The intended use of the Connected OR Hub with Device and Voice Control system is to allow for voice control and remote control of medical device settings by the surgeons or operating room personnel, thereby eliminating the need of manual operation of those devices compatible with the Connected OR Hub, or relying upon verbal communications between the surgeon and other personnel in the operating room in order to adjust the surgical equipment. It also has an additional digital documentation functionality to electronically capture, transfer, store and display medical device data (Class I device function), which is independent of the functions or parameters of any attached device.
Device Story
ConnectedOR Hub acts as a centralized control console for compatible endoscopic and general surgery equipment. Input includes voice commands via headset/base station and manual inputs via handheld IR remote. Device transforms these inputs into control signals to adjust settings of connected surgical devices, eliminating manual adjustment or verbal communication between OR staff. System also performs Class I documentation functions: capturing, transferring, storing, and displaying medical device data independently of connected device parameters. Used in ORs by surgeons and clinical staff. Output allows surgeons to manage equipment settings directly, improving workflow efficiency and reducing reliance on OR personnel. Hardware/software upgrades support higher processing speeds and video resolution compared to predicate.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including device/voice control performance, wireless coexistence (AAMI TIR 69:2017, ANSI C63.27:2017), electrical safety (ANSI/AAMI ES60601-1), EMC (IEC 60601-1-2), usability (IEC 62366-1), and software validation (IEC 62304).
Technological Characteristics
Console-based system with IR remote and voice control headset. Upgraded internal hardware and OS for higher processing/video resolution. Connectivity includes wireless device control. Electrical safety per ANSI/AAMI ES60601-1; EMC per IEC 60601-1-2. No patient-contacting materials.
Indications for Use
Indicated for use with compatible endoscopic and general surgery devices in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery wherever a laparoscope or arthroscope is indicated. Patient population includes those undergoing procedures such as cholecystectomy, hernia repair, appendectomy, pelvic lymph node dissection, hysterectomy, spinal fusion, ACL reconstruction, knee/shoulder/small joint arthroscopy, decompression fixation, wedge resection, lung/pleural biopsy, dorsal sympathectomy, pleurodesis, coronary artery bypass, and cardiac valve replacement. Users include general, gynecological, cardiac, thoracic, plastic, ENT, orthopedic, and urological surgeons.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
Stryker SDC3 HD Information Management System with Wireless Device Control Capability (K160332)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
November 5, 2018
Stryker Jay Nayar Staff Regulatory Affairs Specialist 5900 Optical Court San Jose, California 95138
Re: K181258
Trade/Device Name: ConnectedOR Hub with Device and Voice Control Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ, HRX Dated: September 26, 2018 Received: September 28, 2018
Dear Jay Nayar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/7 description: The image shows a digital signature. The signature is from Long H. Chen -S. The date of the signature is 2018.11.05, and the time is 14:30:52 -05'00'.
Date: 2018.11.05
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181258
#### Device Name
ConnectedOR Hub with Device and Voice Control
#### Indications for Use (Describe)
The ConnectedOR Hub is indicated for use with compatible endoscopic and general surgery devices. The ConnectedOR Hub can be used in general laparoscopy, nasopharyngoscopy, sinuscopy, and plastic surgery wherever a laparoscope, endoscope, or an arthroscope is indicated for use. A few examples of the more common endoscopic surgeries are laparoscopic cholecystectomy, laparoscopic hernia repair, laparoscopic appendectomy, laparoscopic pelvic lymph node dissection, laparoscopically assisted hysterectory, laparoscopic anterior spinal fusion, anterior cruciate ligament reconstruction, knee arthroscopy, small joint arthroscopy, decompression fixation, wedge resection, lung biopsy, pleural biopsy, dorsal sympathectomy, pleurodesis, internal mammary artery dissection for coronary artery bypass, coronary artery bypass grafting where endoscopic visualization is indicated and examination of the evacuated cardiac chamber during performance of valve replacement. ConnectedOR Hub users are general surgeons, gynecologists, cardiac surgeons, thoracic surgeons, orthopedic surgeons, ENT surgeons, and urologists.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are black and the background is white. The word is written in all lowercase letters.
# Section 5. 510(k) Summary
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(c).
#### General Information:
| 510(k) Sponsor: | Stryker Endoscopy |
|-------------------------|------------------------------------------------------------------------------------------------------------|
| Address | 5900 Optical Court<br>San Jose, CA 95138 |
| FDA Registration Number | 2936485 |
| Contact Information | April Malmborg<br>Director, Regulatory Affairs<br>Phone: 408-754-2473<br>Email: april.malmborg@stryker.com |
| Date Prepared | 11/02/2018 |
#### Device Identification:
#### Proposed Device:
| Trade Name | Connected OR Hub with Device and Voice Control |
|-------------------------|------------------------------------------------|
| Common Name | SDC 4K Information Management System |
| Regulatory Class | II |
| Classification Name | Laparoscope, General and Plastic Surgery |
| Regulation Number | 21 CFR 876.1500 |
| Product Code | GCJ |
| Subsequent Product Code | HRX |
#### Predicate Device:
| Trade Name | Stryker SDC3 HD Information Management System with<br>Wireless Device Control Capability |
|---------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Common Name | SDC3 HD Information Management System |
| Regulatory Class | II |
| Classification Name | Laparoscope, General and Plastic Surgery |
| Regulation Number | 21 CFR 876.1500 |
| Product Code | GCJ |
| Subsequent Product Code | HRX |
| This device has not been subjected to a design-related recall | |
{4}------------------------------------------------
#### Device Description:
The Connected OR Hub with Device and Voice Control is a medical device that allows the surgeon to control the state, selection, and settings of any compatible device attached to it. It also has operating room documentation functionalities (Class I device function) to electronically capture, transfer, store and display medical device data independently of the functions or parameters of the connected medical device.
The Connected OR Hub with Device and Voice Control (also referred to as proposed device in the following sections) consists of the following:
- a. A Connected OR Hub console
- b. A Device control package (contains an optional software upgrade and a handheld Infrared (IR) remote control)
- c. A Voice control package (contains an optional software upgrade and a headset and base station)
The proposed device is the latest iteration of the Stryker SDC3 HD Information Management System with wireless device control (hereafter referred to as "predicate device", cleared under K160332). The device and voice control features of the proposed device are identical to the predicate device. The operating room documentation functionalities of electronically capturing, transferring, storing and displaying medical device data (Class I device functions) independently of the functions or parameters of any connected medical device remain similar to the predicate device. The proposed device accesses existing controls within each device and is secondary to the built-in control interface that is already on each device which is identical to the predicate device
Although the device and voice control characteristics of the proposed device remain the same as compared to the predicate device, the operating system and the hardware were altered significantly leading to the need for a traditional submission per the FDA guidance Deciding When to Submit a 510(k) for a Change to an Existing Device and Deciding When to Submit a 510(k) for a Software Change to an Existing Device. This submission hence focuses on providing data regarding substantial equivalence of the proposed device (Stryker Connected OR Hub with Device and Voice Control) to the predicate device (Stryker SDC3 HD Information Management System with wireless device and voice control)
#### Indications for Use:
#### This submission is not proposing any changes from the indications cleared for the predicate device.
The Connected OR Hub is indicated for use with compatible endoscopic and general surgery devices. The Connected OR Hub can be used in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery wherever a laparoscope, or an arthroscope is indicated for use. A few examples of the more common endoscopic surgeries are laparoscopic cholecystectomy, laparoscopic hernia repair, laparoscopic appendectomy, laparoscopic pelvic
{5}------------------------------------------------
lymph node dissection, laparoscopically assisted hysterectomy, laparoscopic and thorascopic anterior spinal fusion, anterior cruciate ligament reconstruction, knee arthroscopy, shoulder arthroscopy, small joint arthroscopy, decompression fixation, wedge resection, lung biopsy, pleural biopsy, dorsal sympathectomy, pleurodesis, internal mammary artery dissection for coronary artery bypass, coronary artery bypass grafting where endoscopic visualization is indicated and examination of the evacuated cardiac chamber during performance of valve replacement. Connected OR Hub users are general surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons, and urologists.
### Intended Use:
# This submission is not proposing any changes from the intended uses cleared for the predicate device.
The intended use of the Connected OR Hub with Device and Voice Control system is to allow for voice control and remote control of medical device settings by the surgeons or operating room personnel, thereby eliminating the need of manual operation of those devices compatible with the Connected OR Hub, or relying upon verbal communications between the surgeon and other personnel in the operating room in order to adjust the surgical equipment. It also has an additional digital documentation functionality to electronically capture, transfer, store and display medical device data (Class I device function), which is independent of the functions or parameters of any attached device.
#### Comparison of Technological Characteristics with the Predicate Device:
The Connected OR Hub with Device and Voice Control has the same technological characteristics and design as the predicate with the exception of the software operating system and internal hardware components. The software and hardware were upgraded to support product sustainment, higher processing speeds and higher video resolution. All other technological characteristics of the proposed device are the predicate. The proposed device uses the same device control communication protocols as the predicate device, employs the same voice recognition software engine and controls the same types of connected devices as the predicate device. The Connected OR Hub has similar technological characteristics as the predicate device in the following areas:
- Operating principles ●
- Software architecture ●
- Electrical characteristics
- Mechanical characteristics
- Communication characteristics
- Performance characteristics ●
- . Compatibility with controllable devices as listed in the product labeling
- Energy source
- Materials (no patient contacting materials) ●
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "Stryker" in a bold, sans-serif font. The letters are black and the background is white. The word is the logo for Stryker Corporation, a medical technology company.
A complete comparison of technological characteristics between the proposed and predicate devices is provided in Section 14- Substantial Equivalence Discussion.
# Summary of Non-Clinical Testing Performed:
| Test | Method | Results |
|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Device and<br>Voice Control<br>performance | Verified if the performance of device and voice control<br>functionalities with compatible devices is in accordance with<br>device performance specifications. | Pass |
| Compatibility<br>with Connected<br>OR Spoke | Verified if proposed device was compatible with Connected<br>OR Spoke in accordance with device performance<br>specification. | Pass |
| Environmental<br>performance | The functionality of the product was verified for the specified<br>operating temperature and humidity range. | Pass |
| Video<br>Compatibility | Verified if the Connected OR Hub is compatible with<br>supported Cameras, Monitors, and peripherals that the Stryker<br>customers use in the field. | Pass |
| Memory<br>Profiling and<br>Performance | Verified if the performance and system response times of the<br>Class II Medical Device functionality (Voice and Device<br>Control functionality) while the Connected OR Hub's Class I<br>documentation functionalities is in use are in accordance with<br>device performance specifications. | Pass |
| Voice<br>Recognition<br>Performance | Verified if the voice recognition performance of the proposed<br>system when voice commands are issued to the system in<br>accordance with device performance specification. | Pass |
| Voice Accuracy<br>Benchmark | Verified if the voice recognition accuracy of the proposed<br>system when voice commands are issued to the proposed<br>system are equivalent or better to the predicate system. | Pass |
| Wireless<br>technology<br>performance | Verified the performance and coexistence of the wireless<br>technology in the proposed device in accordance with AAMI<br>TIR 69:2017, ANSI C63.27:2017 and FDA guidance<br>document Radio Frequency Wireless Technology in Medical<br>Devices. | Pass |
| Electrical Safety | In accordance with ANSI/AAMI ES60601-1:2005/(R)2012<br>and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012<br>(Consolidated Text) | Pass |
| EMC Testing | In accordance with IEC60601-1-2 Edition 4.0 2014-02 | Pass |
| Usability | Product was designed and tested per IEC 60601-1-6 Edition<br>3.1 2013-10 and ANSI/AAMI/IEC 62366-1:2015 | Pass |
| System<br>validation and<br>simulated user<br>testing | Validated system performance in accordance with defined user<br>needs and device performance specifications | Pass |
| Software<br>Validation &<br>Verification | In accordance with IEC 62304 and FDA Guidance Document<br>Principles of Software Validation | Pass |
| Test | Method | Results |
| Cybersecurity<br>verification and<br>Network scan | In accordance with device performance specifications, cybersecurity risk assessment and FDA guidance documents -<br><i>Content of Premarket Submissions for Management of<br/>Cybersecurity in Medical Devices</i> | Pass |
{7}------------------------------------------------
# sti
NOTE: The Connected OR Hub with Device and Voice Control does not require clinical studies to support the determination of substantial equivalence.
## Conclusion:
Based upon the indications for use, intended use, technological characteristic, performance testing and comparison to the predicate device, the Stryker Connected OR Hub with Device and Voice Control raises no new questions of safety and effectiveness as compared to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.