K181009 · Medartis AG · HRS · Aug 24, 2018 · Orthopedic
Device Facts
Record ID
K181009
Device Name
APTUS CMC-I Fusion Plate System
Applicant
Medartis AG
Product Code
HRS · Orthopedic
Decision Date
Aug 24, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The APTUS® CMC-I Fusion Plate System is intended to be used for fusion of the trapezium with the first metacarpal.
Device Story
The APTUS® CMC-I Fusion Plate System is an orthopedic bone fixation device used for the fusion of the trapezium with the first metacarpal. The system consists of an anatomically designed, unalloyed titanium plate (1.3 mm thick, 35 mm x 13 mm) and compatible bone screws (2.0 mm diameter locking, non-locking, and cannulated). The device is used by orthopedic surgeons in a clinical setting to stabilize bone segments during fusion procedures. The plate is designed to be bent to conform to the patient's hand anatomy. By providing rigid fixation, the device facilitates bone healing and fusion, potentially reducing pain and restoring function in the carpometacarpal joint. The device is provided to the end-user for sterilization using established methods.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including four-point bend testing per ASTM F382 and comparative dynamic testing in a simulated fracture model. Biocompatibility is referenced from predicate K051567.
Technological Characteristics
Material: Unalloyed titanium, grade 4 (ASTM F67). Screw material: Ti-6Al-4V alloy (ASTM F136). Design: Anatomical plate, 1.3 mm uniform thickness, 35 mm x 13 mm dimensions. Fixation: Compatible with 2.0 mm locking, non-locking, and cannulated screws. Sterilization: End-user sterilized. Mechanical testing: ASTM F382.
Indications for Use
Indicated for fusion of the trapezium with the first metacarpal in patients requiring internal fixation of the hand and wrist.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 24, 2018
Medartis AG % Kevin A. Thomas, Ph.D. Vice President and Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K181009
Trade/Device Name: APTUS CMC-I Fusion Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances and Accessories Regulatory Class: Class II Product Code: HRS Dated: July 23, 2018 Received: July 24, 2018
Dear Kevin A. Thomas, Ph.D .:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K181009
Device Name
APTUS® CMC-I Fusion Plate System
Indications for Use (Describe)
The APTUS® CMC-I Fusion Plate System is intended to be used for fusion of the trapezium with the first metacarpal.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
# Medartis AG
# APTUS® CMC-I Fusion Plate System
April 17, 2018
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Medartis AG<br>Hochbergerstrasse 60E<br>CH-4057 Basel, Switzerland |
|-------------------|--------------------------------------------------------------------|
| | Telephone: +41 61 633 34 34<br>Fax: +41 61 633 34 00 |
| Official Contact | Andrea Schweizer<br>Head of Quality Management |
| Representative/Consultant | Kevin A. Thomas, PhD |
|---------------------------|------------------------------------------|
| | Floyd G. Larson, MS, MBA |
| | PaxMed International, LLC |
| | 12264 El Camino Real, Suite 400 |
| | San Diego, CA 92130 |
| Telephone: | +1-858-792-1235 |
| Fax: | +1-858-792-1236 |
| Email: | kthomas@paxmed.com<br>flarson@paxmed.com |
## DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | APTUS® CMC-I Fusion Plate System |
|---------------------------|-------------------------------------------------------------------------------|
| Common Name | Plate, fixation, bone |
| Classification Name | Single/multiple component metallic bone fixation appliance<br>and accessories |
| Classification Regulation | 21 CFR 888.3030 |
| Product Code | HRS |
| Classification Panel | Orthopedic Products Panel |
| Reviewing Branch | Joint Fixation Devices Branch Two (JFDB2) |
#### PREDICATE DEVICE INFORMATION
The primary predicate device is K062498, Profyle® System, Howmedica Osteonics Corporation. The reference predicate device is K051567, APTUS® Titanium System, Medartis, Inc.
#### INDICATIONS FOR USE
The APTUS® CMC-I Fusion Plate System is intended to be used for fusion of the trapezium with the first metacarpal.
{4}------------------------------------------------
## SUBJECT DEVICE DESCRIPTION
The subject device plate has an anatomical design appropriate for both the left and right wrist. The plate has a uniform thickness of 1.3 mm and has overall dimensions of approximately 35 mm x 13 mm, before bending to conform to the hand anatomy. The screw holes of the plate are designed to accommodate appropriately sized bone screws presently marketed as part of the APTUS® System and previously cleared under K051567. The subject device plate is compatible with 2.0 mm diameter screws, and is used with TriLock locking screws and cortical (nonlocking) screws. The subject device plate is made of unalloyed titanium conforming to ASTM F67.
## PERFORMANCE DATA
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence include: biocompatibility (referenced from K051567), mechanical testing (according to ASTM F382), and comparative dynamic testing in a simulated fracture model. Clinical data were not provided in this submission.
## EQUIVALENCE TO MARKETED DEVICE
Medartis AG submits the information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, the subject device is substantially equivalent in indications and design principles to the following legally marketed predicate devices:
K062498, Profyle® System, Howmedica Osteonics Corporation; and
K051567, APTUS® Titanium System, Medartis, Inc.
A comparison of the technological characteristics of the subject device and the primary predicate device K062498 is provided in the following table.
| | Subject Device | Primary Predicate Device |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison | APTUS® CMC-I Fusion Plate System<br>Medartis AG | K062498<br>Profyle® System<br>Howmedica Osteonics Corporation |
| Indications for Use<br>Statement | The APTUS® CMC-I Fusion Plate System is intended to<br>be used for fusion of the trapezium with the first<br>metacarpal. | The Profyle® System is intended for use in internal<br>fixation of the bones of hand and wrist. Examples of<br>these procedures may include but are not limited to<br>replantation, lag screw techniques, joint fusions,<br>corrective osteotomies, and the treatment of fractures. |
| Device Characteristics | | |
| Plate Designs | Anatomic plate design<br>One size<br>Single design for right and left hand-wrist;<br>Screw holes accommodate conventional (cortex) and<br>locking screws | Anatomic plate designs<br>Multiple sizes<br>Multiple designs<br>Screw holes accommodate conventional (cortex) and<br>locking screws |
| Plate Overall Dimensions<br>(Approximate) | 13 mm width x 35 mm length | Various |
| Plate Thickness | Uniform thickness, 1.3 mm | Uniform thickness, range from 0.55 mm – 1.5 mm |
| Plate Material | Unalloyed titanium, grade 4 (ASTM F67) | Unalloyed titanium, grade 2 |
| Screw Designs | Previously cleared:<br>Locking screws: cortical thread form, double thread,<br>self-tapping, fully threaded<br><br>Non-locking screws: cortical thread form, single thread,<br>self-tapping, fully threaded<br><br>Cannulated screws: cortical thread form, single thread,<br>self-tapping, partially threaded, cannulated | Cortical thread form<br>Self-tapping, fully threaded<br>Conventional (non-locking) and locking |
| Screw Diameter | Previously cleared:<br>2.0 mm (locking and non-locking)<br>3.0 mm (cannulated) | 1.2 mm to 2.5 mm |
{5}------------------------------------------------
| | Subject Device | Primary Predicate Device |
|----------------|-------------------------------------------------------------------------------------------------------|--------------------------|
| Screw Length | <i>Previously cleared:</i><br>8 mm to 20 mm (conventional and locking)<br>18 mm to 32 mm (cannulated) | 6 mm to 26 mm |
| Screw Material | <i>Previously cleared:</i><br>Ti-6Al-4V alloy, ASTM F136 | Ti-6Al-4V alloy |
The subject device, the primary predicate device K062498, and reference predicate device K051567 have the same intended use for internal fixation of the bones of the hand and wrist, and have the same technological characteristics. The subject device and the plates from both predicate devices (K062498 and K051567) are fabricated from unalloyed titanium. The subject device and the reference predicate device K051567 have similar design characteristics, including the design of screw holes to accommodate locking and non-locking screws. The subject device and both predicate devices (plates) encompass a similar range of physical dimensions (overall width, length, and thickness). The subject device plates are compatible exclusively with screws from the reference predicate device K051567.
The differences between the subject device and primary predicate device K062498 are slight variations in the design of the plates and compatible screws. The differences between the subject device and reference predicate device K051567 plates are the specific design configurations and dimensions. These slight differences among the subject device and predicate devices do not raise of safety or efficacy.
The final finished subject device is manufactured in the same facilities using identical materials and identical manufacturing processes as used for the previously cleared reference predicate device components cleared in K051567, and therefore are substantially equivalent to the reference predicate device regarding biocompatibility. The subject device and reference predicate device K051567 components are packaged using the same materials, and are to be sterilized by the same methods.
The Indications for Use Statement for the subject device is similar to that of the primary predicate device K062498. The slight differences in wording do not affect the intended use for internal fixation of the hand and wrist.
Performance data provided to demonstrate substantial equivalence included four-point bend testing (according to ASTM F382), and comparative dynamic testing in a simulated fracture model.
# CONCLUSION
The subject device plates and the primary predicate device plates have intended use, have similar technological characteristics, encompass a similar range of physical dimensions appropriate to the anatomy and are made of the same material. The subject device and reference predicate device K051567 are packaged in similar materials and are to be sterilized by the end-user using similar methods. Performance data demonstrated the subject device plates to be substantially equivalent to K062498.
The data included in this submission demonstrate substantial equivalence to the predicate devices K062498 and K051567.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.