← Product Code [GEI](/productcode/GEI) · K181085

# LigaSure Maryland Jaw Thoracic Sealer/Divider One-step Sealing, Nano-coated (K181085)

_Covidien · GEI · Jun 21, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K181085

## Device Facts

- **Applicant:** Covidien
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Jun 21, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The LigaSure Sealer/Divider is a bipolar electrosurgical instrument intended for use in minimally invasive or open surgical procedures where ligation of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/ Divider can be used on vessels (arteries, veins, pulmonary veins) up to and including 7 mm. It is indicated for use in general surgery and surgical specialties as urologic, vascular, thoracic, and gynecologic. Procedures may include, but are not limited to, Nissen fundoplication, cholecystectomy, adhesiolysis, hysterectomy, oophorectomy, etc. The LigaSure system has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.

## Device Story

Sterile, single-use, nano-coated bipolar electrosurgical instrument; used in minimally invasive or open surgery. Input: RF energy from compatible Covidien generator. Operation: User clamps tissue between jaws via hand-actuated lever; activates RF energy via button to seal vessels/tissue; user-actuated trigger deploys blade to divide sealed tissue. Used by surgeons for vessel ligation, tissue grasping, and blunt dissection. Output: Sealed and divided tissue. Benefits: Hemostasis in surgical procedures; reduced thermal spread. No software component.

## Clinical Evidence

No clinical data. Evidence includes bench performance testing (mechanical, electrical, functional, ex-vivo vessel burst pressure), acute animal study (sealing performance, lateral thermal spread), and chronic animal study (21-day seal quality/hemostasis). Usability engineering performed per IEC 62366-1.

## Technological Characteristics

Bipolar electrosurgical vessel sealer; nano-coated jaws; 30cm shaft. Energy source: RF from external Covidien generator. Standards: IEC 60601-1 (electrical safety), IEC 60601-2-2 (electrosurgical safety), IEC 60601-1-2 (EMC). Sterile, single-use. No software.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider, Nano-coated (LF4418) ([K162047](/device/K162047.md))
- LigaSure™ Maryland Jaw Sealer/Divider One-step Sealing, Nano-coated (LF1923, LF1937, LF1944) ([K133338](/device/K133338.md), [K141153](/device/K141153.md), [K170869](/device/K170869.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym followed by the full name of the agency, "U.S. Food & Drug Administration," on the right. The FDA acronym and the agency's name are in blue.

June 21, 2018

Covidien Mr. James Davis Senior Specialist, Regulatory Affairs 5920 Longbow Drive Boulder, Colorado 80301

Re: K181085

Trade/Device Name: LigaSure Maryland Jaw Thoracic Sealer/Divider One-step Sealing, Nano-coated (LF1930T) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: April 24, 2018 Received: April 25, 2018

Dear Mr. Davis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely. Jennifer R. Stevenson -For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K181085

#### Device Name

LigaSure™ Maryland Jaw Thoracic Sealer/Divider One-step Sealing, Nano-coated (LF1930T)

#### Indications for Use (Describe)

The LigaSure Sealer/Divider is a bipolar electrosurgical instrument intended for use in minimally invasive or open surgical procedures where ligation of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/ Divider can be used on vessels (arteries, veins, pulmonary veins) up to and including 7 mm. It is indicated for use in general surgery and surgical specialties as urologic, vascular, thoracic, and gynecologic. Procedures may include, but are not limited to, Nissen fundoplication, cholecystectomy, adhesiolysis, hysterectomy, oophorectomy, etc.

The LigaSure system has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the word "COVIDIEN" in a sans-serif font. To the left of the word is a blue square with a white rectangle in the middle. The square is divided into three vertical sections, with the middle section being a lighter shade of blue.

510(k) Summarv Date summary prepared: April 24, 2018

# 510(k) Submitter/Holder

Covidien, Ilc. 5920 Longbow Drive Boulder, CO 80301

#### Contact

James R. Davis Sr. Regulatory Affairs Specialist Telephone: 303-581-6964 Fax: 303-530-6313 Email: james.r.davis(@medtronic.com

# Name of Device

| Trade Name:          | LigaSure™ Maryland Jaw Thoracic Sealer/Divider One-step Sealing,<br>Nano-coated                    |
|----------------------|----------------------------------------------------------------------------------------------------|
| Catalog Numbers:     | LF1930T                                                                                            |
| Common Name:         | Bipolar Vessel Sealing Device                                                                      |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories<br>(21 CFR 878.4400, Class II, GEI) |
| Predicate Devices:   | Primary                                                                                            |
| Trade Name:          | LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider, Nano-<br>coated (LF4418)                  |
| Common Name:         | Bipolar Vessel Sealing Device                                                                      |
| Catalog Number:      | LF4418                                                                                             |
| 510(k) Number:       | K162047 (cleared 9/15/16)                                                                          |
| Manufacturer:        | Covidien                                                                                           |
| Recalls:             | This device has not been subject to a design-related recall                                        |
|                      | Secondary                                                                                          |
| Trade Name:          | LigaSure™ Maryland Jaw Sealer/Divider One-step Sealing, Nano-coate                                 |
| Common Name:         | Bipolar Vessel Sealing Device                                                                      |
| Catalog Number:      | LF1923, LF1937, LF1944                                                                             |
| 510(k) Number:       | K133338 (cleared 12/20/2013), K141153 (cleared 8/14/2014), K170869<br>(cleared 04/21/2017)         |
| Manufacturer:        | Covidien                                                                                           |
| Recalls:             | This device has not been subject to a design-related recall                                        |

No reference devices were used in this submission.

# Device Description

The LigaSure™ Maryland Jaw Thoracic Sealer/Divider One-step Sealing, Nano-coated (LF1930T) device is a sterile, single-use, coated, hand-held bipolar vessel sealing device designed for use with a Covidien electrosurgical generator that include vessel sealing capabilities 

{4}------------------------------------------------

to ligate (seal) and divide (cut) vessels (arteries, veins, pulmonary arteries, pulmonary veins), tissue bundles, and lymphatics clamped between the jaws using radio frequency (RF) energy. The LF1930T subject can also grasp tissue and perform blunt dissection during general surgical procedures (as indicated).

A hand actuated lever allows the user to open or close the instrument jaws during vessel sealing and cutting, with a single activation button incorporated into the body of the handle. The proposed device has a 30cm length shaft and does not contain software.

# Indications for use:

The LigaSure Sealer/Divider is a bipolar electrosurgical instrument intended for use in minimally invasive or open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/ Divider can be used on vessels (arteries, veins, pulmonary arteries, pulmonary veins) up to and including 7 mm. It is indicated for use in general surgery and such surgical specialties as urologic, vascular, thoracic, and gynecologic. Procedures may include, but are not limited to, Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, hysterectomy, oophorectomy, etc.

The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.

# Comparison of Technological Characteristics to the Predicate Devices

The application of radio frequency (RF) energy to target tissue is the fundamental technology for both the subject and predicate devices. A seal is created by application of RF energy to structures (vasculature, lymphatics, and tissue bundles) interposed between the jaws of the instrument. A blade within the instrument is user actuated to divide tissue after the seal is created.

At a high level, the subject and both the primary/secondary predicate devices are based on the following same technological elements:

- . Lever - closes the jaw to grasp the tissue
- Jaws Used to grasp tissue and to deliver RF energy to seal target structures ●
- Activation Button - Allows RF energy to be activated by the user
- Cutting Trigger Allows the user to divide (cut) sealed structures ●
- Device Materials - Materials are identical to the LF19XX secondary predicate devices only

The following technological differences exist between the subject and both the primary/secondary predicate devices:

- . Shaft Length – Provides the jaw assembly access to the surgical site
- Jaw Assembly Specifications Minor restrictions of the jaw assembly specification range ● within the specification range of the LF19XX secondary predicate devices
- . Device Identifiers - Changes to product identifiers on the device, packaging, and labeling

These changes in conjunction with the performance data outlined below support the changes to the indications for use of the subject device and have demonstrated to be substantially equivalent to the predicate devices.

#### Performance Data:

The following performance data were provided in support of the substantial equivalence determination.

{5}------------------------------------------------

# Electrical Safety and Electromagnetic Compatibility (EMC)

Electrical safety testing and EMC testing were conducted on the LigaSure™ Maryland Jaw Thoracic Sealer/Divider One-step Sealing, Nano-coated (LF1930T) or equivalent device as appropriate. The system complies with relevant portions of the IEC 60601-1 and IEC 60601-2-2 standards for electrical safety and IEC 60601-1-2 standard for EMC.

### Bench Performance Testing

Mechanical, electrical, functional, and ex-vivo vessel burst pressure testing was carried out on validated and representative conditions as appropriate to verify that the new device performs as expected. The subject device conforms to the requirements defined in the related design inputs and subsequent product specifications.

# Acute Animal Study

In the acute animal study conducted, subjects underwent procedures to assess acute sealing performance and lateral thermal spread by the subject and predicate devices as appropriate. These studies demonstrated that the subject device is as safe and effective as the predicate devices.

# Chronic Animal Study

A chronic study was conducted to assess seal quality over the course of 21 days. All animals enrolled in the study survived 21 days post-op and all seals maintained chronic hemostasis.

# Human Factors and Usability

The usability engineering process applied to the subject device was in compliance with the requirements of IEC 62366-1 "Medical devices-Application of usability engineering to medical devices". The process included analysis of user needs and potential use errors followed by testing to demonstrate that representative users can use the subject device in a safe and effective manner.

#### Clinical Studies

This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.

#### Summarv

Based on the pre-clinical performance as documented in the performance testing, the LigaSure™ Maryland Jaw Thoracic Sealer/Divider One-step Sealing, Nano-coated (LF1930T) subject device was found to have a safety and effectiveness profile that is similar to the predicate devices.

#### Conclusions

The subject device has the same intended use as the predicate devices and the design changes made to the subject device does not raise different kinds of safety and effectiveness questions. The comparison of device characteristics and review of the performance verification/validation data demonstrate substantial equivalence of the LigaSure™ Maryland Jaw Thoracic Sealer/Divider One-step Sealing, Nano-coated subject device to the legally marketed predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K181085](https://fda-staging.innolitics.com/device/K181085)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
