K180967 · Lazon Medical Laser Co., Ltd. · GEX · Jul 11, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180967
Device Name
Dental Diode Laser System
Applicant
Lazon Medical Laser Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 11, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SOLASE dental diode laser is a surgical and therapeutic device produced by Lazon Medical Laser Co., Ltd., designed for dental soft tissue indications, laser periodontal procedures, as well as teeth whitening and pain relief. It cannot be used for oral hard tissue.
Device Story
Dental diode laser system (models SOLASE-808, SOLASE-976) uses laser diodes to generate infrared light (808nm or 976nm) delivered via optical fiber to oral soft tissues. Device operates in continuous or pulse modes for surgical (ablation, incision, coagulation) and therapeutic (whitening, pain relief) applications. Class 4 laser requires operation by trained/qualified personnel in dental clinical settings. Physician/dentist uses device to perform soft tissue procedures or provide therapeutic heating; output affects tissue via thermal energy. Benefits include precise soft tissue management, hemostasis, and temporary pain relief. Device is not for hard tissue use.
Clinical Evidence
No clinical data; bench testing only. Compliance verified via electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), laser safety (IEC 60601-2-22, IEC 60825-1), and biocompatibility (ISO 10993-5, ISO 10993-10). Performance testing confirmed peak power output meets specifications.
Indicated for incision, excision, vaporization, ablation, and coagulation of oral soft tissues (e.g., fibroma removal, frenectomy, frenotomy, gingival troughing, gingivectomy, gingivoplasty, hemostasis, implant recovery, abscess drainage, operculectomy, pulpotomy, soft tissue crown lengthening, treatment of herpetic/aphthous ulcers); laser periodontal procedures (removal of diseased/inflamed/necrosed tissue, sulcular debridement); light activation for teeth whitening; and topical heating for temporary relief of minor muscle/joint pain, stiffness, arthritis pain, muscle spasm, sprains, and muscular back pain, and temporary increase in local blood circulation/muscle relaxation.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 11, 2018
Lazon Medical Laser Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 Cn
Re: K180967
Trade/Device Name: Dental Diode Laser System, Models: Solace-808 and Solace-976 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 3, 2018 Received: April 13, 2018
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Binita S. Ashar -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180967
Device Name Dental Diode Laser System Models: SOLASE-808 and SOLASE-976
#### Indications for Use (Describe)
The indications for use of the subject device are given below.
1. Dental soft tissue indications
Incision, excision, vaporization, ablation of oral soft tissues including marginal and inter-dental gingival and epithelial lining of free gingiva and the following specific indications:
- fiborma removal
- frenectomy
- frenotomy
- gingival troughing for crown impressions
- gingivectomy
- gingivoplasty
- hemostasis and coagulation
- implant recovery
- incision and drainage of abscess
- operculectomy
- pulpotomy
- soft tissue crown lengthening
- treatment of herpetic and aphthous ulcers of the oral mucosa.
- 2. Laser periodontal procedures
- laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket
- sulcular debridement (removal of diseased, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival bleeding index, probe depth, attachment loss and tooth mobility.)
- 3. Whitening
- light activation for bleaching materials for teeth whitening
- laser-assisted whitening/bleach of teeth
- 4. Pain Relief
- topical heating for the purpose of elevating tissue temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm, minor sprains and minor muscular back pain; the temporary increase in local blood circulation; the temporary relaxation of muscle
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K180967
- 1. Date of Preparation: 7/10/2018
- 2. Sponsor Identification
## Lazon Medical Laser Co., Ltd. No.123, Hezuo Street, Dadong District, Shenyang, 110179, Liaoning, China.
Establishment Registration Number: Not yet registered
Contact Person: Gen Cheng Position: General Manager Tel: +86-24-88327588 Email: gen,cheng@lazonlaser.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Mr. Chengyu Wang (Alternative Contact Person)
Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
- Identification of Proposed Device 4.
Device Name: Dental Diode Laser System Model: SOLASE-808 and SOLASE-976 Common Name: Diode Laser
{5}------------------------------------------------
#### Regulatory Information
Classification Name: Powered Laser Surgical Instrument Classification: II Product Code: GEX Regulation Number: 21 CFR 878.4810 Review Panel: General and Plastic Surgery
Indications for Use:
The indications for use of the subject device are given below.
- 1. Dental soft tissue indications
Incision, excision, vaporization, ablation and coagulation of oral soft tissues including marginal and inter-dental gingival and epithelial lining of free gingiva and the following specific indications:
- fiborma removal
- frenectomy
- frenotomy
- gingival troughing for crown impressions
- gingivectomy
- gingivoplasty
- hemostasis and coagulation
- implant recovery
- incision and drainage of abscess
- operculectomy
- pulpotomy
- soft tissue crown lengthening
- treatment of herpetic and aphthous ulcers of the oral mucosa.
#### 2. Laser periodontal procedures
- laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket
- sulcular debridement (removal of diseased, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival bleeding index, probe depth, attachment loss and tooth mobility.)
#### 3. Whitening
- light activation for bleaching materials for teeth whitening
- laser-assisted whitening/bleach of teeth
- 4. Pain Relief
- topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle
{6}------------------------------------------------
and joint pain and stiffness, minor arthritis pain, or muscle spasm, minor sprains and strains, and minor muscular back pain; the temporary increase in local blood circulation; the temporary relaxation of muscle.
Device Description
The SOLASE dental diode laser is a surgical and therapeutic device produced by Lazon Medical Laser Co., Ltd., designed for dental soft tissue indications, laser periodontal procedures, as well as teeth whitening and pain relief. It cannot be used for oral hard tissue.
The SOLASE laser uses a laser diode as the beam source to radiate invisible infrared light, which is delivered to the operating area by optical fiber. It can generate a short laser pulse up to 10us of interval. The SOLASE laser provides two different models with two different wavelengths - 808nm and 976mm, to meet customers' various requirements.
The SOLASE laser is a Class 4 laser product which may cause injuries in improper handling. Therefore, it MUST be operated only by trained and qualified personnel.
- ડ. Identification of Predicate Device and Reference Device
Predicate Device 510(k) Number: K121286 Product Name: Epic 10 Diode Laser System Model: Epic 10 Manufacturer: Biolase Technology, Inc.
Reference Device 510(k) Number: K103667 Product Name: Sapphire ST Portable Laser Manufacturer: Denmat Holdings LLC.
{7}------------------------------------------------
#### 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- A AAMI/ANSI ES 60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
- A IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility- Requirements and tests
- A IEC 60601-2-22:2012, Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment.
- > IEC 60825-1:2014, Safety of laser products - Part 1: Equipment classification and requirements.
- > ISO 10993-5:2009 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin A sensitization
- A Performance data showing that the difference between the set average peak power and the measured average peak power meets performance criterion for the range of laser pulse specifications.
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
{8}------------------------------------------------
#### 8. Substantially Equivalent (SE) Comparison
| | Proposed Device | Predicate Device | Reference Device |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ITEM | | K121286 | K103667 |
| Product Code | GEX | GEX | GEX |
| Regulation<br>Number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 |
| Class | II | II | II |
| Indications for<br>Use | 1. Dental soft tissue<br>indications<br>Incision, excision,<br>vaporization, ablation<br>and coagulation of oral<br>soft tissues including<br>marginal and inter-dental<br>gingival and epithelial<br>lining of free gingiva<br>and the following<br>specific indications:<br>- fibroma removal<br>- frenectomy<br>- frenotomy<br>- gingival troughing for<br>crown impressions<br>- gingivectomy<br>- gingivoplasty<br>- hemostasis and<br>coagulation<br>- implant recovery<br>- incision and drainage<br>of abscess<br>- operculectomy<br>- pulpotomy-soft tissue<br>crown<br>lengthening-treatme<br>nt of herpetic and<br>aphthous ulcers of | 1. Dental Soft Tissue<br>Indications<br>Incision, excision,<br>vaporization, ablation and<br>coagulation of oral soft<br>tissues including marginal<br>and inter-dental gingival and<br>epithelial lining of free<br>gingiva and the following<br>specific indications:<br>* Excisional and incisional<br>biopsies<br>* Exposure of unerupted<br>teeth<br>* Fibroma removal<br>* Frenectomy<br>* Frenotomy<br>* Gingival troughing for<br>crown impressions<br>* Gingivectomy<br>* Gingivoplasty<br>* Gingival incision and<br>excision<br>* Hemostasis and coagulation<br>* Implant recovery<br>* Incision and drainage of | The Sapphire ST Portable<br>Diode Laser is intended for<br>use in dental intraoral soft<br>tissue general, oral<br>maxilla-facial and cosmetic<br>surgery. It is intended for<br>ablating, incising, excising,<br>vaporizing and coagulation of<br>soft tissues using a fiber optic<br>delivery system. Indications<br>include excision and incision<br>biopsies;, hemostatic<br>assistance; treatment of<br>aphthous ulcers; frenectomy;<br>frenotomy; gingival incision<br>and excision; gingivectomy;<br>gingivoplasty; incising and<br>draining abscesses;<br>operculectomy; oral<br>papillectomy; removal of<br>fibromias; Soft tissue crown<br>lengthening; sulcular<br>debridement (removal of<br>diseased or inflamed soft<br>tissue in the periodontal<br>pocket); tissue retraction |
| | | | |
| the oral mucosa. | abscess | for impression; | |
| | * Leukoplakia | vestibuloplasty, and light | |
| | * Operculectomy | activation of bleaching | |
| | * Oral papillectomnies | materials for teeth whitening, | |
| | * Pulpotomy | laser-assisted bleaching | |
| | * Pulpotomy as an adjunct to | /whitening of teeth. | |
| | root canal therapy | | |
| | * Reduction of gingival | | |
| | hypertrophy | | |
| | * Soft tissue crown | | |
| | lengthening | | |
| | * Treatment of canker sores, | | |
| | herpetic and aphthous ulcers | | |
| | of the oral mucosa. | | |
| | * Vestibuloplasty | | |
| | * Tissue retraction for | | |
| | impression | | |
| 2. Laser periodontal<br>procedures | 2. Laser Periodontal<br>Procedures | | |
| - laser removal of<br>diseased, infected,<br>inflamed and necrosed<br>soft tissue within the<br>periodontal pocket<br>- sulcular debridement<br>(removal of diseased,<br>infected, inflamed and<br>necrosed soft tissue in<br>the periodontal pocket to<br>improve clinical indices<br>including gingival index,<br>gingival bleeding index,<br>probe depth, attachment<br>loss and tooth mobility.) | * Laser soft tissue curettage<br>* Laser removal of diseased,<br>infected, inflamed and<br>necrosed soft tissue within<br>the periodontal pocket<br>* Sulcular debridement<br>(removal of diseased,<br>infected, inflamed and<br>necrosed soft tissue in the<br>periodontal pocket to<br>improve clinical indices<br>including gingival index,<br>gingival bleeding index,<br>probe depth, attachment loss<br>and tooth mobility.) | | |
| | 3. Whitening<br>Light activation for<br>-<br>bleaching materials for<br>teeth whitening<br>*<br>Laser-assisted<br>whitening/bleaching<br>of<br>teeth<br>4. Pain Relief<br>- Topical heating for the<br>purpose of elevating<br>tissue temperature for a<br>temporary<br>relief<br>of<br>minor muscle and joint<br>pain and stiffness, minor<br>arthritis pain, or muscle<br>spasm, minor sprains<br>and strains, and minor<br>muscular back pain; the<br>temporary increase in<br>local blood circulation;<br>the temporary relaxation<br>of muscle. | 3. Whitening<br>* Light activation for<br>bleaching materials for teeth<br>whitening<br>* Laser-assisted<br>whitening/bleaching of teeth<br>4. Pain Relief<br>* Topical heating for the<br>purpose of elevating tissue<br>temperature for a temporary<br>relief of minor muscle and<br>joint pain and stiffness, minor<br>arthritis pain, or muscle<br>spasm, minor sprains and<br>strains, and minor muscular<br>back pain; the temporary<br>increase in local blood<br>circulation; the temporary<br>relaxation of muscle. | |
| Model | SOLASE-808 and<br>SOLASE-976…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.