← Product Code [GBM](/productcode/GBM) · K180824

# VaPro 2 Intermittent Catheter (K180824)

_Hollister Incorporated · GBM · Jun 22, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K180824

## Device Facts

- **Applicant:** Hollister Incorporated
- **Product Code:** [GBM](/productcode/GBM.md)
- **Decision Date:** Jun 22, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5130
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

This intermittent catheter is a flexible tubular device that is inserted through the urethra by male, female and pediatric patients who need to drain urine from the bladder.

## Device Story

VaPro 2 Intermittent Catheter is a single-use, hydrophilic-coated, phthalate-free PVC catheter designed for urinary incontinence management. Device features include two smooth eyelets, a protective sleeve for no-touch insertion, a protective introducer tip to shield against distal urethral bacteria, and a color-coded funnel. Hydration occurs via a sealed water compartment within the packaging that migrates to the catheter. Available in standard and pocket packaging, the latter designed for discretion and storage; both facilitate access for users with limited dexterity. The device is used by patients for self-catheterization to drain urine from the bladder.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including physical performance properties (EN 1618, EN 1616, EN 13868), biocompatibility (ISO 10993-1, -11, -12), and sterilization validation (ISO 11137-1, -2; AAMI/ANSI/ISO 11737-1, -2).

## Technological Characteristics

Hydrophilic-coated (PVP-based) phthalate-free PVC catheter; includes protective sleeve and introducer tip; color-coded funnel; gamma irradiation sterilization. Complies with EN 1618, EN 1616, EN 13868, ISO 10993, and ISO 11137 standards.

## Regulatory Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

## Predicate Devices

- VaPro Intermittent Catheter ([K141642](/device/K141642.md))

## Reference Devices

- VaPro Pocket ([K143594](/device/K143594.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure. The FDA part of the logo has the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 22, 2018

Hollister Incorporated Michelle Schiltz-Taing Regulatory Affairs Manager 2000 Hollister Drive Libertyville, IL 60048

Re: K180824

> Trade/Device Name: Vapro 2 Intermittent Catheter Regulation Number: 21 CFR& 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: GBM Dated: June 8, 2018 Received: June 11, 2018

Dear Michelle Schiltz-Taing:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Timothy Martin -S 2018.06.22 08:48:43 -04'00'

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration            |
| Indications for Use                     |

| Form Approved: OMB No. 0910-0120 |
|----------------------------------|
| Expiration Date: 06/30/2020      |
| See PRA Statement below.         |

| 510(k) Number (if known)       | K180824                                                                                                                                                                       |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                    | Vapro 2 Intermittent Catheter                                                                                                                                                 |
| Indications for Use (Describe) | This intermittent catheter is a flexible tubular device that is inserted through the urethra by male, female and pediatric patients who need to drain urine from the bladder. |

| Type of Use (Select one or both, as applicable) | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

| FORM FDA 3881 (7/17)                      | Page 1 of 2 |
|-------------------------------------------|-------------|
| PSC Publishing Services (301) 443-6740 EF |             |

TET TO BS ADS A

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the Hollister logo. The logo consists of a red, stylized cross-like symbol on the left and the word "Hollister." in gray on the right. The logo is simple and modern.

# 510(k) Summary

| Applicant:                                                                                   | Hollister Incorporated<br>2000 Hollister Drive<br>Libertyville, IL 60048                                                     |
|----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                                                                              | Michelle Schiltz-Taing<br>Hollister Incorporated<br>2000 Hollister Drive<br>Libertyville, IL 60018<br>(t) 847-680-2122       |
| Date Prepared:<br>Trade Name:<br>Common Name:<br>Product Code/Class:<br>Classification Name: | 05 June 2018<br>VaPro 2TM Intermittent Catheter<br>Catheter, Urethral<br>GBM/Class II<br>Urological catheter and accessories |

#### Predicate Device:

CFR:

VaPro 2 Intermittent Catheter is substantially equivalent to its original design: VaPro Intermittent Catheter, K141642 by Hollister Incorporated

21 CRF 876.5130

## Reference Device:

VaPro Pocket, K143594 by Hollister Incorporated for VaPro 2 Pocket.

#### Indications for Use:

This intermittent catheter is a flexible tubular device that is inserted through the urethra by male, female and pediatric patients who need to drain urine from the bladder.

#### Description of Applicant Device:

The VaPro 2 Intermittent Catheter is a hydrophilic coated, single use catheter to be used as a means of managing urinary incontinence by draining urine from the bladder. VaPro 2 has the following features:

- Hydrophilic-coated PVC catheter (phthalate free) ●
- Two smooth catheter eyelets ●
- Protective sleeve ●

{4}------------------------------------------------

- Protective introducer tip as a way to shield the sterile catheter from bacteria in the distal ● urethra during insertion
- Color-coded funnel ●
- Available with a Firmer Style Catheter (Standard packaging only with packaging differentiated as F-Style)

The packaging contains a sealed water compartment chamber from which the water migrates to the catheter compartment and hydrates (lubricates) the catheter. The outer packaging, (Standard and Pocket), was designed to facilitate access for those with limited dexterity. The Pocket version was designed to be discreet and easy to store.

#### Technological Characteristics:

The table below summarizes the technological characteristics of VaPro 2 Intermittent Catheter as compared to the predicate device VaPro.

|                           | VaPro Intermittent Catheter<br>(K141642)                                                                                                                                      | VaPro 2                                  |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| Indication for Use        | This intermittent catheter is a flexible tubular device that is inserted through the urethra by male, female and pediatric patients who need to drain urine from the bladder. |                                          |
| Condition of Use          |                                                                                                                                                                               | Single Use                               |
| Pre-lubricated            |                                                                                                                                                                               | Yes-by water vapor hydration             |
| Ready to use              |                                                                                                                                                                               | Yes                                      |
| End Design                |                                                                                                                                                                               | Funnel                                   |
| Sterile                   |                                                                                                                                                                               | Yes - Gamma Irradiation                  |
| No touch design           |                                                                                                                                                                               | Yes - contains sleeve                    |
| Hydrophilic Coating       |                                                                                                                                                                               | PVP Based (polyvinylpyrollidone) Coating |
| Protective Introducer Tip |                                                                                                                                                                               | Yes                                      |
| Protective Cap            |                                                                                                                                                                               | Yes                                      |
| Catheter Material         |                                                                                                                                                                               | PVC (phthalate free)                     |

#### Brief Description of Non-Clinical Testing:

The physical performance properties of VaPro 2 met all applicable requirements of EN 1618, EN 1616 and EN 13868.

Biocompatibility testing met the requirements of ISO 10993-1. 10993-11 and 10993-12.

Sterilization met all requirements of ISO 11137-1, ISO 11137-2, AAMI/ANSI/ISO 11737-1 and AAMI/ANSI/ISO 11737-2.

## Conclusion:

It is concluded that the information supplied in this submission has demonstrated that VaPro 2 Intermittent Catheter is equivalent to the legally marketed device VaPro Intermittent Catheter.

---

**Source:** [https://fda-staging.innolitics.com/device/K180824](https://fda-staging.innolitics.com/device/K180824)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com