K181089 · Authentic Medical · DSA · Jan 17, 2020 · Cardiovascular
Device Facts
Record ID
K181089
Device Name
SureLead Cable System
Applicant
Authentic Medical
Product Code
DSA · Cardiovascular
Decision Date
Jan 17, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2900
Device Class
Class 2
Indications for Use
The SureLead Disposable Cable System is intended to be used with ECG monitoring devices. The SureLead Disposable Cable System is used to connect electrodes, catheters, and/or sensors placed at appropriate sites on the patient to a monitoring device for generating monitoring and/or diagnostic evaluation by a health care professional.
Device Story
SureLead Disposable Cable System consists of shielded bulk cables with connectors; serves as replacement lead-wires for ECG monitors. Inputs: electrical signals from patient-attached electrodes/sensors. Transformation: passive signal transmission from patient to monitoring device. Output: raw ECG signal for monitor display/analysis. Used in clinical settings by healthcare professionals. Benefits: enables signal acquisition for diagnostic monitoring. Device is disposable; non-sterile; features integral strain relief.
Clinical Evidence
Bench testing only. Compliance verified against ISO 10993-1, ISO 10993-5, ISO 10993-10, AAMI/ANSI EC53:2013, and IEC 60601-1:2005 (including amendments/corrections). No clinical data provided.
Technological Characteristics
Materials: medical grade PVC and PP cable jackets; medical grade PVC and PP over-molded connectors. Sensing: passive electrical signal transmission. Connectivity: wired connection between electrodes and ECG monitor. Standards: ISO 10993, AAMI/ANSI EC53, IEC 60601-1. Configuration: 3, 5, or 6 lead-wire sets; disposable.
Indications for Use
Indicated for use with ECG monitoring devices to connect electrodes, catheters, or sensors to a monitor for diagnostic evaluation by a healthcare professional. No specific age or gender restrictions stated.
Regulatory Classification
Identification
A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 17, 2020
Authentic Medical Ricky Souza CEO 4470 Yankee Hill Rd., Ste 100 Rocklin, California 95677
Re: K181089
Trade/Device Name: SureLead Disposable Cable System Regulation Number: 21 CFR 870.2900 Regulation Name: Patient Transducer and Electrode Cable (Including Connector) Regulatory Class: Class II Product Code: DSA Dated: December 10, 2019 Received: December 11, 2019
Dear Ricky Souza:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Assistant Director (Acting) Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181089
Device Name SureLead Disposable Cable System
#### Indications for Use (Describe)
The SureLead Disposable Cable System is intended to be used with ECG monitoring devices. The SureLead Disposable Cable System is used to connect electrodes, catheters, and/or sensors placed at appropriate sites on the patient to a monitoring device for generating monitoring and/or diagnostic evaluation by a health care professional.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
SureLead Disposable Cable System Traditional 510(k)
## SureLead Disposable Cable System Premarket Notification 510(k) Summary
| DATE PREPARED | 19 APR 2018 |
|--------------------------------------|------------------------------------------------------------------------------|
| MANUFACTURER | Authentic Medical |
| CONTACT PERSON | Ricky Souza |
| | Position: CEO/President |
| | Tel: (916) 952- 6498 |
| | Fax: (916) 644-6015 |
| | Email: rsouza@authenticmed.com |
| PANEL CODE | Cardiovascular |
| CLASSIFICATION | Patient Transducer and Electrode Cable (Including<br>Connector) |
| CLASS | Class II (Class 2) |
| COMMON NAME | Cable / Lead-wire |
| TRADE NAME | SureLead Disposable Cable System |
| PREDICATE DEVICES | K170536 Cable/Lead-Wire (ECG). |
| Identification of<br>Proposed Device | Trade Name: Surelead Disposable cable system<br>Common Name: cable/lead-wire |
### DEVICE DESCRIPTION
The SureLead Disposable Cable System with specific various lengths are the replacements for similar cables manufactured by Original Equipment Manufacturers (OEM) and other third-party after-market manufacturers for their respective monitors. These cables consist of connectors on each cable end and a shielded bulk cable. These cables are used to transfer the signals from the electrodes to the patient monitor. The SureLead Disposable Cable System uses the same type of construction and have the same technological characteristics as the predicate device. The SureLead Disposable Cable System is made of medical grade PVC and PP cable jacket with medical grade PVC and PP over molded connectors with integral relief.
## INDICATIONS FOR USE
The SureLead Disposable Cable System is intended to be used with ECG monitoring devices. The SureLead Disposable Cable System is used to connect electrodes, catheters, and/or sensors placed at appropriate sites on the patient to a monitoring device for generating monitoring and/or diagnostic evaluation by a health care professional.
### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The SureLead Disposable Cable System, is substantially equivalent to the Predicate device (K170536) intended use / indications for use, materials, technological characteristics, and labelling.
{4}------------------------------------------------
## PERFORMANCE DATA
Non clinical tests were conducted to verify that the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate device (K170536). The test results demonstrated that the proposed device complies with the following standards:
- . ISO 10993-1:2009 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process
- . ISO 10993-5:2009 Biological Evaluation of Medical Devices- Part 5: Tests For In Vitro Cytotoxicity.
- . ISO 10993-10:2010 Biological Evaluation of Medical Devices- Part 10: Tests for Irritation and Skin Sensitization.
- AAMI/ANSI EC53: 2013 ECG Trunk Cables and Patient Leadwires .
- IEC 60601-1:2005+CORR.1: 2006+CORR. 2:2007+AM1: 2012, Medical Electrical . Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
# SUBSTANTIAL EQUIVALENCE DISCUSSION
The SureLead Disposable Cable System, (Subject Device) is substantially equivalent to K 170536 ECG Disposable Cable / lead-wire (Predicate Device).
The Subject device and Predicate Device:
- Have nearly Identical Indications for Use
- Are provided non-sterile with steam sterilization instructions
- · Use similar cable lengths and the same materials
Table 1 is provided here to show the comparison.
| | Proposed Device | Predicate Device<br>K170536 | Remarks |
|-------------------------------|--------------------------------------------------------------------|--------------------------------------------------------------------|---------|
| Manufacturer | Authentic Medical | APK Technology<br>Co., Ltd | - |
| 510(k) number | K181089 | K170536<br>Cable/Lead-wire | - |
| Classification<br>Regulation | 21 CFR Part 870.2900 | 21 CFR Part 870.2900 | SE |
| Panel | Cardiovascular | Cardiovascular | SE |
| Common Name | Cable / Lead-wire | Cable / Lead-wire | SE |
| Device<br>Classification Name | Patient transducer<br>and electrode cable<br>(including connector) | Patient transducer<br>and electrode cable<br>(including connector) | SE |
| Product Code | DSA | DSA | SE |
| Class | Class II (Class 2) | Class II (Class 2) | SE |
Table 1: Substantial Equivalence Summary Table
{5}------------------------------------------------
| Indications for Use | The SureLead | The APK ECG | SE |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | Disposable Cable<br>System is intended to<br>be used with ECG<br>monitoring devices.<br>The SureLead Cable<br>System is used to<br>connect electrodes,<br>catheters, and/or<br>sensors placed at<br>appropriate sites on the<br>patient to a monitoring<br>device for generating,<br>monitoring, and/or<br>diagnostic evaluation<br>by a health care<br>professional. | Disposable Lead Wires<br>are intended to be used<br>with ECG. The lead<br>wire is used to connect<br>electrodes placed at<br>appropriate sites on the<br>patient to ECG for<br>general monitoring<br>and/or diagnostic<br>evaluation by health<br>care professional. | |
| Sterility and Shelf-life | Provided non-sterile.<br>No shelf-life | Provided non-sterile.<br>No shelf-life | SE |
| Single Patient Use | Yes | Yes | SE |
| Wire Materials | Shielded&Unshielded<br>Copper with PVC or<br>PP Jacket | Polyvinyl Chloride<br>(PVC) | Analysis 3 |
| Connectors | Medical grade<br>PVC and PP over<br>molded connectors<br>with integral relief. | Medical grade PVC<br>and ABS over<br>molded connectors<br>with integral relief. | Analysis 3 |
| Cable Length<br>(Minimum) | 10 Centimeters | 0.72 Meters | Analysis 2 |
| Cable Length<br>(Maximum) | 11 Foot (3.35 Meters) | 3.6 Meters | Analysis 2 |
| Patient end<br>termination type | Clip | Clip , snap | Analysis 4 |
| Number of leadwires | 3,5,6 | Unknown | Analysis 5 |
| Disposable or<br>reusable | Disposable | Disposable | SE |
The Subject Device and the Predicate Device are substantially equivalent. The minor differences are:
Analysis (2) The minimum SureLead Cable System (Subject Device) has a shorter minimum cable length and a shorter maximum cable length. These differences are negligible.
Analysis (3) The material of proposed device is different from predicate device. However, the biocompatibility for proposed device has been evaluated and the test result conform with requirements of ISO 10993 standards. Therefore, this difference does not affect substantially equivalence.
{6}------------------------------------------------
Analysis (4) The patient end termination type of proposed device is Clip while that of predicated device is Clip or snap, yet the electric performance and safety for proposed device has been tested per IEC 60601-1 and EC 53 and the test result is acceptable. Therefore, this difference does not affect substantially equivalence.
Analysis (5) The subject device is available in 3, 5 ,6 leadwire type, while the leadwire number for predicate device is unknown and the substantially equivalence cannot be determined. However, the electric performance and safety for proposed device has been tested per IEC 60601-1 and EC 53 and the test result is acceptable. Therefore, this difference does not affect substantially equivalence.
## CONCLUSION
The Intended Use and Indications for Use of the SureLead Disposable Cable System (subject device) and the Predicate Device, Cable / Lead-wire (Premarket Notification K170536) are substantially equivalent. The technological characteristics, components and materials used for the Predicate Device and the Subject Device are the same. The Subject Device, SureLead Disposable Cable System is substantially equivalent to the Predicate Device, Cable / Lead-wire (Premarket Notification K170536), a legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.