Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs Aqua Blue (ABLU), Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs Violet Blue (VBLU) - Extended Cuff

K180786 · Hartalega Sdn Bhd · LZA · Nov 2, 2018 · General Hospital

Device Facts

Record IDK180786
Device NameSterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs Aqua Blue (ABLU), Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs Violet Blue (VBLU) - Extended Cuff
ApplicantHartalega Sdn Bhd
Product CodeLZA · General Hospital
Decision DateNov 2, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

The Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs - Violet Blue (VBLU)-Extended Cuff is a sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs. The Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs - Aqua Blue (ABLU) is a sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.

Device Story

Sterile, powder-free nitrile examination gloves; intended for use by healthcare professionals to prevent cross-contamination between patient and examiner. Device provides a physical barrier; tested for resistance to permeation by specific chemotherapy drugs (e.g., Carmustine, Cisplatin, Cyclophosphamide, etc.). Used in clinical settings; worn on examiner's hands. Provides protection against patient-to-examiner and examiner-to-patient contamination; chemotherapy testing informs safe handling duration for hazardous drugs.

Clinical Evidence

No clinical data; bench testing only. Permeation testing performed for various chemotherapy drugs per ASTM standards to determine breakthrough times.

Technological Characteristics

Nitrile rubber material; powder-free; sterile; disposable. Form factor includes extended cuff for VBLU model. Barrier protection principle. No software or electronic components.

Indications for Use

Indicated for use by medical examiners to prevent cross-contamination between patient and examiner. Tested for use with specific chemotherapy drugs.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. November 2, 2018 Hartalega Sdn. Bhd. Nurual Kong Ouality Assurance Senior Manager No. 7. Kawasan Perusahaan Suria Bestari Jaya, 45600 My Re: K180786 Trade/Device Name: Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs Aqua Blue (ABLU), Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs Violet Blue (VBLU) - Extended Cuff Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: October 11, 2018 Received: October 15, 2018 Dear Nurual Kong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Clarence W. Murray lii III -S For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) K180786 #### Device Name Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs - Violet Blue (VBLU)- Extended Cuff ### Indications for Use (Describe) The Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs - Violet Blue (VBLU)-Extended Cuff is a sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs. The list of Chemotherapy Drugs tested (with breakthrough times) is as below: | Carmustine (3.3 mg/ml) | 20.3 minutes | |--------------------------------------|---------------| | Cisplatin (1.0 mg/ml) | > 240 minutes | | Cyclophosphamide (20 mg/ml) | > 240 minutes | | Dacarbazine (10.0 mg/ml) | > 240 minutes | | Doxorubicin Hydrochloride (2.0mg/ml) | > 240 minutes | | Etoposide (20.0 mg/ml) | > 240 minutes | | Fluorouracil (50.0 mg/ml) | > 240 minutes | | Methotrexate (25 mg/ml) | > 240 minutes | | Mitomycin C (0.5 mg/ml) | > 240 minutes | | Paclitaxel (6.0 mg/ml) | > 240 minutes | | Thiotepa (10.0 mg/ml) | 100.6 minutes | | Vincristine Sulfate (1.0 mg/ml) | > 240 minutes | Please note that Carmustine and Thiotepa have extremely low permeation times of 20.3 minutes and 100.6 minutes. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) |× Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) K180786 #### Device Name Sterile Nitrile Powder Free Examination Gloye Tested for Use with Chemotherapy Drugs - Aqua Blue (ABLU) Indications for Use (Describe) The Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs - Aqua Blue (ABLU) is a sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs. The list of Chemotherapy Drugs tested (with breakthrough times) is as below: | Carmustine (3.3 mg/ml) | 43.8 minutes | |--------------------------------------|---------------| | Cisplatin (1.0 mg/ml) | > 240 minutes | | Cyclophosphamide (20 mg/ml) | > 240 minutes | | Dacarbazine (10.0 mg/ml) | > 240 minutes | | Doxorubicin Hydrochloride (2.0mg/ml) | > 240 minutes | | Etoposide (20.0 mg/ml) | > 240 minutes | | Fluorouracil (50.0 mg/ml) | > 240 minutes | | Methotrexate (25 mg/ml) | > 240 minutes | | Mitomycin C (0.5 mg/ml) | > 240 minutes | | Paclitaxel (6.0 mg/ml) | > 240 minutes | | Thiotepa (10.0 mg/ml) | 97.8 minutes | | Vincristine Sulfate (1.0 mg/ml) | > 240 minutes | Please note that Carmustine and Thiotepa have extremely low permeation times of 43.8 minutes. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) × Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...