PURE EP™ is a computerized system intended for acquiring, digitizing, filtering, measuring and calculating, displaying, recording and storing of electrocardiographic and intracardiac signals for patients undergoing electrophysiology (EP) procedures in an EP laboratory under the supervision of licensed healthcare practitioners who are responsible for interpreting the data.
Device Story
PURE EP is a computerized system for electrophysiology (EP) labs; acquires, digitizes, filters, measures, and records electrocardiographic (ECG) and intracardiac (IC) signals. System comprises an amplifier and a personal computer (PC) connected via high-speed fiber optic cable. Patient connects to amplifier via ECG cables and standard IC catheters. Amplifier includes hardware and embedded software for signal acquisition (12-lead ECG and 56 IC signals). PC runs PURE-EP software for real-time visualization and review. Healthcare practitioners use the system to interpret cardiac signals during EP procedures. Output aids in clinical decision-making by providing processed, filtered, and measured cardiac data. Benefits include enhanced signal visualization and data management for diagnostic EP procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including software, electrical, and mechanical verification against industry standards (IEC 60601-1, IEC 62304, etc.).
Technological Characteristics
System consists of an amplifier and PC connected via fiber optic cable. Features 24-bit A/D converter, 2000 samples/sec sampling rate, and up to 64 channels. Software-programmable high-pass and low-pass filters for ECG and IC signals. PC hardware: Intel i7 quad-core, 16 GB RAM, 1 TB storage, Windows 7 OS. Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, and IEC 62304 standards.
Indications for Use
Indicated for patients undergoing electrophysiology (EP) procedures in an EP laboratory. Intended for use under the supervision of licensed healthcare practitioners.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size.
August 8, 2018
BioSig Technologies Tiffini Wittwer Regulatory 12424 Wilshire Blvd. Suite 745 Los Angeles, California 90025
Re: K180805
Trade/Device Name: Pure EP Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DOK Dated: June 13, 2018 Received: June 21, 2018
Dear Tiffini Wittwer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
M.A. Wilhelm
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K180805
Device Name PURE EPTM
Indications for Use (Describe)
PURE EP™ is a computerized system intended for acquiring, digitizing, filtering, measuring and calculating, displaying, recording and storing of electrocardiographic and intracardiac signals for patients undergoing electrophysiology (EP) procedures in an EP laboratory under the supervision of licensed healthcare practitioners who are responsible for interpreting the data.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 16px;">☑</span> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span style="font-size: 16px;">☐</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary K180805
### Table 1: 510(k) Summary
| Submitter: | Biosig Technologies, Inc.<br>12424 Wilshire Blvd, Suite 745<br>Los Angeles, CA 90025 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Tiffini Diage<br>Regulatory Affairs Consultant<br>Phone: 707.799.6732<br>E-mail: tdiage@raechelon.com |
| Date Of Submission: | March 26, 2018 |
| Trade Name: | PURE EP™ |
| Common Name: | Programmable diagnostic computer / electrophysiology system |
| Classification: | Class II, per 21 CFR 870.1425 |
| Product Code: | DQK |
| Predicate Device(s): | The subject device is equivalent to the following devices:<br>• K130626 GE Healthcare; CardioLab (other systems included in cleared 510(k) application) |
| Device Description: | PURE EP™ is a system consists of two main parts: the amplifier, and the personal computer (PC), which are connected over a high-speed fiber optic cable. The patient is connected to the amplifier via a set of supplied ECG cables and commercially available intracardiac (IC) catheters. The amplifier includes hardware and embedded software necessary for acquiring ECG and IC signals from patients. The device is not intended for active patient monitoring. It consists of one ECG and seven IC modules designed to acquire 12-lead ECG and 56 IC signals and send the digitized data to the PC. The PC has preinstalled PURE-EP software and is connected to display monitors. The pre-installed software provides visualization of received data from the amplifier on real-time and review screens. It also includes a signal processing module that provides various filter options for real-time and review screens. |
| Indication for Use: | PURE EP™ is a computerized system intended for acquiring, digitizing, amplifying, filtering, measuring and calculating, displaying, recording and storing of electrocardiographic and intracardiac signals for patients undergoing electrophysiology (EP) procedures in an EP laboratory under the supervision of licensed healthcare practitioners who are responsible for interpreting the data. |
{4}------------------------------------------------
| Item | Predicate Device GE CardioLab | Subject Device PURE EPTM |
|--------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | K130626 | K180805 |
| Class | II | II |
| Product Code | DQK | DQK |
| Intended Use and<br>Indications for Use (IFU) | CardioLab System is intended for the<br>acquiring, filtering, digitizing, simplifying,<br>measuring and calculating, displaying,<br>recording and monitoring of clinical data<br>from adult and pediatric patients.<br>CardioLab System is configurable. Clinical<br>data includes: ECG waveforms,<br>intracardiac signals, stimulus data, ablation<br>data, pulse oximetry (SpO2), respiration<br>rate, CO2 (EtCO2),<br>temperature, and invasive and noninvasive<br>blood pressure. Physiological parameters<br>such as diastolic, systolic, mean pressure,<br>heart rate, and cycle length are derived<br>form the signal data, displayed, and<br>recorded. The data is entered manually or<br>acquired via interfaced devices and/or<br>information systems and may be used for<br>report<br>generation. | PURE EP is a computerized system<br>intended for acquiring, digitizing,<br>amplifying, filtering, measuring and<br>calculating, displaying, recording and<br>storing of electrocardiographic and<br>intracardiac signals for patients<br>undergoing electrophysiology (EP)<br>procedures in an EP laboratory under the<br>supervision of licensed healthcare<br>practitioners who are responsible for<br>interpreting the data. |
| Target Population | Clinical data from adult and pediatric<br>patients. | Patients undergoing EP studies in an EP<br>lab. |
| Anatomical Site | Multiple, including cardiac signals | Cardiac signals |
| Where used | Multiple hospital and clinical settings<br>including EP labs | EP labs |
| Energy<br>Used/Delivered | No energy delivered to patient | No energy delivered to patient |
| Processor/Data<br>Storage | Intel Pentium IV Xeon or greater<br>processor<br>512 MB of Ram<br>2 x 40 GB hard drives<br>OS: Windows XP Professional | Intel i7 quad-core CPU<br>16 GB of Ram<br>1 TB<br>OS: Windows 7 |
| Item | Predicate Device GE CardioLab | Subject Device PURE EPTM |
| Number of Channels | 32/64<br>96/128 | Up to 64 |
| Sampling Rate | 2000 samples per second | 2000 samples per second |
| Analog to Digital<br>Converter | 12 bit | 24 bit |
| Outputs | 12 lead ECG | Same |
| RF Filtering | All inputs | Same |
| Bandwidth | 0.05-1000 Hz | Same |
| ECG high pass<br>filter | 0.05 Hz, 0.5 Hz, 5.0 Hz | Software programmable at: 0.05 Hz, 0.5<br>Hz, 1.0<br>Hz, 5.0 Hz, 20 Hz, 30 Hz,<br>50 Hz, 200 Hz |
| ECG low pass filter | 100 Hz | Software programmable at: 30 Hz, 40<br>Hz, 50 Hz,<br>100 Hz, 150 Hz, 300 Hz,<br>500 Hz; |
| IC high pass filter | 0.05 Hz, 0.5 Hz, 5.0 Hz, 30<br>Hz, 100 Hz | Software programmable 0.05 Hz, 0.5 Hz,<br>1.0 Hz,<br>5.0 Hz, 20 Hz, 30 Hz, 50<br>Hz, 200 Hz |
| IC low pass filter | 150 Hz, 500 Hz, 1000 Hz; | Software programmable at: 30 Hz, 40<br>Hz, 50 Hz,<br>100 Hz, 150 Hz, 300 Hz,<br>500 Hz; |
| Notch Filter | 50/60 Hz | 60 Hz |
| Band Pass Filter | Not available | User selected combination of High Pass,<br>Low Pass and Notch Filter |
| Type CF | 0 -.5 Amps Class I | 0 -.8 Amps Class I |
| Operating Temp<br>range | 0 to 35° C (32/64 channel;) | 10-40° C |
| Temp<br>storage/transport | -15 to 50 C non- condensing | 20-65° C per ISTA 2A<br>2008 |
| Humidity<br>Operating | 95% at 35 C non- condensing | 30-75% non-condensing |
| Humidity<br>Transport/Storage | 95% at 35 C non- condensing | 20-85% per ISTA 2A<br>2008 |
| Patient Source | <10 uA | <10 uA |
| Item | Predicate Device GE CardioLab | Subject Device PURE EPTM |
| Patient Sink | <10 uA | <10 uA |
| Patient Sink (measured<br>at patient leads under<br>single fault<br>condition) | <50 uA | <50 uA |
| Chassis Leakage | <100uA | <100uA |
| Size | Height = 9.5 inches Depth = 13 inches<br>Width = 13 inches (32/64 channel<br>version) | Height = 16 inches Depth = 14.7<br>inches Width = 13 inches |
| Weight | 25 lbs<br>(32/64 channel version) | 38 lbs |
#### Substantial Equivalence Comparison
{5}------------------------------------------------
{6}------------------------------------------------
#### Performance Testing / Safety and Effectiveness:
To verify that device design meets its functional and performance requirements, representative samples of the device underwent software, electrical, and mechanical testing in accordance with the following industry standards.
- ANSI/AAMI/IEC 62366-1:2015; Medical Devices Part 1: Application of Usability ● Engineering to Medical Devices.
- AAMI/ANSI/ISO 14971:2007/(R)2010 Medical devices -- Application of risk management to medical devices.
- IEC 60601-1-6: 2013, Medical Electrical Equipment Part 1-6: General Requirements ● For Basic Safety And Essential Performance - Collateral Standard: Usability
- AAMI/ANSI HE75: 2013, Human Factors Engineering - Design of Medical Devices..
- IEC 62304: 2015, Medical Device Software - Software Life Cycle Processes
- AAMI/ANSI ES60601-1: 2012 Medical Electrical Equipment Part 1: General ● Requirements For Basic Safety And Essential Performance
- AAMI/ANSI IEC 60601-1-2:2014 Medical Electrical Equipment Part 1-2: General ● Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests.
- Guidance for Industry and FDA Staff: ●
- Guidance for the Content of Premarket Submissions for Software Contained in o Medical Devices (Issued 5/11/2005)
- Content of Premarket Submissions for Management of Cybersecurity in Medical o Devices (Issued 10/2/2014)
{7}------------------------------------------------
Summary of Clinical Tests:
Similar to the predicate, the subject of this premarket submission, PURE EPIM, did not require clinical studies to support substantial equivalence.
Conclusion:
PURE EPTM intended use, indication for use, anatomical sites, and fundamental technology is equivalent to those of the predicate device. Both devices acquire ECG and IC signals via instrument amplification. Both devices use A/D converters and send data to a host PC. Software, electrical safety, system performance testing and user validation demonstrates that the differences between the subject device and the predicate do not introduce any new questions of safety and effectiveness. The information demonstrates that the subject device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.