Infrared Thermometer DET series are intended for the intermittent measurement of human body temperature by people of all ages. The devices are reusable for home use only.
Device Story
Hand-held, battery-powered infrared thermometers (models DET-205, DET-206, DET-305, DET-306, DET-215) measure human body temperature via auditory canal or forehead. Operation involves detecting natural thermal radiation from the tympanic membrane or forehead using a thermopile sensor with an integrated thermistor for target readings and a secondary thermistor for ambient temperature compensation. Devices include contact (DET-205, DET-206) and non-contact (DET-305, DET-306) forehead modes, and a multifunction (DET-215) ear/forehead mode. A microcontroller unit processes sensor inputs and triggers a voice integrated circuit to broadcast temperature results in English, which are also displayed on an LCD. Used in home environments by patients or caregivers. Output provides immediate temperature feedback to assist in monitoring health status.
Clinical Evidence
Clinical testing performed on 165 subjects across three age groups: infants (newborn to 1 year), children (>1 to 5 years), and adults (>5 years). Testing conducted per ASTM E1965-98 (2016) requirements. Results demonstrated compliance with clinical performance requirements for infrared thermometers.
Technological Characteristics
Materials: ABS enclosure, ABS/glass probe. Sensing: Thermopile sensor with integrated thermistor. Energy: DC 3V (2x AAA batteries). Form factor: Hand-held. Connectivity: None. Sterilization: Not applicable. Software: Microcontroller-based firmware for signal processing, voice synthesis, and LCD display. Standards: ISO 80601-2-56, IEC 60601-1, IEC 60601-1-2, ASTM E1965-98, ISO 10993-5/10.
Indications for Use
Indicated for intermittent measurement of human body temperature in people of all ages, including infants, children, and adults. For home use only.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Non-contact Clinical Thermometer, Model THB0F (K121428)
RII Multi-function Infrared Thermometer, Model TH52Z (K162083)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services logo. The logo on the right is the FDA logo, which includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 29, 2018
JOYTECH Healthcare Co., Ltd. Yunhua Ren General Manager No. 365 Wuzhou Road, Yuhang Economic Development Zone Hangzhou City Zhejiang, 311100 China
Re: K181239
Trade/Device Name: Infrared Forehead Thermometer, Infrared Ear/Forehead Thermometer; Models: DET-205, DET-206, DET-305, DET-306, DET-215 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: August 29, 2018 Received: September 28, 2018
Dear Yunhua Ren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181239
Device Name
Infrared Thermometers DET-205,DET-206,DET-305,DET-306,DET-215
Indications for Use (Describe)
Infrared Thermometer DET series are intended for the intermittent measurement of human body temperature by people of all ages. The devices are reusable for home use only.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
The assigned 510(k) number is: K181239
#### 1. Date Prepared: 2018.10.25
#### 2. Submitter's Identification:
Name: JOYTECH Healthcare Co.,Ltd. Add.: No,365,Wuzhou Road,Yuhang Economic Development Zone, Hangzhou City,311100 Zhejiang,China Contact Person: Yunhua Ren Phone: +86-571-81957767 Fax: +86-571-81957750 Email: renyh@sejoy.com
#### Name of the Device: 3.
Device Name:Infrared Thermometer
Trade Name:Infrared Forehead Thermometer,Infrared Ear/Forehead Thermometer Including models:DET-205,DET-206,DET-305,DET-306,DET-215 Classification name:Clinical Electronic Thermometer
# 4. Classification Information:
Product Code: FLL Device Class: II Panel: 80 Regulation number:880.2910
#### Predicate Device Information: 5.
Non-contact Clinical Thermometer,Model THB0F(Radiant Innovation Inc,K121428)
{4}------------------------------------------------
RII Multi-function Infrared Thermometer, Model TH52Z. (Radiant Innovation Inc, K162083)
#### Intended use / Indication for Use: 6.
Infrared Thermometer DET series are intended for the intermittent measurement of human body temperature by people of all ages. The devices are reusable for home use only.
### 7. Device Description:
The Infrared thermometers are hand-held, battery powered devices designed to measure human body temperature from the auditory canal or forehead. Its operation is based on measuring the natural thermal radiation from the tympanic membrane or forehead.
The thermometers use a thermopile sensor with integrated thermistor for the target reading, a thermistor mounted in the head of the thermometer for ambient temperature readings, a thermopile collect, the infrared energy emitted from the forehead.
The Infrared thermometer included DET-205,DET-305,DET-305,DET-306 and DET-215. DET-205 and DET-206 units are contact Infrared forehead thermometer,DET-305 and DET-306 units are non-contact Infrared forehead thermometer(measurement distance is 2~3cm), DET-215 unit is multifunction infrared thermometer and it can be selected Ear mode or Forehead mode.
In addition,all models have voice function that used to broadcast the measurement temperature results after complete the measurement. It's working principle is that Microcontroller Unit sends voice commands to the voice integrated circuit and then broadcast the measurement temperature results that display on the LCD.The voice function is use English language and actives after each measurement completed.
# 8.Summary of technological characteristics of device compared to the predicate devices, see the table1 and table2
A comparison of key similarities and differences between the subject devices (Model DET-205,DET-206,DET-305,DET-306,DET-215) and the predicate devices (K121428 and K162083) is provided below.
{5}------------------------------------------------
| able<br>1<br>-<br>1 | 1 |
|---------------------|---|
|---------------------|---|
| SE Comparison | Subject Device | Predicate device(PD) | Note | |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|------------------------------|
| 510k number | Model DET-205,DET-206,<br>DET-305,DET-306 | Radiant THB0F | | |
| 510k number | Present application | K121428 | -- | |
| Regulation<br>number | 21 CFR 880.2910 | 21 CFR 880.2910 | Same | |
| Product Code | FLL | FLL | Same | |
| Intended Use/<br>Indications for use | Infrared Thermometer DET<br>series are intended for the<br>intermittent measurement of<br>human body temperature by<br>people of all ages. The devices<br>are reusable for home use only. | The Non-contact Clinical<br>Thermometer, Model THBOF<br>is an infrared thermometer<br>intended for the intermittent<br>measurement of human body<br>temperature in people of all<br>ages. | Similar | |
| Measure Method | Infrared radiation detection | Infrared radiation detection | Same | |
| Measurement<br>Range | Forehead mode:34.0℃~43.0℃<br>(93.2℉~109.4℉) | Forehead mode:<br>34~42.2℃(93.2~108℉) | Similar<br>(Note1) | |
| Accuracy | Forehead mode:<br>$ \pm 0.2℃ (0.4℉) during 35.5℃~42.0℃ (95.9℉~107.6℉)at 15℃~35℃ (59.0℉~95.0℉)operating temperature range\pm 0.3℃ (0.5℉) for othermeasuring and operatingtemperature range $ | Forehead mode:<br>$ \pm 0.2℃ (0.4℉) within36~39℃(96.8~102℉),\pm 0.3℃(0.5℉) other range $ | Similar<br>(Note1) | |
| Display | 0.1℃(0.1℉) | 0.1℃(0.1℉) | Same | |
| Measurement<br>Place | Forehead | Forehead | Same | |
| Measurement<br>Distance | DET-205,<br>DET-206<br>DET-305,<br>DET-306 | Forehead<br>Contact<br>Non-contact,<br>2~3cm | 2-3cm | Different<br>(Note3)<br>Same |
| Response Time | Approximately 3 seconds | 1s | Similar<br>(Note4) | |
| Sensor Type | Thermopile | Thermopile | Same | |
| Scale Selection | ℃/℉ | ℃/℉ | Same | |
| Display screen | LCD | LCD | Same | |
| Memory | Each 10 sets memories for<br>forehead and object<br>measurements | 60 Sets | Different<br>(Note4) | |
| Auto power off | Yes | Yes | Same | |
{6}------------------------------------------------
| while no operation | | | |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Buzzer | Yes | Yes | Same |
| Voice Function | Yes | Unknown | Different<br>(Note4) |
| Power Source | DC3V(2×AAA battery) | DC3V(2×AAA battery) | Same |
| Patient Contact<br>material | Enclosure;ABS<br>Probe:ABS & Glass | Enclosure;ABS<br>Probe:ABS & Glass | Same |
| Biocompatibility<br>testing | Cytotoxicity, Skin irritation,<br>Skin sensitization | Cytotoxicity, Skin irritation,<br>Skin sensitization | Same |
| Conformance<br>standard | ISO 80601-2-56(Performance)<br>IEC 60601-1(Safety)<br>IEC 60601-1-2(EMC)<br>ASTM E1965-98<br>ISO 10993-5 and ISO 10993-10<br>(Biocompatibility) | ISO 80601-2-56(Performance)<br>IEC 60601-1(Safety)<br>IEC 60601-1-2(EMC)<br>ASTM E1965-98<br>ISO 10993-5 and ISO<br>10993-10 (Biocompatibility) | Same |
# Table 2
| SE Comparison | Subject Device | Pridicate Device | Note |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| | Model DET-215 | Radiant Innovation Inc,<br>Model:TH52Z | |
| 510k number | Present application | K162083 | -- |
| Regulation<br>number | 21 CFR 880.2910 | 21 CFR 880.2910 | Same |
| Product Code | FLL | FLL | Same |
| Intended Use/<br>Indications for use | Infrared Thermometer DET<br>series are intended for the<br>intermittent measurement of<br>human body temperature by<br>people of all ages. The devices<br>are reusable for home use only. | The RII Multi-function Infrared<br>Thermometer,Model TH52Z is<br>intended for the intermittent<br>measurement of human body<br>temperatures.The device is<br>intended for the use at home<br>by people of all ages including<br>neonates and it can be<br>selected Ear mode or<br>Forehead mode. | Similar |
| Measure Method | Infrared radiation detection | Infrared radiation detection | Same |
| Measurement<br>Range | Ear/Forehead mode: 34.0°C<br>~43.0°C (93.2°F~109.4°F) | Ear / Forehead mode:<br>34.0°C~42.2°C (93.2°F~108°F) | Similar<br>(Note1) |
| Accuracy | Ear/Forehead mode:<br>±0.2°C (0.4°F) during 35.5°C<br>~42.0°C (95.9°F~107.6°F)<br>at 15°C~35°C (59.0°F~95.0°F) | ±0.4°F (0.2°C) within<br>95~107.6°F (35~42°C),<br>±0.5°F (0.3°C) for other<br>range | Similar<br>(Note1) |
{7}------------------------------------------------
| | operating temperature range<br>±0.3℃ (0.5°F) for other<br>measuring and operating<br>temperature range | | |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Display | 0.1°C (0.1°F) | 0.1°C (0.1°F) | Same |
| Measurement Place | Forehead<br>Ear | Forehead<br>Ear | Same |
| Response Time | Ear mode/Object mode:<br>Approx. 1s<br>Forehead: Approx. 3s | 1s | Similar<br>(Note4) |
| Sensor Type | Thermopile | Thermopile | Same |
| Scale Selection | °C/°F | °C/°F | Same |
| Display screen | LCD | LCD | Same |
| Memory | Each 10 sets memories for<br>ear,forehead and object<br>measurements | 9 Sets | Different<br>(Note4) |
| Auto power-off<br>while no operation | Yes | Yes | Same |
| Buzzer | Yes | Yes | Same |
| Voice Function | Yes | Unknown | Different<br>(Note4) |
| Power Source | DC3V(2 × AAA battery) | CR2032*1 | Different<br>(Note2) |
| Biocompatibility | Cytotoxicity, Skin irritation,<br>Skin sensitization | Cytotoxicity, Skin irritation,<br>Skin sensitization | Same |
| Conformance<br>standard | ISO 80601-2-56(Performance)<br>IEC 60601-1(Safety)<br>IEC 60601-1-2(EMC)<br>ASTM E1965-98<br>ISO 10993-5 and ISO 10993-10<br>(Biocompatibility) | ISO 80601-2-56(Performance)<br>IEC 60601-1(Safety)<br>IEC 60601-1-2(EMC)<br>ASTM E1965-98<br>ISO 10993-5 and ISO<br>10993-10 (Biocompatibility) | Same |
### Analysis
From the comparison table1 and table2, the subject devices and predicate devices have the similar Intended use & Indications for Use, same measurement place, Scale selection, Display screen, Auto power-off while no operation & Conformance standard. There are slightly differences between the devices and predicate devices as follows,through the verification and validation process, it has been shown that the differences do not raise new questions of safety and effectiveness.
{8}------------------------------------------------
| Note1 | Measurement Range&<br>Accuracy changes | The measurement range and accuracy of subject<br>devices meet the requirements of ISO80601-2-56<br>and ASTM E1965-98. |
|-------|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Note2 | Power Source | Meet the requirement of IEC 60601-1 |
| Note3 | Measurement Distance<br>changes | Meet the requirement of ISO 80601-2-56 clinical test |
| Note4 | Response Time changes<br>from 1s to 3s | Our devices have longer response time to measure<br>the temperature,and all tests passed,so such<br>different do not raise new performance questions. |
| | Memory change from 60<br>sets/9 sets to 10 sets | Software Validation Passed. |
| Note4 | Add voice function | Software Validation Passed. |
# 9. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Infrared Thermometers conforms to applicable standards that include:
- ◆ ASTM E 1965-98 Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
- ◆ IEC 60601-1:2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
- � IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility
- . ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity,
- * ISO 10993-10:2010, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
- � IEC 60601-1-11:2015 medical electrical equipment - part 1-11: general requirements for basic safety and essential performance - collateral standard: requirements for
{9}------------------------------------------------
medical electrical equipment and medical electrical systems used in the home healthcare environment
- ◆ ISO 80601-2-56: 2009 Medical electrical equipment - part 2-56: particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.
- * The software verification and validation was provided in accordance with the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," dated May 11, 2005.
### 12. Discussion of Clinical Tests Performed:
Clinical tests were conducted on the DET-205,DET-206,DET-305,DET-306 and DET-215. The clinical tests evaluated 165 of subjects. Each model was evaluated in three groups 1) infants-newborn to one year; 2) children-greater than one to five years; and 3) adults-greater than five years old.The clinical performance test protocol and data analysis is conducted as the requirement of ASTM E1965-98 (2016). The test report showed the clinical performance of the subject device complied with the requirement of ASTM E1965-98 (2016).
#### 13. Conclusions:
Based on the information provided in this submission, the submit Infrared Thermometers, models: DET-205,DET-206,DET-305,DET-306 and DET-215 are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.