K180797 · Curatia Medical Co. · DQY · Dec 14, 2018 · Cardiovascular
Device Facts
Record ID
K180797
Device Name
Xcess Guiding Catheter
Applicant
Curatia Medical Co.
Product Code
DQY · Cardiovascular
Decision Date
Dec 14, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The guiding catheter is intended to use for intravascular introduction of interventional/diagnostic devices into the coronary or peripheral vascular systems.
Device Story
Xcess Guiding Catheter is a percutaneous catheter used to facilitate the introduction of interventional or diagnostic devices into coronary or peripheral vasculature. The device consists of a multi-layered shaft (Nylon/Pebax, braided stainless steel, PTFE), soft extension, soft tip, strain relief, and luer connector. It is available in 5F and 6F sizes with various distal curve shapes to accommodate different vascular anatomies. The catheter body features a hydrophilic coating. It is operated by clinicians in a clinical setting to provide a pathway for guide wires (up to 0.038" diameter) and other interventional tools. The device aids in navigating the vascular system to reach target sites, supporting diagnostic or therapeutic procedures.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing and biocompatibility evaluation. Bench testing included visual/dimensional inspection, coating integrity, particulate evaluation, usability, shape retention, leakage, torque, pressure integrity, flexibility/kink, tensile strength, corrosion resistance, and radio-detectability. Biocompatibility testing included cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, and hemocompatibility (hemolysis, thrombogenicity, complement activation). All tests met pre-defined acceptance criteria.
Technological Characteristics
Multi-layered catheter shaft: outer Nylon/Pebax, middle braided stainless steel, inner PTFE. Soft tip/extension: outer Nylon/Pebax, inner PTFE. Hydrophilic coating. Sizes: 5F, 6F. Max guide wire: 0.038". Sterilization: Ethylene oxide (ISO 11135). Compliance: ISO 10555-1, ISO 10993 series.
Indications for Use
Indicated for patients requiring intravascular introduction of interventional or diagnostic devices into the coronary or peripheral vascular systems.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 14, 2018
Curatia Medical Co. % Breanne Butler Regulatory Affairs Consultant JCO Consulting 11218 Zest Court NE Blaine, Minnesota 55449
Re: K180797
Trade/Device Name: Xcess Guiding Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: November 15, 2018 Received: November 15, 2018
Dear Breanne Butler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Lydia S. Digitally signed by Lydia S. Glaw -S Date: 2018.12.14 Glaw -S 15:12:17 -05'00'
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180797
Device Name Xcess Guiding Catheter
Indications for Use (Describe)
The guiding catheter is intended to use for introduction of interventional/diagnostic devices into the coronary or peripheral vascular systems.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## 510(k) SUMMARY K180797
| Submitter: | Curatia Medical Co.<br>3227 Kifer Road<br>Santa Clara, CA 95051<br>Establishment Number: 3008361782 |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact Person:<br>Phone:<br>Fax:<br>Email: | Jessica Chiu, VP<br>(408) 414-2188<br>(408) 413-3000<br>jchiu@curatiamed.com |
| Submission Correspondent:<br>Address:<br>Phone:<br>Email: | Breanne Butler, Regulatory Affairs Consultant<br>11218 Zest Ct. NE, Blaine, MN 55449<br>(860)-810-5594<br>breanne.sm.butler@gmail.com |
| Date Prepared: | December 13, 2018 |
| Proprietary Name: | Xcess Guiding Catheter |
| Common Name: | Percutaneous catheter |
| Product Code: | DQY |
| Device Classification: | Class II, 21 CFR 870.1250 - Percutaneous Catheter |
| Predicate Devices: | VISTA BRITE TIP Guiding Catheter (K962830) |
#### Device Description:
The Xcess guiding catheter is available in many different curve shapes and two sizes (5F and 6F). Each device consists of a catheter shaft, soft extension, soft tip, strain relief and luer. The catheter shaft is constructed with 3 layers: an outer layer of radiopaque Nylon/Pebax material, a middle layer of braided stainless steel, and an inner layer of PTFE tubing. The soft extension and soft tip consist of 2 layers: an outer layer of a radiopaque Nylon/Pebax material of differing hardness than the layer in the catheter shaft and an inner layer of PTFE. The strain relief at the proximal end of the catheter prevents kinking during catheter manipulation, and the luer is used to connect with either a hemostasis valve-linked accessory or 3-way connector. The distal ends of the guiding catheters are formed into a variety of shapes required to access differing vasculatures. The guiding catheters are provided with a hydrophilic coating on the catheter body. The guiding catheters are designed to accept a maximum of 0.038" (0.97mm) diameter guide wire.
{4}------------------------------------------------
## Indications for Use:
The guiding catheter is intended to use for intravascular introduction of interventional/diagnostic devices into the coronary or peripheral vascular systems.
### Comparison to Predicate Devices:
The Xcess Guiding Catheter is substantially equivalent in design, construction, and performance characteristics to the following commercially available product: VISTA BRITE TIP Guiding Catheters (CORDIS - FDA registration No: K962830, 1996). Based on comparison and testing of the predicated device and the Xcess Guiding Catheter, the specifications of the Xcess Guiding Catheter (Curatia) are substantially equivalent to the predicate device, VISTA BRITE TIP Guiding Catheter (Cordis). The clinical, technical, material, and biological specifications of the Xcess Guiding Catheter are substantially equivalent to the predicate device.
## Summary of Performance Data and Substantial Equivalence:
The Xcess Guiding Catheters were designed and verified in accordance with the risk analysis and product requirements. All tests confirmed the products met the pre-defined acceptance criteria. Curatia Medical has determined that the Xcess Guiding Catheters are substantially equivalent. The Xcess Guiding Catheters have been tested and shown to be compliant with the following standards documents:
- . ISO 10555-1:2013- Intravascular catheters – Sterile, single-use intravascular catheters – Part 1: General requirements
- ISO 10993-1:2009- Biological evaluation of medical devices – Part 1: Evaluation and testing
- ISO 10993-3:2014- Biological evaluation of medical devices - Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
- . ISO 10993-4:2002+AMD1:2006-Biological testing of medical and dental materials and devices -Part 4: Selection of tests for interactions with blood
- ISO 10993-5: 2009- Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin ● sensitization
- ISO 10993-11:2006- Biological testing of medical and dental materials and devices – Part 11: Tests for systemic toxicity
- . ISO 10993-12:2012- Biological testing of medical and dental materials and devices – Part 12: Sample Preparation And Reference Materials
- . ISO 11135: 2014 Sterilization of health-care products-Ethylene oxide-Requirements for the development, validation and routine control of a sterilization process for medical devices
- EN ISO 11607-1: 2009(R2014)- Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
#### Performance Testing:
To demonstrate substantial equivalence of the Xcess Guiding Catheter to the predicate device, the technological characteristics and performance criteria were evaluated using the bench testing recommendations outlined in the ISO 10555-1 Sterile, single-use intravascular catheters-Part 1: General requirement. The following performance tests were completed:
- . Visual and dimension inspection
- Coating integrity and particulate evaluation ●
- . Catheter usability
{5}------------------------------------------------
- Catheter shape retention ●
- Catheter freedom from leakage ●
- Catheter torque
- . Catheter pressure integrity
- Catheter flexibility and kink ●
- Catheter joints tensile strength ●
- Corrosion resistance
- . Radio-detectability
The results of these tests demonstrate that the technological characteristics and performance criteria of the Xcess Guiding Catheter are adequate for its intended use, and that is substantially equivalent to the predicate device.
## Biocompatibility:
To demonstrate the biocompatibility of the materials in contact with the body and substantial equivalence of the Xcess Guiding Catheter to its predicate device, the following biocompatibility testing was performed in accordance with ISO 10993 and the FDA Guidance Document "Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing"" dated June 2016:
- . Cytotoxicity
- . Sensitization
- Irritation or Intracutaneous Reactivity ●
- Acute Systemic Toxicity ●
- Material-Mediated Pyrogenicity
- Hemocompatibility (Hemolysis, Thrombogenicity, and Complement Activation)
The results from these tests demonstrate that the Xcess Guiding Catheter is biocompatible for its intended use.
#### Conclusion:
Based on comparison of indications for use, technological characteristics, performance testing, biocompatibility testing, the Xcess Guiding Catheter has been shown to be appropriate for its indications for use and is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.