K180960 · S.B.M. Sas Sciences For Bio Materials · MAI · Jul 25, 2018 · Orthopedic
Device Facts
Record ID
K180960
Device Name
FIXIT / ComposiTCP Threaded Anchor System
Applicant
S.B.M. Sas Sciences For Bio Materials
Product Code
MAI · Orthopedic
Decision Date
Jul 25, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The FIXIT®/ ComposiTCP™ threaded anchor system is intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, and elbow, but not limited to, the following procedures: Shoulder: Rotator Cuff Repair, Bankart Repair; SLAP Lesion Repair; Biceps Tenodesis; Acromio-clavicular Separation Repair; Deltoid Repair; Capsule Shift or Capsulolabral Reconstruction; Ankle/Foot: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair/Tendon Repair, Bunionectomy Knee: Anterior Cruciate Ligament Repair; Medial Collateral Ligament Repair; Lateral Collateral Ligament Repair; Patellar Tendon Repair ; Posterior Oblique Ligament Repair; Iliotibial Band Tenodesis ; Wrist/Hand: Scapholunate Ligament Reconstruction; Ulnar or Radial Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction. Biceps Tendon Reattachment; Ulnar or Radial Collateral Ligament Elbow: Reconstruction; Tennis Elbow Repair and Lateral Epicondylitis Repair.
Device Story
FIXIT®/ComposiTCP™ Threaded Anchor System is a bioabsorbable bone anchor used for soft tissue-to-bone fixation. The device consists of a composite anchor (30% β-TCP/70% PLDL) pre-loaded on a disposable screwdriver with two sutures. It is used by surgeons in clinical settings for various orthopedic procedures, including rotator cuff, ligament, and tendon repairs. The anchor is implanted into bone to secure sutures, which then hold soft tissue in place. The device is supplied sterile and is intended to be absorbed by the body over time. It functions as a mechanical fastener to facilitate tissue healing.
Clinical Evidence
No clinical data included. Substantial equivalence is supported by non-clinical performance testing, including mechanical (pull-out strength) and biological (bacterial endotoxin) testing.
Indicated for soft tissue-to-bone fixation in shoulder, foot/ankle, knee, hand/wrist, and elbow orthopedic procedures, including ligament and tendon repairs/reconstructions.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
White/Green BroadBand® Single Flat Suture (K160854)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure with three faces in profile, representing the department's focus on people's health and well-being. The FDA acronym is in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 25, 2018
S.B.M. SAS Sciences for Bio Materials Anne Cospin-Latapie Quality/Regulatory Affairs Manager ZI du Monge Lourdes, France 65100
Re: K180960
Trade/Device Name: FIXIT®/ComposiTCP™ Threaded Anchor System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI Dated: June 20, 2018 Received: June 25, 2018
Dear Ms. Cospin-Latapie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820):
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Laurence D. Coyne -S
For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Science & Bio Materials (SBM). The logo consists of the letters S, B, and M, with the B in a darker color. Below the letters is the text "Science & Bio Materials". To the right of the letters are four right-pointing triangles, with the second triangle from the left having a dotted pattern. The text "K180960 Page 1 of 1" is located in the upper right corner of the image.
# INDICATIONS FOR USE
#### 510(k) Number (if known): K180960
# Device Name: FIXIT® / ComposiTCP™ Threaded Anchor System Indications for Use:
The FIXIT®/ ComposiTCP™ threaded anchor system is intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, and elbow, but not limited to, the following procedures:
Shoulder: Rotator Cuff Repair, Bankart Repair; SLAP Lesion Repair; Biceps Tenodesis; Acromio-clavicular Separation Repair; Deltoid Repair; Capsule Shift or Capsulolabral Reconstruction;
Ankle/Foot: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair/Tendon Repair, Bunionectomy
Knee: Anterior Cruciate Ligament Repair; Medial Collateral Ligament Repair; Lateral Collateral Ligament Repair; Patellar Tendon Repair ; Posterior Oblique Ligament Repair; Iliotibial Band Tenodesis ;
Wrist/Hand: Scapholunate Ligament Reconstruction; Ulnar or Radial Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction.
Biceps Tendon Reattachment; Ulnar or Radial Collateral Ligament Elbow: Reconstruction; Tennis Elbow Repair and Lateral Epicondylitis Repair.
Prescription Use Over-The-Counter Use __ > - Prescription Use - - - - - - - - - - - - - - - - AND/OR - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with the letters S, B, and M in different shades of gray and blue. The letter B is blue, while the letters S and M are gray. Below the letters, the words "Science & Bio Materials" are written in a smaller font. To the right of the letters, there are three right-pointing triangles, with the middle triangle having a blue dotted pattern.
# 510(k) SUMMARY
### 1. SUBMITTER
| S.B.M. SAS |
|-------------------------------------|
| SCIENCE & BIO MATERIALS |
| ZI du Monge |
| F 65100 LOURDES - FRANCE |
| Registration Number: 3004549189 |
| Phone: +33 (0)5 62 42 21 01 |
| Fax: +33 (0)5 62 42 21 00 |
| Contact Person: Anne COSPIN-LATAPIE |
| e-mail : anne.cospin@sbm-fr.com |
| Date prepared: June 14, 2018 |
### 2. DEVICE
| Name of Device | FIXIT®/ ComposiTCP™ Threaded Anchor System |
|----------------------|------------------------------------------------|
| Common or Usual Name | Suture Anchor |
| Classification Name | Fastener, fixation, biodegradable, soft tissue |
| Regulatory Class | II |
| Product Code | MAI |
#### 3. PREDICATE DEVICE
FIXIT Threaded Anchor System (K170868)
Referenced devices:
K070673 Force Fiber® Black Co-braid Polyethylene non-absorbable Suture
K063778 Force Fiber® Polyethylene non-absorbable Suture
K160854 Blue BroadBand® Single Flat Suture
K160854 White/Green BroadBand® Single Flat Suture
#### 4. DEVICE DESCRIPTION
FIXIT® / ComposiTCP™ Threaded Anchor System is composed of a bioabsorbable composite anchor (30% β-TCP/ 70%PLDL) pre-loaded on a disposable screwdriver and 2 sutures. Different variations of the product are available depending on the type of suture combination (2 flat sutures, or 1 round + 1 flat)
For each configuration, the implant is supplied sterile, ready to use.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Science & Bio Materials (SBM). The logo consists of the letters S, B, and M in a light gray color, with the words "Science & Bio Materials" written in a smaller font size underneath the letters. To the right of the letters, there are three gray triangles pointing to the right, with the middle triangle having a blue gradient pattern.
# 5. INDICATIONS FOR USE
The FIXIT®/ ComposiTCP™ threaded anchor system is intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, and elbow, but not limited to, the following procedures:
Shoulder: Rotator Cuff Repair, Bankart Repair; SLAP Lesion Repair; Biceps Tenodesis; Acromio-clavicular Separation Repair; Deltoid Repair; Capsule Shift or Capsulolabral Reconstruction;
Ankle/Foot: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair/Tendon Repair, Bunionectomy
Knee: Anterior Cruciate Ligament Repair; Medial Collateral Ligament Repair; Lateral Collateral Ligament Repair; Patellar Tendon Repair; Posterior Oblique Ligament Repair; Iliotibial Band Tenodesis;
Wrist/Hand: Scapholunate Ligament Reconstruction; Ulnar or Radial Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction.
Elbow: Biceps Tendon Reattachment; Ulnar or Radial Collateral Ligament Reconstruction; Tennis Elbow Repair and Lateral Epicondylitis Repair.
# 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
FIXIT®/ ComposiTCP™ Threaded Anchor System is compared to FIXIT Threaded Anchor System (K170868) manufactured by S.B.M SAS Science and Bio Materials.
The applicant device has the same intended use as the predicate device.
The technological characteristics of this product are believed to be substantially equivalent as those for the predicate device. This device and its predicates use similar performance characteristics, manufacturing materials, and design.
| | FIXIT®/ ComposiTCP™<br>Threaded Anchor<br>System | FIXIT® Threaded Anchor<br>System |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| | Present submission | K170868 (Predicate<br>device) |
| 21CFR | 888.3030 Single/multiple component metallic bone<br>fixation appliances and accessories | |
| Product's code | | MAI |
| Device | Threaded suture anchors | |
| Indications for use | | All these devices include use in soft tissue reattachment<br>procedures in the shoulder, wrist/hand, ankle/foot, elbow, |
| statement | and knee. Specific indications are as follows:<br>Shoulder: Bankart Repair; SLAP Repair; Acromio-<br>clavicular Separation; Rotator Cuff Repair; Capsule Repair<br>or Capsulolabral Reconstruction; Biceps Tenodesis; and<br>Deltoid Repair<br><br>Wrist/Hand: Scapholunate Ligament Reconstruction and<br>Ulnar/Radial Collateral Ligament Reconstruction<br>Ankle/Foot: Lateral Stabilization; Medial Stabilization;<br>Achilles Tendon Repair/Reconstruction; Hallux Valgus<br>Reconstruction; and Mid and Forefoot Reconstruction<br><br>Elbow: Ulnar or Radial Collateral Ligament<br>Reconstruction; Biceps Tendon Reconstruction; and<br>Lateral Epicondylitis Repair<br><br>Knee: Extra-capsular Repair; Medial Collateral Ligament<br>Repair; Lateral Collateral Ligament Repair; Posterior<br>Oblique Ligament Repair; Joint Capsule Closure; Iliotibial<br>Band Tenodesis; Patellar Realignment and Repair;<br>Patellar Ligament/Tendon Repair; and Vastus Medialis<br>Obliquus (VMO) Muscle Advancement | |
| Anchor | β-TCP (30%) / Poly Lactid<br>Copolymer (70%) | β-TCP (30%) / Poly Lactid<br>Copolymer (70%) |
| Resorbable | YES | YES |
| Suture | Pre-loaded on a disposable<br>screwdriver with 2 round<br>sutures | Pre-loaded on a disposable<br>screwdriver with 2 flat<br>sutures, or 1 round + 1 flat |
| Model size | Ø 4.5mm x14.5 mm<br>Ø 5.5mm x17.6 mm<br>Ø 6.5mm x17.6 mm | Ø 4.5mm x14.5 mm<br>Ø 5.5mm x17.6 mm<br>Ø 6.5mm x17.6 mm |
| Mechanical<br>Properties | | |
| -<br>Pull-out<br>strength at 12<br>weeks (≥ 150<br>N) | $229 \pm 19$ N<br>(Cf. RD 1124 Report) | $220 \pm 12$ N<br>(Cf. RD 1204 Report) |
| Sterilization<br>method | Gamma rays 25 kGy<br>minimum | Gamma rays 25 kGy<br>minimum |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Science & Bio Materials, which is represented by the letters S, B, and M. The letter B is blue, while the other letters are gray. To the right of the letters are three gray triangles pointing to the right, with the middle triangle having a blue gradient.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Science & Bio Materials (SBM). The logo consists of the letters S, B, and M, with the B in blue and the S and M in gray. Underneath the letters is the text "Science & Bio Materials". To the right of the letters are three gray triangles pointing to the right, with the middle triangle having a blue gradient.
### 4. PERFORMANCE DATA
#### Non-clinical performance testing
Non-clinical testing including chemical, biological and mechanical performances was performed to determine substantial equivalence. The results indicated that the devices were functional within their intended use and equivalent to the predicate devices.
Bacterial endotoxin testing has been completed and results have demonstrated that the proposed devices meet the endotoxin limits.
#### Clinical performance testing:
Clinical performance data was not included.
### 5. CONCLUSIONS
The FIXIT®/ ComposiTCP™ Threaded Anchor System is substantially equivalent to its predicate device FIXIT® Threaded Anchor System (K170868). Verification and validation tests demonstrate that the FIXIT® Threaded Anchor System is as safe, as effective, and performs as safely and effectively as its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.