K180910 · ETHICON, Inc. · FTL · Jul 2, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180910
Device Name
ULTRAPRO Mesh and ULTRAPRO ADVANCED
Applicant
ETHICON, Inc.
Product Code
FTL · General, Plastic Surgery
Decision Date
Jul 2, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ULTRAPRO™ Mesh is indicated for the repair of abdominal wall hernias and abdominal wall deficiencies that require the addition of a reinforcing material to obtain the desired surgical result. ULTRAPRO ADVANCED™ Mesh is indicated for the repair of abdominal wall hernias and abdominal wall deficiencies that require the addition of a reinforcing material to obtain the desired surgical result.
Device Story
Surgical mesh for abdominal wall hernia repair; provides reinforcement to abdominal wall deficiencies. ULTRAPRO Mesh consists of absorbable poliglecaprone-25 and nonabsorbable polypropylene monofilaments; poliglecaprone-25 absorbs over time, leaving polypropylene structure. ULTRAPRO ADVANCED Mesh consists of polypropylene monofilaments and a twist of polypropylene and poliglecaprone-25 monofilaments; includes blue stripes for orientation. Used by surgeons in clinical settings. Provides mechanical support to abdominal wall; benefits patient by reinforcing tissue during hernia repair.
Clinical Evidence
No clinical data. Substantial equivalence is based on identical technological characteristics and materials to predicate devices.
Technological Characteristics
Partially absorbable surgical mesh. Materials: polypropylene monofilament and poliglecaprone-25 (glycolide/epsilon-caprolactone copolymer). ULTRAPRO ADVANCED includes 3.5 mil polypropylene and 5 mil poliglecaprone-25 fibers. Sterile. No changes to materials, construction, or manufacturing processes compared to predicates.
Indications for Use
Indicated for patients requiring surgical repair of abdominal wall hernias and abdominal wall deficiencies needing reinforcing material.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two main elements: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA acronym and the words "U.S. FOOD & DRUG ADMINISTRATION" are in blue.
Ethicon Inc % Ms. Melina Escobar Regulatory Affairs Specialist P.O. Box 151 U.S. Route 22 West Somerville, New Jersey 08876
July 2, 2018
# Re: K180910
Trade/Device Name: ULTRAPRO™ Mesh and ULTRAPRO ADVANCED™ Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTL Dated: April 5, 2018 Received: April 6, 2018
Dear Ms. Escobar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
K180910
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180910
Device Name ULTRAPRO™ Mesh ULTRAPRO ADVANCED™ Macroporous Partially Absorbable Mesh
#### Indications for Use (Describe)
ULTRAPRO™ Mesh is indicated for the repair of abdominal wall hernias and abdominal wall deficiencies that require the addition of a reinforcing material to obtain the desired surgical result.
ULTRAPRO ADVANCED™M Mesh is indicated for the repair of abdominal wall hernias and abdominal wall deficiencies that require the addition of a reinforcing material to obtain the desired surgical result.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Ethicon, a Johnson & Johnson company. The word "ETHICON" is in large, red, sans-serif font. Below it, in a smaller, cursive font, is the text "a Johnson & Johnson company."
# 510(k) Summary
| Submitter: | Ethicon, Inc. a Johnson & Johnson company<br>P.O. Box 151<br>Route 22 West<br>Somerville, NJ 08876-0151 |
|---------------------|----------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Melina Escobar<br>Regulatory Affairs Specialist<br>Phone: 908-218-2583<br>Fax: 908-218-2595<br>Email: mescob14@its.jnj.com |
| Date Prepared: | April 5, 2018 |
| Device Trade Name: | ULTRAPRO™ Mesh<br>ULTRAPRO ADVANCED™ Macroporous Partially Absorbable<br>Mesh |
| Device Common Name: | ULTRAPRO™ Mesh<br>ULTRAPRO ADVANCED™ |
| Class: | Class II |
| Classification: | 21 CFR 878.3300 – Surgical Mesh |
| Product Code: | FTL |
{4}------------------------------------------------
K180910 Page 2 of 4
Image /page/4/Picture/1 description: The image shows the logo for Ethicon, a Johnson & Johnson company. The word "ETHICON" is written in large, red, sans-serif letters. Below the company name, in a smaller, cursive font, is the text "a Johnson & Johnson company". The logo is simple and clean, with a focus on the company name.
## Predicate Device:
| Device | Company | Product Code | 510(k) Number | Predicate for |
|----------------------|---------------|--------------|---------------|---------------------------------------------------------------------------------------------------------------------------|
| ULTRAPRO™<br>Mesh | Ethicon, Inc. | FTL | K033337 | Fundamental Scientific<br>Technology, Design,<br>Intended Use, Materials,<br>Construction, Performance<br>Characteristics |
| ULTRAPRO<br>ADVANCED | Ethicon, Inc. | FTL | K150906 | Fundamental Scientific<br>Technology, Design,<br>Intended Use, Materials,<br>Construction, Performance<br>Characteristics |
# Device Description:
ULTRAPRO™ Mesh is manufactured from approximately equal parts* of absorbable poliglecaprone-25 monofilament fiber and nonabsorbable polypropylene monofilament fiber. The polymer of the undyed and dyed polypropylene fiber (phthalocyanine blue, Color Index No .: 74160) is identical to the material used for dyed/undyed PROLENE™ suture material. Poliglecaprone-25 fiber consists of a copolymer containing glycolide and &-caprolactone; this copolymer is identical to the material used for MONOCRYL™ suture. After absorption of the poliglecaprone-25 component, only the polypropylene mesh remains. The structure and size of this remaining mesh are optimally designed for the physiological stresses of the abdominal wall.
ULTRAPRO ADVANCED™ Macroporous Partially Absorbable Mesh is a sterile, partially absorbable mesh*, designed for the repair of abdominal wall hernias and other abdominal wall deficiencies. The device is manufactured out of dyed (phtalocyanine blue) and undyed nonabsorbable polypropylene monofilaments (3.5 mil PROLENE™) as well as a twist, composed of dyed (phtalocyanine blue) and undyed non-absorbable polypropylene monofilaments (3.5 mil PROLENE™) and undyed absorbable poliglecaprone 25 monofilaments (5 mil MONOCRYL™). Blue stripes provide orientation.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Ethicon, a Johnson & Johnson company. The word "ETHICON" is written in large, red, serif font. Below it, in a smaller, cursive font, is the phrase "a Johnson & Johnson company."
## Indications for Use:
ULTRAPRO™ Mesh is indicated for the repair of abdominal wall hernias and abdominal wall deficiencies that require the addition of a reinforcing material to obtain the desired surgical result.
ULTRAPRO ADVANCED™ Mesh is indicated for the repair of abdominal wall hernias and abdominal wall deficiencies that require the addition of a reinforcing material to obtain the desired surgical result.
# Summary of Technological Characteristics:
ULTRAPRO Mesh is identical to the ULTRAPRO Mesh (K033337) marketed mesh with respect to technological characteristics. There are no material, construction, manufacturing or sterilization process changes to the currently marketed device differs only in the labeling which certain sections of the Instructions for Use of the subject device have been reworded/ reformatted for clarity to address evolving regulatory expectations and bring the contents of the Instructions for Use up to date.
ULTRAPRO ADVANCED is identical to the ULTRAPRO ADVANCED (K150906) marketed mesh with respect to technological characteristics. There are no material, construction, specification, manufacturing or sterilization process changes to the currently marketed device. The device differs only in the labeling which has been revised to reworded/ reformatted most of the Instructions for Use sections for clarity to address evolving regulatory expectations and bring the contents of the Instructions for Use up to date.
## Substantial Equivalence:
ULTRAPRO™ Mesh and ULTRAPRO ADVANCED are identical to the ULTRAPRO™ Mesh (K033337) and ULTRAPRO ADVANCED (K150906) marketed mesh with respect to technological characteristics. There are no material, construction, specification, manufacturing or sterilization process changes to the currently marketed devices differ only in the labeling (Instructions for Use). Where several sections of the Instructions for Use of the subject devices have been reworded reformatted for clarity to address evolving regulatory expectations and bring the contents of the Instructions for Use up to date and aligned within the Mesh Harmonization project noted in the Cover Letter. The Indication statement of the subject meshes have been modified to add clarity. The Indication of the subject devices do not introduce any new indications or expand the patient population of the predicate meshes.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Ethicon, a Johnson & Johnson company. The word "ETHICON" is written in large, red, sans-serif font. Below it, the phrase "a Johnson+Johnson company" is written in a smaller, cursive font.
# Conclusion:
The subject mesh, ULTRAPRO™ Mesh and ULTRAPRO ADVANCED are identical to the predicate marketed meshes, ULTRAPRO™ Mesh (K033337) and ULTRAPRO ADVANCED (K150906) with respect to intended use, technological characteristics, material, construction, specification, manufacturing and sterilization. In conclusion, the subject ULTRAPRO™ Mesh and ULTRAPRO ADVANCED are substantially equivalent to the predicate meshes.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.