K180986 · Xablecath, Inc. · DQY · May 15, 2018 · Cardiovascular
Device Facts
Record ID
K180986
Device Name
XableCath Support Catheter Product Family
Applicant
Xablecath, Inc.
Product Code
DQY · Cardiovascular
Decision Date
May 15, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K180986 · May 15, 2018
XableCath Support Catheter Product Family
Xablecath, Inc.
Clinical observation/chart review of 26 patients undergoing routine angiography
The sponsor conducted a real-world actual-use assessment of the XableCath Support Catheter (abrasion-tip configuration) in 26 patients to confirm design adequacy and safety during routine peripheral vascular procedures.
26 patients (20 male, 6 female) with peripheral artery disease; mean age 69 years; Sample Size: 26 patients (27 catheters used)
Not applicable for this study
Procedural complications, vessel perforation, major dissection, distal debris embolization, and post-procedural clinical sequelae
Indications for Use
The XableCath catheter is intended to be used to facilitate access to discrete regions of the peripheral vasculature in conjunction with steerable guidewires. This device may be used to facilitate placement and exchange of guidewires and other interventional devices.
Device Story
XableCath Support Catheter is a percutaneous, single-lumen, over-the-wire (OTW) catheter used in peripheral vasculature. It facilitates guidewire placement and exchange during interventional procedures. The device features an abrasion-resistant distal tip (cobalt-chromium alloy) designed to navigate complex arterial segments. It is operated by physicians in hospital or surgical suite settings under fluoroscopic guidance. The catheter is inserted over a steerable guidewire; the physician advances it to the target lesion or vessel segment. The device provides a stable platform for exchanging interventional tools. By enabling safe access to discrete vascular regions, it assists in treating peripheral artery disease (PAD). The device is single-use and sterile.
Clinical Evidence
Clinical validation involved 26 patients (mean age 69, 77% male) with PAD undergoing lower extremity angiography. 27 catheters were used to traverse 80 arterial segments. Primary endpoints focused on procedural safety and passage adequacy. Results showed no procedural complications attributable to the device; one minor dissection occurred due to subsequent balloon angioplasty. No thrombosis, rupture, or embolization occurred. Follow-up at mean 53 days showed no device-related complications.
Technological Characteristics
Single-lumen catheter; Nylon shaft; HDPE proximal hub; Olefin strain relief; Cobalt-chromium alloy distal tip (abrasion configuration); Cobalt-chromium/SS markers. Sterilized via Gamma radiation (SAL 10^-6). Complies with ISO 10555-1. Burst strength ≥300 psi. OTW manual deployment.
Indications for Use
Indicated for patients requiring peripheral vascular access for guidewire placement and exchange. No specific age or gender contraindications noted.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
May 15, 2018
XableCath, Inc. Mr. Rick Gaykowski Chief Regulatory Officer 417 South Wakara Way, Suite 3510 Salt Lake City, Utah 84108-1457
Re: K180986
Trade/Device Name: XableCath Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: April 7, 2018 Received: April 16, 2018
Dear Mr. Gaykowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Nicole G. Ibrahim -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180986
Device Name XableCath Support Catheter
Indications for Use (Describe)
The XableCath catheter is intended to be used to facilitate access to discrete regions of the peripheral vasculature in conjunction with steerable guidewires. This device may be used to facilitate placement and exchange of guidewires and other interventional devices.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif;">✘</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K180986
XableCath
Special 510(k) Special 510(k)
XableCath™ Support Catheter
# 510(k) SUMMARY XableCath™ Support Catheter
| Submitter Information [21 CFR 807.929(a)(1)] | | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | XableCath, Inc. | | |
| Address | 417 S Wakara Way, Suite 3510<br>Salt Lake City, UT 84108-1457<br>USA | | |
| Phone number | 617-447-4000 Mobile | | |
| Fax number | N/A | | |
| Establishment<br>Registration Number | Yet to be secured | | |
| Name of contact<br>person(s) | Rick Gaykowski, Chief Regulatory Officer, Primary<br>Lisa Dunlea, Pres/CEO, Secondary | | |
| Date prepared | April 7, 2018 | | |
| Name of the device [21 CFR 807.92(a)(2)] | | | |
| Trade or proprietary<br>name | XableCath Support Catheter | | |
| Common or usual<br>name | Peripheral Vascular Support Catheter | | |
| Classification name | Class II | | |
| Regulation | 21 CFR 870.1250 | | |
| Product Code(s) | DQY | | |
| Legally marketed<br>device(s) to which<br>equivalence is claimed<br>[21 CFR 807.92(a)(3)] | XableCath Support Catheter, cleared under 510(k) K170041 | | |
| Device description<br>[21 CFR 807.92(a)(4)] | Catheter, Percutaneous | | |
| Indications for use<br>[21 CFR 807.92(a)(5)] | The XableCath catheter is intended to be used to facilitate access to discrete<br>regions of the peripheral vasculature in conjunction with steerable guidewires.<br>This device may be used to facilitate placement and exchange of guidewires and<br>other interventional devices. | | |
| Summary of the technological characteristics of the device compared to the predicate device<br>[21 CFR 807.92(a)(6)] | | | |
| Feature | Subject Device: XableCath Support<br>Catheter | | |
| | Predicate Device: XableCath Support<br>Catheter | | |
| Proximal Luer/Hub | HDPE | | |
| | HDPE | | |
| Strain Relief | Olefin | Olefin | |
| Catheter Shaft | Nylon, Single lumen | Nylon, Single lumen | |
| Markers | Cobalt-chromium/SS | Cobalt-chromium/SS | |
| Distal Tip | Abrasion | Blunt | |
| Distal Tip Material | Cobalt chromium alloy | Cobalt chromium alloy | |
| Coating | N/A | N/A | |
| Infusion<br>Pressure/Burst<br>Strength | ≥300 psi | ≥300 psi | |
| | Outer Diameter | Varies by Model<br>Proximal: 0.043", 0.053", 0.062"<br>Distal: 0.060", 0.070", 0.080"<br>Tip: 0.052", 0.063", 0.071 | Varies by Model<br>Proximal: 0.043", 0.053", 0.062"<br>Distal: 0.060", 0.070", 0.080"<br>Tip: 0.052", 0.063", 0.071 |
| | | Guidewire<br>Compatibility | 0.014", 0.018", 0.035" |
| Guide Catheter (max) | | | ≥4Fr |
| Access Sheath (max) | ≥4Fr | ≥4Fr | |
| Effective Working<br>Length | 65, 90, 145cm | 65, 90, 145cm | |
| | Deployment | OTW - Manual | OTW - Manual |
| Sterilization Method | Gamma<br>(SAL - 10-6) | Gamma<br>(SAL - 10-6) | |
| | Single Use, Sterile | Yes | Yes |
| Labeling | Individual IFU: Warning, Cautions,<br>Contraindications, tables, images,<br>organized outline. | Individual IFU: Warning, Cautions,<br>Contraindications, tables, images,<br>organized outline. | |
| | Packaging | Sterile thermal sealed Tyvek/PET<br>Pouch, SBS Carton | Sterile thermal sealed Tyvek/PET<br>Pouch, SBS Carton |
| Use Environment | | Rx Only - By or on the order of a<br>physician. Hospital, Lab/Surgical<br>Suite | Rx Only - By or on the order of a<br>physician. Hospital, Lab/Surgical<br>Suite |
| | Performance Testing | ISO 10555-1 Second Edition 2013-<br>07-01. Intravascular catheters --<br>Sterile, single-use intravascular<br>catheters -- Part 1: General<br>Requirements | ISO 10555-1 Second Edition 2013-<br>07-01. Intravascular catheters --<br>Sterile, single-use intravascular<br>catheters -- Part 1: General<br>Requirements |
| Indications for Use | | The XableCath catheter is intended to be used to facilitate access to discrete regions of the peripheral vasculature in conjunction with steerable guidewires. This device may be used to facilitate placement and exchange of guidewires and other interventional devices. | The XableCath catheter is intended to be used to facilitate access to discrete regions of the peripheral vasculature in conjunction with steerable guidewires. This device may be used to facilitate placement and exchange of guidewires and other interventional devices. |
| Performance Data [21 CFR 807.92(b)] | | | |
| Summary of non-clinical tests conducted for determination of substantial equivalence<br>[21 CFR 807.92(b)(1)] | | | |
| As required by the risk analysis, the following verification tests have been conducted on the XableCath Support Catheter (please refer to Section 17 for additional information): | | | |
| Direct product bench <i>in-vitro</i> comparison testing has shown the subject & predicate products to be equivalent, via assessment within the following areas: | | | |
| Visual & technical inspections Luer syringe compatibility Sheath compatibility Guidewire retraction/reinsertion Relative radiopacity comparison Torque transmission & capability Simulated use (iliac model) | | Dimensional assessment & comparisons Guidewire compatibility Leak Testing Catheter kink-resistance Tensile testing (proximal & distal) Corrosion testing General packaging, shelf-life/expiry | |
| Full panel biocompatibility was successfully performed in accord with product classification, under GLP rigors, demonstrating that all utilized materials and methods of construction/processing passed biocompatibility rigors. Conducted test included: | | | |
| Cytotoxicity Irritation/Intracutaneous Hemolysis Complement Activation Platelet and Leukocyte Count | Sensitization Reactivity Systemic…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.