The catheters are indicated for short term access to the peripheral venous system for selected intravenous therapies, blood sampling, and power injection of contrast media. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The catheters are suitable for use with power injectors.
Device Story
Vygon Midline Catheters are single-use, EO-sterilized, polyurethane catheters designed for short-term (<30 days) peripheral venous access. The device features a reverse taper shaft with depth indicators, a molded suture wing compatible with StatLock securement, and luer hubs. Available in 3 Fr (single lumen), 4 Fr (single lumen), and 5 Fr (dual lumen) configurations. The device is inserted percutaneously using the Modified Seldinger Technique with a guidewire. The catheter tip terminates distally to the shoulder at or below the axillary line. The device allows for intravenous therapy, blood sampling, and power injection of contrast media. It is used in regulated clinical environments by healthcare professionals. The device benefits patients by providing reliable, short-term peripheral venous access for various therapies and diagnostic imaging procedures.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by bench testing, including dimensional inspections, leak tests, flow tests (gravity/pump/aspiration), tensile/burst/flexural fatigue testing, power injection conditioning, and biocompatibility assessments (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis, and thrombogenicity).
Technological Characteristics
Materials: Polyurethane (shaft, junction, extension leg, luer hub), Acetal (clamp), barium sulfate (radiopacity). Dimensions: 3 Fr, 4 Fr, 5 Fr; 20 cm length. Connectivity: None. Sterilization: 100% Ethylene Oxide (SAL 10^-6). Standards: ISO 9626:2016, ISO 10555-1:2014, ISO 594-1:1986/ISO 80369-7:2016, ASTM F1842-15, ISO 10993 series.
Indications for Use
Indicated for short-term peripheral venous access for intravenous therapy, blood sampling, and power injection of contrast media in any patient population, provided vascular anatomy is adequate.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo.
November 2, 2018
Vygon USA Jillian Mikovich Regulatory Affairs Manager 2750 Morris Rd, Suite A200 Lansdale, Pennsylvania 19460
Re: K181208
Trade/Device Name: Vygon Midline 3 Fr SL Catheter; Vygon Midline 4 Fr SL Catheter; Vygon Midline 5 Fr DL Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: PND Dated: September 24, 2018 Received: September 28, 2018
Dear Jillian Mikovich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K181298
Device Name Vygon Midline Catheters
### Indications for Use (Describe)
The catheters are indicated for short term access to the peripheral venous system intravenous therapies, blood sampling, and power injection of contrast media. These catheters may be used for any patient with consideration given to adequacy of vascular anatomy and appropriateness of the catheters are suitable for use with power injectors.
| Catheter Size | Maximum Power Injection<br>Flow Rate |
|-----------------------|--------------------------------------|
| Vygon Midline 3 Fr SL | 5 ml/s |
| Vygon Midline 4 Fr SL | 7 ml/s |
| Vygon Midline 4 Fr DL | 7 ml/s |
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Vygon logo. The logo consists of the word "VYGON" in white letters on a green rounded rectangle. Above the "V" is a red triangle pointing downwards. Below the logo is the text "Value Life" in a smaller, light gray font.
## 510(k) Summary (21 CFR 807.92) K181208
#### I. SUBMITTER
Submitter Name: Vygon Address: 2750 Morris Road, Suite A200 Lansdale, PA 19446 Telephone: 215-390-2002 Fax: 215-672-6740
Contact Person: Jillian Mikovich jmikovich@vygonus.com Email of Contact: 05/04/2018 Date Prepared:
#### II. DEVICE
Device Name: Common Name: Common/Usual Name: Classification Name: Classification Panel: Regulation Device Name: Regulation Number: Regulatory Class: Product Code:
Vygon Midline Catheters Vygon Midline Catheters Midline Catheter Intravascular Catheter General Hospital Intravascular Catheter 21 CFR 880.5200 2 PND
### III. PREDICATE DEVICE
| Predicate Name: | 3 Fr SL PowerMidline™ Catheter |
|-------------------------|--------------------------------|
| Trade Name: | 3 Fr SL PowerMidline™ Catheter |
| Common/Usual Name: | Midline Catheter |
| Classification Name: | PND – Intravascular Catheter |
| Premarket Notification: | K153393 |
| Manufacturer: | Bard Access Systems. Inc. |
Per the FDA Medical Device Recalls Database, this predicate has not been subject to a design related recall.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Vygon logo. The word "VYGON" is in white letters on a green rounded rectangle. Above the "V" is a small red triangle. Below the logo is the phrase "Value Life" in gray letters.
## IV. DEVICE DESCRIPTION
The Vygon Midline Catheters are EO sterilized, single-use, polyurethane, reverse taper, short-term, peripheral venous system access catheters. They include materials containing barium sulfate for radiopacity.
The device description of the Vygon Midline Catheters is as follows:
- 1. The Vygon Midline Catheters are EO sterilized, single-use, polyurethane catheters. They include materials containing barium sulfate for radiopacity.
- 2. Catheter sizes, lengths, and lumen configurations are in the table that follows. Catheter gauge sizes are printed on the hubs.
- 3. The catheter luer hub material is made of Quadraplast™ (Rigid Thermoplastic Polyurethane (TPU)) and may contain red, white, gray, or purple colorant.
- 4. Purple colorants were added to select catheter materials to indicate power injectable functionality per generally accepted clinical use.
- 5. The catheter extension leg material is made of Quadraflex™ (Aromatic Polyether (ARE-TPU)) and extension leg colors are listed in the table that follows.
- The catheter junction material (Aliphatic Polyether (ALE-TPU)) is molded such that 6. the catheter extension leg and the shaft tubing are joined.
- 7. A molded suture wing is integrated with the over-molded junction and is compatible with StatLock® securement device.
- 8. The catheter extension leg(s) are over-molded over the proximal end(s) of the shaft. Extension leg lengths can be found in the table that follows.
- 9. The catheter extension leg(s), junction, and ID tag(s) on clamp are all printed with markings to include information to facilitate proper use of the device.
- 10. The catheter has a reverse taper design.
- 11. The catheter shaft tubing is marked with depth indicators as a reference for catheter insertion.
- 12. The environment of use of the device is a regulated clinical environment.
- 13. Midlines operate by allowing intravenous access through one or more lumens.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Vygon logo. The logo consists of the word "VYGON" in white letters inside a green rounded rectangle with a red triangle above the "V". Below the rectangle, the words "Value Life" are written in a smaller, gray font.
- 14. Length of contact is short term (<30 days) with intravenous indwelling and skin contact.
| Device Description | | | |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Vygon Midline 3 Fr SL<br>(Subject Device 1) | Vygon Midline 4 Fr SL<br>(Subject Device 2) | Vygon Midline 5 Fr DL<br>(Subject Device 3) |
| Number of<br>Lumens | Single Lumen (one) | Single Lumen (one) | Dual Lumen (two) |
| Lumen Shape | Round | Round | D, D |
| Outer Diameter | 3 Fr (0.044 in.) | 4 Fr (0.055 in.) | 5 Fr (0.069 in.) |
| Inner Diameter | 20 Ga (0.027 in.) | 18 Ga (0.032 in.) | 18 Ga / 18 Ga (0.041 in.<br>H & 0.020 in. W) |
| Effective<br>Length | 20 cm | 20 cm | 20 cm |
| Maximum<br>Power<br>Injection<br>Flow Rate | 5 ml/s | 7 ml/s | 7 ml/s |
| Material | • Luer – QPLAST<br>(Rigid TPU)<br>• Gauge Marking -<br>Marabu Tampastar<br>TPR (970) White<br>• Extension Leg -<br>QFLEX (ARE-TPU)<br>• Extension Leg<br>Printing - Marabu<br>Tampastar TPR (970)<br>White<br>• Clamp - Acetal<br>(POM) Clamp<br>• Clamp Printing -<br>Biocompatible Ink<br>• Junction - QFLEX<br>(ALE-TPU)<br>• Inner Junction – N/A<br>Junction Printing -<br>Marabu Tampastar TPR<br>White (970) | • Luer – QPLAST<br>(Rigid TPU)<br>• Gauge Marking -<br>Marabu Tampastar<br>TPR (970) White<br>• Extension Leg -<br>QFLEX (ARE-TPU)<br>• Extension Leg<br>Printing – Marabu<br>Tampastar TPR (970)<br>White<br>• Clamp - Acetal<br>(POM) Clamp<br>• Clamp Printing -<br>Biocompatible Ink<br>• Junction - QFLEX<br>(ALE-TPU)<br>• Inner Junction – N/A<br>Junction Printing -<br>Marabu Tampastar TPR<br>White (970) | • Luer – QPLAST<br>(Rigid TPU)<br>• Gauge Marking -<br>Marabu Tampastar<br>TPR (970) White<br>• Extension Leg -<br>QFLEX (ARE-TPU)<br>• Extension Leg<br>Printing - Marabu<br>Tampastar TPR (970)<br>White<br>• Clamp - Acetal<br>(POM) Clamp<br>• Clamp Printing -<br>Biocompatible Ink<br>• Junction - QFLEX<br>(ALE-TPU)<br>• Inner Junction – N/A<br>Junction Printing -<br>Marabu Tampastar TPR<br>White (970) |
| Device Description | | | |
| | Vygon Midline 3 Fr SL | Vygon Midline 4 Fr SL | Vygon Midline 5 Fr DL |
| | (Subject Device 1) | (Subject Device 2) | (Subject Device 3) |
| | • Extrusion (Shaft) -<br>QFLEX<br>(ALE-TPU)<br>• Extrusion (Shaft)<br>Printing - Marabu | • Extrusion (Shaft) -<br>QFLEX<br>(ALE-TPU)<br>• Extrusion (Shaft)<br>Printing - Marabu | • Extrusion (Shaft) -<br>QFLEX<br>(ALE-TPU)<br>• Extrusion (Shaft)<br>Printing - Marabu |
| | Tampapur TPU (980)<br>Black | Tampapur TPU (980)<br>Black | Tampapur TPU (980)<br>Black |
| Colorant /<br>Concentrations | • Luer - RV43631224<br>PAN 2607C, 4%<br>(Purple)<br>• Luer Ink - Marabu<br>Tampastar TPR (970)<br>White<br>• Extension Leg - MC-<br>90570TPU, 4%<br>(Trans. Purple Ext.<br>Leg)<br>• Extension Leg Ink -<br>Marabu Tampastar<br>TPR (970) White<br>• Junction - Clariant<br>OM4M664821 4%<br>Purple 2587C<br>• Junction Ink - Marabu<br>Tampastar TPR White<br>(970)<br>• Shaft - Clariant<br>OM4M664821 Purple<br>2587C, 4%<br>• Shaft Ink - Marabu<br>Tampapur TPU (980)<br>Black | • Luer - RV43631224<br>PAN 2607C, 4%<br>(Purple)<br>• Luer Ink - Marabu<br>Tampastar TPR (970)<br>White<br>• Extension Leg - MC-<br>90570TPU, 4%<br>(Trans. Purple Ext.<br>Leg)<br>• Extension Leg Ink -<br>Marabu Tampastar<br>TPR (970) White<br>• Junction - Clariant<br>OM4M664821 4%<br>Purple 2587C<br>• Junction Ink - Marabu<br>Tampastar TPR White<br>(970)<br>• Shaft - Clariant<br>OM4M664821 Purple<br>2587C, 4%<br>• Shaft Ink - Marabu<br>Tampapur TPU (980)<br>Black | • Luer - RV43631224<br>PAN 2607C, 4%<br>(Purple)<br>• Luer Ink - Marabu<br>Tampastar TPR (970)<br>White<br>• Extension Leg - MC-<br>90570TPU, 4%<br>(Trans. Purple Ext.<br>Leg)<br>• Extension Leg Ink -<br>Marabu Tampastar<br>TPR (970) White<br>• Junction - Clariant<br>OM4M664821 4%<br>Purple 2587C<br>• Junction Ink - Marabu<br>Tampastar TPR White<br>(970)<br>• Shaft - Clariant<br>OM4M664821 Purple<br>2587C, 4%<br>• Shaft Ink - Marabu<br>Tampapur TPU (980)<br>Black |
| Ink Material | • Marabu Tampastar<br>TPR White (970)<br>• Marabu Tampapur<br>TPU (980) Black | • Marabu Tampastar<br>TPR White (970)<br>• Marabu Tampapur<br>TPU (980) Black | • Marabu Tampastar<br>TPR White (970)<br>• Marabu Tampapur<br>TPU (980) Black |
| Duration of Use | Short term (<30 days) | Short term (<30 days) | Short term (<30 days) |
| Device Description | | | |
| | Vygon Midline 3 Fr SL<br>(Subject Device 1) | Vygon Midline 4 Fr SL<br>(Subject Device 2) | Vygon Midline 5 Fr DL<br>(Subject Device 3) |
| Tip Placement<br>Location | Peripheral<br>venous<br>system, with<br>catheter tip<br>terminating<br>prior to the<br>axilla | Peripheral<br>venous<br>system, with<br>catheter tip<br>terminating<br>prior to the<br>axilla | Peripheral<br>venous<br>system, with<br>catheter tip<br>terminating<br>prior to the<br>axilla |
| Insertion Site | Peripheral | Peripheral | Peripheral |
| Catheter Base<br>Materials | Shaft Tubing<br>Polyurethane | Shaft Tubing<br>Polyurethane | Shaft Tubing<br>Polyurethane |
| | Catheter Junction<br>Polyurethane | Catheter Junction<br>Polyurethane | Catheter Junction<br>Polyurethane |
| | Extension Leg<br>Polyurethane | Extension Leg<br>Polyurethane | Extension Leg<br>Polyurethane |
| | Luer Hub<br>Polyurethane | Luer Hub<br>Polyurethane | Luer Hub<br>Polyurethane |
| | Extension Leg Clamp<br>Polyurethane | Extension Leg Clamp<br>Polyurethane | Extension Leg Clamp<br>Polyurethane |
| Depth<br>Markings | "0" depth indicator<br>located 1 cm from<br>catheter junction on<br>reverse taper shaft tubing;<br>catheter marked every 1<br>cm, with numeric<br>indicators every 5 cm | "0" depth indicator<br>located 1 cm from<br>catheter junction on<br>reverse taper shaft tubing;<br>catheter marked every 1<br>cm, with numeric<br>indicators every 5 cm | "0" depth indicator<br>located 1 cm from<br>catheter junction on<br>reverse taper shaft tubing;<br>catheter marked every 1<br>cm, with numeric<br>indicators every 5 cm |
| Power<br>Injection<br>Maximum<br>Flow Rate | 5 ml/sOpen Ended | 7 ml/sOpen Ended | 7 ml/sOpen Ended |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Vygon. The logo consists of the word "VYGON" in white letters on a green background. Above the "V" is a small red triangle pointing downwards. Below the green background are the words "Value Life" in a light gray color.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the word "VYGON" in white letters inside of a green rounded rectangle. Above the "V" is a red triangle pointing downwards. Below the green rectangle is the phrase "Value Life" in gray letters.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the Vygon logo. The logo consists of the word "VYGON" in white letters on a green rounded rectangle. Above the "V" is a red triangle pointing downwards. Below the rectangle are the words "Value Life" in a smaller font.
#### INDICATIONS FOR USE V.
The catheters are indicated for short term access to the peripheral venous system for selected intravenous therapies, blood sampling, and power injection of contrast media. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The catheters are suitable for use with power injectors.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
All characteristics of the Vygon Midline Catheters are substantially equivalent to the predicate device, the Bard Access System 3 Fr SL PowerMidline™ Catheter. Both the subject and the predicate catheters are sterile, single-use, polyurethane Midlines with the same intended use, indications for use, similar design, similar sizes, similar or the same materials, and similar properties. See comparison table below.
| Attribute | Predicate Device | Subject Device 1 | Subject Device 2 | Subject Device 3 | Substantially Equivalent or Identical to Predicate |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| Device Name | 3 Fr SL<br>PowerMidline™<br>Catheter | Vygon<br>Midline 3 Fr<br>SL (Subject<br>Device 1) | Vygon<br>Midline 4 Fr<br>SL (Subject<br>Device 2) | Vygon<br>Midline 5 Fr<br>DL (Subject<br>Device 3) | |
| 510(k)<br>Number | K153393 | K181208 | K181208 | K181208 | N/A |
| 510(k)<br>Submitter | Bard Access<br>Systems (BAS) | Vygon USA | Vygon USA | Vygon USA | N/A |
| Trade Name | 3 Fr SL<br>PowerMidline™<br>Catheter | Vygon<br>Midline 3 Fr<br>SL Catheter | Vygon<br>Midline 4 Fr<br>SL Catheter | Vygon<br>Midline 5 Fr<br>DL Catheter | Substantially Equivalent to Predicate |
| Common<br>Name | Midline Catheter | Midline<br>Catheter | Midline<br>Catheter | Midline<br>Catheter | Identical to Predicate |
| Classification<br>Name | Midline Catheter | Midline<br>Catheter | Midline<br>Catheter | Midline<br>Catheter | Identical to Predicate |
| Regulatory<br>Class | 2 | 2 | 2 | 2 | Identical to Predicate |
| Classification<br>Regulation | 21 CFR<br>880.5200 | 21 CFR<br>880.5200 | 21 CFR<br>880.5200 | 21 CFR<br>880.5200 | Identical to Predicate |
| Attribute | Predicate Device | Subject<br>Device 1 | Subject<br>Device 2…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.