K181165 · Connected Sensing- A Division of Philips Medical Systems · DRT · Mar 7, 2019 · Cardiovascular
Device Facts
Record ID
K181165
Device Name
Philips wearable biosensor-G5 Solution
Applicant
Connected Sensing- A Division of Philips Medical Systems
Product Code
DRT · Cardiovascular
Decision Date
Mar 7, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
The Philips wearable biosensor-G5 is indicated for single patient use whenever heart rate measurement is needed in non-critical hospital settings. The Philips wearable biosensor-G5 solution is used as a higher resolution heart rate log by nurses or physicians retrospectively as an aid in making non-life threatening therapeutic decisions. The biosensor is intended for patients who are 18 years of age or older. The G5 Biosensor is intended only for patients with a baseline narrow QRS complex (less than 100 ms).
Device Story
Philips wearable biosensor-G5 Solution is a single-use, battery-operated, chest-worn device; captures surface electrical waveforms to derive heart rate; transmits data wirelessly via Bluetooth to G5 application. Used in non-critical hospital settings by nurses or physicians for retrospective review of heart rate logs; aids in non-life-threatening therapeutic decisions. Device provides higher resolution heart rate data; assists clinicians in monitoring patients who are stationary or ambulatory. System includes biosensor and data visualization software; output is a user interface and exportable file for analysis.
Clinical Evidence
Bench testing only. Feasibility wear study conducted on 28 healthy volunteers to validate adhesive performance per AAMI ANSI EC12:2000/(R)2010, establishing a 48-hour maximum wear duration. System verified for electrical safety (IEC 60601-1, IEC 60601-2-27) and EMC (IEC 60601-1-2).
Technological Characteristics
Single-lead ECG sensing; two electrodes with biocompatible adhesive. Battery-powered; Bluetooth telemetry. Dimensions/form factor: body-worn chest sensor. Software: firmware-based signal processing. Standards: IEC 60601-1, IEC 60601-2-27, IEC 60601-2-47, AAMI ANSI EC12:2000/(R)2010, ISO 10993-1 (biocompatibility).
Indications for Use
Indicated for patients 18+ years old with baseline narrow QRS complex (<100 ms) requiring heart rate monitoring in non-critical hospital settings.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 7, 2019
Connected Sensing-Division of Philips Medical Systems Katie Pacheco Regulatory Affairs Specialist 4 2 Canal Park Cambridge, Massachusetts 02141
Re: K181165
Trade/Device Name: Philips wearable biosensor-G5 Solution Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: DRT Dated: February 1. 2019 Received: February 4, 2019
Dear Katie Pacheco:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica E. Paulsen -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K181165
Device Name Philips wearable biosensor-G5 Solution
### Indications for Use (Describe)
The Philips wearable biosensor-G5 is indicated for single patient use whenever heart rate measurement is needed in noncritical hospital settings. The Philips wearable biosensor-G5 solution is used as a higher resolution heart rate log by nurses or physicians retrospectively as an aid in making non-ife threatening therapeutic decisions. The biosensor is intended for patients who are 18 years of age or older.
The G5 Biosensor is intended only for patients with a baseline narrow QRS complex (less than 100 ms).
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------|-----------------------------------------------|
|------------------------------------------------|-----------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
(per 21 CFR 807.92)
#### SUBMITTER I.
Connected Sensing-Division of Philips Medical Systems. 2 Canal Park Cambridge, MA 02141 Phone: 617-218-0802 Fax: 617-218-0802 Contact Person: Katie Pacheco Date Prepared: March 6, 2019
#### II. DEVICE
Name of Device: Philips wearable biosensor-G5 Solution Common or Usual Name: Wearable Biosensor Classification Name: Monitor, Cardiac (Incl. Cardiotachometer and rate alarm) (21 CFR 870.2300) Regulatory Class: II Product Code: DRT
#### PREDICATE DEVICE III.
BioModule 3-MI, K123658 This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
Philips wearable biosensor-G5 Solution is a physiological sensing solution that gathers and stores a patient's heart rate. Philips wearable biosensor-G5 Solution is comprised of the:
- Philips wearable biosensor-G5
- and data visualization application "G5 application" ●
The Philips wearable biosensor-G5 is a battery operated, single-use device, measuring heart rate by continuously acquiring surface electrical waveforms related to cardiac excitations and measuring beat-to-beat intervals when a patient is stationary or ambulatory. The sensor functions by capturing and then sending physiological data wirelessly to the software application. The sensor's frequency of data collection and transmission is configurable. The G5 application receives and displays data from the Philips wearable biosensor-G5 providing a user interface and exportable file for retrospective review and analysis.The G5 Biosensor is intended only for patients with a baseline narrow QRS complex (less than 100 ms).
#### V. INDICATIONS FOR USE
The Philips wearable biosensor-G5 is indicated for single patient use whenever heart rate measurement is needed in non-critical hospital settings. The Philips wearable biosensor-G5 solution is used as a higher resolution heart rate log by nurses or physicians retrospectively as an aid in making non-critical or non-life threatening therapeutic decisions. The biosensor is intended for patients who are 18 years of age or older. The G5 Biosensor is intended only for patients with a baseline narrow QRS complex (less than 100 ms).
{4}------------------------------------------------
The Indications for Use statement for the Philips wearable biosensor-G5 Solution is not identical to the predicate device; however, the differences do not alter the intended therapeutic use of the subject device nor do they affect the safety and effectiveness of the subject device to the predicate. Both the subject and predicate devices have the same intended use of collecting and transmitting heart rate measurement, by acquiring electrical signals from the skin surface.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
Acquiring surface electrical waveforms related to the contractile activity of the heart is the technological principle for both the subject and predicate devices. It is based on the use of electrode technology to capture electrical waveforms via the chest to derive physiological signals processed by the system firmware.
At a high level, the subject and predicate devices are based on the following same technological elements:
| Similarities | |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Scientific Concept | Single-Lead ECG used to derive heart rate measurements. |
| Electrodes | Two electrodes and biocompatible adhesive materials used for<br>adhesion and electrical contact= |
| Device Placement Location | Devices are both placed on the chest |
| Analog or digital technology | Analog physiological signals converted to digital |
| Computer Processing | On sensor signal processing through firmware |
| Storage of recorded signals | Data storage and transfer capabilities |
| Radio frequency telemetry | Bluetooth® Transmitter/Receiver |
| Power | Battery operated |
| Alarm management | Not an alarming device |
| IEC 60601-1 | Portable, Body-worn, CF-Applied Part Internally powered |
| IEC 60601-1-2 | RF emission CISPR 11: Group 1, Class B |
| Heart Rate Resolution | ±1bpm |
| Heart Rate Range | The predicate device has heart rate range of 30 - 240 bpm per<br>IEC 60601-2-47, while the subject device has a heart rate range<br>of 30 - 220 bpm per IEC 60601-2-27. |
| Heart Rate Accuracy | IEC 60601-2-47, subject device also meets IEC 60601-2-27 |
| Operational Relative Humidity<br>Range | 15% to 95% non-condensing |
| Differences | |
| Reusable, Single-Use | The subject device is a fully disposable single- use device with<br>encapsulated electrode technology and puck. The predicate device<br>is comprised of single-use electrodes with a reusable puck. |
| Electrodes | Fully encapsulated electrode technology and puck. The predicate<br>device is comprised of single-use electrodes with a reusable puck. |
| Parameters Measured | The subject device provides Heart Rate while the predicate device<br>provides multiple parameters Heart Rate, ECG, Respiration Rate,<br>Activity. |
| Wear Duration | The predicate electrodes can be worn up <72 hours, while the<br>subject device can be worn for 48 hours. While the wear<br>duration is shorter in the subject device, there are no new<br>questions of safety and efficacy raised with the shorter wear<br>duration. Adhesive performance of the subject device and<br>predicate device demonstrated by compliance to recognized<br>standard AAMI ANSI EC12:2000/(R) 2010. |
| Shelf Life | The predicate electrodes have a 24 month shelf life, while the<br>subject device have a 3 month shelf life. While the shelf life is<br>shorter in the subject device, there are no new questions of<br>safety and efficacy raised with the shorter shelf-life duration.<br>Both the subject and predicate device address the risk of<br>degradation of the device during storage and out of pouch. |
| Atmospheric Range | The predicate electrodes have an atmospheric range of 12kPa to<br>107kPa, while the subject device have an atmospheric range of<br>50 kPa -106kPa. While the while the atmospheric lower range<br>is higher in the subject device, there are no new questions of<br>safety and efficacy for in hospital use. |
| Operational Temperature Range | The predicate electrodes have an operational temperature range<br>of 0° C to +45° C, while the subject device have an operational<br>temperature range of 15° C to +35° C. While the while the<br>operational temperature range is different in the subject device,<br>there are no new questions of safety and efficacy for in hospital<br>use. |
{5}------------------------------------------------
1
{6}------------------------------------------------
# VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility testing
The biocompatibility evaluation for the Philips wearable biosensor-G5 Solution was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process"(2016), and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity
- Sensitization .
- Irritation ●
The Philips wearable biosensor-G5 Solution is considered skin contacting for a duration of more than 24 hours.
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Philips wearable biosensor-G5 Solution. The system complies with the applicable requirements within IEC 60601-1, and IEC 60601-2-27 standards for safety and the IEC 60601-1-2 standard for EMC.
### Performance Testing
Performance testing bench testing was performed using voluntary standards IEC 60601-2-27, IEC 60601-2-47, EC57, and EC12. Testing to applicable requirements were conducted and the system was found to be compliant.
### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software could directly result in minor injury to the patient or operator.
### Wear Duration Study
Testing of the Philips wearable biosensor-G5 Solution included a feasibility wear study of 28 normal healthy volunteers. Data collected from this study was used to determine the adhesive performance (duration of use) as calculated per the ANSVAAMI EC12:2000/@2010 Disposable ECG Electrodes. The study established a 48 hour maximum duration of use for the Philips wearable biosensor-G5 Solution.
# VIII. CONCLUSIONS
The results of the verification and validation testing demonstrate that the Philips wearable biosensor-G5 Solution is safe and effective and should perform as intended in the specified use conditions. The non-clinical and clinical testing performed demonstrate that the Philips wearable biosensor-G5 Solution is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.