FORA NAS100 Electronic Nasal Aspirator, Electronic Nasal Aspirator
K180863 · Taidoc Technology Corporation · BTA · Dec 19, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180863
Device Name
FORA NAS100 Electronic Nasal Aspirator, Electronic Nasal Aspirator
Applicant
Taidoc Technology Corporation
Product Code
BTA · General, Plastic Surgery
Decision Date
Dec 19, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The FORA NAS100 Electronic Nasal Aspirator (Electronic Nasal Aspirator, TD- 7601), NAS100 is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment.
Device Story
Portable, battery-operated (2x1.5V AA) suction device; utilizes motor pump to generate negative pressure (52-60 Kpa); nasal secretions collected in device container via silicone tips. Used in home environment by parents/caregivers for children aged 2-12. Device provides non-invasive removal of mucus; lower noise level (45 dBA) compared to predicate; smaller form factor. Healthcare providers do not interact with output; device is for symptom relief/hygiene.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via electrical safety (IEC 60601-1), home healthcare environment safety (IEC 60601-1-11), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993-1, -5, -10, -12).
Technological Characteristics
Materials: ABS, PC, Silicone (KE-941U). Principle: Motor-driven negative pressure suction. Power: 3V DC (2x1.5V AA batteries). Dimensions: 41x41x200 mm. Weight: 175g. Connectivity: None. Ingress protection: IP22. Applied part: Type BF. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993.
Indications for Use
Indicated for intermittent removal of nasal secretions and mucus in children aged 2-12 years in a home environment.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 19, 2018
TaiDoc Technology Corporation Sylvia Liu Regulatory Affairs Specialist 6F, No.127, Wugong 2nd Rd., Wugu District New Taipei City, 24888 TW
Re: K180863
Trade/Device Name: FORA NAS100 Electronic Nasal Aspirator, Electronic Nasal Aspirator Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: BTA Dated: March 31, 2018 Received: April 2, 2018
Dear Sylvia Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Srinivas Nandkumar -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K180863
#### Device Name
FORA NAS100 Electronic Nasal Aspirator (Model No: NAS100), Electronic Nasal Aspirator (Model No: TD-7601)
Indications for Use (Describe)
The FORA NAS100 Electronic Nasal Aspirator (Electronic Nasal Aspirator, TD- 7601), NAS100 is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) information is being submitted in accordance with the requirements of 21 CFR 807.92.
| The Assigned 510(k) number is: K180863 | | | | |
|----------------------------------------|--|--|--|--|
|----------------------------------------|--|--|--|--|
| 1. Submitter Information | | |
|--------------------------|------------------------------------------------------------|--|
| Company Name: | TaiDoc Technology Corporation | |
| Contact Person: | Sylvia Liu | |
| Title: | Regulatory Affairs Specialist | |
| Address: | B1-7F., No. 127, Wugong 2nd Rd., Wugu District, New Taipei | |
| | City 24888, TAIWAN | |
| Phone: | +886-2-6625-8188 #6134 | |
| Fax: | +886-2-6625-0288 | |
| E-mail: | Sylvia.liu@taidoc.com.tw | |
| Prepared Date: | December 10th, 2018 | |
## 2. Device Name
| Proprietary Name: | FORA NAS100 Electronic Nasal Aspirator, NAS100<br>(Electronic Nasal Aspirator, TD-7601) |
|-------------------|-----------------------------------------------------------------------------------------|
| Common Name: | Powered Suction Pump |
| Product Code: | BTA |
| Review Panel: | General & Plastic Surgery |
| Device Class: | Class II |
21 CFR § 878.4780
#### Regulation Number: 3. Predicate Device
| Proprietary Name: | Avita Nasal Aspirator, Model NS1 |
|-------------------|----------------------------------|
| Manufacturer | AVITA Corporation |
| 510(K) no. | K090379 |
#### 4. Device Descrip tion
#### The FORA NAS100 Electronic Nasal Aspirator (Electronic Nasal Aspirator, TD-
7601), NAS100 is a portable device which is intended for suction of nasal passages in children 2-12 years of age. The motor pump provides a negative pressure which removes nasal secretions. Two different shapes of silicone nasal tips are provided to enable easier and more effective removal of the nasal mucus.
#### Principle of Oneration
#### The FORA NAS100 Electronic Nasal Aspirator (Electronic Nasal Aspirator, TD-
7601), NAS100, utilizes a motor pump to generate negative pressure in the suction system, which allows nasal secretions to flow into the device container.
{4}------------------------------------------------
# 5. Indications for Use
#### The FORA NAS100 Electronic Nasal Aspirator (Electronic Nasal Aspirator, TD-
7601), NAS100 is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment.
### 6. Technological Characteristics
Table of Specification Comnarison
| Item | Predicate device | Proposed device | |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510K Number | K090379 | K180863 | |
| | General Information | | |
| Appearance | Image: Avita Nasal Aspirator | Image: FORA NAS100 Electronic Nasal Aspirator | |
| Device Trade Name<br>/ Proprietary name | Avita Nasal Aspirator, Model NS1 | FORA NAS100 Electronic Nasal<br>Aspirator (Model No: NAS100);<br>Electronic Nasal Aspirator<br>(Model No: TD-7601) | |
| Manufacturer | AVITA Corporation | TaiDoc Technology Corporation | |
| Common<br>/Classification<br>Name | Powered Suction Pump | Same as the predicate | |
| Device Class | II | Same as the predicate | |
| Product Code | BTA | Same as the predicate | |
| Classification Panel | General & Plastic Surgery | Same as the predicate | |
| Regulation Number | 21 CFR § 878.4780 | Same as the predicate | |
| Indication for Use | This device is designed for using<br>Intermittent suction to remove nasal<br>secretion and mucus in Children (age 2-<br>12 years old) at home environment. | | The FORA NAS100 Electronic<br>Nasal Aspirator (Electronic Nasal<br>Aspirator, TD- 7601), NAS100 is<br>intended for intermittent removal<br>of nasal secretions and mucus from<br>children (age 2-12 years old). This<br>device is used in a home<br>environment. |
| Item | Predicate device | Proposed device | |
| 510K Number | K090379 | K180863 | |
| Population | Age 2-12 years old | Same as the predicate | |
| Intended<br>Environment | Home use | Same as the predicate | |
| Device Description | AVITA NS1 Nasal Aspirator is a<br>portable, DC powered device Intended to<br>provide the suction function to aspirate<br>children's nasal secretion. The device<br>consist of a pump that is driven by Two<br>(2) 1.5V, AA size alkaline batteries, soft<br>aspiration tip, collection cup and Music<br>IC with 12 Chord Melody. | The FORA NAS100 Electronic Nasal<br>Aspirator (Electronic Nasal Aspirator,<br>TD-7601), NAS100 is a portable<br>device which is intended for suction<br>of nasal passages in children 2-12<br>years of age. The motor pump<br>provides a negative pressure which<br>removes nasal secretions. Two<br>different shapes of silicone nasal tips<br>are provided to enable easier and<br>more effective removal of the nasal<br>mucus. | |
| General Functions | | | |
| Vacuum pressure | 52Kpa | 52-60 Kpa | |
| Noise Level | 75-80dB/22mm 0.25w/1M | 45 dBA | |
| Power consumption | - | 3W | |
| Device Dimension | 93.5(L) x 39.9 (W) x 148(H) mm | 41 (L) x 41 (W) x 200 (H) mm | |
| Tips Dimension (ψ) | Type1: 6 (OD)/ 2(ID)<br>Type2: 4.2 (OD)/2.6 (ID) | Type1: 5.5 (OD)/ 3(ID)<br>Type2: 4.5 (OD)/2.5 (ID) | |
| Weight | 250(g) | 175(g) | |
| Motor Type | 3V DC | Same as the predicate | |
| Power Source | 2x1.5V AA | Same as the predicate | |
| Material | ABS, PC, Silicone | Same as the predicate | |
| Operating condition | 60.8°F to 95°F; up to 85% R.H.<br>(non condensing) | 41°F to 104°F; 15% to 93% R.H.<br>(non condensing) | |
| Storage condition | -13°F to 131°F; up to 85% R.H | -13°F to 158°F;10% to 95% R.H. | |
| Expected service life | - | 2 years | |
| Type BF applied part | - | Type BF Applied part | |
| Safety | - | IEC 60601-1 | |
| EMC | | IEC 60601-1-2 | |
| Water-resistance | | IP22 | |
| Biocompatibility Information | | | |
| Item | Predicate device | Proposed device | |
| 510K Number | K090379 | K180863 | |
| Description | Unknown | NAS100 (TD-7601) operates in<br>conjunction with silicon nasal<br>aspiration tips, which come into the<br>contact with nasal skin and mucosa<br>for less than 24 hours. Two different<br>shapes of silicone nasal tips are<br>provided to enable easier and more<br>effective removal of the nasal mucus. | |
| Standard | Unknown | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-12 | |
| Contacted Parts | Silicone Tip | Same as predicate | |
| Material | - | KE-941U | |
{5}------------------------------------------------
TaiDoc TechnologyCorp.
{6}------------------------------------------------
The conceptual design of the proposed device (K180863) is similar to the predicate device (K090379). Both have the same principle of operation, general function, and indications for use. The proposed device (K180863) has been designed to have smaller dimensions, which makes it more convenient to operate. Another advantage of the proposed device (K180863) is a lower noise level, which decreases hearing discomfort during device use.
Additionally, the silicone nasal tips of the proposed device have undergone biocompatibility evaluation. The device complies with general requirements for basic safety and essential performance, and electromagnetic compatibility according to IEC 60601-1-2, IEC60601-11, and IEC 60601-1.
{7}------------------------------------------------
# 7. Performance Data
## Safety and Effectiveness Characteristics:
| Standard | Title | Intended Use | Acceptance Criteria | Results | |
|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| IEC 60601-1 | General<br>requirements for<br>basic safety and<br>essential<br>performance | This study is to test for the<br>basic safety and essential<br>performance of medical<br>electrical equipment and<br>medical electrical systems<br>which are intended by their<br>manufacturer for use. | The pass/fail criteria is<br>to evaluate the basic<br>safety and essential<br>performance of<br>medical electrical<br>equipment and medical<br>electrical systems. | 1. General requirements, Pass.<br>2. General requirements for testing<br>ME equipment, Pass.<br>3. Classification of MEequipment<br>and MEsystems, Pass.<br>4. ME equipment identification,<br>marking and documents, Pass.<br>5. Protection against excessive<br>temperatures and other hazards,<br>Pass.<br>6. Accuracy if controls and<br>instruments and protection<br>against hazardous outputs, Pass.<br>7. Construction of MEequipment,<br>Pass.<br>8. Protection against strangulation<br>or asphyxiation, Pass.<br>9. Additional requirements for<br>electromagnetic emissions of ME<br>equipment and MEsystems,<br>Pass.<br>10. Additional requirements for<br>alarm systems if ME equipment<br>and MEsystems, Pass. | |
| IEC 60601-1-11 | General<br>requirements for<br>basic safety and<br>essential<br>performance –<br>Collateral<br>Standard: | This study is to evaluate to<br>the basic safety and<br>essential performance of<br>medical electrical<br>equipment and medical<br>electrical systems which<br>are intended by their | The pass/fail criteria is<br>to evaluate the basic<br>safety and essential<br>performance of<br>medical electrical<br>equipment and medical<br>electrical systems for | 1. General requirements, Pass.<br>2. General requirements for testing<br>ME equipment, Pass.<br>3. Classification of MEequipment<br>and MEsystems, Pass.<br>4. ME equipment identification,<br>marking and documents, Pass.<br>5. Protection against excessive<br>temperatures and other hazards | |
| | Requirements<br>for medical<br>electrical<br>equipment and<br>medical<br>electrical<br>systems used in<br>the home<br>healthcare<br>environment | manufacturer for use in the<br>home healthcare<br>environment, regardless of<br>whether the medical<br>electrical equipment or<br>medical electrical system is<br>intended for use by a lay<br>operator or by trained<br>healthcare personnel. | use in the home<br>healthcare<br>environment. | Pass.<br>6. Accuracy if controls and<br>instruments and protection<br>against hazardous outputs, Pass.<br>7. Construction of MEequipment,<br>Pass.<br>8. Protection against strangulation<br>or asphyxiation, Pass.<br>9. Additional requirements for<br>electromagnetic emissions of ME<br>equipment and MEsystems,<br>Pass.<br>10. Additional requirements for<br>alarm systems if MEequipment<br>and MEsystems, Pass. | |
| IEC 60601-1-2 | EMC Test<br>Report | IEC 60601-1-2:2014<br>applies to the basic safety<br>and essential performance<br>of Medical Equipment<br>(ME) equipment and ME<br>systems in the presence of<br>electromagnetic<br>disturbances and to<br>electromagnetic<br>disturbances emitted by me<br>equipment and me systems. | The pass/fail criteria is<br>limited to maintain the<br>Essential Performance<br>and Basic Safety of<br>EMC requirements. | 1. Enclosure port, Pass<br>2. ESD, Pass<br>3. RS, Pass<br>4. RF, Pass<br>5. PFMF, Pass | |
| ISO 14971 | Risk<br>Management<br>Report | The failure of the function<br>resulting in wrong<br>analytical result which may<br>have serious impairment to<br>the health of a patient may<br>happen. Control measures<br>were taken to reduce the | The risk criteria were<br>established when<br>setting the context, the<br>level of risk would<br>against this criteria in<br>order to determine<br>whether the risk is | The risk distribution is shown the<br>before/after risk analysis to be sure<br>that all hazards are controlled on<br>"acceptable region". | |
| | | risk to as minimum as<br>possible. Safety and<br>effectiveness of use about<br>the system was then<br>verified. | acceptable. | | |
| | | | | | |
TaiDoc TechnologyCorp.
{8}------------------------------------------------
TaiDoc TechnologyCorp.
{9}------------------------------------------------
Biocompatibility & Clinical Test Report
| Standard | Title | Intended Use | Acceptance Criteria | Results |
|------------------------------------------------------------|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-12 | Biocompatibility<br>Test Report for<br>Thermometer | When assessing medical<br>device, the sponsor<br>should specifically state<br>if the medical device<br>does not result in any<br>risk of direct or indirect<br>tissue-contacting<br>components. Thus,<br>performing the<br>biocompatibility test to<br>indicate the safety of<br>device, which includes<br>with Cytotoxicity Test,<br>Skin Sensitization Test,<br>and Irritation Test. | Cytotoxicity Test:<br>If cell viability is reduced to <<br>70% of the reagent control<br>extract, a cytotoxic potential<br>exists.<br>Skin Sensitization Test:<br>Grades of 1 or greater observed<br>in the test group generally<br>indicated sensitization,<br>provided that grades of less<br>than 1 were observed on the<br>control animals.<br>Irritation Test:<br>For each rabbit, the irritation<br>score for test area was<br>calculated by adding together<br>the scores of erythema and<br>edema at each time point and<br>divide the sum by the total<br>number of observation. | 1. Irritation index for the test<br>result was calculated to be 0.<br>2. There were not found<br>abnormal clinical<br>symptoms except skin<br>reactions in rabbit.<br>3. There were normal weight<br>change.<br>4. The positive/negative rate<br>of all sample extract animal<br>is 0%<br>5. The sample extract showed<br>no significant evidence of<br>causing skin sensitization in<br>the guinea pig.<br>6. Under the conditions of the<br>MTT assay, the test article<br>Breast Pump extract did not<br>show potential toxicity to L-<br>929 cells. |
| ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-12 | Biocompatibility<br>Test Report for<br>Stainless Steel<br>Cap | When assessing medical<br>device, the sponsor<br>should specifically state<br>if the medical device<br>does not result in any<br>risk of direct or indirect<br>tissue-contacting<br>components. Thus,<br>performing the<br>biocompatibility test to<br>indicate the safety of<br>device, which includes<br>with Cytotoxicity Test,<br>Skin Sensitization Test,<br>and Irritation Test. | Cytotoxicity Test:<br>If cell viability is reduced to <<br>70% of the reagent control<br>extract, a cytotoxic potential<br>exists.<br>Skin Sensitization Test:<br>Grades of 1 or greater observed<br>in the test group generally<br>indicated sensitization,<br>provided that grades of less<br>than 1 were observed on the<br>control animals.<br>Irritation Test:<br>For each rabbit, the irritation<br>score for test area was<br>calculated by adding together<br>the scores of erythema and<br>edema at each time point and<br>divide the sum by the total<br>number of observation. | 1. Irritation index for the test<br>result was calculated to be 0.<br>2. There were not found<br>abnormal clinical<br>symptoms except skin<br>reactions in rabbit.<br>3. There were normal weight<br>change.<br>4. The positive/negative rate<br>of all sample extract animal<br>is 0%<br>5. The sample extract showed<br>no significant evidence of<br>causing skin sensitization in<br>the guinea pig.<br>6. Under the conditions of the<br>MTT assay, the test article<br>Breast Pump extract did not<br>show potential toxicity to L-<br>929 cells. |
| ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-12 | Biocompatibility<br>Test Report for<br>Elastic Band | When assessing medical<br>device, the sponsor<br>should specifically state<br>if the medical device<br>does not result in any<br>risk of direct or indirect<br>tissue-contacting<br>components. Thus,<br>performing the<br>biocompatibility test to<br>indicate the safety of<br>device, which includes<br>with Cytotoxicity Test,<br>Skin Sensitization Test,<br>and Irritation Test. | Cytotoxicity Test:<br>If cell viability is reduced to <<br>70% of the reagent control<br>extract, a cytotoxic potential<br>exists.<br>Skin Sensitization Test:<br>Grades of 1 or greater observed<br>in the test group generally<br>indicated sensitization,<br>provided that grades of less<br>than 1 were observed on the<br>control animals.<br>Irritation Test:<br>For each rabbit, the irritation<br>score for test area was<br>calculated by adding together<br>the scores of erythema and<br>edema at each time point and<br>divide the sum by the total<br>number of observation. | 1. Irritation index for the test<br>result was calculated to be 0.<br>2. There were not found<br>abnormal clinical<br>symptoms except skin<br>reactions in rabbit.<br>3. There were normal weight<br>change.<br>4. The positive/negative rate<br>of all sample extract animal<br>is 0%<br>5. The sample extract showed<br>no significant evidence of<br>causing skin sensitization in<br>the guinea pig.<br>6. Under the conditions of the<br>MTT assay, the test article<br>Breast Pump extract did not<br>show potential toxicity to L-<br>929 cells. |
| ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-12 | Biocompatibility<br>Test Report for<br>Silicon belt | When assessing medical<br>device, the sponsor<br>should specifically state<br>if the medical device<br>does not result in any<br>risk of direct or indirect<br>tissue-contacting<br>components. Thus,<br>performing the<br>biocompatibility test to<br>indicate the safety of<br>device, which includes<br>with Cytotoxicity Test,<br>Skin Sensitization Test,<br>and Irritation Test. | Cytotoxicity Test:<br>If cell viability is reduced to <<br>70% of the reagent control<br>extract, a cytotoxic potential<br>exists.<br>Skin Sensitization Test:<br>Grades of 1 or greater observed<br>in the test group generally<br>indicated sensitization,<br>provided that grades of less<br>than 1 were observed on the<br>control animals.<br>Irritation Test:<br>For each rabbit, the irritation<br>score for test area was<br>calculated by adding together<br>the scores of erythema and<br>edema at each time point and<br>divide the sum by the total<br>number of observation. | 1. Irritation index for the test<br>result was calculated to be 0.<br>2. There were not found<br>abnormal clinical<br>symptoms except skin<br>reactions in rabbit.<br>3. There were normal weight<br>change.<br>4. The positive/negative rate<br>of all sample extract animal<br>is 0%<br>5. The sample extract showed<br>no significant evidence of<br>causing skin sensitization in<br>the guinea pig.<br>6. Under the conditions of the<br>MTT assay, the test article<br>Breast Pump extract did not<br>show potential toxicity to L-<br>929 cells. |
{10}------------------------------------------------
## 8. Conclusions
#### The FORA NAS100 Electronic Nasal Aspirator (Electronic Nasal Aspirator),
NAS100 (TD-7601) is designed to help users remove the nasal mucus in their children (age 2-12 years old) in a home environment.
The proposed device (K180863) has been designed to have smaller dimensions, which makes it more convenient to operate.
Another advantage of the proposed device (K180863) is a lower noise level, which decreases hearing discomfort during the device use.
Also, the NAS100 (TD-7601) has been evaluated according to IEC 60601-1-2, IEC60601-11, IEC 60601-1, and ISO 10993. The performance testing demonstrated that the NAS100 (TD-7601) is substantially equivalent to the legally marketed predicate device (K090379).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.