← Product Code [NBW](/productcode/NBW) · K181131

# Accu-Chek Guide Me Blood Glucose Monitoring System (K181131)

_Roche Diabetes Care · NBW · Dec 13, 2018 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K181131

## Device Facts

- **Applicant:** Roche Diabetes Care
- **Product Code:** [NBW](/productcode/NBW.md)
- **Decision Date:** Dec 13, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 862.1345
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry
- **Attributes:** Pediatric

## Indications for Use

The Accu-Chek Guide Me Blood Glucose Monitoring System is intended for single-patient use only.

## Device Story

System measures blood glucose via Accu-Chek Guide test strips and meter; uses amperometric detection with FAD-GDH enzyme. Input: 600 nanoliter capillary blood sample. Output: quantitative glucose concentration (20-600 mg/dL) displayed on fixed-segment LCD. Used by patients with diabetes for self-monitoring; intended for single-patient use. Meter features 3-button interface, BLE connectivity for data transfer (fixed PIN pairing), and USB port. Device aids in monitoring glucose control effectiveness; results inform patient self-management decisions. No strip ejector, strip light, or pattern detection included compared to predicate.

## Clinical Evidence

Bench testing only. No clinical data presented. Substantial equivalence supported by comparison of technological characteristics and performance specifications to the predicate device.

## Technological Characteristics

Glucose monitoring system; fixed segment LCD display; BLE connectivity (single device pairing); external meter materials validated for chemical resistance to Super Sani-Cloth Wipes. Fundamental sensing technology unchanged from predicate.

## Regulatory Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

## Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Accu-Chek Guide Blood Glucose Monitoring System (k160944)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

SPECIAL 510(K): DEVICE MODIFICATION OIR DECISION SUMMARY

510(k) Number: k181131

This 510(k) submission contains information/data on modifications made to the applicant’s own class II or class I devices requiring 510(k). The following items are present and acceptable:

1. The name and 510(k) number of the applicant’s previously cleared device: Accu-Chek Guide Blood Glucose Monitoring System (k160944). (For a preamendments device, a statement to this effect has been provided.)

2. Applicant’s statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).

3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user’s and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.

The changes were for the following items:

- Removal of the target range indication, pattern detection, and meal indication features.
- Removal of the beep associated with reminders
- Removal of the test strip ejector and test strip port light
- Modified the LCD display from dot matrix to fixed segment and removed the backlight
- Reduced the number and type of meter buttons
- Reduced the BLE pairing capability of the meter from paring with 5 devices at a time to one device
- Change in external meter materials.

4. Comparison Information (similarities and differences) to applicant’s legally marketed predicate device including, labeling, intended use, physical characteristics, and specifications.

5. A Design Control Activities Summary which includes:

a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied

The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter’s description of the particular modification(s) and the comparative information between the modified and unmodified

1

{1}

2

devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.

The Accu-Chek Guide Me Blood Glucose Monitoring System is intended for single-patient use only. Disinfection efficacy studies were conducted on external meter materials and demonstrated complete inactivation of hepatitis B Virus (HBV) with Super Sani-Cloth Wipes, EPA Registration No. 9480-4. The sponsor also conducted robustness studies and demonstrated that there was no change in performance or in the external materials of the meter after 260 cleaning and disinfection cycles representing 5 years of single patient use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.

---

**Source:** [https://fda-staging.innolitics.com/device/K181131](https://fda-staging.innolitics.com/device/K181131)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).