BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use, BIO-VENTURE Rapid Amphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Oxazepam Test Cassette for Rx Use, BIO-VENTURE Rapid Cocaine Test Cassette for Rx
BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use, BIO-VENTURE Rapid Amphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Oxazepam Test Cassette for Rx Use, BIO-VENTURE Rapid Cocaine Test Cassette for Rx
Applicant
Shanghai Venture Bio-Tech Co., Ltd.
Product Code
DJC · Clinical Toxicology
Decision Date
Dec 17, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3610
Device Class
Class 2
Indications for Use
BIO-VENTURE Rapid Amphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Oxazepam Test Cassette for Rx Use, BIO-VENTURE Rapid Cocaine Test Cassette for Rx Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Morphine Test Cassette for Rx Use, BIO-VENTURE Rapid Marijuana Test Cassette for Rx Use is a rapid lateral flow immunoassay for the qualitative detection of d-Amphetamine, Oxazepam, Benzoylecgonine, d-Methamphetamine, Morphine and Delta-9-THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows: Drug Identifier Cut-off Level Amphetamine(AMP) 1000ng/mL Oxazepam 300 ng/mL Cocaine(COC) 300 ng/mL Marijuana(THC) 50 ng/mL Methamphetamine(MET) 1000 ng/mL Morphine 300 ng/mL The cassette test may be configured as single drug test in any drug analytes listed in the table. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The test is intended for prescription use. BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use is a rapid lateral flow immunoassay for the qualitative detection of d-Amphetamine, Oxazepam, Benzoylecgonine, d-Methamphetamine, Morphine and Delta-9-THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows: Drug Identifier Cut-off Level Amphetamine(AMP) 1000ng/mL Oxazepam 300 ng/mL Cocaine(COC) 300 ng/mL Marijuana(THC) 50 ng/mL Methamphetamine(MET) 1000 ng/mL Morphine 300 ng/mL The cassette test may be configured as single drug test in any drug analytes listed in the table. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The test is intended for over-the-counter use.
Device Story
Lateral flow immunochromatographic assay; detects drugs of abuse in human urine. Cassette format houses single drug test strip. Principle: competitive binding; target drugs in urine compete with immobilized drug-conjugate for binding sites on gold-conjugated monoclonal antibodies. Absence of test line indicates positive result (drug concentration exceeds cutoff); presence of test line indicates negative result. Control line confirms proper migration. Used in clinical or home settings; operated by healthcare professionals or lay users. Provides preliminary results; requires confirmatory testing via GC/MS for clinical decision-making. Benefits: rapid, qualitative screening for drug presence.
Clinical Evidence
Bench testing only. Precision/reproducibility evaluated across 3 lots with 6 operators over 25 days. Analytical specificity (cross-reactivity and interference) and matrix effects (pH, specific gravity) demonstrated. Lay user study (n=840) confirmed ease of use and performance across diverse demographics. Method comparison against GC/MS performed for all analytes.
Technological Characteristics
Lateral flow immunochromatographic assay; mouse monoclonal antibodies on gold-conjugated nanoparticles; anti-IgG polyclonal antibody control line. Single-use cassette format. No external energy source. Stable at 2-30°C for 24 months.
Indications for Use
Indicated for qualitative detection of d-Amphetamine, Oxazepam, Benzoylecgonine, d-Methamphetamine, Morphine, and Delta-9-THC-COOH in human urine. Intended for prescription or over-the-counter use as a preliminary screening tool. Requires confirmatory testing (GC/MS).
Regulatory Classification
Identification
A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.
Special Controls
*Classification.* Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC Use) (k173303)
INSTANT-VIEW plus Multi-Drug Urine Test - Simple Cup (Prescription Use) (k173303)
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k180878
B. Purpose for Submission:
New device
C. Measurand:
Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine
D. Type of Test:
Qualitative, lateral flow immunochromatographic assay
E. Applicant:
Shanghai Venture Bio-Tech CO., Ltd.
F. Proprietary and Established Names:
BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use
BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use
BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use
BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use
BIO-VENTURE Rapid Morphine Test Cassette for OTC Use
BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use
BIO-VENTURE Rapid Amphetamine Test Cassette for Rx Use
BIO-VENTURE Rapid Oxazepam Test Cassette for Rx Use
BIO-VENTURE Rapid Cocaine Test Cassette for Rx Use
BIO-VENTURE Rapid Methamphetamine Test Cassette for Rx Use
BIO-VENTURE Rapid Morphine Test Cassette for Rx Use
BIO-VENTURE Rapid Marijuana Test Cassette for Rx Use
{1}
G. Regulatory Information:
| Product Codes | Class II | Regulation Section | Regulation Name | Panel |
| --- | --- | --- | --- | --- |
| DKZ, NFT | | 862.3100 | Amphetamine Test System | Toxicology (91) |
| DIO, NFY | | 862.3250 | Cocaine and Cocaine metabolites test system | |
| JXM, NFV | | 862.3170 | Benzodiazepine test system | |
| DJC, NGG | | 862.3610 | Methamphetamine Test System | |
| DNK, NGI | | 862.3640 | Morphine Test System | |
| LDJ, NFW | | 862.3870 | Cannabinoid test system | |
H. Intended Use:
1. Intended use(s):
Refer to Indication for Use.
2. Indication(s) for use:
BIO-VENTURE Rapid Amphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Oxazepam Test Cassette for Rx Use, BIO-VENTURE Rapid Cocaine Test Cassette for Rx Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Morphine Test Cassette for Rx Use, BIO-VENTURE Rapid Marijuana Test Cassette for Rx Use is a rapid lateral flow immunoassay for the qualitative detection of d-Amphetamine, Oxazepam, Benzoylecgonine, d-Methamphetamine, Morphine and Delta-9-THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
| Drug Identifier | Cut-off Level |
| --- | --- |
| Amphetamine(AMP) | 1000ng/mL |
| Oxazepam | 300 ng/mL |
| Cocaine(COC) | 300 ng/mL |
| Marijuana(THC) | 50 ng/mL |
| Methamphetamine(MET) | 1000 ng/mL |
| Morphine | 300 ng/mL |
The cassette test may be configured as single drug test in any drug analytes listed in the table. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method.
Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. The test is intended for prescription use.
{2}
BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use is a rapid lateral flow immunoassay for the qualitative detection of d-Amphetamine, Oxazepam, Benzoylecgonine, d-Methamphetamine, Morphine and Delta-9-THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
| Drug Identifier | Cut-off Level |
| --- | --- |
| Amphetamine(AMP) | 1000ng/mL |
| Oxazepam | 300 ng/mL |
| Cocaine(COC) | 300 ng/mL |
| Marijuana(THC) | 50 ng/mL |
| Methamphetamine(MET) | 1000 ng/mL |
| Morphine | 300 ng/mL |
The cassette test may be configured as single drug test in any drug analytes listed in the table. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The test is intended for over-the-counter use.
3. Special conditions for use statement(s):
The BIO-VENTURE Rapid Amphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Oxazepam Test Cassette for Rx Use, BIO-VENTURE Rapid Cocaine Test Cassette for Rx Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Morphine Test Cassette for Rx Use, BIO-VENTURE Rapid Marijuana Test Cassette for Rx Use are for prescription use.
The BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use are for over the counter use.
4. Special instrument requirements:
Not applicable.
{3}
I. Device Description:
The BIO-VENTURE Rapid Amphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Oxazepam Test Cassette for Rx Use, BIO-VENTURE Rapid Cocaine Test Cassette for Rx Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Morphine Test Cassette for Rx Use, BIO-VENTURE Rapid Marijuana Test Cassette for Rx Use and the BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use differ in intended use (prescription use or over the counter use), but share the same test cassette, with each cassette housing only one of 6 different drug test strips (for the qualitative detection of d-Methamphetamine, Morphine, 11-Nor-Δ9-Tetrahydrocannabinol-9-COOH, d-Amphetamine, Cocaine or Oxazepam in human urine).
The BIO-VENTURE test cassettes (noted in the above paragraph) are immunochromatographic assays that use a lateral flow system for the qualitative detection of d-Methamphetamine, Morphine, 11-Nor-Δ⁹-Tetrahydrocannabinol-9-COOH, d-Amphetamine, Cocaine and Oxazepam in human urine. The tests are available in single drug test cassette formats. BIO-VENTURE test cassettes are coated with mouse monoclonal antibodies, each raised against specific drug analytes (deposited on the test line), and a anti-IgG polyclonal antibody (deposited on the control line). The cassettes are provided in individual aluminum foil bags.
J. Substantial Equivalence Information:
1. Predicate device name(s):
INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC Use)
INSTANT-VIEW plus Multi-Drug Urine Test - Simple Cup (Prescription Use)
2. Predicate 510(k) number(s):
k173303
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Devices | Predicate – k173303 |
| Intended Use | Same | For the qualitative determination of drugs of abuse in human urine. |
| Methodology | Same | Competitive binding, lateral flow immunochromatographic assays based on the principle of antigen antibody |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Devices | Predicate – k173303 |
| | | immunochemistry. |
| Specimen type | Same | Urine |
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate – k173303 |
| Formats | Cassette | Cup |
# K. Standard/Guidance Document Referenced (if applicable):
None referenced.
# L. Test Principle:
The BIO-VENTURE Rapid Amphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Oxazepam Test Cassette for Rx Use, BIO-VENTURE Rapid Cocaine Test Cassette for Rx Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Morphine Test Cassette for Rx Use, BIO-VENTURE Rapid Marijuana Test Cassette for Rx Use and the BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, and BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use are rapid tests for the qualitative detection of Methamphetamine, Morphine, 11-Nor- $\Delta$ 9-Tetrahydrocannabinol-9-COOH, d-Amphetamine, Benzoylecgonine and Oxazepam in urine samples. The tests are lateral flow chromatographic immunoassays. During testing, a urine specimen migrates across the test strip by capillary action. If target drugs present in the urine specimen are below the cut-off concentration, they will not saturate the binding sites of specific anti-target drug monoclonal mouse antibodies coated on gold-conjugated nanoparticles, allowing the antibody-coated nanoparticles to be captured by immobilized drug-conjugate on the test line, and a visible colored line will show up in the test line region. The colored line in the test line region will not form if the target drug level exceeds its cutoff concentration because it will saturate all the binding sites of the antibody-coated nanoparticles, thus competing for binding to the test line region.
A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample to indicate that sample has migrated past the test and control line regions.
# M. Performance Characteristics (if/when applicable):
1. Analytical performance:
{5}
a. Precision/Reproducibility:
The BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, and BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use are identical physical devices to the BIO-VENTURE Rapid Amphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Oxazepam Test Cassette for Rx Use, BIO-VENTURE Rapid Cocaine Test Cassette for Rx Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Morphine Test Cassette for Rx Use, BIO-VENTURE Rapid Marijuana Test Cassette for Rx Use, but have different labeling appropriate for OTC users. Precision testing was conducted using the The BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, and BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use devices, but represents expected performance from both the Rx and OTC devices.
Precision was evaluated by analyzing spiked urine samples at concentrations of -100%, -75%, -50%, -25%, cutoff, +25%, +50%, +75%, and +100%. Three lots were evaluated and the results were read by six operators, each operator performs 9 samples for 25 days in a randomized order. Results were as follows:
Methamphetamine:
| MET | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | cutoff | +25% cutoff | +50% cutoff | +75% cutoff | +100% cutoff |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 10-/40+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 9-/41+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 10-/40+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
Morphine:
| MOP | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | cutoff | +25% cutoff | +50% cutoff | +75% cutoff | +100% cutoff |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 17-/33+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 19-/31+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 15-/35+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
{6}
Marijuana:
| THC | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | cutoff | +25% cutoff | +50% cutoff | +75% cutoff | +100% cutoff |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 36-/14+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 38-/12+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 35-/15+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
Amphetamine:
| AMP | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | cutoff | +25% cutoff | +50% cutoff | +75% cutoff | +100% cutoff |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 28-/22+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 26-/24+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 26-/24+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
Cocaine:
| COC | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | cutoff | +25% cutoff | +50% cutoff | +75% cutoff | +100% cutoff |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 27-/23+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 24-/26+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 26-/24+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
Oxazepam:
| OXAZ | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | cutoff | +25% cutoff | +50% cutoff | +75% cutoff | +100% cutoff |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 28-/22+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 25-/25+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 28-/22+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
b. Linearity/assay reportable range:
Not applicable. These devices are intended for qualitative use only.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
This device has internal process controls. A control line appears in the control region confirming that sufficient sample volume has been applied to the test strip and that the sample has migrated correctly on the test strip. Users are informed that the test is invalid if the colored line does not appear in the control
{7}
region. There are no external controls supplied with the device.
The sponsor claims that the devices are stable at 2-30 °C for 24 months based on a real time stability study at room temperature.
d. Detection limit:
Not applicable.
e. Analytical specificity:
The BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, and BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use are identical physical devices to the BIO-VENTURE Rapid Amphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Oxazepam Test Cassette for Rx Use, BIO-VENTURE Rapid Cocaine Test Cassette for Rx Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Morphine Test Cassette for Rx Use, BIO-VENTURE Rapid Marijuana Test Cassette for Rx Use, but have different labeling appropriate for OTC users. Cross reactivity and interference testing was conducted using the The BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, and BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use devices, but represents expected cross-reactivity and interference performance from both the Rx and OTC devices.
Cross-reactivity:
Cross-reactivity was evaluated by spiking in potential cross-reactants and titrating down to find the lowest concentration of cross-reactant that produced a positive result with the assays. The percent cross-reactivity was calculated as:
(cutoff concentration / lowest concentration causing a positive result) x 100
Results are summarized below.
Methamphetamine
| Potential cross-reactant | Lowest Concentration producing a positive result (ng/ml) | % Cross Reactivity |
| --- | --- | --- |
| D-Amphetamine | 100,000 | 1 |
{8}
9
| Potential cross-reactant | Lowest Concentration producing a positive result (ng/ml) | % Cross Reactivity |
| --- | --- | --- |
| L-Amphetamine | 100,000 | 1 |
| D/L-Amphetamine | 100,000 | 1 |
| p-Hydroxyamphetamine | 50,000 | 2 |
| MDA | 20,000 | 5 |
| MDMA | 3000 | 33.3 |
| MDEA | 5000 | 20 |
| D/L-Methamphetamine | 3000 | 33.3 |
| D-Methamphetamine | 1000 | 100 |
| L-Methamphetamine | 10,000 | 10 |
Morphine
| Potential cross-reactant | Lowest Concentration producing a positive result (ng/ml) | % Cross Reactivity |
| --- | --- | --- |
| Acetylmorphine | 6000 | 5 |
| Hydromorphone | 3000 | 10 |
| Dimethylmorphine | 3000 | 10 |
| Hydrocodone | 50,000 | 0.6 |
| Oxycodone | 50,000 | 0.6 |
| Levorphanol | 1500 | 20 |
| Morphine-3- Glucuronide | >100,000 | <0.3 |
| Morphine | 300 | 100 |
| Codeine | 300 | 100 |
| Heroin | 300 | 100 |
| O^{6}- Monoacetylmorphine | 300 | 100 |
| Ethylmorphine | >100,000 | <0.3 |
Marijuana
| Potential cross-reactant | Lowest Concentration producing a positive result (ng/ml) | % Cross Reactivity |
| --- | --- | --- |
| (-)-11-Nor-Δ^{9}-Tetrahydrocannabinol-9-COOH | 50 | 100 |
| 11-Hydroxy-Δ^{9}-Tetrahydrocannabinol | 10,000 | 0.5 |
| 11-Nor-Δ^{8}-Tetrahydrocannabinol-9-COOH | 2500 | 2 |
| Cannabinol | 5000 | 1 |
| 11-Nor-Δ^{9}-THC-carboxy glucuronide | 5000 | 1 |
| (-)-11-nor-9-carboxy-Δ^{9}-THC | 10,0000 | 0.5 |
{9}
10
| Potential cross-reactant | Lowest Concentration producing a positive result (ng/ml) | % Cross Reactivity |
| --- | --- | --- |
| Δ^{8}-Tetrahydrocannabinol | >50,000 | 0.1 |
| Δ^{9}-Tetrahydrocannabinol | >50,000 | 0.1 |
| Cannabidiol | >50,000 | 0.1 |
## Amphetamine
| Potential cross-reactant | Lowest Concentration producing a positive result (ng/ml) | % Cross Reactivity |
| --- | --- | --- |
| D-Amphetamine | 1000 | 100.0 |
| D/L-Amphetamine | 1500 | 66.7 |
| Hydroxyamphetamine | 10,000 | 10.0 |
| Phentermine | 3000 | 33.3 |
| MDA | 10,000 | 10.0 |
| MDMA | >100,000 | <1 |
| MDE | >100,000 | <1 |
| Ephedrine | >100,000 | <1 |
| Pseudoephedrine | >100,000 | <1 |
| D/L-Methamphetamine | >100,000 | <1 |
| D-Methamphetamine | >100,000 | <1 |
| L-Amphetamine | 3000 | 33.3 |
## Cocaine
| Potential cross-reactant | Lowest Concentration producing a positive result (ng/ml) | % Cross Reactivity |
| --- | --- | --- |
| Cocaine HCl | 5000 | 6 |
| Norcocaine | 25,000 | 1.2 |
| Ecgonine HCl | 50,000 | 0.6 |
| Benzoylecgonine | 300 | <0.3 |
| Cocaethylene | >100,000 | 100 |
## Oxazepam
| Potential cross-reactant | Lowest Concentration producing a positive result (ng/ml) | % Cross Reactivity |
| --- | --- | --- |
| Diazepam | 1000 | 30 |
| Alprazolam | 1000 | 30 |
| α-Hydroxyalprazolam | 5000 | 6 |
| Bromazepam | 10000 | 3 |
| Chlordiazepoxide | 10000 | 3 |
| Clobazam | 500 | 60 |
| Clonazepam | 3000 | 10 |
{10}
11
| Potential cross-reactant | Lowest Concentration producing a positive result (ng/ml) | % Cross Reactivity |
| --- | --- | --- |
| Delorazepam | 5000 | 6 |
| Estazolam | 10000 | 3 |
| Flunitrazepam | 5000 | 6 |
| Midazolam | 50000 | 0.6 |
| Nitrazepam | 500 | 60 |
| Nordiazepam | 5000 | 6 |
| Temazepam | 500 | 60 |
| Triazolam | 10000 | 3 |
| Lorazepam | 25000 | 1.2 |
| Oxazepam | 300 | 100 |
| Clorazepate dipotassium | >100,000 | <0.3 |
| Desalkylflurazepam | >100,000 | <0.3 |
| Norchlordiazepoxide | >100,000 | <0.3 |
## Interferences:
Interference was evaluated by spiking the potential interferents below into urine samples at $\pm 25\%$ of the cut-off for the AMP, COC, BNZ, MET, MOP, and THC assays.
All potential interferents were tested at a concentration of $100~\mu \mathrm{g/mL}$ except for THC which was tested at $10~\mu \mathrm{g/mL}$ and $1\%$ Ethanol. There were no deviations from the expected results for any of the three formats.
## Methamphetamine
| Methadone | Naltrexone | Naloxone | Morphine |
| --- | --- | --- | --- |
| Ephedrine | Pseudo Ephedrine | Gatifloxacin | Procaine |
| Amitriptyline | Chlorpheniramine Maleate | Promethazine | Amoxicillin |
| Methoxyphenamine | Ketamine Hydrochloride | Ranitidine | Tramadol |
| Buprenorphine | Phenobarbital | Nifedipine | Diazepam |
| Dextromethorphan | Cocaine | Theophylline | Aspirin |
| Acetaminophen | Δ9-THC | Hydrocortisone | Oxycodone |
| Ascorbic Acid | Hemoglobin | Bilirubin | Ibuprofen |
| Dolantin | Hydroxyketone | EDDP | Phencyclidine |
| Propoxyphene | Aminopyrine | Cotinine | Fentanyl |
| Benzoylecgonine | Secobarbital | Angiotensin | Adrenaline |
| Oxazepam | Uric Acid | Triglyceride | Oxalic Acid |
| Prednisolone acetate | Cholesterol | Ethanol | |
{11}
Morphine
| Methadone | Naltrexone | Naloxone | Cocaine |
| --- | --- | --- | --- |
| Ephedrine | Pseudo Ephedrine | Gatifloxacin | Procaine |
| Amitriptyline | Chlorpheniramine Maleate | Promethazine | Amoxicillin |
| Methoxyphenamine | Ketamine Hydrochloride | Ranitidine | Tramadol |
| Buprenorphine | Phenobarbital | Nifedipine | Methamphetamine |
| Dextromethorphan | Diazepam | Theophylline | Aspirin |
| Acetaminophen | Δ9-THC | Cholesterol | Hydrocortisone |
| Ascorbic Acid | Hemoglobin | Bilirubin | Ibuprofen |
| Dolantin | Hydroxyketone | EDDP | Phencyclidine |
| Propoxyphene | Aminopyrine | Cotinine | Fentanyl |
| Benzoylecgonine | Secobarbital | Angiotensin | Adrenaline |
| Oxazepam | Uric Acid | Triglyceride | Oxalic Acid |
| Prednisolone acetate | Amphetamine | MDMA | Ethanol |
THC
| Methadone | Naltrexone | Naloxone | Morphine |
| --- | --- | --- | --- |
| Ephedrine | Pseudo Ephedrine | Gatifloxacin | Procaine |
| Amitriptyline | Chlorpheniramine Maleate | Promethazine | Amoxicillin |
| Methoxyphenamine | Ketamine Hydrochloride | Ranitidine | Tramadol |
| Buprenorphine | Phenobarbital | Nifedipine | Methamphetamine |
| Dextromethorphan | Diazepam | Theophylline | Aspirin |
| Acetaminophen | Cocaine | Amphetamine | MDMA |
| Ascorbic Acid | Hemoglobin | Bilirubin | Ibuprofen |
| Dolantin | Hydroxyketone | EDDP | Phencyclidine |
| Propoxyphene | Aminopyrine | Cotinine | Fentanyl |
| Benzoylecgonine | Secobarbital | Angiotensin | Adrenaline |
| Oxazepam | Uric Acid | Triglyceride | Oxalic Acid |
| Prednisolone acetate | Cholesterol | Hydrocortisone | Oxycodone |
| Ethanol | | | |
AMP
| Methadone | Naltrexone | Naloxone | Morphine |
| --- | --- | --- | --- |
| Gatifloxacin | Procaine | Amitriptyline | Chlorpheniramine Maleate |
| Promethazine | Amoxicillin | Methoxyphenamine | Ketamine Hydrochloride |
| Ranitidine | Tramadol | Buprenorphine | Phenobarbital |
| Nifedipine | Diazepam | Dextromethorphan | Cocaine |
| Theophylline | Aspirin | Acetaminophen | Δ9-THC |
| Ascorbic Acid | Hemoglobin | Bilirubin | Ibuprofen |
| Dolantin | Hydroxyketone | EDDP | Phencyclidine |
| Propoxyphene | Aminopyrine | Cotinine | Fentanyl |
{12}
13
| Benzoylecgonine | Secobarbital | Angiotensin | Adrenaline |
| --- | --- | --- | --- |
| Oxazepam | Uric Acid | Triglyceride | Oxalic Acid |
| Prednisolone acetate | Cholesterol | Hydrocortisone | Oxycodone |
| Ethanol | | | |
Oxazepam
| Methadone | Naltrexone | Naloxone | Morphine |
| --- | --- | --- | --- |
| Ephedrine | Pseudo Ephedrine | Gatifloxacin | Procaine |
| Amitriptyline | Chlorpheniramine Maleate | Promethazine | Amoxicillin |
| Methoxyphenamine | Ketamine Hydrochloride | Ranitidine | Tramadol |
| Buprenorphine | Phenobarbital | Nifedipine | Methamphetamine |
| Dextromethorphan | Cocaine | Theophylline | Aspirin |
| Acetaminophen | Δ9-THC | Hydrocortisone | Oxycodone |
| Ascorbic Acid | Hemoglobin | Bilirubin | Ibuprofen |
| Dolantin | Hydroxyketone | EDDP | Phencyclidine |
| Propoxyphene | Aminopyrine | Cotinine | Fentanyl |
| Benzoylecgonine | Secobarbital | Angiotensin | Adrenaline |
| MDMA | Uric Acid | Triglyceride | Oxalic Acid |
| Prednisolone acetate | Cholesterol | Amphetamine | Ethanol |
Cocaine
| Methadone | Naltrexone | Naloxone | Morphine |
| --- | --- | --- | --- |
| Ephedrine | Pseudo Ephedrine | Gatifloxacin | Procaine |
| Amitriptyline | Chlorpheniramine Maleate | Promethazine | Amoxicillin |
| Methoxyphenamine | Ketamine Hydrochloride | Ranitidine | Tramadol |
| Buprenorphine | Phenobarbital | Nifedipine | Methamphetamine |
| Dextromethorphan | Diazepam | Theophylline | Aspirin |
| Acetaminophen | Δ9-THC | Hydrocortisone | Oxycodone |
| Ascorbic Acid | Hemoglobin | Bilirubin | Ibuprofen |
| Dolantin | Hydroxyketone | EDDP | Phencyclidine |
| Propoxyphene | Aminopyrine | Cotinine | Fentanyl |
| Secobarbital | Angiotensin | Adrenaline | Oxalic Acid |
| Oxazepam | Uric Acid | Triglyceride | Prednisolone acetate |
| Cholesterol | MDMA | Amphetamine | Ethanol |
The effect of pH was evaluated by preparing samples at $\pm 25\%$ of the cutoff for the MET, MOR, THC, AMP, COC, and OXAZ assays and adjusting the pH to 4.00 to 9.00, in increments of $1\mathrm{pH}$ unit. There were no deviations from the expected results.
The effect of specific gravity was evaluated by preparing samples at $\pm 25\%$ of
{13}
the cutoff for the MET, MOR, THC, AMP, COC, and OXAZ assays and adjusting the specific gravity to 1.002, 1.016, 1.018, 1.023, 1.033, and 1.036. There were no deviations from the expected results.
f. Assay cut-off:
For characterization of how the device performs analytically around the claimed cutoff concentration, please refer to section M.1.a., above.
2. Comparison studies:
a. Method comparison with predicate device:
The BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, and BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use are identical physical devices to the BIO-VENTURE Rapid Amphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Oxazepam Test Cassette for Rx Use, BIO-VENTURE Rapid Cocaine Test Cassette for Rx Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Morphine Test Cassette for Rx Use, BIO-VENTURE Rapid Marijuana Test Cassette for Rx Use, but have different labeling appropriate for OTC users. Method comparison testing was conducted using the The BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, and BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use devices, but represents expected performance from both the Rx and OTC devices.
Method comparison studies for the BIO-VENTURE Rapid Single Drug Test Cassettes for Amphetamine, Oxazepam, Cocaine and Cocaine metabolites, Marijuana, Methamphetamine, Morphine were performed in-house with three laboratory assistants for each device. Operators ran unaltered clinical samples; 87 samples for amphetamine, 80 samples for cocaine, 80 samples for Oxazepam, 81 samples for Methamphetamine, 81 samples for Morphine and 82 samples for Marijuana. The samples were blind-labeled and compared to results obtained by GC/MS. The results are presented in the tables below:
14
{14}
15
| Methamphetamine – Cassette | | Negative | (less than -50% by LC/MS) | Near Cutoff Negative by LC/MS (Between -50% and cut-off) | Near Cutoff Positive by LC/MS (Between the cut-off and +50%) | High Positive by LC/MS (greater than +50%) |
| --- | --- | --- | --- | --- | --- | --- |
| Operator 1 | Negative | 17 | 16 | 7 | 0 | 0 |
| | Positive | 0 | 0 | 1 | 11 | 29 |
| Operator 2 | Negative | 17 | 16 | 8 | 3 | 0 |
| | Positive | 0 | 0 | 0 | 8 | 29 |
| Operator 3 | Negative | 17 | 16 | 7 | 0 | 0 |
| | Positive | 0 | 0 | 1 | 11 | 29 |
Discordant results
| Drug in urine | | Operator | Cut-off (ng/mL) | GC/MS results(ng/mL) | Test results |
| --- | --- | --- | --- | --- | --- |
| Methamphetamine | 2 | 1000 | 1025 | - | |
| | | | 2 | 1033 | - |
| | | | 1 | 984 | + |
| | | | 3 | 984 | + |
| | | | 2 | 1055 | - |
| Morphine – Cassette | | Negative | (less than -50% by LC/MS) | Near Cutoff Negative by LC/MS (Between -50% and cut-off) | Near Cutoff Positive by LC/MS (Between the cut-off and +50%) | High Positive by LC/MS (greater than +50%) |
| --- | --- | --- | --- | --- | --- | --- |
| Operator 1 | Negative | 19 | 9 | 12 | 0 | 0 |
| | Positive | 0 | 0 | 1 | 15 | 25 |
| Operator 2 | Negative | 19 | 9 | 13 | 3 | 0 |
| | Positive | 0 | 0 | 0 | 12 | 25 |
| Operator 3 | Negative | 19 | 9 | 12 | 0 | 0 |
| | Positive | 0 | 0 | 1 | 15 | 25 |
{15}
Discordant results
| Drug in urine | Operator | Cut-off (ng/mL) | GC/MS results(ng/mL) | Test results |
| --- | --- | --- | --- | --- |
| Morphine | 2 | 300 | 323 | - |
| | 2 | | 325 | - |
| | 2 | | 335 | - |
| | 1 | | 293 | + |
| | 3 | | 293 | + |
| THC Cassette | | Negative | <-50% cut off | -50% cut off~cut off | Cut off~+50% cut off | >+50% cut off |
| --- | --- | --- | --- | --- | --- | --- |
| Operator 1 | Negative | 18 | 11 | 13 | 4 | 0 |
| | Positive | 0 | 0 | 0 | 12 | 24 |
| Operator 2 | Negative | 18 | 11 | 13 | 0 | 0 |
| | Positive | 0 | 0 | 0 | 16 | 24 |
| Operator 3 | Negative | 18 | 11 | 13 | 1 | 0 |
| | Positive | 0 | 0 | 0 | 15 | 24 |
Discordant results
| Drug in urine | Operator | Cutoff(ng/mL) | GC/MS Results (ng/mL) | Test Results |
| --- | --- | --- | --- | --- |
| Marijuana | 1 | 50 | 61 | - |
| | 1 | | 59 | - |
| | 3 | | 59 | - |
| | 1 | | 61 | - |
| | 1 | | 59 | - |
| Amphetamine- cassette | | Negative | (less than -50% by LC/MS) | Near Cutoff Negative by LC/MS (Between -50% and cut-off) | Near Cutoff Positive by LC/MS (Between the cut-off and +50%) | High Positive by LC/MS (greater than +50%) |
| --- | --- | --- | --- | --- | --- | --- |
| Operator 1 | Positive | 19 | 16 | 11 | 1 | 0 |
| | Negative | 0 | 0 | 0 | 13 | 27 |
| Operator 2 | Positive | 19 | 16 | 10 | 1 | 0 |
| | Negative | 0 | 0 | 1 | 13 | 27 |
| Operator 3 | Positive | 19 | 16 | 10 | 0 | 0 |
{16}
17
| Amphetamine- cassette | | Negative | (less than -50% by LC/MS) | Near Cutoff Negative by LC/MS (Between -50% and cut-off) | Near Cutoff Positive by LC/MS (Between the cut-off and +50%) | High Positive by LC/MS (greater than +50%) |
| --- | --- | --- | --- | --- | --- | --- |
| | Negative | 0 | 0 | 1 | 14 | 27 |
Discordant results
| Drug in urine | Operator | Cut-off (ng/mL) | GC/MS results(ng/mL) | Test results |
| --- | --- | --- | --- | --- |
| Amphetamine | 2 | 1000 | 989 | + |
| | 3 | | 989 | + |
| | 2 | | 1035 | - |
| | 1 | | 1062 | - |
| Oxazepam - cassette | | Negative | (less than -50% by LC/MS) | Near Cutoff Negative by LC/MS (Between -50% and cut-off) | Near Cutoff Positive by LC/MS (Between the cut-off and +50%) | High Positive by LC/MS (greater than +50%) |
| --- | --- | --- | --- | --- | --- | --- |
| Operator 1 | Positive | 15 | 10 | 15 | 0 | 0 |
| | Negative | 0 | 0 | 0 | 20 | 20 |
| Operator 2 | Positive | 15 | 10 | 13 | 0 | 0 |
| | Negative | 0 | 0 | 2 | 20 | 20 |
| Operator 3 | Positive | 15 | 10 | 13 | 0 | 0 |
| | Negative | 0 | 0 | 2 | 20 | 20 |
Discordant results
| Drug in urine | Operator | Cut-off (ng/mL) | GC/MS results(ng/mL) | Test results |
| --- | --- | --- | --- | --- |
| Oxazepam | 2 | 300 | 283 | + |
| | 2 | | 288 | + |
| | 3 | | 288 | + |
| | 3 | | 298 | + |
{17}
| Cocaine - cassette | | Negative | (less than -50% by LC/MS) | Near Cutoff Negative by LC/MS (Between -50% and cut-off) | Near Cutoff Positive by LC/MS (Between the cut-off and +50%) | High Positive by LC/MS (greater than +50%) |
| --- | --- | --- | --- | --- | --- | --- |
| Operator 1 | Positive | 18 | 13 | 11 | 0 | 0 |
| | Negative | 0 | 0 | 0 | 17 | 21 |
| Operator 2 | Positive | 18 | 13 | 9 | 0 | 0 |
| | Negative | 0 | 0 | 2 | 17 | 21 |
| Operator 3 | Positive | 18 | 13 | 9 | 0 | 0 |
| | Negative | 0 | 0 | 2 | 17 | 21 |
Discordant results
| Drug in urine | Operator | Cut-off (ng/mL) | GC/MS results(ng/mL) | Test results |
| --- | --- | --- | --- | --- |
| Cocaine | 2 | 300 | 278 | + |
| | 3 | | 278 | + |
| | 2 | | 285 | + |
| | 3 | | 285 | + |
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
A lay user study was conducted to assess the BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, and BIO-
{18}
VENTURE Rapid Marijuana Test Cassette for OTC Use devices with 840 lay persons with diverse educational and professional backgrounds and ranging in age from 20 to <50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug(s) into drug free-pooled urine specimens. The concentrations of the samples were confirmed by GC/MS, and the results provided below are from each lay user, who tested only one device at one test concentration. Lay-users were also given surveys on the ease of understanding the package insert instructions (Flesch-Kincaid reading analysis revealed a reading Grade Level of 7 for the device labeling). All lay users indicated that the device instructions could be easily followed.
BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use
| % of cut off | Number of samples | d-Amphetamine Concentration by GC/MS(ng/mL) | Number of Positive | Number of Negative | Percentage of Correct Results (%) |
| --- | --- | --- | --- | --- | --- |
| -100% cut off | 20 | 0 | 0 | 20 | 100% |
| -75% cut off | 20 | 248 | 0 | 20 | 100% |
| -50% cut off | 20 | 492 | 0 | 20 | 100% |
| -25% cut off | 20 | 741 | 2 | 18 | 90% |
| +25% cut off | 20 | 1262 | 19 | 1 | 95% |
| +50% cut off | 20 | 1506 | 20 | 0 | 100% |
| +75% cutoff | 20 | 1744 | 20 | 0 | 100% |
BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use
| % of cut off | Number of samples | Cocaine Concentration by GC/MS(ng/mL) | Number of Positive | Number of Negative | Percentage of Correct Results(%) |
| --- | --- | --- | --- | --- | --- |
| -100% cut off | 20 | 0 | 0 | 20 | 100% |
| -75% cut off | 20 | 62 | 0 | 20 | 100% |
| -50% cut off | 20 | 135 | 0 | 20 | 100% |
| -25% cut off | 20 | 212 | 2 | 18 | 90% |
| +25% cut off | 20 | 365 | 20 | 0 | 100% |
| +50% cut off | 20 | 444 | 20 | 0 | 100% |
| +75% cut off | 20 | 533 | 20 | 0 | 100% |
{19}
BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use
| % of cut off | Number of samples | Oxazepam Concentration by GC/MS(ng/mL) | Number of Positive | Number of Negative | Percentage of Correct Results(%) |
| --- | --- | --- | --- | --- | --- |
| -100% cut off | 20 | 0 | 0 | 20 | 100% |
| -75% cut off | 20 | 64 | 0 | 20 | 100% |
| -50% cut off | 20 | 141 | 0 | 20 | 100% |
| -25% cut off | 20 | 221 | 1 | 19 | 95% |
| +25% cut off | 20 | 362 | 20 | 0 | 100% |
| +50% cut off | 20 | 440 | 20 | 0 | 100% |
| +75% cut off | 20 | 530 | 20 | 0 | 100% |
BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use
| % of cut off | Number of samples | d-Methamphetamine Concentration by GC/MS(ng/mL) | | Number of Positive | Number of Negative | Percentage of Correct Results (%) |
| --- | --- | --- | --- | --- | --- | --- |
| -100% cut off | 20 | 0 | 0 | 20 | 100% | |
| -75% cut off | 20 | 248 | 0 | 20 | 100% | |
| -50% cut off | 20 | 489 | 0 | 20 | 100% | |
| -25% cut off | 20 | 734 | 2 | 18 | 90% | |
| +25% cut off | 20 | 1284 | 20 | 0 | 100% | |
| +50% cut off | 20 | 1535 | 20 | 0 | 100% | |
| +75% cut off | 20 | 1783 | 20 | 0 | 100% | |
BIO-VENTURE Rapid Morphine Test Cassette for OTC Use
| % of cut off | Number of samples | Morphine Concentration by GC/MS(ng/mL) | Number of Positive | Number of Negative | Percentage of Correct Results(%) |
| --- | --- | --- | --- | --- | --- |
| -100% cut off | 20 | 0 | 0 | 20 | 100% |
| -75% cut off | 20 | 70 | 0 | 20 | 100% |
| -50% cut off | 20 | 136 | 0 | 20 | 100% |
| -25% cut off | 20 | 228 | 2 | 18 | 90% |
| +25% cut off | 20 | 378 | 20 | 0 | 100% |
| +50% cut off | 20 | 445 | 20 | 0 | 100% |
| +75% cut off | 20 | 510 | 20 | 0 | 100% |
{20}
BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use
| % of cut off | Number of samples | Marijuana Concentration by GC/MS(ng/mL) | Number of Positive | Number of Negative | Percentage of Correct Results(%) |
| --- | --- | --- | --- | --- | --- |
| -100% cut off | 20 | 0 | 0 | 20 | 100% |
| -75% cut off | 20 | 10 | 0 | 20 | 100% |
| -50% cut off | 20 | 24 | 0 | 20 | 100% |
| -25% cut off | 20 | 36 | 0 | 20 | 100% |
| +25% cut off | 20 | 60 | 20 | 0 | 100% |
| +50% cut off | 20 | 73 | 20 | 0 | 100% |
| +75% cut off | 20 | 82 | 20 | 0 | 100% |
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Not applicable.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.