← Product Code [GEI](/productcode/GEI) · K180872

# SCARLET SRF (K180872)

_VIOL Co., Ltd. · GEI · Oct 11, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K180872

## Device Facts

- **Applicant:** VIOL Co., Ltd.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Oct 11, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The SCARLET SRF is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

## Device Story

SCARLET SRF is a radiofrequency (RF) electrosurgical system consisting of a main unit, handpiece with consumable tip, and footswitch. The device generates bipolar RF energy delivered to target tissue via penetrating needle electrodes in the consumable tip. The tip is held at right angles to the skin; RF energy creates an electro-thermal reaction for tissue coagulation and hemostasis. Used in clinical settings by healthcare professionals. The device is substantially equivalent to the INFINI Radiofrequency System (K121481).

## Clinical Evidence

Preclinical animal study conducted on three Yucatan Mini-pigs to evaluate safety and effectiveness for dermal treatment and hemostasis. Histopathology and dermal Draize scores were monitored over 22 days. Results showed transient erythema resolving by day 3, with expected minimal necrosis and subsequent dermal fibrosis. Bench testing included sterilization validation (ISO 11135-1), shelf-life (ASTM F1980-07), biocompatibility (ISO 10993), and electrical safety/EMC (IEC 60601 series).

## Technological Characteristics

Bipolar RF electrosurgical system. Consumable tips feature gold-plated surgical stainless steel needle electrodes. Depth range 0.5-3.5 mm. Spot size 10x10 mm. EO gas sterilization. Connectivity via footswitch/finger switch. Software is firmware-based. Complies with IEC 60601-1, IEC 60601-2-2, and EN 60601-1-2 standards.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- INFINI Radiofrequency System ([K121481](/device/K121481.md))

## Reference Devices

- VIVACE Electrosurgical ([K150409](/device/K150409.md))
- CELFIRM ([K172023](/device/K172023.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

#### October 11, 2018

VIOl Co., Ltd. % Mina Joo Assistant Manager BT Solutions, Inc. 91-14 Seolleung-ro #402 Gangnam-gu Seoul, Korea 06150

Re: K180872

Trade/Device Name: SCARLET SRF Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: September 11, 2018 Received: September 13, 2018

Dear Mina Joo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Lo
20
-C

Long H. Chen -S 2018.10.11 07:58:57 -04'00'

for

Binita S. Ashar. M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K180872

Device Name SCARLET SRF

Indications for Use (Describe)

The SCARLET SRF is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

| Type of Use (Select one or both, as applicable)                                                                                    |                                                            |
|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)         </span> </div> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
  
Food and Drug Administration
  
Office of Chief Information Officer
  
Paperwork Reduction Act (PRA) Staff
  
PRAStaff@fda.hhs.gov

*“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of*

of conduct or sponsor, and a person is not required to respons
 information unless it displays a currently valid OMB number." તિ

{3}------------------------------------------------

# 5. 510(k) Summary

# 1. General Information

| Applicant/Submitter: | VIOL Co., Ltd.                                                                                                                       |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Address:             | C-808, 809, Bundang Technopark C,<br>744, Pangyo-ro, Bundang-gu, Seongnam-si,<br>Gyeonggi-do (13510), South Korea                    |
|                      | Tel) +82-31-8017-7893<br>Fax) +82-31-8016-7894                                                                                       |
| Contact Person:      | Mina Joo, BT Solutions, Inc.                                                                                                         |
| Address:             | 91-14 Seolleung-ro, #402<br>Gangnam-gu Seoul, 06150<br>Republic of Korea<br>Tel) +82.2.538.9140<br>Email) smanager@btsolutions.co.kr |
| Preparation Date:    | September-5-2018                                                                                                                     |

## 2. Device Name and Code

| Device Trade Name:   | SCARLET SRF                                          |
|----------------------|------------------------------------------------------|
| Common Name:         | Radiofrequency System                                |
| Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories |
| Product Code:        | GEI                                                  |
| Regulation Number:   | 878.4400                                             |
| Classification:      | Class II                                             |
| Review Panel:        | General & Plastic Surgery                            |

# 3. Predicate Devices

Scarlet SRF is substantially equivalent to the following devices

|  | Table 5.1 Predicate devices |  |
|--|-----------------------------|--|
|--|-----------------------------|--|

| Applicant            | Device Name                  | 510(k) Number |
|----------------------|------------------------------|---------------|
| Lutronic Corporation | INFINI Radiofrequency System | K121481       |

{4}------------------------------------------------

### Scarlet SRF 510(k) Summary

Table 5.2 Reference devices *

| Applicant               | Device Name            | 510(k) Number |
|-------------------------|------------------------|---------------|
| Sung Hwan E&B Co., Ltd. | VIVACE Electrosurgical | K150409       |

*This reference device is referred, of which similarity of having smaller number of non-insulated needles when compare to the same predicate device.

#### 4. Device Description

SCARLET SRF is basically an identical device with the 510(k)-cleared device CELFIRM™ (K172023).

The SCARLET SRF includes a system main device, a hand-piece equipped with a tip, and a footswitch. The RF signal is generated from the main device which is then delivered to the handpiece and then to consumable tip. The RF signal is delivered to the target tissue using penetrating needle electrodes in the tip. The tip is placed in light contact with the epidermis while the handpiece is being held at right angles to the target tissue. As the RF signal passes through the skin, it generates an electro thermal reaction which is capable of coagulating the tissue. Using the consumable tip, SCARLET SRF creates heat within the target skin tissue via needle electrodes from the consumable tip.

#### 5. Indications / Intended Use

The SCARLET SRF is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

#### 6. Technical Characteristics in Comparison to Predicate Devices

|                                         | Predicate Device                                                                                                                                                                                              | Proposed Device                                                                                                                     |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) Number                           | K121481                                                                                                                                                                                                       | N/A                                                                                                                                 |
| Manufacturer                            | Lutronic Corporation                                                                                                                                                                                          | VIOL Co., Ltd.                                                                                                                      |
| Device Name                             | INFINI Radiofrequency System                                                                                                                                                                                  | SCARLET SRF                                                                                                                         |
| Clearance Date:                         | Jun 25,2013                                                                                                                                                                                                   | N/A                                                                                                                                 |
| Classification /<br>Regulation          | Class II/21 CFR 878.4400<br>Electrosurgical, Cutting & Coagulation<br>& Accessories                                                                                                                           | Class II/21 CFR 878.4400<br>Electrosurgical, Cutting & Coagulation<br>& Accessories                                                 |
| Product Code                            | GEI                                                                                                                                                                                                           | GEI                                                                                                                                 |
| Intended Use/<br>Indications for<br>Use | The INFINI Radiofrequency System is<br>intended for use in dermatologic and<br>general surgical procedures for<br>electrocoagulation and hemostasis, and<br>the percutaneous treatment of facial<br>wrinkles. | The SCARLET SRF is intended for use<br>in dermatologic and general surgical<br>procedures for electrocoagulation and<br>hemostasis. |
| Intended for                            | Prescription Use                                                                                                                                                                                              | Prescription Use                                                                                                                    |
| RF Output Type                          | BiPolar                                                                                                                                                                                                       | BiPolar                                                                                                                             |

{5}------------------------------------------------

#### Scarlet SRF 510(k) Summary

| Electrode<br>exposure<br>(Depth) |                                        |                                        |
|----------------------------------|----------------------------------------|----------------------------------------|
| Output Control                   | Foot switch                            | Foot / Finger switch                   |
| Electrode<br>Material            | Gold plated (Surgical stainless steel) | Gold plated (Surgical stainless steel) |
| No. of usage                     | Single use                             | Single use                             |
| Sterilization                    | EO gas sterilization                   | EO gas sterilization                   |
| Spot size                        | $10 \times 10$ (mm)                    | $10 \times 10$ (mm)                    |
|                                  | 0.5-3.5 mm (Unit: 0.1)                 | 0.5-3.5 mm (Unit: 0.25)                |

## 7. Performance Data

SCALRET SRF is basically an identical device with the 510(k)-cleared device CELFIRM™.

Sterilization Validation, Shelf life Validation, Biocompatibility, Software Validation, EMC & Electrical Safety Test, Max Output Energy, and Animal Test were conducted on the subject device.

#### Sterilization Validation Test

To verify the sterility assurance level (10-6) for EtO sterilization and biological indicator (BI) overkill method was used in accordance to ISO 11135-1, ISO 10993-1, ISO 10993-7, ISO 11138-1, ISO 11138-2, ISO 11737-1, ISO 11737-2, AAMI TIR 15, and ISO 13485:2012.

#### Shelf life Validation Test

The tests to validate the shelf life of the device through the proposed shelf life were conducted using the accelerated aging method in accordance to ASTM F1980-07 (2011) Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. The referenced standards for the testing are ISO 11607-1, ISO 11607-2, ISO 11737-2, ASTM F1929, ASTM F 88, and USPNF.

#### Biocompatibility Test

The patient contacting component of the subject device is the Consumable Tip and the biocompatibility tests were in accordance with ISO 10993-5. ISO 10993-7. ISO 10993-10, ISO 10993-11, and USP 39 <151>.

#### Software Validation Test

The level of concern of the software (firmware) of the SCARLET SRF is moderate and the software validation tests were performed.

## EMC & Electrical Safety Test

{6}------------------------------------------------

#### Scarlet SRF

#### 510(k) Summary

The subject device has been evaluated for electromagnetic compatibility and electrical safety testing per applicable standards of IEC 60601-1:2005, IEC 6061-1-6:2010, IEC 60601-2-2:2009, EN 60601-1-2:2007, EN 6100-3-2:2006, and EN 6100-3-3:2008, and all results were passing.

#### Max Output Energy

The test on Max Output Energy per Electrode was performed on the subject device to evaluate the max energy per electrode. The test results were compared and evaluated with the standard values of maximum output energy per Electrode of the device. The test results met the pre-set criteria and validated the device performance specifications for max energy.

#### Preclinical Test

The study was designed to evaluate the safety and effectiveness of the subject device as a radiofrequency(RF) heating device to treat dermal conditions and hemostasis. Three (3) Yucatan Mini-pigs were used in this study. Animal 2162 was euthanized 3 h after treatment on Day 1;

Animal 2163 was euthanized on Day 4; and Animals 2164 was euthanized on Day 22. The various treatment sites were then harvested and stored in a 10% NBF prior to histopathology evaluation. During the in-life stage, mortality, clinical observations and body weights were monitored and recorded. In addition, dermal Draize scores were recorded within 3-4 h of treatment, and on Days 3, 5, 13, and 22 in surviving animals.

Treatment with the subject device resulted in transient very slight to well-defined erythema noted within 3-4 h of treatment but fully resolved by the next Draize evaluation on Day 3. In addition, minimal to mild necrosis was noted in animals euthanized on Days 1 and 5 with the animal euthanized on Day 22 showing dermal fibrosis at the treatment sites. There were no other application-related findings.

Test results including histology data demonstrated that the subject device is substantially equivalent to the predicate devices in the market for its intended use.

#### 8. Substantial Equivalence

The intended use of the SCARLET SRF is within the scope of the predicate device. SCARLET SRF, from both a design and clinical perspective, uses similar or identical technology as the cited predicate device and has the same intended uses. Based upon the predicted overall performance characteristics for the SCARLET SRF, VIOL Co., Ltd., believes that no significant differences exist in usage of its underlying technological principles between SCARLET SRF and the cited predicate device.

#### 9. Conclusions

Based on the information provided in this premarket notification. VIOL Co., Ltd. concludes that the SCARLET SRF is substantially equivalent to the predicate device as described herein.

---

**Source:** [https://fda-staging.innolitics.com/device/K180872](https://fda-staging.innolitics.com/device/K180872)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
