DIO CAD/CAM Abutment

K181037 · Dio Corporation · NHA · Dec 21, 2018 · Dental

Device Facts

Record IDK181037
Device NameDIO CAD/CAM Abutment
ApplicantDio Corporation
Product CodeNHA · Dental
Decision DateDec 21, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3630
Device ClassClass 2
AttributesTherapeutic

Indications for Use

DIO CAD/CAM Abutment is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.

Device Story

Device includes Hybrid Link Abutments (two-piece: titanium alloy base + CAD/CAM coping) and Patient-Specific Abutments (titanium alloy). Inputs: clinician-prescribed design parameters based on patient anatomy. Transformation: CAD/CAM software designs patient-specific superstructures/abutments; DIO Corporation milling center fabricates components. Output: custom dental abutments. Used in dental clinics; operated by clinicians. Output used to support prosthetic restorations (crowns/bridges); affects clinical decision-making by providing patient-specific fit for implant-supported prostheses. Benefits: improved prosthetic fit and aesthetics for edentulous patients.

Clinical Evidence

No clinical data. Substantial equivalence established via bench testing, including fatigue testing (ISO 14801:2016) on worst-case scenarios (smallest diameter, maximum angulation), steam sterilization validation (ISO 17665), and biocompatibility testing (ISO 10993).

Technological Characteristics

Materials: Ti-6Al-4V ELI (ASTM F136), Zirconia Oxide. Design: Hybrid Link (two-piece) or Patient-Specific Abutment. Connection: Internal hex/non-hex. Manufacturing: CAD/CAM milling at validated center. Sterilization: Non-sterile, user-performed steam sterilization. Connectivity: CAD/CAM software-driven design.

Indications for Use

Indicated for partially or fully edentulous patients requiring single or multiple tooth prostheses supported by dental implants in the maxilla or mandible.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 21, 2018 DIO Corporation % April Lee Consultant Withus Group Inc 106 Superior Irvine, California 92620 Re: K181037 Trade/Device Name: DIO CAD/CAM Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: November 19, 2018 Received: November 26, 2018 Dear April Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Andrew I. Steen -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K181037 Device Name DIO CAD/CAM Abutment Indications for Use (Describe) #### Indication for Use DIO CAD/CAM Abutment is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. | Implant System Compatibility | Implant Diameter (mm) | Platform Diameter (mm) | |------------------------------|---------------------------------|---------------------------------| | UF(II) Narrow Implant System | 3.0/3.3 | 3.0/3.3 | | UF Sub merged Implant System | 3.8/4.0/4.5/5.0/5.5/6.0/6.5/7.0 | 3.8/4.0/4.5/5.0/5.5/6.0/6.5/7.0 | | UF(II) Implant System | 3.8/4.0/4.5/5.0/5.5 | 3.8/4.0/4.5/5.0/5.5 | Patient specific abutment is intended for use with the UF Implant Systems provided in the chart. All digitally designed abutments for use with DIO CAD/CAM Abutments are intended to be manufactured at a DIO Corporation validated milling center. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary #### Submitter DIO Corporation Jiae, Park 66, Centum Seo-ro, Haeundae-gu Busan, 48058 Republic of Korea Email: japark@dio.co.kr Phone: +82-51-745-7836 # Official Correspondent Withus Group Inc. April Lee 106 Superior, Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122 ## Device Information - Trade Name: DIO CAD/CAM Abutment - Common Name: Endosseous dental implant abutment - Classification Name: Abutment, Implant, Dental, Endosseous . - Product Code: NHA - . Panel: Dental - Regulation Number: 21 CFR 872.3630 ● - . Device Class: Class II - Date prepared: 12/21/2018 ● ### Predicate Devices: The subject device is substantially equivalent to the following predicate device: Primary Predicate - Medentika CAD/CAM Abutment (K150203) Reference Devices - Sirona Dental CAD/CAM System (K111421) - Zimmer® patient Specific Abutment, Internal Hex, Titanium(K143505) ● - Encode Patient Specific Abutment for Straumann Bone Level Connection(K112730) ● - . UF(II) Narrow Implant System (K161987) - UF Sub Merged Implant System (K122519) ● - UF(II) Implant System (K170608) - Self-Adhesive Resin Cement (K073173) - . ZIRMON SERIES (K131117) {4}------------------------------------------------ # Device Description The DIO CAD/CAM Abutment includes two CAD/CAM abutment designs, Hybrid Link Abutment and Patient-Specific Abutment. ## 1) Hybrid Link abutment ## Hybrid Link Abutment Hybrid Link abutment is intended to provide support for customized prosthetic restorations such as crowns and bridges. The hybrid link abutment is composed of two-piece abutment that is a hybrid link at the bottom and a coping (CAD/CAM patient specific superstructure) at the top. The hybrid link abutments are pre-manufactured (stock) abutments, made from a titanium alloy conforming to ASTM F136. The diameters of Hybrid Link Abutment are 4.0/4.5/5.5mm. Hybrid Link abutment is provided non-sterile therefore must be sterilized after the cementation of the patient-specific superstructure on the Hybrid Link Abutment. The proposed devices are compatible with the following device. #### Dental Implants | Proprietary Name | UF(II) Narrow<br>Implant System | UF Sub merged Implant<br>System | UF(II) Implant System | |--------------------------------------------------|---------------------------------|-------------------------------------|----------------------------------| | Compatible Implants<br>(Knumber) | K161987 | K122519 | K170608 | | Implant diameter size | 3.0/3.3 | 3.8/4.0/4.5/5.0/5.5/6.0/6.5<br>/7.0 | 3.8/4.0/4.5/5.0/5.5 | | Implant Interface<br>Connection<br>Type/Size(mm) | Internal connection<br>type/2.3 | Internal connection<br>type/3.35 | Internal connection<br>type/3.35 | | Type of Implant-<br>Abutment Connection | Hex/Non Hex | Hex/Non Hex | Hex/Non Hex | Raw material blanks - K131117 ZIRMON SERIES manufactured by KUWOTECH CO., LTD . ## Raw material cement - . K073173 Self-Adhesive Resin Cement manufactureed by Densply is the cement The coping that composes the final abutment should be designed and milled through the CAD/CAM software, according to the prosthetic planning and patient clinical situation. The coping and crowns designed using these or more recent versions of the CAD/CAM System, within the design limits as defined within the design software, are compatible with the Hybrid link abutment. The coping would be manufactured by DIO Corporation only with design input using CAD/CAM Software from and by DIO Corporation milling center. Design Limitation for Superstructure: | Super structure design Limitation (Unit: mm) | | | | |----------------------------------------------|----------------|---------------------------|---------------| | Range (Diameter) | Range (Height) | Range (Wall<br>thickness) | Range (Angle) | {5}------------------------------------------------ | Min .<br>4.0mm<br> | land of the comments of the comments of the comments of the comments of the contribution of the contribution of the contribution of the contribution of the contribution of th<br>Min<br>. Omm | U.Smm<br>ا ا 1 1 11 | 1 C ( .<br>रे | |--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|---------------| |--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|---------------| #### 2) Patient-Specific Abutment Patient-specific abutment is made from titanium alloy conforming to ASTM F136 titanium abutment to be used in fabricating patient-specific abutments. The subject abutments are indicated for cemented or "Screw-and Cement-Retained Prosthesis(SCRP)" restorations. Each patient-specific abutment is individually prescribed by the clinician. The diameters of patient-specific Abutment are 3.0/3.3/3.8/4.0/4.5/5.0/5.5/6.0/6.5/7.0 and two connection designs (Hex, Non-hex). Patient-Specific Abutment is compatible with following Implant Systems. | Proprietary Name | UF(II) Narrow Implant<br>System | UF Sub merged<br>Implant System | UF(II) Implant<br>System | |--------------------------------------------------|---------------------------------|-------------------------------------|----------------------------------| | Compatible Implants<br>(Knumber) | K161987 | K122519 | K170608 | | Implant diameter size | 3.0/3.3 | 3.8/4.0/4.5/5.0/5.5/6.0<br>/6.5/7.0 | 3.8/4.0/4.5/5.0/5.5 | | Implant Interface<br>Connection<br>Type/Size(mm) | Internal connection<br>type/2.3 | Internal connection<br>type/3.35 | Internal connection<br>type/3.35 | | Type of Implant-<br>Abutment Connection | Hex/Non Hex | Hex/Non Hex | Hex/Non Hex | Patient-specific abutments are supplied with an abutment screw previous cleared device as K122519 and K161987 and provided non-sterile. | Model Name | Range (Diameter) | Range (Height) | Range (Angle) | |----------------|------------------|----------------|---------------| | CASSH<br>CASSN | 3.9~12.0 | 3.1~14.9 | 0~29.8° | | CAUNH<br>CAUNN | 3.9~12.0 | 3.1~14.9 | 0~29.3° | #### Indication for Use DIO CAD/CAM Abutment is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. | Implant System Compatibility | Implant Diameter (mm) | Platform Diameter (mm) | |------------------------------|---------------------------------|---------------------------------| | UF(II) Narrow Implant System | 3.0/ 3.3 | 3.0/ 3.3 | | UF Sub merged Implant System | 3.8/4.0/4.5/5.0/5.5/6.0/6.5/7.0 | 3.8/4.0/4.5/5.0/5.5/6.0/6.5/7.0 | | UF(II) Implant System | 3.8/4.0/4.5/5.0/5.5 | 3.8/4.0/4.5/5.0/5.5 | Patient specific abutment is intended for use with the UF Implant Systems provided in the chart. All digitally designed abutments for use with DIO CAD/CAM Abutments are intended to be manufactured at a DIO Corporation validated milling center. {6}------------------------------------------------ # Summaries of Technological Characteristics The subject device is substantially equivalent to the currently cleared devices. They are substantially equivalent in intended use, material and connection interfaces to the implants are identical for each individual diameter and connection type. Comparison demonstrating Substantial Equivalence follows: # 1) Hybrid Link Abutment | | Subject Device | Primary Predicate Device | Reference Device | |----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Applicant | DIO Corporation | Medentika GmbH | Sirona Dental Systems GmbH | | Trade Name | DIO CAD/CAM Abutment | Medentika CAD/CAM Abutment | Sirona Dental CAD/CAM System<br>TiBase Abutment | | 510(K) No. | K181037 | K150203 | K111421 | | Classification Name | Endosseous Dental Implant, Abutment<br>(872.3630) | Endosseous Dental Implant, Abutment<br>(872.3630) | Endosseous Dental Implant, Abutment<br>(872.3630) | | Product Code | NHA | NHA | NHA | | Class | II | II | II | | Material | Ti-6Al-4V ELI (ASTM F136)<br>Zirconia Oxide | Titanium Alloy | Ti-6Al-4V ELI (ASTM F136)<br>Zirconia Oxide | | Design | Titanium Base | Titanium Base | Titanium Base | | Diameters (mm) | 4.0/4.5/5.5 | 3.25-7.0 | 3.5~6.5 | | Superstructure<br>Angulation (°) | 0~15 | Up to 30 | Up to 20 | | Superstructure<br>material | Zirconia | Zirconia | Zirconia | | Sterile | Steam Sterilization by user<br>(Delivered non sterile) | Steam Sterilization by user<br>(Provided Non-Sterile) | Steam Sterilization by user<br>(Delivered non sterile) | | Type of Retention | Screw-retained or cement retained | Cement-retained | Screw-retained or cement retained | | Indications For Use | DIO CAD/CAM Abutment is intended for use<br>with dental implants as a support for single or<br>multiple tooth prostheses in the maxilla or<br>mandible of a partially or fully edentulous patient.<br> | Medentika Preface CAD/CAM Abutments are intended for<br>use with dental implants as a support for single or multiple<br>tooth prostheses in the maxilla or mandible of a partially or<br>fully edentulous patient. | The Sirona Dental CAD/CAM System is<br>intended for use in partially or fully<br>edentulous mandibles and maxillae in<br>support of single or multiple-unit cement<br>retained restorations. For the titanium<br>bases SSO 3.5L and SBL 3.3L, the<br>indication is restricted for replacement of<br>single lateral incisors in the maxilla and | {7}------------------------------------------------ | | Narrow<br>Implant<br>System | | | Nobel Biocare<br>Replace™ Select | E | 3.5, 4.3,<br>5.0, 6.0 | 3.5, 4.3,<br>5.0, 6.0 | lateral and central incisors in the mandible.<br>The system consists of three major parts:<br>TiBase, inCoris mesostructured, and<br>CAD/CAM software. Specifically, the<br>inCoris mesostructured and TiBase<br>components make up a two-piece<br>abutment which is used in conjunction<br>with endosseoous dental implants to<br>restore the function and aesthetics in the<br>oral cavity. The inCoris mesostructured<br>may also be used in conjunction with the<br>Camlog Titanium base CAD/CAM(types<br>K2244.xxxx)(K083496) in the Camlog<br>Implant System. The CAD/CAM software<br>is intended to design and fabricate the<br>inCoris mesostructure. The inCoris<br>mesostructured and TiBase two-piece<br>abutment is compatible with the following<br>implants systems: | |------------------------------------------|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|-------------------------------|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | UF Sub<br>merged<br>Implant<br>System | 3.8/4.0/4.5/<br>5.0/5.5/6.0/<br>6.5/7.0 | 3.8/4.0/4.5/<br>5.0/5.5/6.0/<br>6.5/7.0 | Nobel Biocare<br>Nobel Active™ | F | 3.5, 4.3,<br>5.0 | 3.5, 3.9<br>(4.3), 3.9<br>(5.0) | | | | UF(II)<br>Implant<br>System | 3.8/4.0/4.5/<br>5.0/5.5 | 3.8/4.0/4.5/<br>5.0/5.5 | Biomet 3i<br>Osseotite® | H | 3.25, 4.0,<br>5.0 | 3.4, 4.1,<br>5.0 | Nobel Biocare Replace(K020646) | | | | | | Certain<br>Biomet 3i<br>Osseotite® | I | 3.25, 3.75,<br>4.0, 5.0 | 3.4, 4.1,<br>5.0 | Nobel Biocare Branemark(K022562) | | | | | | Nobel Biocare<br>Branemark | K | 3.3, 3.75,<br>4.0, 5.0 | 3.5, 4.1,<br>4.1, 5.1 | Friadent Xive(K013867) | | | | Patient specific abutment is intended for use with<br>the UF Implant Systems provided in the chart. All<br>digitally designed abutments for use with DIO<br>CAD/CAM Abutments are intended to be<br>manufactured at a DIO Corporation validated<br>milling center. | | Straumann Bone<br>Level | L | 3.3, 4.1,<br>4.8 | 3.3, 4.1,<br>4.8 | Biomet 3i Osseotite(K980549) | | | | | | Straumann<br>Stnadard | N | 3.3, 4.1,<br>4.8 | 3.5(NNC),<br>4.8, 6.5 | Astra Tech Osseospeed(K091239) | | | | | | Zimmer Tapered<br>Screw-vent® | R | 3.3, 3.7,<br>4.1, 4.7,<br>6.0 | 3.5, 4.5,<br>5.7 | Zimmer Tapered Screw-Vent(K061410) | | | | | | Austra Tech<br>OsseoSpped™ | S | 3.5, 4.0,<br>4.5, 5.0 | 3.5, 4.0,<br>4.5, 5.0 | Straumann SynOcta(K061176) | | | | | | Dentsply<br>Friadent®<br>Frialit/XiVE® | T | 3.4, 3.8,<br>4.5, 5.5 | 3.4, 3.8,<br>4.5, 5.5 | Straumann Bone Level(K053088,<br>K062129, K060958) | | | | | | Dentsply<br>Friadent®<br>Ankylos® | Y | 3.5, 4.5,<br>5.5, 7.0 | 3.5, 4.5,<br>5.5, 7.0 | Biomet 3i Certain(K014235, K061629) | | | | | | Medentika PreFace is intended for use with the<br>Straumann® CARES® System. All digitally designed<br>abutments for use with Medentika CAD/CAM Abutments<br>are intended to be manufactured at a Straumann®<br>CARES® validated milling center. | | | | Nobel Biocare Active(K071370) | | Substantial<br>Equivalence<br>Comparison | | | | | | | | The subject Hybrid Link Abutment is substantially equivalent in designs, dimensions, material, indications, and technological characteristics<br>with the identified primary predicate device. The DIO CAD/CAM Abutment is similar in fundamental scientific technology to the<br>predicate device in that they all have been designed, manufactured and tested in compliance with FDA's Class II special controls<br>guidance document root-food endosseous dental implants and endosseous dental implant abutments.<br><br>The diameters of the subject device are slightly different from the predicate devices. However, the subject diameters are in the range of<br>diameters of predicates and this dimensional difference doesn't affect device safety and effectiveness.<br>The Indications for Use of the subject and primary predicate device are identical other than the compatible implant bodies. This difference is<br>mitigated by fatigue testing, reverse engineering, dimensional analysis and identification of reference device for compatible implant bodies | {8}------------------------------------------------ Both the predicate and subject devices are intended to be milled into patient specific abutments using CAD/CAM technology under the manufacturing control of the sponsor. Any differences in technology characteristics are accompanied by information that demonstrated the device is substantially equivalent as the predicate and do not raise different questions of safety and effectiveness than the predicate. # 2) Patient-Specific Abutment | | Subject Device | Primary<br>Predicate Device | Reference Device | | |-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|-------------------------------------------------------------------------------| | Applicant | DIO Corporation | Medentika GmbH | Zimmer Dental, Inc. | BIOMET 3i | | Trade Name | DIO CAD/CAM Abutment | Medentika CAD/CAM Abutment | Zimmer® patient<br>Specific Abutment,<br>Internal Hex, Titanium | Encode Patient<br>Specific Abutment for<br>Straumann Bone Level<br>Connection | | 510(K) No. | Not yet assigned | K150203 | K143505 | K112730 | | Classification<br>Name | Endosseous Dental Implant, Abutment<br>(872.3630) | Endosseous Dental Implant, Abutment<br>(872.3630) | Endosseous Dental<br>Implant, Abutment<br>(872.3630) | Endosseous Dental<br>Implant, Abutment<br>(872.3630) | | Product Code | NHA | NHA | NHA | NHA | | Class | II | II | II | II | | Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM<br>F136) | Ti-6Al-4V ELI<br>(ASTM F136) | | Design | CAD/CAM Patient-Specific Abutment | CAD/CAM Blank | CAD/CAM Patient-<br>Specific Abutment | CAD/CAM Patient-<br>Specific Abutment | | Diameters (mm) | CAD/CAM Patient-Specific Abutment<br>: 3.0/3.3/3.8/4.0/4.5/5.0/5.5/6.0/6.5/7.0 | 3.0-7.0 | 3.5/ 4.5/ 5.7 | 3.3/ 4.1/ 4.8 | | Sterile | Steam Sterilization by user<br>(Provided Non-Sterile) | Steam Sterilization by user<br>(Provided Non-Sterile) | Steam Sterilization by<br>user<br>(Provided Non-Sterile) | Steam Sterilization by<br>user<br>(Provided Non-Sterile) | | Type of Retention | Screw-retained or cement retained | Cement-retained | Screw-retained or<br>cement retained | Screw-retained or<br>cement retained | | Abutment Seat | Sits on Taper | Sits on Taper | Sits on Taper | Sits on Taper | | Anatomical Site | Oral Cavity | Oral Cavity | Oral Cavity | Oral Cavity | | Constructions | Machined | Machined | Machined | Machined | | Indications For<br>Use/<br>Intended Use | DIO CAD/CAM Abutment is intended for use<br>with dental implants as a support for single or<br>multiple tooth prostheses in the maxilla or | Medentika Preface CAD/CAM Abutments are intended<br>for use with dental implants as a support for single or<br>multiple tooth prostheses in the maxilla or mandible of | The Zimmer® Patient<br>Specific Abutment is<br>used as a terminal or | The 3i Patient-Specific<br>Dental Abutment is<br>intended for use as an | {9}------------------------------------------------ | mandible of a partially or fully edentulous patient. | | | a partially or fully edentulous patient. | | | intermediate abutment<br>for a cemented<br>prosthesis. The<br>abutment can be used<br>for a single of multiple-<br>unit restoration. | accessory to an<br>endosseous dental<br>implant to support a<br>prosthetic device in a<br>partially or edentulous<br>patient. It is intended<br>for use to support<br>single and multiple<br>tooth prostheses, in the<br>mandible or maxilla.<br>The prostheses can be<br>screw or cement<br>retained to the<br>abutment. | | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | Implant<br>System<br>Compatibility | Implant<br>Diameter<br>(mm) | Platform<br>Diameter<br>(mm) | Implant System<br>Compatibility | Series | Implant<br>Diameter<br>(mm) | Platform<br>Diameter<br>(mm) | | | | UF(II)<br>Narrow<br>Implant<br>System | 3.0/ 3.3 | 3.0/ 3.3 | Nobel Biocare<br>ReplaceTM Select | E | 3.5, 4.3, 5.0,<br>6.0 | 3.5, 4.3,<br>5.0, 6.0 | | | | UF Sub<br>merged<br>Implant<br>System | 3.8/4.0/4.5/<br>5.0/5.5/6.0/<br>6.5/7.0 | 3.8/4.0/4.5/<br>5.0/5.5/6.0/<br>6.5/7.0 | Nobel Biocare<br>Nobel ActiveTM | F | 3.5, 4.3, 5.0 | 3.5, 3.9<br>(4.3), 3.9<br>(5.0) | | | | | | | Biomet 3i<br>Osseotite®<br>Certain | H | 3.25, 4.0, 5.0 | 3.4, 4.1,<br>5.0 | | | | | | | Biomet 3i<br>Osseotite® | I | 3.25, 3.75,<br>4.0, 5.0 | 3.4, 4.1,<br>5.0 | | | | | | | Nobel Biocare<br>Branemark | K | 3.3, 3.75,<br>4.0, 5.0 | 3.5, 4.1,<br>4.1, 5.1 | | | | UF(II)<br>Implant<br>System | 3.8/4.0/4.5/<br>5.0/5.5 | 3.8/4.0/4.5/<br>5.0/5.5 | Straumann Bone<br>Level | L | 3.3, 4.1, 4.8 | 3.3, 4.1,<br>4.8 | | | | | | | Straumann<br>Stnadard | N | 3.3, 4.1, 4.8 | 3.5(NNC),<br>4.8, 6.5 | | | | | | | Zimmer Tapered<br>Screw-vent® | R | 3.3, 3.7, 4.1,<br>4.7, 6.0 | 3.5, 4.5,<br>5.7 | | | | | | | Austra Tech<br>OsseoSppedTM | S | 3.5, 4.0, 4.5,<br>5.0 | 3.5, 4.0,<br>4.5, 5.0 | | | |…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...