← Product Code [CAC](/productcode/CAC) · K180831

# CATSmart (K180831)

_Fresenius Kabi AG · CAC · Dec 10, 2018 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K180831

## Device Facts

- **Applicant:** Fresenius Kabi AG
- **Product Code:** [CAC](/productcode/CAC.md)
- **Decision Date:** Dec 10, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5830
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The CATSmart System by Fresenius Kabi is an autotransfusion device indicated for the processing of autologous shed blood collected intraoperatively and postoperatively to obtain washed packed red blood cells for reinfusion. Additionally, it can be used for perioperative separation of blood into Packed Red Cells (PRC), Plasma (PLS) and Platelet Rich Plasma (PRP).

## Device Story

CATSmart is an automated continuous flow centrifuge system for intraoperative and postoperative autotransfusion. It processes anticoagulated shed blood collected in a sterile reservoir; performs continuous washing to remove plasma, non-erythrocytic components, activated coagulation factors, fibrinolysis products, and anticoagulants; and collects washed packed red cells in a reinfusion bag. The system also performs Plasma Sequestration (PSQ) to separate blood into Packed Red Cells, Plasma, and Platelet Rich Plasma based on density and particle size in a centrifugal field. Operated by clinical staff in surgical or trauma settings, the device provides processed blood components for patient reinfusion. It utilizes specific disposable tubing sets and includes an integrated hematocrit monitoring feature. The system aims to reduce the need for allogeneic blood transfusion by recovering and processing a patient's own shed blood.

## Clinical Evidence

Bench testing only. In vitro blood quality studies compared the CATSmart device to the predicate C.A.T.S. system. Validation focused on the PSQ procedure using the PSQ set and AT3 tubing set. Results demonstrated that the new PSQ wash program passed validation tests.

## Technological Characteristics

Continuous flow centrifuge; utilizes disposable tubing sets (AT1, AT3, ATS, ATY, ATO, ATR40, ATR120, ATV-70, ATV-180, ATF40, ATF120, ATV-F140, PSQ set). Features integrated hematocrit monitoring. Operates via physical separation of cellular components in a centrifugal field based on density and particle size. Software-controlled wash programs.

## Regulatory Identification

An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.

## Predicate Devices

- C.A.T.S. Continuous Autotransfusion System ([K971274](/device/K971274.md))

## Reference Devices

- C.A.T.S. ([K160735](/device/K160735.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and name on the right. The Department of Health and Human Services logo is a stylized depiction of an eagle. The FDA acronym is in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 10, 2018

Fresenius Kabi AG % Anju Kurian Manager, Regulatory Affairs Fresenius Kabi USA Three Corporate Drive Lake Zurich, Illinois 60047

Re: K180831

Trade/Device Name: CATSmart Autotransfusion System Regulation Number: 21 CFR 868.5830 Regulation Name: Autotransfusion apparatus Regulatory Class: Class II Product Code: CAC Dated: October 26, 2018 Received: October 29, 2018

Dear Anju Kurian:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Fernando Aguel Fernando Aguel -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K180831

Device Name CATSmart

Indications for Use (Describe)

The CATSmart System by Fresenius Kabi is an autotransfusion device indicated for the processing of autologous shed blood collected intraoperatively and postoperatively to obtain washed packed red blood cells for reinfusion. Additionally, it can be used for perioperative separation of blood into Packed Red Cells (PRC), Plasma (PLS) and Platelet Rich Plasma (PRP).

| Type of Use (Select one or both, as applicable) |                                              |
|-------------------------------------------------|----------------------------------------------|
| Prescription Use (Part 21 CER 801 Subnart D)    | Over-The-Counter Use (21 C.FR 801 Submart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Fresenius Kabi logo. The logo consists of three blue wavy lines on the left and the words "FRESENIUS KABI" in blue on the right. Below the words "FRESENIUS KABI" is the phrase "caring for life" in a smaller blue font.

# 510(K) SUMMARY

Date Prepared: March 29, 2018

#### I. Submitter

Fresenius Kabi 3 Corporate Drive Lake Zurich, IL 60047

Contact Person: Barry G. Hicks Manager, Regulatory Affairs Phone: 847-550-7981 Fax: 847-550-2690 Email: barry.hicks(@fresenius-kabi.com

#### II. Device

| Trade Name:                | CATSmart                                                                                                                      |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Common or Usual Name:      | Automated Blood Processing Autotransfusion System                                                                             |
| Product Code:              | CAC                                                                                                                           |
| Classification Regulation: | 21 CFR § 868.5830                                                                                                             |
| Classification Name:       | Apparatus, Autotransfusion                                                                                                    |
| Regulation Description:    | An autotransfusion apparatus is a device used to collect and<br>reinfuse the blood lost by a patient due to surgery or trauma |
| Review Panel:              | Anesthesiology                                                                                                                |
| Device Class:              | Class II                                                                                                                      |

#### III. Predicate Device

Trade Name: Common or Usual Name: 510(k) Number: Date Cleared: Product Code: Device Class:

C.A.T.S. Continuous Autotransfusion System Automated Blood Processing Autotranfusion System K971274 June 13, 1997 CAC Class II

# IV. Device Description

The Fresenius Kabi CATSmart device is an intraoperative autotransfusion system for intra- and/or postoperative processing of blood lost through surgery or trauma. This version of CATSmart is also capable of perioperative separation of blood into PRP and PLS.

#### Fresenius Kabi USA, LLC

Three Corporate Drive Lake Zurich, Illinois 60047 T 847-550-2300 T 888-391-6300 www.fresenius-kabi.us

{4}------------------------------------------------

The CATSmart device operates on the principle of a continuous flow centrifuge, comparable to continuous systems for hemapheresis which for decades, have been widely used in blood banks.

In a typical CATSmart procedure, the shed blood, which is anticoagulated and collected in a sterile reservoir, is processed in a continuous washing process to obtain washed packed red cells for reinfusion to the patient. During this process all plasmatic and non-erythrocytic cellular components of the collected blood, and thus activated coagulation factors, products of fibrinolysis and cell trauma as well as the anticoagulant are removed. The packed red cells are collected in a reinfusion bag from which they can be reinfused to the patient via a transfusion set when needed.

In the Plasma Sequestration (PSQ) procedure, the patient's blood is separated into packed red cells (PRCs), plasma (PLS) and platelet rich plasma (PRP). The principle of separation during plasma sequestration is the same as it is for autotransfusion i.e. physical separation of cellular components in the centrifugal field based on the differences in density and particle size. There is no change to system hardware or disposable sets described previously in K160735.

The system includes the disposables AT1. AT3. ATS. ATY. ATO. ATR40 and ATR120. ATV-70 and ATV-180, ATF40 and ATF120. An ATV-F140 disposable set has been created (Class I 510(k) exempt). Also, PSQ disposable cleared under K971274 has also been qualified for use with the CATSmart System for the PSQ procedure.

An additional wash program has been developed and added to the software to manage the PSO procedure.

# V. Indications for Use

The CATSmart System by Fresenius Kabi is an autotransfusion device indicated for the processing of autologous shed blood collected intraoperatively and postoperatively to obtain washed red blood cells for reinfusion.

Additionally, it can be used for perioperative separation of blood into Packed Red Cells (PRC), Plasma (PLS) and Platelet Rich Plasma (PRP).

The indications for use of the CATSmart System, subject of this 510(k), is the same as the predicate device C.A.T.S Autotransfusion System cleared under K971274.

# VI. Comparison of Technological Characteristics with the Predicate Device

The technological characteristics of the CATSmart device used for PSQ procedure remains the same as the predicate C.A.T.S device. The only functional difference of CATSmart compared to C.A.T.S./C.A.T.S?140 is the monitoring of the hematocrit value as an additional technical feature. This feature was evaluated and cleared under K160735.

# VII. Performance Data

The performance of the CATSmart device was tested by in vitro blood quality studies in direct comparison to the predicate device C.A.T.S. (currently marketed as C.A.T.S.MM). The focus of the validation study was the PSQ procedure using CATSmart. Overall it was judged that the validation tests passed for the new PSO wash program on the CATSmart device with the PSO set and AT3 tubing set.

{5}------------------------------------------------

#### VIII. Conclusions

The performance testing demonstrates that CATSmart is substantially equivalent to the predicate device C.A.T.S Autotransfusion System.

---

**Source:** [https://fda-staging.innolitics.com/device/K180831](https://fda-staging.innolitics.com/device/K180831)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
