The Sonicision™ reusable sterilization tray is intended to provide storage for the Sonicision™ reusable generator and Sonicision™ reusable battery insertion guides during sterilization, storage, and transportation within the hospital environment. The Sonicision™ reusable sterilization tray is only intended to maintain sterility of the enclosed components during autoclave sterilization when used in conjunction with an FDA-cleared sterilization wrap or rigid container (validated with Aesculap™ rigid container model JN741/JK789). The sterilization tray has only been evaluated for a non-stacked configuration.
Device Story
Sonicision Reusable Sterilization Tray serves as storage accessory for Sonicision System components (reusable generator, battery insertion guides). Device facilitates steam sterilization (pre-vacuum/gravity) within hospital environments; requires external FDA-cleared sterilization wrap or rigid container to maintain sterility. Tray features perforated design for steam penetration; constructed from polyphenylsulfone base and stainless steel lid. Used by hospital staff to organize and protect components during reprocessing cycles. Benefits include standardized, validated storage and transport solution ensuring component integrity through repeated autoclave cycles.
Clinical Evidence
No clinical data. Evidence consists of bench testing: cleaning and sterilization validation per ANSI/AAMI/ISO 17665-1, AAMI TIR30:2011, and AAMI TIR12:2010; mechanical/reliability testing for repeated reprocessing; and human factors validation per IEC 62366-1.
Technological Characteristics
Materials: Polyphenylsulfone (base), stainless steel (lid). Design: Perforated tray/lid for steam sterilization. Dimensions: 8.5 x 11.0 x 3.0 inches. Sterilization: Steam (pre-vacuum and gravity). Standards: ANSI/AAMI/ISO 17665-1, AAMI TIR30:2011, AAMI TIR12:2010, IEC 62366-1.
Indications for Use
Indicated for storage of Sonicision reusable generator and battery insertion guides during sterilization, storage, and transport in hospital settings. Intended for use with FDA-cleared sterilization wraps or specific rigid containers (Aesculap JN741/JK789) in non-stacked autoclave configurations.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Covidien Candice Bautista-Biddle Associate Regulatory Affairs Specialist 5920 Longbow Drive Boulder, Colorado 80301
Re: K180915
Trade/Device Name: Sonicision Reusable Sterilization Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: June 26, 2018 Received: June 27, 2018
Dear Candice Bautista-Biddle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Clarence W. Murray lii III -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180915
Device Name Sonicision™ Reusable Sterilization Tray
#### Indications for Use (Describe)
The Sonicision™ reusable sterilization tray is intended to provide storage for the Sonicision™ reusable generator and Sonicision™ reusable battery insertion guides during sterilization, storage, and transportation within the hospital environment. The Sonicision™ reusable sterilization tray is only intended to maintain sterility of the enclosed components during autoclave sterilization when used in conjunction with an FDA-cleared sterilization wrap or rigid container (validated with Aesculap™ rigid container model JN741/JK789). The sterilization tray has only been evaluated for a non-stacked configuration.
The validated sterilization cycle parameters are as follows:
### Steam Sterilization Cycles
| Cycle Type | Exposure<br>Temperature | Exposure Time<br>(minutes) | Minimum<br>Dry Time1<br>(minutes) | Compatible Sterile Barriers |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|----------------------------|-----------------------------------|------------------------------|
| Pre-vacuum | 132 °C (270 °F) | 4 | 40 | Wrap(s) and rigid container2 |
| Pre-vacuum | 135 °C (275 °F) | 3 | 40 | Wrap(s) and rigid container2 |
| Gravity | 132 °C (270 °F) | 15 | 40 | Wrap(s) |
| Gravity | 135 °C (275 °F) | 10 | 30 | Wrap(s) |
| 1Due to variations in autoclave sterilization equipment, actual dry times may vary from the minimum<br>dry times validated for this product.<br>2Validated with Aesculap™* rigid container model JN741/JK789. | | | | |
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the word "COVIDIEN" in a bold, dark blue font. To the left of the word is a square logo. The logo is a dark blue square with a lighter blue vertical line and a light blue horizontal rectangle in the center.
# 510(k) Summary: K180915
Date summary prepared: 7/16/2018
#### 510(k) Submitter/Holder
Covidien 5920 Longbow Drive Boulder, CO 80301
#### Contact
Candice Bautista-Biddle Associate Regulatory Affairs Specialist Telephone: 303-530-6448 Fax: 303-530-6313 Email: candice.bautista-biddle@medtronic.com
#### Name of Device
| Trade Name: | Sonicision™ Reusable Sterilization Tray |
|----------------------|--------------------------------------------------------------------------------------------------------|
| Common Name: | Sterilization Tray |
| Catalog Numbers: | SCSTA |
| Classification Name: | Sterilization Wrap Containers, Trays, Cassettes and Other Accessories (21 CFR 880.6850, Class II, KCT) |
#### Predicate Device
| Trade Name: | Signia™ Sterilization Tray |
|----------------------|--------------------------------------------------------------------------------------------------------|
| Catalog Numbers: | SIGTRAY |
| Common Name: | Sterilization Tray |
| Classification Name: | Sterilization Wrap Containers, Trays, Cassettes and Other Accessories (21 CFR 880.6850, Class II, KCT) |
| 510(k) Number: | K161347 |
| Manufacturer: | Covidien |
#### Device Description
The Sonicision Sterilization Tray is a reusable, optional accessory to the Sonicision System (cleared through K180149). This device is specially made to enclose up to one Sonicision Reusable Generator and up to two Sonicision Reusable Battery Insertion Guides in order to provide storage during sterilization, storage, and transportation within the hospital environment. The tray does not provide a sterile barrier on its own: it can only maintain sterility using an FDA-cleared sterilization wrap or rigid container.
#### Indications for Use
The Sonicision™ reusable sterilization tray is intended to provide storage for the Sonicision™ reusable generator and Sonicision™ reusable battery insertion guides during sterilization, storage, and transportation within the hospital environment. The Sonicision™ reusable sterilization tray is only intended to maintain sterility of the enclosed components during autoclave sterilization when used in conjunction with an FDA-cleared sterilization wrap or rigid container (validated with Aesculan™M* Rigid Container model JN741/JK789). The sterilization tray has only been evaluated for a non-stacked configuration.
The validated sterilization cycle parameters are as follows:
{4}------------------------------------------------
## Steam Sterilization Cycles
| Cycle Type | Exposure<br>Temperature | Exposure Time<br>(minutes) | Minimum Dry<br>Time1 (minutes) | Compatible Sterile<br>Barriers |
|----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|----------------------------|--------------------------------|---------------------------------|
| Pre-vacuum | 132 °C (270 °F) | 4 | 40 | Wrap(s) and rigid<br>container2 |
| Pre-vacuum | 135 °C (275 °F) | 3 | 40 | Wrap(s) and rigid<br>container2 |
| Gravity | 132 °C (270 °F) | 15 | 40 | Wrap(s) |
| Gravity | 135 °C (275 °F) | 10 | 30 | Wrap(s) |
| 1Due to variations in autoclave sterilization equipment, actual dry times may vary from the minimum dry times<br>validated for this product. | | | | |
| 2Validated with Aesculap Rigid Container model JN741/JK789 | | | | |
## Comparison of Technological Characteristics with the Predicate Device
Both the proposed and predicate devices have the same intended use of providing storage for components of their respective systems during sterilization, storage, and transportation within the hospital environment. Additionally, both devices have the same basic design of tray and lid with perforations to allow for steam sterilization of the enclosed devices.
| | Subject Device<br>Sonicision Reusable Sterilization<br>Tray | Predicate Device K161347<br>Signia sterilization tray |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Enclosed devices<br>for sterilization | The tray can contain at a maximum:<br>One (1) Sonicision Reusable Generator,<br>Two (2) Sonicision Reusable Battery<br>Insertion Guides. | The tray can contain at a maximum:<br>One (1) Signia Adapter,<br>One (1) Signia Reusable Insertion<br>Guide and<br>One (1) Signia Manual Retraction Tool. |
| Dimensions | Approx. 8.5 x 11.0 x 3.0 (H) inches | Approx. 10.0 x 21.4 x 3.0 (H) inches |
| Sterilization<br>methods | Steam (Pre-vacuum and Gravity) | Steam (Pre-vacuum) |
| Steam sterilization<br>parameters | <b>132 °C Pre-vacuum Steam Cycle</b><br>Exposure temperature: 270 °F (132 °C)<br>Exposure time: 4 minutes<br>Minimum dry time: 40 minutes<br><br><b>135 °C Pre-vacuum Steam Cycle</b><br>Exposure temperature: 275 °F (135 °C)<br>Exposure time: 3 minutes<br>Minimum dry time: 40 minutes<br><br><b>132 °C Gravity Steam Cycle</b><br>Exposure temperature: 270 °F (132 °C)<br>Exposure time: 15 minutes<br>Minimum dry time: 40 minutes<br><br><b>135 °C Gravity Steam Cycle</b><br>Exposure temperature: 275 °F (135 °C)<br>Exposure time: 10 minutes<br>Minimum dry time: 30 minutes | <b>132 °C Pre-vacuum (Hi Vac) Steam<br/>Cycle</b><br>Exposure temperature: 270 °F (132 °C)<br>Exposure time: 4 minutes<br>Vacuum dry time: 20-40 minutes<br><br><b>134 °C Pre-vacuum (Hi Vac) Steam<br/>Cycle</b><br>Exposure temperature: 273 °F (134 °C)<br>Exposure time: 3 minutes<br>Vacuum dry time: 20-40 minutes<br><br>Not applicable |
{5}------------------------------------------------
| | Subject Device<br>Sonicision Reusable Sterilization<br>Tray | Predicate Device K161347<br>Signia sterilization tray |
|---------------------------|----------------------------------------------------------------|---------------------------------------------------------------|
| Base and Lid<br>Materials | Base: Polyphenylsulfone<br>Lid: Stainless Steel | Base and Lid: Stainless Steel |
| Stacking | Do not stack loaded/wrapped<br>sterilization trays in chamber. | Do not stack cases and trays in the<br>sterilization chamber. |
| Max. Load<br>Capacity | < 1800 grams (3.97 pounds, including<br>tray) | 10 pounds |
#### Performance Data
The following performance data were provided in support of the substantial equivalence determination.
#### Cleaning and Sterilization Testing
Cleaning and sterilization testing was performed in accordance with ANSI/AAMI/ISO 17665-1, AAMI TIR30:2011, AAMI TIR12:2010, and FDA Guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff" published March 17, 2015.
#### Mechanical/Reliability Testing
Mechanical and reliability testing was carried out to verify that the proposed device performs as expected at initial use and after repeated cleaning and sterilization cycles.
#### Human Factors Validation Testing
Human factors testing was performed in accordance with IEC 62366-1:2015 along with FDA Guidance, "Applying Human Factors and Usability Engineering to Medical Devices" published February 3, 2016 to verify that intended users are able to safely and correctly use the device.
#### Conclusion
The performance data support that the Sonicision Reusable Sterilization Tray is as safe, as effective, and performs as well or better than the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.