AIM2 is a silicone impression material for taking oral impressions.
Device Story
AIM2 is a vinyl polysiloxane (VPS) silicone impression material used by dental professionals to capture oral anatomy. The device consists of a base silicone paste and a catalyst silicone paste housed in a cartridge; these are extruded and automixed via a mixing tip. The material undergoes an addition silicone polymerization reaction catalyzed by a platinum complex, with palladium included as a hydrogen scavenger. The clinician injects the mixture onto prepared teeth and into an impression tray, which is then seated in the patient's mouth. After a 3-minute setting time, the impression is removed, cleaned, and disinfected for model pouring. The device facilitates the creation of accurate dental models for restorative or diagnostic purposes.
Clinical Evidence
Bench testing only. The device was tested for physical properties including linear dimensional change, elastic recovery, and strain-in compression, confirming conformance to ISO 4823 standards. Biocompatibility was established based on data from the predicate device.
Indicated for taking oral impressions in patients requiring dental impressions.
Regulatory Classification
Identification
Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
GC America Inc. Mark Heiss, DDS Director. Regulatory Affairs 3737 W. 127th Street Alsip, Illinois 60803
July 12, 2018
## Re: K181011
Trade/Device Name: AIM2 Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: Class II Product Code: ELW Dated: May 17, 2018 Received: May 21, 2018
Dear Mark Heiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Section 4 - Indications for Use Statement
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>Indications for Use | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. |
|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| 510(k) Number (if known) | |
| K181011 | |
| Device Name | |
AIM 2
Indications for Use (Describe)
AIM2 is a silicone impression material for taking oral impressions.
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the letters 'GC' in a stylized, sans-serif font. The letters are a teal color, and there are single quotation marks before and after the letters. The letters are slightly italicized, giving them a dynamic appearance.
GC AMERICA INC
3737 W 127th STREET
ALSIP, ILLINOIS 60803
TEL (708) 597-0900
FAX (708) 926-9100
www.gcamerica.com
Date Prepared: April 16, 2018
- 1. Submitter Information:
- GC AMERICA INC. 3737 W. 127th Street Alsip, IL 60803
| Contact Person: | Mark Heiss, D.D.S. |
|--------------------|--------------------|
| Phone: | (708) 926-3090 |
| Alternate Contact: | Lori Rietman |
| Phone: | (708) 926-3092 |
| Fax: | (708) 926-9100 |
- 2. Device Name:
- Proprietary Name: AIM2 Classification Name: Material, Impression CFR Regulation: 21CFR 872.3660 Device Classification: Class II Product Code: ELW
- 3. Predicate Devices:
| Company | Device | 510(k) No. | Date Cleared |
|-----------------|------------------------------------|------------|---------------|
| GC America Inc. | EXAFLEX/EXAMIX Impression Material | K955932 | March 11,1996 |
- 4. Description of Device:
AIM2 is a VPS silicone impression material. The components consist of the base silicone paste and the catalyst silicone paste that are extruded from a cartridge and are automixed with a mixing tip. AIM2 is available in one viscosity.
The cartridge is made of high density polyethylene and the cap is made of polypropylene. The mixing tip is made of polypropylene.
- న. Indications for Use:
AIM2 is a silicone impression material for taking oral impression.
- 6. Performance Bench Tests
It is confirmed that the device conforms to the required specifications according to ISO 4823 and is suitable for its intended use.
#### Packaging 11.
- a) AIM2, 2-Pk; contains two cartridges and 6 mixing tips
- b) Mixing Tip II: L
#### 7. Shelf Life Evaluation and Storage Conditions:
- Shelf Life 2 years ।
- Recommended for optimal performance, store at room temperature. 15-25°C (59.0 77.0°F) (Relative । humidity: 50%).
{4}------------------------------------------------
| | Applicant device | Predicate device |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product<br>category | Hydrophilic vinyl polysiloxane impression<br>material | Hydrophilic vinyl polysiloxane impression<br>material |
| Trade name | AIM2 | EXAMIX |
| Manufacturer | GC Corporation | GC Corporation |
| Indications for<br>use | AIM2 is a silicone impression material for taking<br>oral impression. | EXAMIX is a silicone impression material for<br>taking oral impression. |
| Product<br>description | The device consists of the base paste and the<br>catalyst paste Including VPS.<br>The material sets by mixing the catalyst paste<br>with the base paste. | The device consists of the base paste and the<br>catalyst paste Including VPS.<br>The material sets by mixing the catalyst paste<br>with the base paste. |
| Instructions<br>for use | Note: The cartridge and mixing tip provided are<br>compatible with GC Cartridge Dispenser 2.<br><b>CARTRIDGE LOADING AND DISPENSING</b><br>1. Lift the release lever of the CARTRIDGE<br>DISPENSER 2. (referred to as the dispenser<br>hereafter) and pull the piston plunger all the way<br>back into the dispenser. Lift the cartridge holder<br>of the dispenser and load the cartridge, ensuring<br>that the V-shaped notch on the flange of the<br>cartridge is facing down. Push the cartridge<br>holder down to hold the cartridge firmly in place.<br>2. Lift the release lever and push the piston<br>plunger forward until it engages into the<br>cartridge.<br>3. Remove the cartridge cap by rotating 1/4 turn<br>anti-clockwise. Tilt the cap downward and peel it<br>away from the cartridge. Gently squeeze the<br>dispenser handle to extrude a small amount of<br>material from the two openings at the end of<br>cartridge. Make sure that base and catalyst<br>come out evenly.<br>4. Align the V-shaped notch on the rim of mixing<br>tip to the V-shaped notch between the cartridge<br>barrels. Push firmly to attach the mixing tip.<br>Then rotate the colored collar of the mixing tip<br>1/4 turn clockwise to the end of the cartridge.<br>The dispenser is now ready for use.<br>5. Squeeze the handle several times to extrude<br>the material. After use, do not remove the<br>mixing tip as this will become the storage cap<br>until next use. When replacing the mixing tip,<br>rotate the collar on the mixing tip 1/4 turn anti-<br>clockwise to align the V-shaped notch on the<br>cartridge. Tilt the mixing tip downward and peel<br>it away from the cartridge.<br>6. Remove and replace the old mixing tip<br>immediately prior to next use. Before attaching a<br>new tip, gently extrude a small amount of<br>material to ensure that base and catalyst are<br>flowing evenly from both openings. If the<br>materials should fail to extrude, remove any<br>hardened materials from the end of the<br>cartridge.<br>7. To replace the cartridge, lift the release lever<br>and retract the piston plunger fully. Remove the<br>empty cartridge by lifting the cartridge holder<br>and load a new one into the dispenser.<br><b>USING THE IMPRESSION MATERIAL</b><br>Note: Prior to use, leave the material to stand at<br>room temperature. Cooled or refrigerated<br>material may delay setting. | Note: The cartridge and mixing tip provided are<br>compatible with GC Cartridge Dispenser 2.<br><b>CARTRIDGE LOADING AND DISPENSING</b><br>1. Lift the release lever of the CARTRIDGE<br>DISPENSER 2. (referred to as the dispenser<br>hereafter) and pull the piston plunger all the way<br>back into the dispenser. Lift the cartridge holder<br>of the dispenser and load the cartridge, ensuring<br>that the V-shaped notch on the flange of the<br>cartridge is facing down. Push the cartridge<br>holder down to hold the cartridge firmly in place.<br>2. Lift the release lever and push the piston<br>plunger forward until it engages into the<br>cartridge.<br>3. Remove the cartridge cap by rotating 1/4 turn<br>anti-clockwise. Tilt the cap downward and peel it<br>away from the cartridge. Gently squeeze the<br>dispenser handle to extrude a small amount of<br>material from the two openings at the end of<br>cartridge. Make sure that base and catalyst<br>come out evenly.<br>4. Align the V-shaped notch on the rim of mixing<br>tip to the V-shaped notch between the cartridge<br>barrels. Push firmly to attach the mixing tip.<br>Then rotate the colored collar of the mixing tip<br>1/4 turn clockwise to the end of the cartridge.<br>The dispenser is now ready for use.<br>5. Squeeze the handle several times to extrude<br>the material. After use, do not remove the mixing<br>tip as this will become the storage cap until next<br>use. When replacing the mixing tip, rotate the<br>collar on the mixing tip 1/4 turn anti-clockwise to<br>align the V-shaped notch on the cartridge. Tilt<br>the mixing tip downward and peel it away from<br>the cartridge.<br>6. Remove and replace the old mixing tip<br>immediately prior to next use. Before attaching a<br>new tip, gently extrude a small amount of<br>material to ensure that base and catalyst are<br>flowing evenly from both openings. If the<br>materials should fail to extrude, remove any<br>hardened materials from the end of the<br>cartridge.<br>7. To replace the cartridge, lift the release lever<br>and retract the piston plunger fully. Remove the<br>empty cartridge by lifting the cartridge holder<br>and load a new one into the dispenser.<br><b>USING THE IMPRESSION MATERIAL</b><br>Note: Prior to use, leave the material to stand at<br>room temperature. Cooled or refrigerated<br>material may delay setting. |
| | 1. Attach an intraoral tip to the mixing tip or load<br>material into a syringe. 2. Inject the mixture onto the prepared teeth. 3. Load the tray with the material. | 1. Attach an intraoral tip to the mixing tip or load<br>material into a syringe. 2. Inject the mixture onto the prepared teeth. 3. Load the tray with the appropriate material (putty, monophase, heavy body or regular - depending on technique being used). |
| | 4. Seat the loaded tray in the mouth within<br>specified working times. 5. Wait for set 3 minutes in the mouth). 6. The obtained impression should be cleaned, then disinfected, utilizing a 2.5% or 3.4% glutaraldehyde, or other appropriate disinfectant, according to the manufacturers label recommendations. 7. Remove the impression and pour a model immediately (if desired). Maximum time for pouring the model is 14 days. | 4. Seat the loaded tray in the mouth within<br>specified working times. 5. Wait for set 4 minutes in the mouth). 6. The obtained impression should be cleaned, then disinfected, utilizing a 2.5% or 3.4% glutaraldehyde, or other appropriate disinfectant, according to the manufacturers label recommendations. 7. Remove the impression and pour a model immediately (if desired). Maximum time for pouring the model is 14 days. |
| Technological<br>Characteristics | The curing mechanism is described as an<br>addition silicone reaction. It is an addition<br>polymerization between vinyl polysiloxanes and<br>poly-me,-siloxanes hydrogen terminated. A<br>platinum complex is used as a catalyst. A side<br>reaction may occur and release small quantity of<br>hydrogen gas. Therefore, palladium is used as a<br>scavenger. A surfactant is added to allow<br>wettability/hydrophilicity.<br>Setting for AIM2 is 3 minutes in the mouth.<br>Results of benchtop testing indicate the physical<br>properties such as linear dimensional change,<br>elastic recover and strain-in compression met<br>the specification set in ISO 4823.<br>Biological safety test data shows the<br>biocompatibility of the predicate device. | The curing mechanism is described as an<br>addition silicone reaction. It is an addition<br>polymerization between vinyl<br>dimethylpolysiloxane and methylhydrogen<br>dimethylpolysiloxane. A platinum complex is<br>used as a catalyst. A surfactant is added to allow<br>wettability/hydrophilicity.<br>Setting time for Examix is 4 minutes in the<br>mouth.<br>Results of benchtop testing indicate the physical<br>properties such as linear dimensional change,<br>elastic recovery and strain-in compression meet<br>the specification set in ISO 4823.<br>Biological safety test data shows the<br>biocompatibility of the predicate device. |
## Table 5.1.1 Comparison Table
{5}------------------------------------------------
#### 8. Substantial equivalence:
The applicant and predicate device are the same in function and intended use. The curing mechanism of the applicant and predicate device is substantially equivalent in principle. This supports that the compatibility and safety of the applicant device is substantially equivalent to the predicate device.
Although there are slight differences in bench top performance tests, results meet requirements as outlined in ISO 4823:2015.
#### 9. Differences
The following differences are noted: setting time for AIM2 is 3 minutes in the mouth and Examix is 4 minutes in mouth.
## 10. Conclusion
Based on similarities in intended use, mode of action, chemical composition, and performance testing, AIM2 is substantially equivalent to the selected predicate Examix (K955932).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.