← Product Code [GEI](/productcode/GEI) · K181071

# SIRA RFA Electrosurgical Device (K181071)

_Innoblative Designs, Inc. · GEI · Jan 16, 2019 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K181071

## Device Facts

- **Applicant:** Innoblative Designs, Inc.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Jan 16, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The SIRA™ RFA Electrosurgical Device supplies energy for use in electrosurgery and is indicated for use in intraoperative coaqulation and ablation of soft tissue. The SIRA™ RFA Device is to be used in conjunction with a radiofrequency (RF) electrosurgical generator for use in open abdominal surgical procedures. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).

## Device Story

SIRA RFA Electrosurgical Device delivers bipolar radiofrequency (RF) energy and 0.9% saline to soft tissue for coagulation and ablation. Device comprises 4cm diameter electrode array, handle, co-extruded cable, integrated RF switch/timer, and fluid administration set. Used in open abdominal surgery by surgeons; requires external RF generator, IV pole, and saline bag. Saline flow controlled via manual regulator (5-300 ml/hr). Device provides simultaneous energy/saline delivery to target tissue; larger electrode array compared to predicate allows for larger contact surface area per lesion. Equivalent treatment depth achieved compared to predicate. Benefits include effective soft tissue coagulation/ablation during open procedures.

## Clinical Evidence

Nonclinical performance testing included software verification/validation, electrical safety (IEC 60601-1, IEC 60601-2-2), electromagnetic compatibility (IEC 60601-1-2), biocompatibility, sterility validation, and mechanical/functional bench testing. In vivo animal model testing evaluated safety and effectiveness. Usability testing confirmed intended users can operate the device without patterns of use error.

## Technological Characteristics

Bipolar RF electrosurgical device; 4cm diameter electrode array; manual saline flow regulator (5-300 ml/hr); integrated RF switch and timer. Sterile, single-patient use, disposable. Connects to external RF generator. Complies with IEC 60601-1, IEC 60601-2-2, and IEC 60601-1-2.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Medtronic Aquamantys Malleable Bipolar Sealer with Light ([K073495](/device/K073495.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the words "U.S. Food & Drug Administration" on the right. The FDA part of the logo is in blue, with the acronym in a square and the words in a smaller font below.

January 16, 2019

Innoblative Designs, Inc. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street, 23rd Floor Philadelphia, Pennsylvania 19103

Re: K181071

Trade/Device Name: SIRA RFA Electrosurgical Device Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: December 18, 2018 Received: December 18, 2018

Dear Janice Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Image /page/1/Picture/5 description: The image shows a digital signature. The signature is by Long H. Chen -S. The date of the signature is 2019.01.16 16:05:02 -05'00'.

for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement on last page

_ Over-The-Counter Use (21 CFR 801

510(k) Number (if known)

# K181071

Device Name

SIRA™ RFA Electrosurgical Device

Indications for Use (Describe)

The SIRA™ RFA Electrosurgical Device supplies energy for use in electrosurgery and is indicated for use in intraoperative coaqulation and ablation of soft tissue. The SIRA™ RFA Device is to be used in conjunction with a radiofrequency (RF) electrosurgical generator for use in open abdominal surgical procedures. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).

Type of Use (Select one or both, as applicable)

& Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

> FORM FDA 3881 (7/17) Page 1 of 1 FDA

{3}------------------------------------------------

#### 510(k) SUMMARY

# Innoblative Designs, Inc.'s SIRA™ Device

# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Tyler Wanke, President and CEO Innoblative Designs, Inc. 4660 Ravenswood Avenue Chicago, IL 60640 Phone: 920-450-7529

Date Prepared: January 14, 2019

## Name of Device

SIRA™ RFA Electrosurgical Device

## Common or Usual Name

Electrosurgical Device

## Classification

21 CFR 878.4400, Class II, product code GEI

#### Predicate Device

Medtronic Aquamantys Malleable Bipolar Sealer with Light (K073495)

#### Intended Use / Indications for Use

The SIRA™ RFA Electrosurgical Device supplies energy for use in electrosurgery and is indicated for use in intraoperative coaqulation and ablation of soft tissue. The SIRA™ RFA Device is to be used in conjunction with a radiofrequency (RF) electrosurgical generator for use in open abdominal surgical procedures. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).

#### Device Description

The SIRA™ RFA Electrosurqical Device is used for the simultaneous administration of RF energy and saline for the coagulation and ablation of soft tissue in open intraoperative procedures. The SIRA™ RFA Electrosurgical Device is comprised of a 4cm diameter electrode array, a handle (shaft), a co-extruded cable, an integrated electrical RF switch and timer, a main cable, and a fluid administration set.

{4}------------------------------------------------

# Technological Characteristics

The SIRA™ RFA Electrosurgical Device has similar technological characteristics as its predicate device.

The designs of both the SIRA™ RFA Electrosurgical Device and Aquamantys predicate consist of an electrode tip/shaft assembly extended out of an ergonomically designed body. Both the SIRA™ RFA and the Aquamantys devices deliver bipolar radiofrequency energy coupled with 0.9% saline to coagulate soft tissue. Both devices are sterile, single-patient use, and disposable. Further, both the subject and predicate device consist of an integrated cable assembly that connects the devices to an electrosurgical generator. Furthermore, both the subject and predicate devices are electrically insulated, except for the shaft that is electrically active. Both the subject device and Aquamantys predicate allow the user to control the level of RF energy and duration. In addition, both the subject and predicate devices are required to be used with similar accessories, including a 0.9% saline bag, an IV pole, and an external electrosurgical generator.

Although the saline flow for the subject device is controlled by a manual flow regulator with settings of 5-300 ml/hr, whereas the saline flow for the predicate Aquamantys is controlled by an automatic saline pump with low, medium, and high settings, this minor difference does not raise new questions of safety or effectiveness because both devices provide an appropriate mechanism for saline infusion. In addition, the overall size of the working end of the SIRA™ RFA device is larger than the predicate device's electrodes, which can result in larger contact surface area than the predicate with each ablation lesion delivery. However, the differences in dimensions do not raise new questions of safety or effectiveness. Furthermore, comparative testing demonstrates that the depth of treatment created by both devices is equivalent when energy is applied at similar durations, with similarly applied energy levels.

Therefore, the SIRA™ RFA Electrosurgical Device has very similar technological characteristics as its predicate device and is substantially equivalent.

# Performance Data

The following nonclinical performance testing has been conducted to support the substantial equivalence of the SIRA™ RFA Electrosurgical Device to its predicate. In all instances, the subject device functioned as intended.

{5}------------------------------------------------

- . Software verification and validation was performed, and demonstrated that the software performs as intended.
- Electrical safety (IEC 60601-1 and IEC 60601-2-2) and electromagnetic compatibility . (IEC 60601-1-2) testing was conducted and results were passing.
- Biocompatibility of the patient-contacting components of the device was established. .
- . Sterility validation was established for all required components.
- Performance bench testing, including predicate comparative testing, mechanical testing, . dimensional testing, functional tests.
- In vivo testing in an animal model was performed to evaluate and establish the safety and effectiveness of the subject device.
- . Usability testing to demonstrate that the intended users of the device can use the device as intended to coagulate/ablate soft tissues without patterns of perceptible use errors or difficultly that could result in serious harm to the user or patient.

## Substantial Equivalence

The SIRA™ RFA Electrosurgical Device has the same indications for use and similar indications for use, as well as very similar technological characteristics, and principles of operation as its predicate device. The minor technological differences between the subject and the predicate device do not raise different questions of safety or effectiveness. Performance testing of the device has demonstrated that the device performs as intended and thus, is substantially equivalent.

## Conclusion

The SIRA™ RFA Electrosurgical Device has been evaluated in performance testing. Testing demonstrates that the device performs as intended. The SIRA™ RFA Electrosurgical Device is substantially equivalent to its predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K181071](https://fda-staging.innolitics.com/device/K181071)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
