The device is intended to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities(arm) and lower extremities (leg) due to strain from exercise or normal household work activities. And to be used for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
Device Story
Wireless, two-channel wearable TENS device; delivers electrical stimulation to muscles/joints via self-adhesive hydrogel electrodes. Input: user-selected TENS programs (23 total) via mobile app (iOS/Android) or dedicated remote control. Communication: Bluetooth Low Energy (4.0). Operation: device generates biphasic, rectangular pulses; controlled by patient at home. Output: electrical pulses for pain relief. Healthcare provider role: none; patient self-use. Benefits: non-invasive, convenient pain management for chronic/acute muscle/joint pain.
Clinical Evidence
Bench testing only. No clinical data. Performance supported by biocompatibility (ISO 10993-1, -5, -10), electrical safety (IEC 60601-1, -11), EMC (IEC 60601-1-2), and output waveform verification (IEC 60601-2-10). Software verification and validation conducted per FDA guidance.
Technological Characteristics
Wireless TENS stimulator; ABS housing; rechargeable Li-ion battery (main) and AAA batteries (remote). Sensing/actuation: current-controlled biphasic rectangular pulses. Connectivity: Bluetooth Low Energy (4.0). Sterilization: N/A (non-sterile). Software: microprocessor-controlled with 23 TENS programs. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10, ISO 10993.
Indications for Use
Indicated for adults of all genders for temporary relief of pain in shoulder, waist, back, arms, and legs due to exercise or household strain; symptomatic relief/management of chronic, intractable pain; and relief of pain associated with arthritis.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 15, 2018
Pfizer Inc. % Siping Yuan R.A. Specialist Shenzhen Dongdixin Technology Co., Ltd No. 3 Building, Xilibaimang Xusheng Industrial Estate Nanshan, Shenzhen, Cn 518108
Re: K181102
Trade/Device Name: ThermaCare Quick Therapy TENS, TC001 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NYN Dated: April 20, 2018 Received: April 26, 2018
Dear Siping Yuan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Pamela D. Scott -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K181102
Device Name ThermaCare® Quick Therapy TENS Model: TC001
#### Indications for Use (Describe)
The device is intended to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities(arm) and lower extremities (leg) due to strain from exercise or normal household work activities. And to be used for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <span> <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> </span> <span> <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
005_510 (k) Summary Version: 1.1
## 510(k) SUMMARY
as required by section 21 CFR 807.92
ThermaCare® Quick Therapy TENS
| Date of Submission: | 04/23/2018 |
|-----------------------------------|-----------------------------------------------------------------------------------------|
| Submitter/510(k)<br>Owner's Name: | Pfizer Inc. |
| Address: | 235 East 42nd Street, New York, New York 10017,<br>USA |
| Contact Person: | Nicola Romano |
| E-mail: | nicola.romano@pfizer.com |
| Tel: | 973-660-5858 |
| Correspondent's Name: | Shenzhen Dongdixin Technology Co., Ltd. |
| Address: | No.3 Building Xilibaimang Xusheng Industrial Estate,<br>Nanshan, Shenzhen, CHINA 518108 |
| Correspondent<br>Contact: | Siping Yuan |
| E-mail: | ra@dundex.com |
| Tel: | +86(755) 27652471 |
| FAX: | +86(755) 27652674 |
{4}------------------------------------------------
#### 1. Proposed Device:
| Device Name: | ThermaCare® Quick Therapy TENS |
|-------------------------------|------------------------------------------------------------|
| Model: | TC001 |
| Device classification Name: | Stimulator, Nerve, Transcutaneous, Over-The-Counter |
| Regulation Description: | Transcutaneous electrical nerve stimulator for pain relief |
| Regulation Medical Specialty: | Neurology |
| Review Panel: | Neurology |
| Regulation Number: | 882.5890 |
| Product Code: | NUH |
| Device Class: | II |
| Device classification Name: | Stimulator, Electrical, Transcutaneous, For Arthritis |
| Regulation Description: | Transcutaneous electrical nerve stimulator for pain relief |
| Regulation Medical Specialty: | Neurology |
| Medical Specialty: | Neurology |
|--------------------|-----------|
| Review Panel: | Neurology |
| Regulation Number: | 882.5890 |
| Product Code: | NYN |
| Device Class: | II |
#### 2. Predicate Device:
| Legally Marketed Device: | Wireless Pain Relieve Device |
|--------------------------|----------------------------------------|
| | LT5018C |
| 510(k) Number: | K173462 |
| Manufacturer: | Shenzhen Dondixin Technology Co., Ltd. |
#### 3. Description of Proposed Device:
The ThermaCare® Quick Therapy TENS device is a wireless, two channels wearable electrotherapy device that is designed to alleviate chronic muscle and joint pain on multiple body locations. It delivers TENS (Transcutaneous Electrical Nerve Stimulation) technology through the simple, convenient use of the dedicated APP (iOS or Android) or remote control. Reusable, self-adhesive pads allow for discreet and convenient placement on multiple pain locations on the body. The ThermaCare® Quick Therapy TENS contains 23 TENS programs. It is designed to be used at home, by adults of all genders.
The ThermaCare® Quick Therapy TENS device contains one main TENS unit which is rechargeable and one remote control which is powered by 2 AAA batteries. The pads with attached TENS unit can then be applied to intact skin at the desired location for pain relief. Control of the ThermaCare® Quick Therapy TENS device is completed through the available TENS APP (iOS or Android)or remote control. The
{5}------------------------------------------------
communication is done via Bluetooth Low Energy.
The device will be packaged with an Instruction Manual which provides details on setting up the device for use, installing of the App, setting and controlling therapy programs, and troubleshooting. The ThermaCare® Quick Therapy TENS consists of the following elements:
• Main device: generates the TENS pulses; connects to the electrode; controlled by APP on smart phone and remote control.
• Electrodes: Self-adhesive re-usable hydrogel electrodes connects to the device.
• APP on mobile device: The APP controls the functions of the main device via a Bluetooth (4.0) Low Energy connection between the mobile device and the main device.
• Remote control: The remote controls the functions of the main device via a Bluetooth (4.0) Low Energy connection between the remote control and the main device.
· Adapter (optional): for charging the battery unit via a micro-USB socket which connected with the adapter, and insert the adapter to the supply mains.
#### 4. Proposed Device Intended for Use Statement:
Device Name: ThermaCare® Quick Therapy TENS, Model: TC001
#### Indications for Use:
The device is intended to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities(arm) and lower extremities (leg) due to strain from exercise or normal household work activities. And to be used for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
{6}------------------------------------------------
#### 5. Technological Characteristics and Substantial Equivalence
Both the TC001 and Predicate Device utilize the application of electrical current through electrodes placed on the impulses are generated by the device and delivered through electrodes on the skin in direct proximity to the (painful) muscles to be stimulated. Basic technological characteristics, new device vs. Predicate device.
| | | New device | Predicate device | S.E. Discussion |
|----------|----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | 510K# | To be assigned | K173462 | N/A |
| 2 | Device Name and<br>Model | ThermaCare® Quick Therapy TENS<br>Mode: TC001 | Wireless Pain Relieve Device<br>Mode: LT5018C | N/A |
| 3 | Applicant | Pfizer Inc. | Shenzhen Dongdixin Technology Co., Ltd | N/A |
| 4 | Manufacturer | Shenzhen Dongdixin Technology Co., Ltd | Shenzhen Dongdixin Technology Co., Ltd | Same |
| 5 | Intended for use | TENS: The device is designed to be used<br>for temporary relief of pain associated<br>with sore and aching muscles in the<br>shoulder, waist, back, upper extremities<br>(arm) and lower extremities (leg) due to<br>strain from exercise or normal household<br>work activities. And to be used for the<br>symptomatic relief and management of<br>chronic, intractable pain and relief of pain<br>associated with arthritis. | TENS: The device is designed to be used<br>for temporary relief of pain associated<br>with sore and aching muscles in the<br>shoulder, waist, back, neck, upper<br>extremities (arm) and lower extremities<br>(leg) due to strain from exercise or<br>normal household work activities. And<br>to be used for the symptomatic relief<br>and management of chronic, intractable<br>pain and relief of pain associated with<br>arthritis.<br>EMS: The device is designed to be used<br>for stimulate healthy muscles in order to<br>improve and facilitate muscle<br>performance. | Applicant is not pursuing<br>EMS use statement for the<br>new device. |
| 6 | Power Source | DC 3.7V Li-ion Battery(main device)<br>DC 3.0V, 2 x AAA Batteries (Remote<br>control) | DC 3.7V Li-ion Battery(main device)<br>DC 3.0V, 2 x AAA Batteries (Remote<br>control) | Same |
| | -Method of<br>Line | N/A | N/A | Same |
| | current isolation | | | |
| | - Patient Leakage<br>Current (μΑ)<br>-Normal condition<br>-Single fault<br>condition | d.c:0 uA. a.c.:29 uA max<br>d.c:0 uA. a.c.:69 uA max | 0uA<br>2.0uA | Different, TC 001 have been<br>evaluated and passed the test<br>according to the requirement<br>of IEC 60601-1. |
| 7 | Average DC current<br>through electrodes<br>when device is on but<br>no pulses are being<br>applied (μΑ) | 0 | 0 | Same |
| 8 | Number of Output<br>Modes | 1 (TENS) | 2 (TENS/EMS(including MASSAGE)) | Different, TC001 only has<br>TENS mode and TC001<br>remove indications for use of<br>EMS mode accordingly. |
| 9 | Number of Output Channels | | | |
| | Synchronous or<br>Alternating? | Alternating | N/A | Different, TC001 has two<br>channels and each channel<br>outputs have been evaluated<br>and passed the test according<br>to the requirement of IEC<br>60601-2-10. |
| | Method of Channel<br>Isolation | By Electrical Circuit and Software | N/A | Different, TC001 has two<br>channels and have been<br>evaluated and passed the test<br>according to the requirement<br>of IEC 60601-1. |
| 10 | Regulated Current or<br>Regulated Voltage? | Current control | Current control | Same |
| 11 | Software/Firmware/<br>Micro processor<br>Control? | Yes | Yes | Same |
| 12 | Automatic Overload<br>Trip | Yes | Yes | Same |
| 13 | Automatic No Load<br>contact Trip | Yes | Yes | Same |
| 14 | Automatic Shut off | Yes | Yes | Same |
| 15 | User Override<br>Control? | Yes<br>Power [on/off] button on the device<br>Power [on/off] button on the remote<br>control<br>[Stop session] button in the APP software | Yes<br>Power on/off button on the device<br>Power on/off on the remote control<br>Power on/off in the APP software | Same function |
| 16 | Indicator Display: | | | |
| | On/Off Status? | Yes | Yes | Same |
| | Low Battery? | Yes | Yes | Same |
| | Voltage/<br>Current<br>Level? | Yes | Yes | Same |
| 17 | Timer<br>Range<br>(minutes) | 5-30 minutes | 30 minutes | Different, new device have<br>been evaluated and passed<br>the test according to the<br>requirement of IEC<br>60601-2-10.<br>User can set the treatment<br>time according their needs. |
| 18 | Weight (grams.) | 32grams (Main device)<br>55 grams (Remote control without<br>batteries) | 36grams (Main device)<br>65 grams (Remote control) | Different, the new device<br>have been evaluated and |
| 19 | Dimensions<br>(cm.)<br>H*W * L | 80.5(L)x45(W)x13.5(H)mm (Main<br>device)<br>127(L)x48.6(W)x24.4(H)mm (Remote<br>control) | 360(L)x59(W)x11.5(H)mm<br>(Main<br>device)<br>115(L)x53(W)x25(H)mm<br>(Remote<br>control) | passed the testing according<br>to the requirement of IEC<br>60601-1. |
| 20 | Housing Materials & | ABS | ABS | Same |
| | | | | |
| | Construction | | | |
| 21 | Compliance with 21<br>CFR 898 | Yes | Yes | Same |
| 22 | Compliance<br>with | IEC 60601-1:2005+A1:2012 | IEC 60601-1:2005+A1:2012 | Same |
| | Voluntary Standards? | IEC 60601-1-2:2014 | IEC 60601-1-2:2014 | |
| | | IEC 60601-2-10: 2012 | IEC 60601-2-10: 2012 | |
| | | IEC60601-1-11: 2015 | IEC60601-1-11: 2015 | |
| | | ISO 10993-1:2009 | ISO 10993-1:2009 | |
| | | ISO 10993-5:2009 | ISO 10993-5:2009 | |
| | | ISO 10993-10:2010 | ISO 10993-10:2010 | |
| | Table 2: Output Specification TENS mode | | | |
| | | New device | Predicate device | S.E. Discussion |
| 1 | 510K# | To be assigned | K173462 | N/A |
| 2 | Device Name<br>or | ThermaCare® Quick Therapy TENS | Wireless Pain Relieve Device | N / A |
| | Program Name | Mode:TC001 | Mode:LT5018C | |
| 4 | Waveform | Biphasic | Biphasic | Same |
| 5 | Shape | Rectangular | Rectangular | Same |
| 6 | Max Output Voltage (V) ±20% | | | |
| 7 | 500Ω | 30.4 | 31 | Similar, the new device have |
| 8 | 2kΩ | 70 | 66 | been evaluated and passed |
| 9 | 10kΩ | 70 | 66 | the test according<br>to the |
| | | | | requirement<br>IEC<br>of…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.