K181279 · Siemens Medi Cal Solutions, Inc. · KPR · Jun 13, 2018 · Radiology
Device Facts
Record ID
K181279
Device Name
Ysio Max
Applicant
Siemens Medi Cal Solutions, Inc.
Product Code
KPR · Radiology
Decision Date
Jun 13, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Ysio Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image. Ysio Max enables radiographic and tomographic exposures of the whole body and may be used on pediatric, adult and bariatric patients. It can also be used for emergency applications. Ysio Max is not for mammography examinations.
Device Story
Stationary X-ray system for radiography; utilizes X-ray generator, tube, collimator, and ceiling rail system. Inputs: X-ray patterns captured by Solid State X-ray Imagers (SSXIs) (mobile or fixed in Bucky tray). Transformation: syngo FLC system processes signals into visible images using Diamond View Plus algorithms. Output: Diagnostic radiographic/tomographic images. Used in hospitals/clinics by healthcare professionals. Software version VF10 introduces Windows 10, enhanced cybersecurity, and expanded pediatric programs. System monitors performance; blocks functions and displays error messages upon failure. Benefits: Enables diagnostic imaging for diverse patient populations, including trauma/critical care, with improved IT integration and cybersecurity.
Clinical Evidence
No clinical data provided. Bench testing only. Performance verified via technical comparison of SSXI detectors, software verification/validation, and compliance with IEC/ISO standards (e.g., IEC 60601-1, 60601-1-2, 62304).
Technological Characteristics
Stationary X-ray system. Components: Polydoros 65/80 kW generator, OPTITOP 150/40/80/HC-100 tube, Digital Multileaf Collimator N. Detectors: Trixell Pixium a-Si/CsI SSXIs. Connectivity: DICOM, Hospital/Radiology Information System integration. Software: syngo FLC (VF10) on Windows 10. Standards: IEC 60601-1 (3.1), IEC 60601-1-2 (4th ed), IEC 62304, ISO 14971.
Indications for Use
Indicated for visualization of anatomical structures via X-ray imaging. Used for radiographic and tomographic exposures of the whole body in pediatric, adult, and bariatric patients, including emergency applications. Contraindicated for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 13, 2018
Siemens Medical Solutions, Inc. % Ms. Denise Adams Regulatory Affairs Specialist 40 Liberty Boulevard 65-1A MALVERN PA 19355
Re: K181279
Trade/Device Name: Ysio Max Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: May 14, 2018 Received: May 15, 2018
Dear Ms. Adams:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
K181279
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Hol 2. Mild
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K181279
Device Name
Ysio Max
Indications for Use (Describe)
Y sio Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image. Ysio Max enables radiographic and tomographic exposures of the whole body and may be used on pediatric, adult and bariatric patients. It can also be used for emergency applications.
Ysio Max is not for mammography examinations.
Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, tor
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid QMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots.
# Special 510(k) Summary: Ysio Max
Company:
Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355
Date Prepared: June 12, 2018
This 510(k) summary of safety and effectiveness information is being submitted in accordance with 21 CFR § 807.92.
## 1. General Information:
Importer / Distributor: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Establishment Registration Number: 2240869
Location of Manufacturing Site:
Siemens Healthcare GmbH Siemensstr. 1 91301 Forchheim, Germany Establishment Registration Number: 3004977335
# 2. Contact Person:
Denise Adams, RAC Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65 - 1A Malvern, PA 19355 adams.denise@siemens-healthineers.com
# Alternate Contact Person:
Patricia D. Jones Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65 - 1A Malvern, PA 19355 patricia.d.jones@siemens-healthineers.com
#### 3. Device Name and Classification:
Trade Name: Ysio Max
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots.
Classification Name: Stationary x-ray System Classification Panel: Radiology Classification Regulation: 21 CFR §892.1680 Device Class: II Product Codes: KPR
## 4. Legally Marketed Predicate Device:
Trade Name: Ysio Max 510(k) Number: K133259 Device Classification Name: Stationary x-ray System Regulation Medical Specialty: Radiology Review Panel: Radiology Product Code: KPR Submission Type: Special 510(k) Regulation Number: 892.1680 Device Class: II
## 5. Device Description:
The Ysio Max is a stationary X-ray system for radiography. The main components are the X-ray generator, the X-ray tube with collimator supported by a ceiling rail system. The Ysio Max comes with SSXIs (Solid State X-ray Imagers) that can be mobile or fixed in the Bucky tray. An image processing and data management system (syngo FLC) completes the radiographic suite.
The purpose of this submission is the upgrade to a new software version VF10 and minor hardware changes. The modified Yisio Max will introduce the following new features:
- The operating system will be MS Windows 10 ●
- New cybersecurity features
- Additional pediatric programs ●
- Implementing a "Virtual Machine" that supports hospital IT ●
- The SSXIs have been updated ●
- The EMC (Electromagnetic Compatibility was tested according to the IEC . 4th edition)
- . The image processing algorithms (Diamond view Plus) will be used for exposures without grid.
#### 6. Indications for Use:
Y sio Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image. Ysio Max enables radiographic and tomographic exposures of the whole body and may be used on pediatric, adult and bariatric patients. It can also be used for emergency applications.
Ysio Max is not for mammography examinations.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a cluster of orange dots.
#### 7. Substantial Equivalence:
The new system software VF10 does not alter the intended use or the fundamental scientific technology. The device remains within the same classification regulation for the same intended use as the predicate device. The new system software design was completed in accordance with Siemens Quality Management System Design Controls comparable to the processes available for the predicate device. The scope of internationally recognized standards compliance was updated to the standards recognized at the time of the design of the new software. Verification and Validation testing was the same or similar to the testing being used with the predicate device.
The new system software version VF10 controls solid state X-ray imagers (SSXI) similar to the imagers being used with the predicate device. The updated MAX wi-D and Max mini detectors have been previously cleared in K173639.
- 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device:
The subject device with the new software version VF10 uses the same X-ray generator, the same X-ray tube and similar SSXIs with the same digital imaging system and similar image processing software as the predicate device. There are no changes in the patient environment or the type of user interface.
The following tables compare the main performance data of the subject device with the predicate device to substantiate equivalence.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
| | Subject Device | Predicate Device | |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Feature | Ysio Max | Ysio Max<br>K133259 | Comparis<br>on Results |
| Indications for Use Comparison | | | |
| Indications<br>for Use<br>statement | Ysio Max is a device<br>intended to visualize<br>anatomical structures<br>by converting an X-<br>ray pattern into a<br>visible image. Ysio<br>Max enables<br>radiographic and<br>tomographic<br>exposures of the<br>whole body and may<br>be used on pediatric,<br>adult and bariatric<br>patients. It can also be<br>used for emergency<br>applications.<br>Ysio Max is not<br>for<br>mammography<br>examinations. | The Ysio Max is a radio-<br>graphic system used in<br>hospitals, clinics, and<br>medical practices. Ysio<br>Max enables radiographic<br>and tomographic expo-<br>sures of the whole body<br>including: skull, chest,<br>abdomen, and extremities<br>and may be used on<br>pediatric, adult and<br>bariatric patients. It can<br>also be used for intrave-<br>nous, small interventions<br>(like biopsy, punctures,<br>etc.) and emergency<br>(trauma, critical ill)<br>applications.<br>Exposures may be taken<br>with the patient sitting,<br>standing, or in the prone<br>position.<br>The Ysio Max system is<br>not meant for<br>mammography.<br>The Ysio Max uses<br>integrated or portable<br>digital detectors for<br>generating diagnostic<br>images by converting x-<br>rays into electronic<br>signals. Ysio Max is also<br>designed to be used with<br>conventional film/screen<br>or Computed Radiography | Intended<br>use is the<br>same.<br>Indications<br>for use<br>have been<br>reworded<br>to<br>simplify. |
# Table 1: Comparison of the Subject Device Ysio Max to the Predicate
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots.
| Technical Comparison | | | |
|-------------------------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| X-ray | | | |
| Generator | Polydoros<br>65/80 kW | Polydoros<br>65/80 kW | |
| X-ray tube | OPTITOP<br>150/40/80/HC-100 | OPTITOP<br>150/40/80/HC-100 | Same |
| Collimator | Digital Multileaf<br>Collimator N | Digital Multileaf<br>Collimator N | |
| Air kerma | Kerma X | Kerma X | |
| X-ray techniques | Radiography | Radiography | |
| Organ programs | X-ray parameters<br>Imaging processing<br>parameters | X-ray parameters<br>Imaging processing<br>parameters | |
| Testing | | | |
| IEC Compliance | IEC 4th edition for<br>EMC testing | IEC 3rd edition for<br>EMC testing | Testing according to<br>current IEC test scope |
| Digital Imaging | | | |
| SSXI for Rad<br>imaging | Trixell Pixium<br>MAX wi-D =<br>3543EZh<br>MAX mini = 2430EZ<br>MAX Static =<br>4343RC | Trixell Pixium<br>MAX wi-D =<br>3543EZh<br>MAX mini = 2430EZ<br>MAX Static =<br>4343RC | Image Chain<br>is similar to predicate<br>device<br>Detector Performance<br>data according To<br>SSXI guidance are<br>provided |
| Digital imaging<br>system | Fluorospot Compact<br>aka syngo FLC | Fluorospot Compact | |
| Image processing | Diamond View Plus | Diamond View Plus | Same processing<br>made user friendly |
| MS Operating<br>system | Windows 10 | Windows 7 | New Operating<br>system |
| Cybersecurity | Security package<br>based on MS Win 10 | Security package<br>based on MS Win 7 | Improved |
| Integration of<br>hospital IT | Hospital/Radiology<br>Information System | Hospital/Radiology<br>Information System | Improved |
| Pediatric package | Increased number of<br>organ programs | Organ programs | Increased number and<br>variability |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is an orange graphic of several circles.
| Table 2: Comparison of Trixell Pixium 4343RCE to the predicate Pixium 4343RC | |
|------------------------------------------------------------------------------|--|
| (MAX static) | |
| Technical<br>Specifications | Trixell Pixium<br>4343RCE detector<br>(MAX static) (fixed)<br>(subject) | Trixell Pixium<br>4343RC detector<br>(fixed) (predicate) | Comparison<br>Results |
|-----------------------------|-------------------------------------------------------------------------|----------------------------------------------------------|-------------------------------|
| Dimensions | 423.3mm x 425.4 mm | 424.6 mm x 425.5 mm | Difference not<br>significant |
| Resolution | 2860 x 2874 pixels | 2869 x 2874 pixels | Difference not<br>significant |
| Pixel size | 148 um | 148 um | Same |
| Semiconductor<br>Material | Amorphous silicon,<br>a-Si | Amorphous silicon,<br>a-Si | Same |
| Scintillator | Cesium iodide (CsI) | Cesium iodide (CsI) | Same |
| Acquisition<br>denth | 16 bit | 16 bit | Same |
| DQE (Detective | DQE @ 0.05 lp/mm (2<br>uGy), 67% | DQE @ 0.05 lp/mm (2<br>uGy), 65% | Difference not<br>significant |
| nlations | MTF @ 1 lp/mm, 62% | MTF @ 1 lp/mm, 63% | Difference not<br>significant |
| Table 3: Comparison of Trixell Pixium 3543 EZh to the predicate Pixium 3543EZh | |
|--------------------------------------------------------------------------------|--|
| (MAX wi-D) | |
| Technical<br>Specifications | Trixell Pixium 3543<br>EZh detector<br>(wireless)(subject) | Trixell Pixium 3543<br>EZh detector (MAX<br>wi-D) (wireless)<br>(predicate) | Comparison<br>Results |
|----------------------------------------------|------------------------------------------------------------|-----------------------------------------------------------------------------|-------------------------------|
| Dimensions | 348 mm x 424 mm | 349 mm x 425 mm | Difference not<br>significant |
| Resolution | 2350 x 2866 pixels | 2356 x 2872 pixels | Difference not<br>significant |
| Pixel size | 148 µm | 148 µm | Same |
| Semiconductor<br>Material | Amorphous silicon,<br>a-Si | Amorphous silicon,<br>a-Si | Same |
| Scintillator | Cesium iodide (CsI) | Cesium iodide (CsI) | Same |
| Acquisition depth | 16 bit | 16 bit | Same |
| DQE (Detective<br>Quantum<br>Efficiency) | DQE @ 1 lp/mm (2<br>µGy), 51% | DQE @ 1 lp/mm (2<br>µGy), 50% | Difference not<br>significant |
| MTF<br>(Modulations<br>transfer<br>function) | MTF @<br>1 lp/mm,63% | MTF @<br>1 lp/mm,61% | Difference not<br>significant |
Siemens Medical Solutions USA, Ino.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a graphic of orange dots. The logo is simple and modern.
| Table 4: Comparison of Trixell Pixium 2430 EZ to the predicate Pixium 2430 EZ | |
|-------------------------------------------------------------------------------|--|
| (MAX mini) | |
| Technical<br>Specifications | Trixell Pixium<br>2430<br>EZ detector<br>(wireless)<br>(subject) | Trixell<br>Pixium 2430<br>EZ detector<br>(MAX mini)<br>(wireless)<br>(predicate) | Comparison<br>Results |
|----------------------------------------------|------------------------------------------------------------------|----------------------------------------------------------------------------------|-------------------------------|
| Dimensions | 225 mm x 284<br>mm | 228 mm x 284<br>mm | Difference not<br>significant |
| Resolution | 1520 x 1920<br>pixels | 1560 x 1920<br>pixels | Difference not<br>significant |
| Pixel size | 148 µm | 148 µm | Same |
| Semiconductor<br>Material | Amorphous<br>silicon, a-Si | Amorphous<br>silicon, a-Si | Same |
| Scintillator | Cesium iodide<br>(CsI) | Cesium iodide<br>(CsI) | Same |
| Acquisition<br>depth | 16 bit | 16 bit | Same |
| DQE<br>(Detective<br>Quantum<br>Efficiency) | DQE @ 1<br>lp/mm (2 µGy),<br>50% | DQE @ 1<br>lp/mm (2 µGy),<br>50% | Same |
| MTF<br>(Modulations<br>transfer<br>function) | MTF @ 1<br>lp/mm,61% | MTF @ 1<br>lp/mm,61% | Same |
# Summary of Non-Clinical Test Data:
The software VF10 design was completed in accordance with Siemens Quality Management System Design Controls and verification and validation testing were successfully conducted. Non-clinical performance data are provided for the updated detectors. The new IEC 4th edition has been applied for the compliance to the IEC 60601-1-2 electromagnetic compatibility standard.
The tests were performed on the new software VF10 demonstrating that the device is safe and effective, performs comparably to the predicate device, and is substantially equivalent to the predicate device.
Documentation provided demonstrates compliance of the subject device to all FDA requirements stated in the Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, including results of verification/validation tests of software requirements and software risk hazards.
Performance testing confirmed that the five devices comply with 21 CFR 1020.30-31 Federal Performance Standards for X-Ray Fluoroscopic equipment
The software VF10 controls solid state X-ray imagers (SSXI) that are similar in performance to the predicates. Performance data are provided according to the "Guidance
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a cluster of orange dots of varying sizes.
for the Submission of 510(k)'s for Solid State X-ray Imaging Devices, issued on: September 1, 2016''. X-ray geometry and techniques are the same so that clinical image comparisons involving SSXI were not conducted.
Together, the verification/validation activities successfully confirmed that the software requirements have been fulfilled and that system functionality is consistent with the user needs and intended uses. The VF10 software correctly performs as designed and raises no new questions regarding safety or effectiveness. Therefore, when compared to the predicate device the Ysio Max with the new software VF10 support a determination of substantial equivalence to the predicate device.
## 9. Summary of Non-Clinical Tests
The devices operating with software VF10 comply with the voluntary standards as listed in the following table:
| Development<br>Organization and<br>Reference Number | Title of Standard |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1:2012,<br>Edition 3.1 | Medical Electrical Equipment - Part 1: General<br>Requirements for Safety |
| IEC 60601-1-2:2007<br>Edition 4.0 | Medical Electrical Equipment - Part 1-2: General<br>requirements for basic safety and essential<br>performance - Collateral Standard: Electromagnetic<br>Compatibility -- Requirements and Tests |
| IEC 60601-1-3: 2012,<br>Edition 2.1 | Medical electrical equipment - Part 1-3: General<br>requirements for basic safety and essential<br>performance - Collateral Standard: Radiation<br>protection in diagnostic X-ray equipment |
| IEC 62366:2014<br>Edition 1.1 | Medical devices – Application of usability<br>engineering to medical devices |
| ISO 14971, 2007 | Medical devices – application of risk management<br>to medical devices |
| IEC 62304 Ed. 1.0,<br>2006 | Medical device software - Software life cycle<br>processes |
| IEC 60601-2-28:2010<br>Edition 2.0 | Medical electrical equipment - Part 2-28: Particular<br>requirements for the basic safety and essential<br>performance of X-ray tube assemblies for medical<br>diagnosis |
| IEC 60601-2-54, 2009<br>Edition 1.0 | Medical electrical equipment - Part 2-54: Particular<br>requirements for the basic safety and essential<br>performance of X-ray equipment for radiography<br>and radioscopy |
| IEC 61910-1, 2014 | Medical electrical equipment - Radiation dose<br>documentation - Part 1: Radiation dose structured<br>reports for radiography and radioscopy (IEC 61910-<br>1:2014) |
#### Table 5: Standards
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots arranged in a circular pattern.
| NEMA PS 3.1 - 3.20,<br>2011 | Digital Imaging and Communications in Medicine<br>(DICOM) Set |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------|
| ISO 10993-1, 2009 | Biological evaluation of medical devices -- Part 1:<br>Evaluation and testing within a risk management<br>process |
Documentation provided demonstrates compliance of the subject devices to all FDA requirements stated in the Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, including results of verification/validation tests of software requirements and software risk hazards.
Performance testing confirmed that the Ysio Max with VF10 complies with 21 CFR 1020.30-31 Federal Performance Standards for X-Ray equipment. The applicable sections include:
| 1020.30(c) | Manufacturer's Responsibility (Certification) |
|-------------------------|--------------------------------------------------------------------------------------------------|
| 1020.30(e) | Identification of X-ray components |
| 1020.30(g) | Information to be provided to assemblers |
| 1020.30(h) | Information to be provided to users |
| 1020.31(k) | Leakage Radiation |
| 1020.30(m) | Beam Quality |
| 1020.31(a) | Peak Tube Potential |
| 1020.31 (d)(2)(i, iii) | Alignment of visually defined X-ray fields and contrast<br>of light defined X-ray field. |
| 1020.31 (e)(1)(i) | Alignment of the center of the radiographic X-ray field<br>with the center of the image receptor |
| 1020.31 (e)(1)(ii, iii) | Radiography X-ray field size and image receptor size |
The software VF10 controls solid state X-ray imagers (SSXI) that are similar in performance to the predicates. Performance data are provided according to the "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices, issued on: September 1, 2016".
#### 10. Summary of Clinical Tests:
For the subject of this premarket submission, Siemens did not do an evaluation of the clinical image quality as X-ray technology; geometry and SSXI changes are minor.
#### 11. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner. Several safety features including visual and audible warnings are incorporated into the system design. In addition the devices are continually monitored, and if an error occurs, the system functions will be blocked and an error message will be displayed.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a cluster of orange dots of varying sizes.
Furthermore the operators are health care professionals familiar with and responsible for the X-ray examinations to be performed. To minimize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing.
## 12. Conclusion as to Substantial Equivalence:
The Ysio Max with VF10 software has the same intended use as the predicate device. The operating environment is the same and the technology differences do not affect safety and effectiveness. Siemens concludes via the documentation provided in this 510(k) submission that the radiology device Ysio Max with VF10 is substantially equivalent to the predicate device Ysio Max.
## 13. Guidance documents
The following FDA guidance documents were utilized in the documentation of this Premarket Notification:
- Guidance for Industry and FDA Staff ●
Bundling Multiple Devices or Multiple Indications in a Single Submission Document issued on: June 22, 2007
- Content of Premarket Submissions for Management of Cybersecurity in . Medical Devices
Guidance for Industry and Food and Drug Administration Staff Document Issued on: October 2, 2014
- Information to Support a Claim of Electromagnetic Compatibility ● (EMC) of Electrically-Powered Medical Devices
Guidance for Industry and Food and Drug Administration Staff Document issued on July 11, 2016. The draft of this document was issued on November 2, 2015.
Pediatric Information for X-ray ● Imaging Device Premarket Notifications Guidance for Industry and Food and Drug Administration Staff Document issued on November 28, 2017.
. Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices 1 Guidance for Industry and Food and Drug Administration Staff Document issued on: September 1, 2016
> ● Guidance for Industry and FDA Staff
Guidance for the Content of Premarket Submissions for Software contained in Medical Devices, Document issued on: May 11, 2005
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a cluster of orange dots.
Guidance for Industry and FDA Staff . Recognition and Use of Consensus Standards Document issued on: September 17, 2007
- The 510(k) Program: Evaluating Substantial Equivalence in Premarket . Notifications [510(k)]
Guidance for Industry and Food and Drug Administration Staff Document issued on: July 28, 2014
Radio Frequency Wireless Technology in Medical Devices . Guidance for Industry and Food and Drug Administration Staff Document issued on: August 14, 2013
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.