← Product Code [EKX](/productcode/EKX) · K180953

# AnyCheck IMT-100 (K180953)

_Dms Co., Ltd. · EKX · Feb 20, 2019 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K180953

## Device Facts

- **Applicant:** Dms Co., Ltd.
- **Product Code:** [EKX](/productcode/EKX.md)
- **Decision Date:** Feb 20, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4200
- **Device Class:** Class 1
- **Review Panel:** Dental

## Indications for Use

The AnyCheck IMT-100 is indicated for use in measuring the stability of implants in the oral cavity and maxillofacial region.

## Device Story

AnyCheck IMT-100 is a portable, handheld dental instrument for non-invasive assessment of dental implant fixation stability at the alveolar bone. The device operates by generating mechanical vibrations via a tapping rod that contacts the implant abutment. The system measures the resulting resonance frequency to calculate an implant stability test (iST) value, which is displayed numerically on an OLED screen; higher values indicate greater stability. Used by dental professionals in clinical settings, the device provides objective data to assist in evaluating implant integration. It requires no complex setup menus. The device benefits patients by providing a non-invasive method to monitor implant health and stability, aiding clinical decision-making regarding loading protocols or implant success.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by non-clinical performance testing, including measurement accuracy, output scale, reproducibility, electrical safety (IEC 60601-1), and biocompatibility (ISO 10993-5, ISO 10993-10, ISO 10993-12).

## Technological Characteristics

Handheld, battery-powered device. Materials: ABS, PEEK, Titanium Alloy, Aluminum, Silicon, polycarbonate resin, Acetal, Trimrite, Stainless Steel. Sensing principle: mechanical tapping rod generating vibrations to determine resonance frequency. Output: OLED display. Connectivity: standalone. Biocompatibility: ISO 10993 compliant. Electrical safety: IEC 60601-1 compliant.

## Regulatory Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

## Predicate Devices

- Osstell® ISQ Implant Stability Meter ([K082523](/device/K082523.md))
- Osstell® IDx ([K142358](/device/K142358.md))
- Tellos ISQ Buddy ([K143445](/device/K143445.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

February 20, 2019

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out as "U.S. Food & Drug Administration" below it.

DMS Co., Ltd. % Soojung Moon Consultant Allura Medical Solution, Inc. 8141 Lampson Ave. #4 Garden Grove, California 92841

Re: K180953

Trade/Device Name: AnyCheck IMT-100 Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I Product Code: EKX Dated: January 15, 2019 Received: January 18, 2019

Dear Soojung Moon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Mary S. Runner -S3 -s3 Date: 2019.02.20 18:26:54 -05'00

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K180953

Device Name AnyCheck IMT-100

Indications for Use (Describe)

The AnyCheck IMT-100 is indicated for use in measuring the stability of implants in the oral cavity and maxillofacial region.

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

-

## 510(k) Summary K180953

| Submitter                                                                           | DMS Co., Ltd.<br>#201, #202 & #204, Oriental Medical Industry Development Center,<br>Sangji University, 83, Sangjidae-gil, Wonju-si, Gangwon-do, 26339<br>Republic of Korea<br>Phone. +82-33-742-3210<br>Fax. +82-33-742-3219                                                                                                                                                                                                                                                                                                                                                                         |
|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person                                                                      | HWI JOON PARK<br>8141 Lampson Ave. #4, Garden Grove, CA 92841, USA<br>Phone: +1-972-800-0044<br>Fax: +1-210-899-0079                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Submission Date                                                                     | Jan 18, 2019                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Trade / Proprietary name                                                            | AnyCheck IMT-100                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Common / Usual Name                                                                 | Dental Implant Mobility Measuring Instrument                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Classification Name<br>Classification Code<br>Regulatory Class<br>Regulation Number | Handpiece, Direct Drive, AC-Power<br>EKX<br>Class I<br>872.4220                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Predicate Device                                                                    | Osstell® ISQ Implant Stability Meter (Osstell AB)<br>(K082523 - Primary predicate)<br>Osstell® IDx (Osstell AB)<br>(K142358 - Reference predicate)<br>Tellos ISQ Buddy (Tellos Medical AB)<br>(K143445 - Reference predicate)                                                                                                                                                                                                                                                                                                                                                                         |
| Description of Device                                                               | The AnyCheck IMT-100 Dental Implant Mobility Measuring Instrument<br>is a system designed to measure dental implant mobility (stability) in the<br>oral cavity and maxillofacial region. The AnyCheck IMT-100 is a<br>portable, handheld instrument to observe non-invasively the fixation<br>stability of an implant at alveolar bone. The implant stability test (iST) is<br>performed using vibration generated by mechanical contacting of a<br>tapping rod. The AnyCheck IMT-100 numerically indicates a scale result<br>of iST. The higher number means better stability of the target implant. |
| Indication for Use                                                                  | The AnyCheck IMT-100 is indicated for use in measuring the stability of<br>implants in the oral cavity and maxillofacial region.                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

{4}------------------------------------------------

| Comparison of<br>Technological<br>Characteristics with<br>the Predicate Devices | Comparison of Indication for Use<br>The AnyCheck IMT-100 and the predicate devices have identical<br>indications for use statements.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|---------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                                 | Materials<br>ABS, PEEK, Titanium Alloy, Aluminum, Silicon, polycarbonate resin,<br>Accetal, Trimrite, Stainless Steel.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                                 | Design<br>The subject device has no "device setup menu", which is similar to Tellos ISQ<br>Buddy (K143445), while other predicates need various parameters to set up<br>before use. The weight and power scale of the subject device is within the<br>range of all the predicate devices (3-12VA).                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                                                                 | Technological Characteristics<br>The result of the measurement of the subject device and all other predicate<br>devices is calculated from the resonance frequency. For measuring, predicate<br>devices (K082523 and K142358) can be attached either to the implant or<br>abutment, while the subject device should be attached only to the abutment.<br>But the performance data shows the output values between the subject device<br>and the predicate devices are similar and the difference is negligible.<br>Also, OLED display does not raise any issues or concerns of safety and<br>effectiveness for the use of the subject device, compared to the predicate<br>devices which use LED or LCD display.                                          |
|                                                                                 | Clinical Tests<br>This submission does not include clinical performance data, similar to the<br>predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                                                                 | Non-clinical Tests<br>Non-clinical testing was performed and comparative performance testing<br>was conducted in order to validate the product against the company's<br>specified design requirement. Measurement accuracy, output scale,<br>reproducibility and electrical safety tests according to IEC60601-1 were<br>performed.<br>Also, cytotoxicity, oral mucosa irritation and skin sensitization test were<br>conducted with respect to biocompatibility according to ISO 10993-5, ISO<br>10993-10 and ISO 10993-12. The laboratory certified that the insolubility<br>is in compliance with the requirements of the standard. There is no<br>evidence that effects hazardous to the patient will arise by leachable<br>ingredients/contaminants. |
| Conclusion                                                                      | Based upon the similarities of indications for use, and similarities<br>in technological characteristics, including materials and design,<br>together with the results of non-clinical performance testing, we<br>find that the AnyCheck IMT-100 Dental Implant Mobility<br>Measuring Instrument is substantially equivalent to the predicate<br>devices.                                                                                                                                                                                                                                                                                                                                                                                                 |

---

**Source:** [https://fda-staging.innolitics.com/device/K180953](https://fda-staging.innolitics.com/device/K180953)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).