K181086 · Cousin Biotech Sas · MBI · Jan 9, 2019 · Orthopedic
Device Facts
Record ID
K181086
Device Name
Cortical Fixation Systems
Applicant
Cousin Biotech Sas
Product Code
MBI · Orthopedic
Decision Date
Jan 9, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cortical Fixation Systems are used for fixation of tendons and ligaments during orthopedic reconstruction procedures such as anterior cruciate ligament (ACL) reconstruction.
Device Story
Cortical Fixation Systems consist of two models: a fixed-length loop system and an adjustable-length loop system. Both utilize a titanium cortical button to anchor soft tissue to bone. An optional Extended Plate increases the contact surface area against the cortical bone. The graft is passed through the loop, and the button provides tensioning and fixation. Used in orthopedic reconstruction procedures (e.g., ACL repair) by surgeons in clinical settings. The device provides mechanical fixation of soft tissue to bone, facilitating healing and structural stability.
Clinical Evidence
Bench testing only. Performance testing included mechanical testing, biocompatibility evaluation per ISO 10993-1, and sterilization validation per ISO 11137 and ISO 11135. Packaging validation was performed per ISO 11607 and ASTM standards. No clinical data was provided.
Technological Characteristics
Materials: Ti-6Al-4V or Ti-6Al-4V ELI titanium alloy (ISO 5832-3), Polyethylene Terephthalate (PET), and Ultra-High Molecular Weight Polyethylene (UHMWPE). Form factor: Cortical button with loop and optional extended plate. Sterilization: Gamma radiation or Ethylene Oxide. Non-active surgical implant.
Indications for Use
Indicated for fixation of tendons and ligaments during orthopedic reconstruction procedures, such as anterior cruciate ligament (ACL) reconstruction, in patients requiring soft tissue to bone fixation.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 9, 2019
Cousin Biotech S.A.S. Franck Pelletier Regulatory Affairs Director 8 Rue De L'abbe Bonpain Wervicg-Sud. 59117 France
Re: K181086
Trade/Device Name: Cortical Fixation Systems Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI, GAT Dated: December 11, 2018 Received: December 12, 2018
Dear Mr. Pelletier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurence D. Coyne -S
For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
### Page 1 of 1
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
K181086
Device Name Cortical Fixation Systems
Indications for Use (Describe)
The Cortical Fixation Systems are used for fixation of tendons and ligaments during orthopedic reconstruction procedures such as anterior cruciate ligament (ACL) reconstruction.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary provided in accordance with 21 CFR §807.92(C)
| Submission Date: | April 23, 2018 |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Cousin Biotech S.A.S.<br>8 Rue De L'abbe Bonpain<br>59117 Wervicq-Sud, France |
| Submitter Contact: | Ms. Mathilde Collet<br>Regulatory Affairs Engineer<br>Phone: +011 33 (3) 20144130<br>Email: m.collet@cousin-biotech.com |
| Application<br>Correspondent: | Mr. Franck Pelletier<br>Regulatory Affairs Director<br>Phone: +011 33 (3) 20144119<br>Email: f.pelletier@cousin-biotech.com |
| Manufacturing Site: | Cousin Biotech S.A.S.<br>Allée des roses<br>F-59117 Wervicq-Sud, France |
| Model Name: | Cortical Fixation System<br>Adjustable Cortical Fixation System<br>Extended Plate for Cortical Fixation Systems |
| Classification Name: | Fastener, Fixation, Nondegradable, Soft Tissue |
| Classification<br>Regulation: | 21 CFR §888.3040<br>21 CFR §878.5000 |
| Product Code: | MBI<br>GAT |
| Substantially<br>Equivalent Devices: | K980155 – Smith & Nephew, Inc. – EndoButton CL Ultra (Primary)<br>K100652 - Arthrex, Inc. - ACL TightRope (Additional)<br>K083070 - Biomet Sports Medicine - ToggleLocTM System<br>(Additional) |
{4}------------------------------------------------
## 510(k) Summary provided in accordance with 21 CFR §807.92(C)
| Device Description: | the system. | loop). The second model is the Adjustable Cortical Fixation System<br>the cortical bone. These products are anchor devices with titanium<br>cortical button and are intended to be used as a soft tissue to bone | The Cortical Fixation Systems are available in two (2) models. The first<br>one is the Cortical Fixation System (cortical button with fixed length<br>(cortical button with adjustable length loop). These devices can be used<br>with an Extended Plate which allows a larger surface of contact with<br>fixation, especially for the ACL repair. The graft is passed through the<br>loop and the titanium button ensures the tensioning and the fixation of | | |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Intended Use:<br>Technology | The Cortical Fixation Systems are used for fixation of tendons and<br>ligaments during orthopedic reconstruction procedures such as anterior<br>cruciate ligament (ACL) reconstruction.<br>The Cousin Biotech's Cortical Fixation Systems employ the same | | | | |
| Comparison: | technological characteristics as the primary predicate device. | | | | |
| | Characteristic | Smith & Nephew, Incorporated /<br>EndoButton CL Ultra<br>(K980155) | Cousin Biotech S.A.S.<br>Cortical Fixation Systems<br>(Proposed Device) | | |
| | Design | Titanium button with loop in PET<br>preloaded with two (2) polyester<br>braids sutures (1 green, 1 white) | Titanium button with loop in PET<br>preloaded with two (2) polyester braids<br>sutures (1 green, 1 white) | | |
| | Sizes | EndoButton:<br>width: 4 - 6 mm<br>length: 12 - 18 mm | The titanium button size is 4 mm<br>(width) x 12 mm (length) x 1.5 mm<br>(height).<br>The extended plate size is 7 mm (width)<br>x 20 mm (length) x 1.5 mm (height). | | |
| | | Suture loop span length:<br>10 – 80 mm | The non-adjustable loop size is<br>available from 15 mm to 40 mm.<br>The adjustable loop size is capable of<br>provided a loop from 15 mm to 40 mm. | | |
Ti-6Al-4V or Ti-6Al-4V ELI
Polyethylene Terephthalate (PET)
titanium alloy.
Polyester (PET)
Yes
Button
Loop
Traction
threads
Shipped Sterile
Any size is possible between both
Ti-6Al-4V or Ti-6Al-4V ELI titanium
Polyethylene Terephthalate (PET) for
Polyethylene (UHMWPE) for the
Polyester (PET) and colorant D&C
the non-adjustable loop
adjustable loop
values.
alloy.
Green
Yes
{5}------------------------------------------------
# 510(k) Summary provided in accordance with 21 CFR §807.92(C)
## Summary of Performance Testing:
| Sterilization | The Cortical Fixation System and Extended Plate are gamma radiation sterilized and were validated to a sterility assurance level of 10-6 in accordance with the following standards: |
|---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | ISO 11137-1: 2006, Am1: 2013, Sterilization of health care products - Requirements for validation and routine control – Radiation sterilization; ISO 11137-2:2013, Sterilization of health care products – Radiation - Part 2: Establishing the sterilization dose; ISO 11737-1: 2006, Cor1: 2007, Sterilization of medical devices - Microbiological methods - Part 1: Determination of a population of microorganisms on products [Including: Technical Corrigendum 1 (2007)]; and ISO 11737-2: 2009, Sterilization of medical devices -- Microbiological methods -- Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process. |
| | Validation results indicate that the Cortical Fixation System and Extended Plate comply with the standards. |
| | The Adjustable Cortical Fixation System is Ethylene Oxide sterilized and was validated in accordance with the following standard: |
| | ISO 11135:2014, Sterilization of health-care products -- Ethylene oxide -- Requirements for the development, validation and routine control of a sterilization process for medical devices; and ISO 10993-7: 2008, Cor1: 2009, Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals [Including: Technical Corrigendum 1 (2009)]. |
| | Validation results indicate that the Adjustable Cortical Fixation System complies with the standards. |
| | The Cortical Fixation System, Adjustable Cortical Fixation System, and Extended Plate were tested for pyrogenicity as recommended by the following FDA guidance document: |
| | Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile, 21 Jan 16. |
| | Validation results indicate that the Cortical Fixation System, Adjustable Cortical Fixation System, and Extended Plate are non-pyrogenic in accordance with the FDA guidance document. |
{6}------------------------------------------------
# provided in accordance with 21 CFR §807.92(C)
| Shelf Life | The packaging for the Cortical Fixation Systems was validated in accordance with the following standards:<br>ISO 11607-1: 2006 Packaging for terminally sterilized medical devices – Part 1: requirements for materials, sterile barrier systems and packaging systems; ISO 11607-2: 2006 Packaging for terminally sterilized medical devices – Part 2: Validation requirements for forming, sealing and assembly processes; ASTM D4169-16, Standard Practice for Performance Testing of Shipping Containers and Systems; ASTM F1929-12, Standard Test Method for Detecting Seal Leaks In Porous Medical Packaging By Dye Penetration; and ASTM F1980-16, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. Validation results indicate that the packaging for the Cortical Fixation Systems complies with the standards. |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | The Cortical Fixation Systems patient contact materials was verified in accordance with the following standard:<br>ISO 10993-1: 2009, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process. Verification results indicated that the Cortical Fixation Systems patient contact materials comply with the standard. |
| Performance Testing – Bench | The Cortical Fixation Systems were tested for performance in accordance with its predetermined specifications and the following standards:<br>IEC 62366: 2007, Medical devices – Application of usability engineering to medical devices; ISO 5832-3: 2016, Implants for surgery - Metallic materials – Part 3: Wrought titanium-6 aluminum-4 vanadium alloy; and ISO 14630: 2012, Non-active surgical implants – General requirements. Test results indicate that the Cortical Fixation Systems comply with its predetermined specifications and with the standards. |
{7}------------------------------------------------
## 510(k) Summary provided in accordance with 21 CFR §807.92(C)
Conclusion Verification and validation activities were conducted to establish the performance and safety characteristics of Cortical Fixation System, Adjustable Cortical Fixation System and Extended Plate. The results of these activities demonstrate that the Cortical Fixation Systems are as safe, as effective, and perform as well as or better than the predicate devices.
Therefore, the Cortical Fixation Systems are considered substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.