AIM2

K181011 · GC America, Inc. · ELW · Jul 12, 2018 · Dental

Device Facts

Record IDK181011
Device NameAIM2
ApplicantGC America, Inc.
Product CodeELW · Dental
Decision DateJul 12, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3660
Device ClassClass 2

Indications for Use

AIM2 is a silicone impression material for taking oral impressions.

Device Story

AIM2 is a vinyl polysiloxane (VPS) silicone impression material used by dental professionals to capture oral anatomy. The device consists of a base silicone paste and a catalyst silicone paste housed in a cartridge; these are extruded and automixed via a mixing tip. The material undergoes an addition silicone polymerization reaction catalyzed by a platinum complex, with palladium included as a hydrogen scavenger. The clinician injects the mixture onto prepared teeth and into an impression tray, which is then seated in the patient's mouth. After a 3-minute setting time, the impression is removed, cleaned, and disinfected for model pouring. The device facilitates the creation of accurate dental models for restorative or diagnostic purposes.

Clinical Evidence

Bench testing only. The device was tested for physical properties including linear dimensional change, elastic recovery, and strain-in compression, confirming conformance to ISO 4823 standards. Biocompatibility was established based on data from the predicate device.

Technological Characteristics

VPS silicone impression material; addition polymerization curing mechanism; platinum catalyst; palladium hydrogen scavenger; surfactant for hydrophilicity. Cartridge: high-density polyethylene; cap/mixing tip: polypropylene. Conforms to ISO 4823. Setting time: 3 minutes. Storage: 15-25°C.

Indications for Use

Indicated for taking oral impressions in patients requiring dental impressions.

Regulatory Classification

Identification

Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.

Special Controls

*Classification.* Class II (Special Controls).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. GC America Inc. Mark Heiss, DDS Director. Regulatory Affairs 3737 W. 127th Street Alsip, Illinois 60803 July 12, 2018 ## Re: K181011 Trade/Device Name: AIM2 Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: Class II Product Code: ELW Dated: May 17, 2018 Received: May 21, 2018 Dear Mark Heiss: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Mary S. Runner -S For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Section 4 - Indications for Use Statement | DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>Indications for Use | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. | |------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------| | 510(k) Number (if known) | | | K181011 | | | Device Name | | AIM 2 Indications for Use (Describe) AIM2 is a silicone impression material for taking oral impressions. Type of Use (Select one or both, as applicable) [X] Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17) PSC Publishing Services (301) 443-6740 EF {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the letters 'GC' in a stylized, sans-serif font. The letters are a teal color, and there are single quotation marks before and after the letters. The letters are slightly italicized, giving them a dynamic appearance. GC AMERICA INC 3737 W 127th STREET ALSIP, ILLINOIS 60803 TEL (708) 597-0900 FAX (708) 926-9100 www.gcamerica.com Date Prepared: April 16, 2018 - 1. Submitter Information: - GC AMERICA INC. 3737 W. 127th Street Alsip, IL 60803 | Contact Person: | Mark Heiss, D.D.S. | |--------------------|--------------------| | Phone: | (708) 926-3090 | | Alternate Contact: | Lori Rietman | | Phone: | (708) 926-3092 | | Fax: | (708) 926-9100 | - 2. Device Name: - Proprietary Name: AIM2 Classification Name: Material, Impression CFR Regulation: 21CFR 872.3660 Device Classification: Class II Product Code: ELW - 3. Predicate Devices: | Company | Device | 510(k) No. | Date Cleared | |-----------------|------------------------------------|------------|---------------| | GC America Inc. | EXAFLEX/EXAMIX Impression Material | K955932 | March 11,1996 | - 4. Description of Device: AIM2 is a VPS silicone impression material. The components consist of the base silicone paste and the catalyst silicone paste that are extruded from a cartridge and are automixed with a mixing tip. AIM2 is available in one viscosity. The cartridge is made of high density polyethylene and the cap is made of polypropylene. The mixing tip is made of polypropylene. - న. Indications for Use: AIM2 is a silicone impression material for taking oral impression. - 6. Performance Bench Tests It is confirmed that the device conforms to the required specifications according to ISO 4823 and is suitable for its intended use. #### Packaging 11. - a) AIM2, 2-Pk; contains two cartridges and 6 mixing tips - b) Mixing Tip II: L #### 7. Shelf Life Evaluation and Storage Conditions: - Shelf Life 2 years । - Recommended for optimal performance, store at room temperature. 15-25°C (59.0 77.0°F) (Relative । humidity: 50%). {4}------------------------------------------------ | | Applicant device | Predicate device | |----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product<br>category | Hydrophilic vinyl polysiloxane impression<br>material | Hydrophilic vinyl polysiloxane impression<br>material | | Trade name | AIM2 | EXAMIX | | Manufacturer | GC Corporation | GC Corporation | | Indications for<br>use | AIM2 is a silicone impression material for taking<br>oral impression. | EXAMIX is a silicone impression material for<br>taking oral impression. | | Product<br>description | The device consists of the base paste and the<br>catalyst paste Including VPS.<br>The material sets by mixing the catalyst paste<br>with the base paste. | The device consists of the base paste and the<br>catalyst paste Including VPS.<br>The material sets by mixing the catalyst paste<br>with the base paste. | | Instructions<br>for use | Note: The cartridge and mixing tip provided are<br>compatible with GC Cartridge Dispenser 2.<br><b>CARTRIDGE LOADING AND DISPENSING</b><br>1. Lift the release lever of the CARTRIDGE<br>DISPENSER 2. (referred to as the dispenser<br>hereafter) and pull the piston plunger all the way<br>back into the dispenser. Lift the cartridge holder<br>of the dispenser and load the cartridge, ensuring<br>that the V-shaped notch on the flange of the<br>cartridge is facing down. Push the cartridge<br>holder down to hold the cartridge firmly in place.<br>2. Lift the release lever and push the piston<br>plunger forward until it engages into the<br>cartridge.<br>3. Remove the cartridge cap by rotating 1/4 turn<br>anti-clockwise. Tilt the cap downward and peel it<br>away from the cartridge. Gently squeeze the<br>dispenser handle to extrude a small amount of<br>material from the two openings at the end of<br>cartridge. Make sure that base and catalyst<br>come out evenly.<br>4. Align the V-shaped notch on the rim of mixing<br>tip to the V-shaped notch between the cartridge<br>barrels. Push firmly to attach the mixing tip.<br>Then rotate the colored collar of the mixing tip<br>1/4 turn clockwise to the end of the cartridge.<br>The dispenser is now ready for use.<br>5. Squeeze the handle several times to extrude<br>the material. After use, do not remove the<br>mixing tip as this will become the storage cap<br>until next use. When replacing the mixing tip,<br>rotate the collar on the mixing tip 1/4 turn anti-<br>clockwise to align the V-shaped notch on the<br>cartridge. Tilt the mixing tip downward and peel<br>it away from the cartridge.<br>6. Remove and replace the old mixing tip<br>immediately prior to next use. Before attaching a<br>new tip, gently extrude a small amount of<br>material to ensure that base and catalyst are<br>flowing evenly from both openings. If the<br>materials should fail to extrude, remove any<br>hardened materials from the end of the<br>cartridge.<br>7. To replace the cartridge, lift the release lever<br>and retract the piston plunger fully. Remove the<br>empty cartridge by lifting the cartridge holder<br>and load a new one into the dispenser.<br><b>USING THE IMPRESSION MATERIAL</b><br>Note: Prior to use, leave the material to stand at<br>room temperature. Cooled or refrigerated<br>material may delay setting. | Note: The cartridge and mixing tip provided are<br>compatible with GC Cartridge Dispenser 2.<br><b>CARTRIDGE LOADING AND DISPENSING</b><br>1. Lift the release lever of the CARTRIDGE<br>DISPENSER 2. (referred to as the dispenser<br>hereafter) and pull the piston plunger all the way<br>back into the dispenser. Lift the cartridge holder<br>of the dispenser and load the cartridge, ensuring<br>that the V-shaped notch on the flange of the<br>cartridge is facing down. Push the cartridge<br>holder down to hold the cartridge firmly in place.<br>2. Lift the release lever and push the piston<br>plunger forward until it engages into the<br>cartridge.<br>3. Remove the cartridge cap by rotating 1/4 turn<br>anti-clockwise. Tilt the cap downward and peel it<br>away from the cartridge. Gently squeeze the<br>dispenser handle to extrude a small amount of<br>material from the two openings at the end of<br>cartridge. Make sure that base and catalyst<br>come out evenly.<br>4. Align the V-shaped notch on the rim of mixing<br>tip to the V-shaped notch between the cartridge<br>barrels. Push firmly to attach the mixing tip.<br>Then rotate the colored collar of the mixing tip<br>1/4 turn clockwise to the end of the cartridge.<br>The dispenser is now ready for use.<br>5. Squeeze the handle several times to extrude<br>the material. After use, do not remove the mixing<br>tip as this will become the storage cap until next<br>use. When replacing the mixing tip, rotate the<br>collar on the mixing tip 1/4 turn anti-clockwise to<br>align the V-shaped notch on the cartridge. Tilt<br>the mixing tip downward and peel it away from<br>the cartridge.<br>6. Remove and replace the old mixing tip<br>immediately prior to next use. Before attaching a<br>new tip, gently extrude a small amount of<br>material to ensure that base and catalyst are<br>flowing evenly from both openings. If the<br>materials should fail to extrude, remove any<br>hardened materials from the end of the<br>cartridge.<br>7. To replace the cartridge, lift the release lever<br>and retract the piston plunger fully. Remove the<br>empty cartridge by lifting the cartridge holder<br>and load a new one into the dispenser.<br><b>USING THE IMPRESSION MATERIAL</b><br>Note: Prior to use, leave the material to stand at<br>room temperature. Cooled or refrigerated<br>material may delay setting. | | | 1. Attach an intraoral tip to the mixing tip or load<br>material into a syringe. 2. Inject the mixture onto the prepared teeth. 3. Load the tray with the material. | 1. Attach an intraoral tip to the mixing tip or load<br>material into a syringe. 2. Inject the mixture onto the prepared teeth. 3. Load the tray with the appropriate material (putty, monophase, heavy body or regular - depending on technique being used). | | | 4. Seat the loaded tray in the mouth within<br>specified working times. 5. Wait for set 3 minutes in the mouth). 6. The obtained impression should be cleaned, then disinfected, utilizing a 2.5% or 3.4% glutaraldehyde, or other appropriate disinfectant, according to the manufacturers label recommendations. 7. Remove the impression and pour a model immediately (if desired). Maximum time for pouring the model is 14 days. | 4. Seat the loaded tray in the mouth within<br>specified working times. 5. Wait for set 4 minutes in the mouth). 6. The obtained impression should be cleaned, then disinfected, utilizing a 2.5% or 3.4% glutaraldehyde, or other appropriate disinfectant, according to the manufacturers label recommendations. 7. Remove the impression and pour a model immediately (if desired). Maximum time for pouring the model is 14 days. | | Technological<br>Characteristics | The curing mechanism is described as an<br>addition silicone reaction. It is an addition<br>polymerization between vinyl polysiloxanes and<br>poly-me,-siloxanes hydrogen terminated. A<br>platinum complex is used as a catalyst. A side<br>reaction may occur and release small quantity of<br>hydrogen gas. Therefore, palladium is used as a<br>scavenger. A surfactant is added to allow<br>wettability/hydrophilicity.<br>Setting for AIM2 is 3 minutes in the mouth.<br>Results of benchtop testing indicate the physical<br>properties such as linear dimensional change,<br>elastic recover and strain-in compression met<br>the specification set in ISO 4823.<br>Biological safety test data shows the<br>biocompatibility of the predicate device. | The curing mechanism is described as an<br>addition silicone reaction. It is an addition<br>polymerization between vinyl<br>dimethylpolysiloxane and methylhydrogen<br>dimethylpolysiloxane. A platinum complex is<br>used as a catalyst. A surfactant is added to allow<br>wettability/hydrophilicity.<br>Setting time for Examix is 4 minutes in the<br>mouth.<br>Results of benchtop testing indicate the physical<br>properties such as linear dimensional change,<br>elastic recovery and strain-in compression meet<br>the specification set in ISO 4823.<br>Biological safety test data shows the<br>biocompatibility of the predicate device. | ## Table 5.1.1 Comparison Table {5}------------------------------------------------ #### 8. Substantial equivalence: The applicant and predicate device are the same in function and intended use. The curing mechanism of the applicant and predicate device is substantially equivalent in principle. This supports that the compatibility and safety of the applicant device is substantially equivalent to the predicate device. Although there are slight differences in bench top performance tests, results meet requirements as outlined in ISO 4823:2015. #### 9. Differences The following differences are noted: setting time for AIM2 is 3 minutes in the mouth and Examix is 4 minutes in mouth. ## 10. Conclusion Based on similarities in intended use, mode of action, chemical composition, and performance testing, AIM2 is substantially equivalent to the selected predicate Examix (K955932).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...