The CARDIOSAVE IABP (Hybrid Model) is an electromechanical system used to inflate and deflate intra-aortic balloons. The IABP provides temporary support to the left ventricle via the principle of counterpulsation. The intra-aortic balloon is placed in the descending aorta, just distal to the left subclavian artery. Once the balloon is positioned, the pump is adjusted to trigger in synchrony with the ECG or arterial pressure waveform to ensure that inflation and deflation occur at the appropriate points during the cardiac cycle. The target population is adult. The balloon pump is intended for use in the health care facility setting.
Device Story
Cardiac assist device; supports left ventricle via counterpulsation; increases coronary perfusion and reduces left ventricular work. Inputs: ECG or arterial pressure waveforms. Operation: synchronizes inflation/deflation of intra-aortic balloon (IAB) in descending aorta with cardiac cycle; IAB inflates during diastole and deflates during systole. Used in healthcare facilities by clinicians. Modification: addition of Top Protection Cover to prevent liquid ingress into internal circuit components, connector receptacles, and air vents; prevents electrical shorts or device shutdown during therapy. Benefits: maintains continuous cardiac support by protecting electronics from fluid spills.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Electromechanical intra-aortic balloon pump. Modification includes a physical Top Protection Cover to prevent liquid ingress. System utilizes ECG or arterial pressure waveform triggering for counterpulsation. No changes to core sensing, energy source, or software algorithms.
Indications for Use
Indicated for adult patients with acute coronary syndrome, cardiac or non-cardiac surgery, or complications of heart failure.
Regulatory Classification
Identification
An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.
Special Controls
*Classification.* (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device;
(ii) Software verification, validation, and hazard analysis must be performed;
(iii) The device must be demonstrated to be biocompatible;
(iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components;
(v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and
(vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
May 31, 2018
Datascope Corp. Hemang Kotecha Regulatory Affairs Specialist II 1300 MacArthur Blvd. Mahwah, New Jersey 07430
Re: K181122
Trade/Device Name: CARDIOSAVE Intra-Aortic Balloon Pump (Hybrid Model) Regulation Number: 21 CFR 870.3535 Regulation Name: Intra-aortic balloon and control system Regulatory Class: Class II Product Code: DSP Dated: April 27, 2018 Received: April 30, 2018
Dear Hemang Kotecha:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Nicole G. Ibrahim -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K181122
Device Name CARDIOSAVE IABP (Hybrid Model)
Indications for Use (Describe)
The CARDIOSAVE IABP (Hybrid Model) is indicated for acute coronary syndrome, cardiac surgery, or complications of heart failure.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
## CARDIOSAVE INTRA-AORTIC BALLOON PUMP (HYBRID MODEL)
[Prepared in accordance with 21 CFR Part 807.92]
| Submitter's Name<br>and Address: | Datascope Corp.<br>1300 MacArthur Blvd.<br>Mahwah, NJ 07430 |
|------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person<br>for this<br>submission: | Hemang Kotecha<br>Regulatory Affairs Specialist II<br>Datascope Corp<br>1300 MacArthur Blvd.<br>Mahwah, NJ 07430<br>Phone: (973) 709-7376<br>Fax: (973) 629-1482<br>hemang.kotecha@getinge.com |
| Date Prepared: | April 27, 2018 |
| Device<br>Information: | Trade Name: CARDIOSAVE IABP (Hybrid Model)<br>Common/Generic Name: Intra-Aortic Balloon Pump<br>Classification Name: Intra-Aortic Balloon and control system<br>Regulation Number: 21 CFR 870.3535<br>Regulatory Class: Class II<br>Product Code: DSP |
| Predicate Device<br>Information: | CARDIOSAVE Intra-Aortic Balloon Pump (K112372) |
| Device<br>Description: | The CARDIOSAVE IABP (Hybrid Model) is a cardiac assist<br>device. It supports the heart's left ventricle by increasing<br>coronary perfusion and reducing left ventricular work. Coronary<br>perfusion is increased by augmenting blood pressure during<br>the diastolic phase of the cardiac cycle. This increase in aortic<br>pressure promotes more blood flow through the coronary |
| Datascope Corp. | Special 510(k) : Corrective Action Being Affected<br>CARDIOSAVE IABP Modification: Top Protection Cover |
| | arteries. Left ventricular work is reduced by decreasing aortic<br>end-diastolic pressure and reducing resistance to ventricular<br>ejection, resulting in a decrease in blood pressure during the<br>systolic phase of the cardiac cycle. |
| | These beneficial effects are caused by the inflation and<br>deflation of an intra-aortic balloon (IAB) in the patient's<br>descending aorta. The balloon's inflation and deflation must be<br>properly synchronized with the cardiac cycle. IAB inflation is<br>initiated at the onset of diastole at the dicrotic notch and<br>remains inflated through diastole. The IAB is then deflated at,<br>or just prior to, the onset of systole and the balloon remains<br>deflated throughout systole. Hence, the therapy is also referred<br>to as counterpulsation. This is the same intended use as other<br>IABPs. |
| Indications For<br>Use: | The CARDIOSAVE IABP (Hybrid Model) is indicated for acute<br>coronary syndrome, cardiac and non-cardiac surgery, or<br>complications of heart failure. |
| Intended Use: | The CARDIOSAVE IABP (Hybrid Model) is an<br>electromechanical system used to inflate and deflate intra-<br>aortic balloons. The IABP provides temporary support to the left<br>ventricle via the principle of counterpulsation. The intra-aortic<br>balloon is placed in the descending aorta, just distal to the left<br>subclavian artery. Once the balloon is positioned, the pump is<br>adjusted to trigger in synchrony with the ECG or arterial<br>pressure waveform to ensure that inflation and deflation occur<br>at the appropriate points during the cardiac cycle. |
| | The target population is adult. The balloon pump is intended for<br>use in the health care facility setting. |
| Comparison of<br>Technological<br>Characteristics<br>with the Predicate<br>Device: | The proposed CARDIOSAVE IABP (Hybrid Model), which is<br>the subject of this submission, contains the same<br>technologically characteristics as the predicate CARDIOSAVE<br>Intra-Aortic Balloon Pump (K112372). |
| | The proposed modification is the addition a new Top Protection<br>Cover that shall be installed on top of the CARDIOSAVE cart. |
| | The Top Protection Cover provides protection against liquid<br>spills, such as saline, that may occur on top of the unit. It is |
| Datascope Corp. | Special 510(k) : Corrective Action Being Affected<br>CARDIOSAVE IABP Modification: Top Protection Cover |
| | designed to remove any paths for the liquid that is spilled on top<br>of the unit and prevent flow of the liquid into the interfacing<br>panels, connector receptacles and air vents. Such liquid spills are<br>hazardous as they can create bridges of resistance between the<br>circuit components; causing the circuit not to function as<br>intended; potentially damage the unit and possibly causing it to<br>shut down during operation. This may impact the initiation or<br>continuation of therapy. |
| Performance<br>Data: | Datascope Corp. development process required that the<br>following activities be completed during the development of the<br>new Top Protection Cover:<br>• Risk Assessment<br>• Verification testing |
| | The modification proposed within this submission for<br>CARDIOSAVE IABP (Hybrid Model) does not impact the<br>performance characteristics of the predicate device. |
| | The results of the tests conducted demonstrate that the<br>proposed device met its respective acceptance criteria.<br>According to the risk assessment performed, the modification<br>does not introduce any new risks and does not raise any new<br>questions of safety and effectiveness. |
| Clinical Tests: | There was no clinical testing deemed necessary for the<br>proposed device. |
| Conclusion: | Based upon the information submitted in this Special 510(k)<br>submission, the proposed device is substantially equivalent to<br>the currently marketed CARDIOSAVE IABP (Hybrid Model). |
{4}------------------------------------------------
{5}------------------------------------------------
Protection Cover is similar to the predicate device in the intended use and the fundamental scientific technology of the device. The design verification testing established that the CARDIOSAVE IĂBP (Hybrid Model) with the new Top Protection Cover performs same as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.