Invacare® Solara® 3G and Solara® 3G Spree Manual Wheelchairs
K181090 · Invacare Corporation · IOR · Nov 14, 2018 · Physical Medicine
Device Facts
Record ID
K181090
Device Name
Invacare® Solara® 3G and Solara® 3G Spree Manual Wheelchairs
Applicant
Invacare Corporation
Product Code
IOR · Physical Medicine
Decision Date
Nov 14, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Indications for Use
The Invacare® Solara® 3G Manual Wheelchair is intended to provide mobility to persons limited to a sitting position.
Device Story
Mechanical, non-folding, attendant-operated or self-propelled manual wheelchair for adults. Features 'Tilt in Space' (seat/back tilt -5° to 50°) for stability, comfort, and head control; includes Low Shear Recline system to reduce shear forces. Operated by patient or attendant in clinical or home settings. Provides mobility to users with limited sitting ability. Output is physical mobility and postural support. Benefits include improved stability, pressure management, and ease of caregiving.
Clinical Evidence
No clinical data. Bench testing only. Performance demonstrated via ANSI/RESNA WC-1:2009 standards (static stability, brake effectiveness, dimensions, maneuvering space, seating/wheel dimensions, static/impact/fatigue strength, coefficient of friction). Biocompatibility testing per ISO 10993-5 and ISO 10993-10. Flammability testing per CAL 117 and ISO 8191.
Technological Characteristics
Mechanical manual wheelchair; rigid non-folding frame. Materials include steel tubing and steel plates. Features Tilt-in-Space (dual cable trigger or foot pedal) and Low Shear Recline (locking gas cylinders). Weight capacities: 200 lbs (Spree), 300 lbs, 400 lbs (Heavy Duty). Tested per ANSI/RESNA WC-1:2009. Biocompatible per ISO 10993.
Indications for Use
Indicated for adults limited to a sitting position requiring mobility assistance. Includes patients needing tilt-in-space features for stability, comfort, head control, or as an attendant aid for care/feeding.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Invacare® Solara® Jr. Manual Wheelchairs (K012370)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of it is the FDA logo, with the letters 'FDA' in a blue square, followed by 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
November 14, 2018
Invacare Corporation Elijah Wreh Regulatory Affairs Manager One Invacare Way Elyria, Ohio 44035
Re: K181090
Trade/Device Name: Invacare® Solara® 3G and Solara® 3G Spree Manual Wheelchairs Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: IOR Dated: August 13, 2018 Received: August 16, 2018
Dear Elijah Wreh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vivek J. Pinto -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181090
#### Device Name
Invacare® Solara® 3G and Solara® 3G Spree Manual Wheelchairs
Indications for Use (Describe)
The Invacare® Solara® 3G Manual Wheelchair and Invacare® Solara® 3G Spree Manual Wheelchair are intended to provide mobility to adults limited to a sitting position.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Invacare logo. The logo consists of the word "INVACARE" in white, bold, sans-serif font, set against a blue oval background. The oval is horizontally oriented and slightly tilted upwards to the right.
# Yes, you can:
# 510(k) Summary - K181090
# SUBMITTER per 21 CFR 807.92(a)(1)
Invacare Corporation One Invacare Way Elyria, OH 44035
| PRIMARY CONTACT<br>PERSON: | Elijah Wreh<br>Regulatory Affairs Manager<br>Phone: (440) 329-6840<br>Email: <a href="mailto:ewreh@invacare.com">ewreh@invacare.com</a> |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| SECONDARY CONTACT<br>PERSON: | Kevin Malinowski<br>Senior Quality Engineer<br>Phone: (440) 329-6179<br>Email: <a href="mailto:kmalinowski@invacare.com">kmalinowski@invacare.com</a> |
| MANUFACTURER: | INVAMEX S.DE R.L. DE C.V.<br>Avenida El Puente S/N<br>Parque Industrial El Puente Manimex<br>Reynosa Tamaulipas, Mexico C.P. 88783 |
### Date Prepared per 21 CFR 807.92(a)(1)
November 14, 2018
#### DEVICE INFORMATION per 21 CFR 807.92(a)(2)
Name of Device: Common or Usual Name: Regulation Description: Regulation Number: Review Panel: Device Class: Product Code:
Solara® 3G and Solara® 3G Spree Manual Wheelchairs Wheelchair, Mechanical Mechanical Wheelchair 21 CFR §890.3850 Physical Medicine 1 IOR
{4}------------------------------------------------
#### PRIMARY PREDICATE Invacare® Solara® Manual Wheelchairs (K984447) DEVICE:
#### SECONDARY PREDICATE Invacare® Solara® Jr. Manual Wheelchairs (K012370) DEVICE:
#### DEVICE DESCRIPTION per 21 CFR 808.92(a)(4)
The Invacare® Solara® 3G Manual Wheelchair is an attendant operated and self-propelled, mechanical wheelchair for adults. The indication for use of the Invacare® Solara® 3G Manual Wheelchair is to provide mobility to adults limited to a sitting position. The indication for use of the Invacare® Solara® 3G Spree Manual Wheelchair is to provide mobility to person limited to a sitting position. These devices are rigid, "non-folding" type wheelchairs that incorporate a solid seating surface (seat pan).
Below is the subject devices weight capacity:
- Invacare® Solara® 3G Manual Wheelchair weight capacity is 300 lbs. .
- Invacare® Solara® 3G Manual Wheelchair with Heavy Duty weight capacity is 400 . lbs.
Invacare® Solara® 3G Spree Manual Wheelchair weight capacity is 200 lbs. . Invacare® Solara® 3G and Invacare® Solara® 3G Spree Manual Wheelchair include a "Tilt in Space" feature which allows the seat and back of the wheelchair to be tilted backward. This feature is used for those patients who require a tilt feature for stability, comfort or head control. It also serves as an attendant aid in those situations where the patient needs to be fed or receive care. Both predicate devices include the "Tilt in Space" feature.
Both subject devices include a Low Shear Recline which was developed to reduce the amount of shear experienced by the user versus the single-pivot recline system incorporated into the predicate devices.
The associated model and accessories include:
- . Invacare® Solara® 3G Manual Wheelchair o weight capacity is 300 lbs.
- . Invacare® Solara® 3G Manual Wheelchair with Heavy Duty
- o weight capacity is 400 lbs.
- Invacare® Solara® 3G Spree Manual Wheelchair ●
- o weight capacity is 200 lbs.
{5}------------------------------------------------
- . Stealth™ Headrest
- . Matrx™ Cushions
- Straps (seat, hook and loop, airline buckle)
- . Loops (Heal, Toe)
- . Anti-Tippers
- . Telescoping Rod
# INTENDED USE per 21 CFR 807.92(A)(5)
The Invacare® Solara® 3G Manual Wheelchair is intended to provide mobility to persons limited to a sitting position.
### INDICATIONS FOR USE per FORM FDA 3881
The Invacare® Solara® 3G Manual Wheelchair and Invacare® Solara® 3G Spree Manual Wheelchair are intended to provide mobility to adults limited to a sitting position.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES per 21 CFR 807.92(a)(6)
The device comparison showed that the subject device is substantially equivalent in intended use, design and operational principles to the previously Invacare® Solara® Jr. Manual Wheelchairs (K012370) and Invacare® Solara® Manual Wheelchairs (K984447). The subject device is substantially equivalent to the predicate devices regarding intended use, design, materials, and operational principles to provide mobility to persons limited to a sitting position.
# BASIS OF SUBSTANTIAL EQUIVALENCE per 21 CFR 807.100(b)(2)(ii)(A)
The substantial equivalence of the subject device was determined as per the FDA guidance document, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" and the technological characteristics which include materials, design, energy source, and other device features, as defined in section 513(i)(1)(B) of the FD&C Act and 21 CFR 807.100(b)(2)(ii)(A).
The subject device components are as safe and effective as the predicate devices and do not raise different questions of safety and effectiveness.
The major frame components for the subject Invacare® Solara® 3G and Invacare® Solara® 3G Spree Manual Wheelchair and the previously cleared Invacare® Solara® Jr. Manual
{6}------------------------------------------------
Wheelchairs (K012370) and Invacare® Solara® Manual Wheelchairs (K984447) are the same. The data generated from the subject device test report supports a finding of substantial equivalence regarding the device comparison, dimensional analysis, device specifications, design characteristics and to provide mobility to persons limited to a sitting position. The difference in weight capacity between the subject Invacare® Solara® 3G and Invacare® Solara® 3G Spree Manual Wheelchair and the previously cleared Invacare® Solara® Jr. Manual Wheelchairs (K012370) and Invacare® Solara® Manual Wheelchairs (K984447) do not raise new issues of safety or effectiveness.
{7}------------------------------------------------
#### Design Comparison – Invacare® Solara® 3G Manual Wheelchair with 300 lbs Weight Capacity
| Device | Invacare® Solara® 3G Manual Wheelchair | Invacare® Solara® Manual Wheelchairs |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| 510(k) Number | Subject Device<br>Pending Submission | Predicate Device<br>(K984447) |
| Intended Use | The Invacare® Solara® 3G Manual Wheelchair is intended to provide<br>mobility to persons limited to a sitting position. | Intended to provide mobility to persons limited to a sitting position. |
| Indications for Use | The Invacare® Solara® 3G Manual Wheelchair and Invacare® Solara® 3G<br>Spree Manual Wheelchair are intended to provide mobility to adults limited<br>to a sitting position. | To provide mobility to adults limited to a sitting position |
| Design | | |
| Weight Limit | 300 lbs. | 250 lbs.<br>300 lbs. (With Heavy Duty Package) |
| Seat Width | 12"-24" | 14"-20" |
| Seat Depth | 12"-22" | 14"-18" |
| Back Style | Standard Back Cane<br>1Low Sheer Recline (not available on 400 lbs. weight limit wheelchair) | Standard Back Cane |
| Back Height | 17"-20" Adjustable: 20" or 24" fixed. | 20" and 24" |
| Anti-Tippers | Standard | Standard |
| Arm Types | Full Adjustable Height, Desk Cantilever | Full Adjustable Height, Desk Cantilever |
| Wheel Locks | Push-to-Lock (Standard) or Pull-to-Lock, Foot lock | Push-to-Lock (Standard) |
| Turning Radius | 25" | 25" |
| Rear Wheel Sizes | 12"-24" | 12"-22" |
| Front riggings | Swing away | Swing away |
| Product Weight | 34.5 lbs. without accessories | 50 lbs. |
| Tilt Range (Tilt in Space Feature) | -5° to 50° | -5° to 50° |
| Tilt-in-Space | Dual Cable Trigger Mechanism or Foot Pedal Mechanism (Standard) | Dual Cable Trigger Mechanism |
1 Note: The Low Shear Recline is designed to reace the anom of shear experienced by the user versus the system incorporated into the predicate device. The Low Sheur Recline for similar construction as the single point recline design incorporated into both predicate devices, brase, bras, locking gas cylinders and nutbolt feature is operated by two triggers that enggedissenge the locking cylinders wa adjustable ables. This is the same operating neckanism as wed in the predicate devices. Design Verfication testing demonstrates that the fective ind found to be substantially equivalent to the predicate devices throughout this 510(k) submission.
{8}------------------------------------------------
#### Design Comparison - Invacare® Solara® 3G Manual Wheelchair 400 lbs Weight Capacity
| Device | Invacare® Solara® 3G Manual Wheelchair | Invacare® Solara® Manual Wheelchairs |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| 510(k) Number | Subject Device<br>Pending Submission | Predicate Device<br>(K984447) |
| Intended Use | The Invacare® Solara® 3G Manual Wheelchair is intended to provide<br>mobility to persons limited to a sitting position. | Intended to provide mobility to persons limited to a sitting position. |
| Indications for Use | The Invacare® Solara® 3G Manual Wheelchair and Invacare® Solara® 3G<br>Spree Manual Wheelchair are intended to provide mobility to adults limited<br>to a sitting position. | To provide mobility to adults limited to a sitting position |
| Design | | |
| Weight Limit | 1400 lbs. (With Heavy Duty Package) | 250 lbs.<br>300 lbs. (With Heavy Duty Package) |
| Seat Width | 12"-24" | 14"-20" |
| Seat Depth | 12"-22" | 14"-18" |
| Back Style | Standard Back Cane<br>2Low Sheer Recline (not available on 400 lbs. weight limit wheelchair) | Standard Back Cane |
| Back Height | 17"-20" Adjustable: 20" or 24" fixed. | 20" and 24" |
| Anti-Tippers | Standard | Standard |
| Arm Types | Full Adjustable Height, Desk Cantilever | Full Adjustable Height, Desk Cantilever |
| Wheel Locks | Push-to-Lock (Standard) or Pull-to-Lock, Foot lock | Push-to-Lock (Standard) |
| Turning Radius | 25" | 25" |
| Rear Wheel Sizes | 12"-24" | 12"-22" |
| Front riggings | Swing away | Swing away |
| Product Weight | 34.5 lbs. without accessories | 50 lbs. |
| Tilt Range (Tilt in Space Feature) | -5° to 50° | -5° to 50° |
| Tilt-in-Space | Dual Cable Trigger Mechanism or Foot Pedal Mechanism (Standard) | Dual Cable Trigger Mechanism |
' Note: Invacare® Solara® 3G Manual Wheelcht copacity components and accessories are substantially equivalent to the predicate device itentified throughout this submission based on the technological che includes materials, design, operational principles, and neckanical properties per 21 CFR 807.100(b)(2)(ii).). The basic opentional principe device as well as the is to provide mobility to persons linited to a sitting position. Design Verfication testing according to ANSINESNA device components and accessories are substantially equivalent to the predicated over regarding static stubility, effectiveness of brakes, dimensivering space, neasurement of secting and wheel dimensions, static inpact and fairger strences of ficion, information and abeling. The subject device's risk profile remains unchanged and the ligher weight appacity as and raise different questions of safety and effectiveness.
4 Note: The Low Shear Recline for anount of thear experienced by the user versus the single-pivot recline system incorporated into the predicate device. The Low Shear Recline is of sinilar construction as the signing recined into both predicate devices (seel uling, steel plant, locking as cylinders and nutbolt feature is operated by two triggers that enggeddissenges the locking cylinders wa adjustable. This is the same operating mechanism as wed in the single of the predicate devices. Design Verfication testing demonstrates that the fective in 1 found to be substantially equivalent to the predicate devices throughout this 510(k) submission.
{9}------------------------------------------------
| Device | Invacare® Solara® 3G Spree Manual Wheelchair | Invacare® Solara® Jr. Manual Wheelchairs |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|
| 510(k) Number | Subject Device<br>Pending Submission | Predicate Device<br>(K012370) |
| Intended Use | The Invacare® Solara® 3G Manual Wheelchair is intended to provide<br>mobility to persons limited to a sitting position. | To provide mobility to adults limited to a sitting position. |
| Indications for Use | The Invacare® Solara® 3G Manual Wheelchair and Invacare® Solara® 3G<br>Spree Manual Wheelchair are intended to provide mobility to adults limited<br>to a sitting position. | Persons who may be restricted to a sitting position. |
| Design | | |
| Weight Limit | 1200 lbs. | 150 lbs. |
| Seat Width | 12"-18" | 10"-16" |
| Seat Depth | 12"-19" | 12"-18" |
| Back Style | Standard Back Cane<br>2Low Sheer Recline | Standard Back Cane<br>Recliner and Adjustable Angle |
| Back Height | 17"-20" Adjustable<br>20" and 24" 3Fixed | 20" or 24"<br>Adjustable |
| Arm Types | Full Adjustable Height, Desk Cantilever | Full Adjustable Height, Desk Cantilever |
| Rear Wheel Sizes | 12"-24" | 12"-22" |
| Front Riggings | Swing away | Swing away |
| Product Weight | 32 lbs. without accessories | 27.5 lbs. |
| Tilt Range (Tilt in Space Feature) | -5° to 50° | -5° to 50° |
#### Design Comparison - Invacare® Solara® 3G Spree Manual Wheelchair with 200 lbs Weight Capacity
- Note: Imacare® Solara® 3G Manual Wheelchar 2001 s weight components and accessories are substantially equivalent o the predicate device itentified throughout this submission based on the technistics which includes naterials, design, qperational principles, and nechanical proveties per CFR 807.100(b)(2)(ii).A). The basic overtional prices as vell as the subject devices is to provide no a sitting position. Design Verfication testing according to ANSIPESMA demonstrates that the components and accessories are substantially equivalent to the prearling static stability, effectivenes of brakes, and manewering space, neasurement of secting and wheel dimensions, static inpact and fairger strences of cofficient of ficion, information and abeling. The subject device's risk profile remains underged and the higher weight appear one not not not not not not not not not not not not not not n of safety and effectiveness.
4 Note: The Low Shear Recline for anount of thear experienced by the user versus the single-pivot recline system incorporated into the predicate device. The Low Shear Recline is of sinilar construction as the sign incryporated into both predicate clevices (steel tubing, steel plans, locking gas cylinders and nutbolt feature is operated by two triggers that enggeddissnage the locking cylinders via adjustable. This is the same operating mechanism as used in the predicate devices. Design Verfication testing demonstrates that the feature is safe and effective and found to be substantially equivalent to the predicate devices throughout these 510(k) submissions.
1 Note: The subject device fixed back height is similate back height because the fixed back height was the norst-case seencrio used for the testing. The Design Verification testing device fixed back height met the requirements acceptance criteria and do not raise diferent questions of safet and effectiveness. The only difference between the fixed back height is that the adjustable back height has the ability to increase the leight of the back cane.
{10}------------------------------------------------
# PERFORMANCE DATA
# Non-Clinical Test per 21 CFR 807.92(b)(1)
Rehabilitation Engineering and Assistive Technology Society of North America (RESNA) testing was performed to demonstrate that the subject device meets the performance requirements and is substantially equivalent to the predicate devices identified throughout this submission.
The following testing was performed:
- . ANSI/RESNA WC-1:2009 Section 1: Determination of Static Stability
- ANSI/RESNA WC-1:2009 Section 3: Determination of Effectiveness of Brakes
- ANSI/RESNA WC-1:2009 Section 5: Determination of Dimensions, Mass . and Maneuvering Space
- ANSI/RESNA WC-1:2009 Section 7: Method of Measurement of Seating and . Wheel Dimensions
- ANSI/RESNA WC-1:2009 Section 8: Requirements and Test Methods for Static, . Impact and Fatigue Strengths
- ANSI/RESNA WC-1 :2009 Section 15: Requirements for Information . Disclosure, Documentation and Labeling
- ANSI/RESNA WC-1:2009 Section 13: Determination of Coefficient of . Friction of Test Surfaces
- ANSI/RESNA WC-1:2009 Section 11: Test Dummies .
- CAL 117:2013, Section 1: Flammability Test
- · ISO 8191-1:1987 & 8191-2:1988: Flammability Testing
Design Verification Testing demonstrated that the subject Invacare® Solara® 3G Manual Wheelchair with weight capacities of 300 and 400 lbs. and Invacare® Solara® 3G Spree Manual Wheelchair with capacity of 200 lbs. are substantially equivalent to the marketed predicate devices.
### Biocompatibility Testing
The biocompatibility evaluation for the subject device was conducted in accordance with the FDA Blue Book Memorandum #G95 – 1 "Use of International Standard ISO – 10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993 – 1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
{11}------------------------------------------------
- . AAMI / ANSI / ISO 10993-5:2009, Biological Evaluation of Medical Devices - Part 5: Tests for in vitro Cytotoxicity
- AAMI / ANSI / ISO 10993-10:2010, Biological Evaluation of Medical Devices Part 10: . Tests for Skin Irritation
#### Risk Management
Risk Management has been conducted in accordance with ISO 14971:2012 - Medical Devices -Application of Risk Management to Medical Devices" for the subject Invacare® Solara® 3G Manual Wheelchair with weight capacities of 300 and 400 lbs. and Invacare® Solara® 3G Spree Manual Wheelchair with capacity of 200 lbs. The Category Hazard Analysis (CHA) provides guidance on the principal factors to consider in conducting a risk-based assessment to determine the Device Hazard Analysis (DHA) of the subject device's risk profile. The risk assessment Device Hazard Analysis (DHA)involves describing the relationships between a hazard (a potential source of harm) and the ultimate consequences in terms of physical injury or damage. Based on the risk-based assessment Device Hazard Analysis (DHA) performed, the subject device's risk profile remained unchanged and there is no significant impact on the safety or effectiveness of the subject device.
#### Animal Study
Animal testing was not required for this submission.
# Clinical Testing
Clinical testing was not required for this submission.
### CONCLUSIONS per 21 CFR 807.92(b)(3)
The subject device has the same intended use and similar technological characteristics as the predicate devices. The performance testing, device comparison, and dimensional analysis demonstrates that the subject device components are substantially equivalent to the predicate devices regarding static stability, effectiveness of brakes, dimensions, mass and maneuvering space, measurement of seating and wheel dimensions, static impact and fatigue strength, coefficient of friction, information disclosures, documentation and labeling. The non-clinical laboratory data support the safety and performance of the subject device and demonstrate that the subject device should perform as intended in the specified use conditions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.