K181217 · Caremed Supply, Inc. · JOW · Aug 5, 2018 · Cardiovascular
Device Facts
Record ID
K181217
Device Name
VesoFlow Lite DVT Compression Device
Applicant
Caremed Supply, Inc.
Product Code
JOW · Cardiovascular
Decision Date
Aug 5, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Caremed Supply Inc. VesoFlow Lite Deep Vein Thrombosis (DVT) Compression Device is intended to increase venous blood flow in at risk patients in order to help prevent deep vein thrombosis.
Device Story
VesoFlow Lite is an intermittent pneumatic compression system for DVT prophylaxis. It consists of an air pump and single-patient-use compression sleeves (foot, calf, or thigh). The device operates by inflating sleeves in a wavelike, milking motion to augment venous blood flow, reduce stasis, and stimulate fibrinolysis. The pump is pre-set with fixed pressure (120mmHg for foot; 40mmHg for calf/thigh) and timing (12s inflation, 48s deflation). It is used in clinical settings to complement other measures like antiembolic stockings or anticoagulants. The system includes a control unit with a power switch, alarm light, and LED status display. Healthcare providers attach the sleeves to the patient's lower extremities and monitor the system for operation and fit. The device helps prevent DVT and associated pulmonary embolism by emptying pooled blood from femoral vein valve cusps.
Clinical Evidence
No clinical test data was used to support the decision of substantial equivalence. Evidence is based on non-clinical laboratory studies, including biocompatibility, software validation, electromagnetic compatibility, electrical safety, reliability, performance, and usability testing.
Technological Characteristics
Intermittent pneumatic compression system. Components: air pump, single-patient-use sleeves (foot, calf, thigh), air hoses. Energy: AC 100-240V, 50/60Hz. Pressure: 120mmHg (foot), 40mmHg (calf/thigh). Cycle: 12s inflation, 48s deflation. Class II, Type BF. Ingress protection: IP21. Operation: continuous. Software: embedded control logic for pump timing and alarm monitoring.
Indications for Use
Indicated for at-risk patients to increase venous blood flow and help prevent deep vein thrombosis (DVT).
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
Predicate Devices
VESOFLOW PLUS DVT Compression Devices Model Name: IPCS/SQS (K141064)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, and the FDA logo is on the right. The FDA logo includes the agency's name, "U.S. Food & Drug Administration," in blue text.
August 5, 2018
Caremed Supply, Inc. Tsung-Hsuan Liu General Manager 7F., No. 2. Lane 235, Baoqiao Rd., Xindien, Dist. New Taipei City, 23145 Tw
Re: K181217
Trade/Device Name: VesoFlow Lite DVT Compression Device Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: April 26, 2018 Received: May 7, 2018
#### Dear Tsung-Hsuan Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Fernande Aguel Fernando Aguel -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K181217
Device Name VesoFlow Lite DVT Compression Device
Indications for Use (Describe)
The Caremed Supply Inc. VesoFlow Lite Deep Vein Thrombosis (DVT) Compression Device is intended to increase venous blood flow in at risk patients in order to help prevent deep vein thrombosis.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Traditional 510(k) Section 5 - 510 (k) Summary
# 510(k) SUMMARY
| 5.1 Type of Submission: | Traditional |
|-------------------------|--------------------------------------------------------------------------------------------|
| 5.2 Date of Summary: | May 3, 2018 |
| 5.3 Submitter: | Caremed Supply, Inc. |
| Address: | 7F., No. 2, Lane 235, Baoqiao Rd., Xindien Dist.,<br>New Taipei City 23145, Taiwan, R.O.C. |
| Phone: | +886-2-2917-9808 |
| Fax: | +886-2-2918-6505 |
| Contact: | TSUNG-HSUAN LIU<br>(oscar@caremed.com.tw) |
## 5.4 Identification of the Device:
| Proprietary/Trade name: | VesoFlow Lite DVT Compression<br>Device |
|------------------------------|-----------------------------------------|
| Classification Product Code: | JOW |
| Regulation Number: | 870.5800 |
| Regulation Description: | Compressible limb sleeve |
| Review Panel: | Cardiovascular |
| Device Class: | II |
# 5.5 Identification of the Predicate Device:
| Predicate Device Name: | VESOFLOW PLUS DVT Compression<br>Devices<br>Model Name: IPCS/SQS |
|------------------------------|------------------------------------------------------------------|
| Manufacturer: | Caremed Supply, Inc. |
| Classification Product Code: | JOW |
| Regulation number: | 870.5800 |
| Device Class: | II |
| 510(k) Number: | K141064 |
{4}------------------------------------------------
Traditional 510(k) Section 5 - 510 (k) Summary
## 5.6 Identification of the Reference Device:
| Reference Device Name: | VesoPress DVT System, Pump Model<br>VP500D |
|------------------------------|--------------------------------------------|
| Applicant: | Compression Therapy Concepts, Inc. |
| Manufacturer: | Caremed Supply, Inc. |
| Classification Product Code: | JOW |
| Regulation number: | 870.5800 |
| Device Class: | II |
| 510(k) Number: | K061814 |
### 5.7 Intended Use/ Indications for Use of the Device
The Caremed Supply Inc. VesoFlow Lite Deep Vein Thrombosis (DVT) Compression Device is intended to increase venous blood flow in at risk patients in order to help prevent deep vein thrombosis.
# 5.8 Device Description
The VesoFlow Lite DVT Compression Device is an Intermittent Pneumatic Compression System that counteracts blood stasis and coagulation changes; two of the three major factors that promote deep vein thrombosis (DVT) a potentially life threatening condition which can lead to pulmonary embolism.
The VesoFlow Lite is a non-invasive mechanical prophylactic system that massages the feet or legs in a wavelike, milking motion that promotes blood flow and deters thrombosis, helping to empty pooled or static blood from the valve cusps of the femoral vein. Fibrinolytic activity is increased, stimulating the release of a plasminogen activator. This therapy typically complements other prophylactic measures, such as antiembolic stockings and anticoagulants.
Some exciting features include:
- o User friendly master control unit that is designed so that its functions are self-explanatory
- Power micro switch (ON/OFF)
{5}------------------------------------------------
- . An alarm light displays a fault status with an audio alarm and LED display
- . LED display that can monitor errors, sleeves statuses the device
- User friendly hanging bracket that provides easy attachment
The system consists of an air pump and single-patient-use compression sleeve one for the foot, one for the calf and one for the thigh. The air pump and compression sleeve may be used on one lower extremity. If the air pump is used simultaneously on both lower extremities, it is important that the identical sleeve type be used on both legs (i.e. foot sleeve with foot sleeve, calf sleeve with calf sleeve, or thigh sleeve with thigh sleeve). The inflation pressure settings are pre-set by the manufacturer and cannot be changed. The compression level for the foot sleeve is 120mmHg (16kPa). The compression level for the calf and thigh sleeve is 40mmHg. The sleeve inflates and maintains compression for 12 seconds. Then, the sleeve deflates for a period of 48 seconds. As the sleeve inflates, it compresses the foot, calf or thigh which augments venous blood flow, thereby reducing stasis. This process also stimulates fibrinolysis: thus, reducing the risk of early clot formation. Regularly check the system while in use, assuring pump operation and sleeve fit.
#### 5.9 Non-clinical Testing
A series of safety and performance tests were conducted on the subject device, VesoFlow Lite DVT Compression Device.
- . Biocompatibility
- . Software Validation
- Electromagnetic compatibility and electrical safety ●
- . Reliability
- Performance ●
- . Usability
All the test results demonstrate VesoFlow Lite DVT Compression Device meets the requirements of its pre-defined acceptance criteria and intended use, and is substantially equivalent to the predicate device.
{6}------------------------------------------------
# 5.10 Clinical Testing
No clinical test data was used to support the decision of substantial equivalence.
# 5.11 Substantial Equivalence Determination
The VesoFlow Lite DVT Compression Device submitted in this 510(k) file is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the cleared VESOFLOW PLUS DVT Compression Devices Model Name: IPCS/SQS (K141064), especially IPCS. We conducted comparative performance test with IPCS, and also compared with referenced VesoPress DVT System, Pump Model VP500D (K061814) for the specifications of Pressure, Inflation time, and Deflation time. Differences between the devices cited in this section do not raise any new issue of substantial equivalence.
| Item | Subject device | Predicate device | Substantial<br>equivalence<br>determination |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Proprietary Name | VesoFlow Lite DVT<br>Compression Device | VESOFLOW PLUS DVT<br>Compression Devices | |
| Model Name | VesoFlow Lite | IPCS | |
| 510(k) No. | (to be assigned) | K141064 | |
| Intended Use | The Caremed Supply Inc.<br>VesoFlow Lite Deep Vein<br>Thrombosis (DVT) Compression<br>Device is intended to increase<br>venous blood flow in at risk<br>patients in order to help prevent<br>deep vein thrombosis. | The Caremed Supply Inc.<br>VESOFLOW® PLUS SQS and<br>IPCS Deep Vein Thrombosis (DVT)<br>Compression Devices are intended to<br>increase venous blood flow in at risk<br>patients in order to help prevent deep<br>vein thrombosis. | Same |
| Type of use | Prescription Use | Prescription Use | Same |
| For Control Unit | | | |
| Size | 9.83" x 4.37" x 8.28" | 7.54" x 5.12" x 7.95" | Different but does<br>not raise any new<br>issue of substantial<br>equivalence |
| | | | |
| Weight | 2.4 Kg | 2.8 Kg | Different but does<br>not raise any new<br>issue of substantial<br>equivalence |
| Pressure (mmHg) | Calf/Thigh: 40;<br>Foot: 120 | Calf/Thigh: 40;<br>Foot: 130 | Different but<br>Same as Reference<br>Device K061814<br>(Calf/Thigh: 40;<br>Foot: 80-120) |
| Input Rating | AC 100-240V, 50/60Hz | AC 100-240V, 50/60Hz | Same |
| Fuse Rating | T2AL 250V | 1A/250V | Different but does<br>not raise any new<br>issue of substantial<br>equivalence |
| IEC Classification | Class II, Type BF<br>Not AP or AGP type | Class II, Type BF<br>Not AP or AGP type | Same |
| Ingress of Water<br>Protection | IP21 | IP22 | Different but does<br>not raise any new<br>issue of substantial<br>equivalence |
| Operation Humidity | 30% to 75% | 30 - 75% | Same |
| Operation<br>Temperature | 15°C - 40°C | 15°C - 35°C | Different but does<br>not raise any new<br>issue of substantial<br>equivalence |
| Operation<br>Atmospheric<br>Pressure Range | 700 hPa to 1060 hPa | 700 hPa to 1060 hPa | Same |
| Mode of Operation | Continuous | Continuous | Same |
| Applied Part | Sleeve and Air Hose | Garment and Air Hose | Same |
| Applied Mode of<br>Pressure | Intermittent | Intermittent | Same |
| | | | |
| Inflation time<br>per chamber | 12 seconds | 12 seconds for Calf/Thigh,<br>3 seconds for Foot | Different but<br>Same as Reference<br>Device K061814<br>(all 12 seconds) |
| Deflation time<br>per chamber | 48 seconds | 48 seconds for Calf/Thigh,<br>30 seconds for Foot | Different but<br>Same as Reference<br>Device K061814<br>(all 48 seconds) |
| Application | Non-invasive / external | Non-invasive / external | Same |
| Battery Pack | No | Yes | Different but does<br>not raise any new<br>issue of substantial<br>equivalence |
| Software /<br>Control panel | w/o Pressure and Timer display<br>w/o Mute, Caution, and Battery<br>symbols display<br>w/o Timer Reset function<br>w/o Alarm mute button<br>w/o Maintenance alarm light | w/ Pressure and Timer display<br>w/ Mute, Caution, and Battery<br>symbols display<br>w/ Timer Reset function<br>w/ Alarm mute button<br>w/ Maintenance alarm light | Different but does<br>not raise any new<br>issue of substantial<br>equivalence |
| For Applied Part | | | |
| Calf Sleeves<br>(by calf<br>circumference) | Small: up to 12"<br>Medium: up to 18"<br>Large: up to 24"<br>XL: up to 32" | XS: up to 12"<br>M: up to 18"<br>L: up to 24"<br>B: up to 32" | Same |
| Thigh Sleeves<br>(by thigh<br>circumference) | Medium: up to 29"<br>Large: up to 36"<br>XL: up to 42" | M: up to 29"<br>L: up to 36"<br>B: up to 42" | Same |
| Foot Sleeves<br>(by foot length) | Medium: 13"<br>Large: 16" | U: 13"<br>L: 16" | Same |
| Air Hose | extension of 60" (pair), standard<br>extension of 118" (pair) | extension of 60" (pair), standard<br>extension of 118" (pair)<br>extension of 177" (pair) | Same |
{7}------------------------------------------------
Section 5 - Traditional 510(k)
Section 5 - 510 (k) Summary
{8}------------------------------------------------
# Section 5 - 510 (k) Summary
{9}------------------------------------------------
#### 5.12 Similarity and Difference
The VesoFlow Lite DVT Compression Device has been compared with "VESOFLOW PLUS DVT Compression Devices Model Name: IPCS" and referred to "VesoPress DVT System, Pump Model VP500D". The subject device has same intended use, principle of operation and similar technological characteristics as model: IPCS of the predicate device and reference device. Although there are some specifications that are different between two devices, the performance test has been completed to demonstrate that the differences between these parameters would not impact the safety and effectiveness of the subject device, and the specifications of Pressure. Inflation time, and Deflation time are the same as those of reference device. The subject device has also undergone safety and performance tests, and the results complied with the test requests. Therefore, the differences between the subject device and the predicate device do not raise any problem of substantial equivalence. The subject device is substantially equivalent to the predicate device in intended use, design, and performance claims.
# 5.13 Conclusion
After analyzing non-clinical laboratory studies and safety testing data, it can be concluded that the VesoFlow Lite DVT Compression Device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.