5D Viewer is a software application for the display and 3D visualization of US Volume Data measured by an ultrasound system. This application is designed to allow the user to examine and analyze images based on volume data received from the specified ultrasound diagnostic system. It may also be used to search, and store the volume data.
Device Story
5D Viewer is standalone software installed on Windows 7 PCs/laptops. It processes 3D ultrasound volume data (DICOM files) exported from Samsung Medison ultrasound systems via storage media (USB). The software enables clinicians to review, edit, measure, and visualize volume data offline, away from the ultrasound system. Key functions include multi-planar reconstruction (MPR), volume rendering (e.g., Realistic Vue, Crystal Vue), and specialized analysis tools (e.g., 5D Heart, 5D Follicle, 5D CNS+). Some advanced features require a USB-type hardware dongle. By offloading data review from the ultrasound system, the device increases clinical workflow efficiency. Output is displayed on the PC monitor for physician interpretation to support clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing and verification/validation of software features.
Technological Characteristics
Standalone software application for Windows 7. Minimum requirements: Intel Core i5 CPU, 1GB Video Memory (Nvidia, Shader Model 3.0), 4GB RAM. Supports DICOM volume data files (*.mvl, *.sty, *.dcm) from Samsung Medison systems. Features include MPR, 3D/4D rendering, and measurement tools. Software life-cycle processes follow IEC 62304; risk management follows ISO 14971. HIPAA compliant.
Indications for Use
Indicated for the display, 3D visualization, examination, analysis, editing, measurement, and storage of ultrasound volume data acquired from specified Samsung Medison diagnostic ultrasound systems. Intended for use by clinicians.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA acronym is in a blue square, and the full name, "U.S. Food & Drug Administration," is in blue text.
Samsung Medison Co., Ltd. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street. NW BUFFALO MN 55313
Re: K180883
Trade/Device Name: 5D Viewer Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: May 9, 2018 Received: May 10, 2018
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
May 15, 2018
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180883
Device Name 5D Viewer
Indications for Use (Describe)
5D Viewer is a software application for the display and 3D visualization of US Volume Data measured by an ultrasound system.
This application is designed to allow the user to examine and analyze images based on volume data received from the specified ultrasound diagnostic system. It may also be used to search, and store the volume data.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
### 1. Submitter's Information:
SAMSUNG MEDISON CO., LTD.
145, Pangyoyeok-ro, Bundang-gu, Seongnam-si, Gyeonggi-do, 13530, Korea
#### 1.1. Contact Person
Ji Yea Lee Regulatory Affairs Specialist SAMSUNG MEDISON CO., LTD. Phone: 82.2.2194.1594 FAX: 82.2.556.3974 E-mail: jiyea722.lee@samsungmedison.com
Data Prepared: November 15, 2017
### 2. Name of the device:
Common/Usual Name: Picture Archiving Communications System Proprietary Name: 5D Viewer Classification Names: FR Number Product Code Picture Archiving Communications System 892.2050 LLZ
### 3. Identification of the predicate or legally marketed device:
- -5D Viewer (K161955)
- WS80A Diagnostic Ultrasound System (K171070) -
### 4. Device Description:
5D Viewer is standalone software that can be installed in laptops/PCs with Microsoft Windows 7.
This product lets users use their computers to review, analyze, edit, and measure the volume data exported from ultrasound equipment via storage media such as USB drive.
Since this product reads 3D volume data, users can review the test results of patients more quickly and
{4}------------------------------------------------
easily. This function allows them to check 3D image results without using an ultrasound system, helping them conduct more tests with the ultrasound system.
5D Viewer only supports the DICOM files containing the volume data created by Samsung Medison's Diagnostic Ultrasound Systems. In other words, you only use the volume data in DICOM files. 5D Viewer is not compatible with the DICOM files from third parties.
#### Main operational functions
- Display and edit volume data set
- · Save data (video and volume data)
- · Support simple caliper (Distance, Ellipse, or 3 Distance Volume)
#### Functions that require a USB-type dongle
- HDVI
- · 5D Heart
5D Viewer uses the volume data of Samsung Medison's WS80A, and HS70A Ultrasound systems. The Volume Data contains the raw data obtained during the scan as well as the information required for rendering (e.g., Geometry, View mode, etc.). The patient information is not relevant to the rendering, so it is not included in the Volume Data.
### 5. Intended Uses:
5D Viewer is a software application for the display and 3D visualization of Ultrasound volume data derived from Ultrasound system. It is designed to allow the user to observe images and perform analysis using the volume data acquired with specified diagnostic ultrasound systems. It is intended to be used for viewing, analyzing, editing, measuring and storing of volume data.
### 6. Technological Characteristics:
5D Viewer is substantially equivalent with respect to safety, effectiveness, and functionality to the 5D Viewer (K161955) and WS80A Diagnostic Ultrasound System (K171070) as a reference. The device is a software application and does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention interprets volume data being displayed. It is designed to allow the user to observe images and perform analysis using the volume data acquired with specified diagnostic ultrasound systems. It is intended to be used for viewing, analyzing, editing, measuring and storing of volume data.
These are described in detail in the technological characteristics comparison table as below.
{5}------------------------------------------------
| Feature | Subject Device | The Primary predicate device | The Reference predicate devices |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 5D Viewer<br>(Under Review) | 5D Viewer<br>(K161955) | WS80A<br>(K171070) |
| Indications for Use | 5D Viewer is a software application for the display and 3D visualization of US Volume Data measured by an ultrasound system.<br>This application is designed to allow the user to examine and analyze images based on volume data received from the specified ultrasound diagnostic system. It may also be used to search, analyze, edit, measure, and store the volume data. | 5D Viewer is a software application for the display and 3D visualization of US Volume Data measured by an ultrasound system.<br>This application is designed to allow the user to examine and analyze images based on volume data received from the specified ultrasound diagnostic system. It may also be used to search, analyze, edit, measure, and store the volume data. | The WS80A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.<br>The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal(Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Peripheral vessel and Intra-operative. |
| Computer platform<br>(minimum requirements) | - CPU: Intel Core i5<br>- VGA: Nvidia graphic card supporting Shader Model 3.0 with 1GB Video Memory<br>- Resolution: 1280x720<br>- Memory: 4GB | - CPU: Intel Core i5<br>- VGA: Nvidia graphic card supporting Shader Model 3.0 with 1GB Video Memory<br>- Resolution: 1280x720<br>- Memory: 4GB | Not applicable |
| Computer Operating System | Windows 7 | Windows 7 | Not applicable |
| Opening and saving files | *.mvl, *.sty, *.dcm<br>※ 5D Viewer only supports the DICOM files containing the volume data created by Samsung Medison. In other words, you only use the volume data in DICOM files. 5D Viewer is not compatible with the DICOM files from third parties. | *.mvl, *.sty, *.dcm<br>※ 5D Viewer only supports the DICOM files containing the volume data created by Samsung Medison. In other words, you only use the volume data in DICOM files. 5D Viewer is not compatible with the DICOM files from third parties. | *.mvl, *.sty |
| Functionality | (Similarity)<br>Functionality of 5D Viewer is same with the previously cleared 510(k) submission 5D Viewer (K161955) and WS80A (K171070) that there is not difference in functionality between them.<br>(Difference)<br>In the 5D Viewer, some functionality (Crystal Vue Flow) that was previously cleared on WS80A (K171070) has been added with 5D Viewer.<br>It doesn't impact on device safety and effectiveness. It is substantially equivalent with respect to safety, effectiveness, and functionality to the 5D Viewer (K161955) and WS80A (K171070) in | | |
| Feature | Subject Device | The Primary predicate device | The Reference predicate devices |
| | 5D Viewer<br>(Under Review) | 5D Viewer<br>(K161955) | WS80A<br>(K171070) |
| regards to the device with 5D Viewer. | | | |
| - MPR | Yes | Yes | Yes |
| - Slub 3D | Yes | Yes | Yes |
| - Accept ROI | Yes | Yes | Yes |
| - FAD | Yes | Yes | Yes |
| - Curved ROI | Yes | Yes | Yes |
| - Mirror View | Yes | Yes | Yes |
| - MagiCut | | | |
| - Smooth Cut | Yes | Yes | Yes |
| - Volume Slice | Yes | Yes | Yes |
| - MSV | Yes | Yes | Yes |
| - Oblique View | Yes | Yes | Yes |
| - Volume CT | Yes | Yes | Yes |
| - VOCAL | Yes | Yes | Yes |
| - XI VOCAL | Yes | Yes | Yes |
| - Cine View | | | |
| 3D Cine | Yes | Yes | Yes |
| 4D Cine | Yes | Yes | Yes |
| - 5D Functions | | | |
| 5D NT | Yes | Yes | Yes |
| 5D CNS+<br>(old name: 5D<br>CNS) | Yes | Yes | Yes |
| 5D Follicle | Yes | Yes | Yes |
| 5D LB | Yes | Yes | Yes |
| 5D Heart | Yes | Yes | Yes |
| 5D Limb Vol | Yes | No | Yes |
| - Measurement | | | |
| Distance | Yes | Yes | Yes |
| Ellipse | Yes | Yes | Yes |
| 3 Distance<br>Volume | Yes | Yes | Yes |
| - Render Setup<br>functions | | | |
| Realistic Vue | Yes | Yes | Yes |
| VSI | Yes | Yes | Yes |
| ClearVision | Yes | No | Yes |
| Crystal Vue | Yes | No | Yes |
| Crystal Vue<br>Flow | Yes | No | Yes |
| Natural Vue | Yes | No | No |
| - Post Processing<br>functions | | | |
| Post Gain | Yes | Yes | Yes |
| Feature | Subject Device | The Primary<br>predicate device | The Reference<br>predicate devices |
| | 5D Viewer<br>(Under Review) | 5D Viewer<br>(K161955) | WS80A<br>(K171070)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.