patients where the humeral head, neck and glenoid vault are of sufficient bone stock and there is an intact or reconstructable rotator cuff. The design requires minimal bone resection, thus giving the patient an alternative to other total shoulder designs where more bone is removed.
{4}------------------------------------------------
This submission adds larger, standard (non-spherical) humeral components to the CSR Shoulder System. Like the previously cleared CSR humeral components, the Size H and Size I CSR standard humeral components are manufactured from cast Co-Cr-Mo alloy conforming to ASTM F75. They have a non-spherical polished surface for articulation with the glenoid component or the glenoid cavity of the scapula. The humeral components incorporate 4 pegs which assist with alignment and provide rotational stability. The four plane geometry of the back side of the humeral component matches four cut surfaces on the humeral head to recreate the geometry and thickness of the removed bone.
The CSR humeral components are compatible with previously cleared CSR 3 Peg glenoid components.
#### Intended Use / Indications:
The Catalyst CSR Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint where hemi- or total shoulder arthroplasty is determined by the surgeon to be the preferred method of treatment. The Catalyst CSR Shoulder System is intended for use in patients with the following conditions where the humeral head, neck and glenoid vault are of sufficient bone stock and the rotator cuff is intact or reconstructable.
- . Osteoarthritis
- Avascular Necrosis ●
- Rheumatoid Arthritis
- Post-traumatic Arthritis
- Correction of functional deformity
Both components of the Catalyst CSR Shoulder System are intended for cemented use only.
### Summary of Technologies/Substantial Equivalence:
The additional Catalyst CSR humeral components are substantially equivalent to the predicate devices in regards to intended use and indications, material and design. The larger sizes do not raise new types of safety and effectiveness questions, nor are there new technological issues.
### Non-Clinical Testing:
Static lever-out testing, endotoxin testing, size comparison and radial mismatch comparisons were performed on worst case CSR humeral components. The results indicate that the performance of the Catalyst CSR humeral components is adequate for their intended use.
#### Clinical Testing:
Clinical testing was not necessary to demonstrate substantial equivalence of the new Catalyst CSR humeral components to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.