Cytal Wound Particulate
K180776 · Acell, Inc. · KGN · May 11, 2018 · SU
Device Facts
| Record ID | K180776 |
| Device Name | Cytal Wound Particulate |
| Applicant | Acell, Inc. |
| Product Code | KGN · SU |
| Decision Date | May 11, 2018 |
| Decision | SESE |
| Submission Type | Special |
| Device Class | Class U |
| Attributes | Therapeutic |
Indications for Use
Cytal® Wound Particulate is intended for the management of wounds including: partial and full-thickness wounds, pressure ulcers, venous ulcers, diabetic ulcers, chronic vascular ulcers, tunnel/undermined wounds, surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, lacerations, second-degree burns, and skin tears), and draining wounds. The device is intended for onetime use.
Device Story
Cytal Wound Particulate is a resorbable, porcine-derived extracellular matrix (ECM) scaffold composed of urinary bladder matrix (UBM); supplied as dry, absorbent particulate (<1000µm) in amber glass vials. Applied to wound bed either dry or pre-hydrated with sterile saline; can be used with other sheet-based ECM scaffolds. Device undergoes enzymatic degradation, cellular infiltration, and capillary growth, leading to remodeling into host tissue or sloughing. Used in clinical settings for wound management; intended for one-time use.
Clinical Evidence
Bench testing only. Performance data included particle size characterization, dispensing/deployability testing, sterilization validation, and endotoxin testing. No clinical data presented.
Technological Characteristics
Porcine-derived urinary bladder matrix (UBM) collagen scaffold; dry particulate form (<1000µm); terminally sterilized via electron beam irradiation; packaged in amber glass vials with butyl stoppers.
Indications for Use
Indicated for management of partial/full-thickness wounds, pressure/venous/diabetic/chronic vascular ulcers, tunnel/undermined wounds, surgical wounds (donor sites, grafts, post-Mohs/laser, podiatric, dehiscence), trauma wounds (abrasions, lacerations, second-degree burns, skin tears), and draining wounds.
Predicate Devices
- ACell Inc. Cytal® Wound Matrix (K152721)
Reference Devices
- ACell Inc. MicroMatrix® (K172399)
- Cook Biotech Inc. Cook® ECM Powder (K152033)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 11, 2018
ACell, Inc. c/o John Smith Hogan Lovells 555 Thirteenth St., NW Washington, District of Columbia 20004
Re: K180776
Trade/Device Name: Cytal Wound Particulate Regulatory Class: Unclassified Product Code: KGN Dated: April 13, 2018 Received: April 13, 2018
Dear John Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may: therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180776
Device Name Cytal® Wound Particulate
#### Indications for Use (Describe)
Cytal® Wound Particulate is intended for the management of wounds including: partial and full-thickness wounds, pressure ulcers, venous ulcers, diabetic ulcers, chronic vascular ulcers, tunnel/undermined wounds, surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, lacerations, second-degree burns, and skin tears), and draining wounds. The device is intended for onetime use.
Type of Use (Select one or both, as applicable)
| <span style="font-size: large;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: large;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY ACell Inc.'s Cytal® Wound Particulate
| Submitter: | ACell, Inc.<br>6640 Eli Whitney Drive<br>Columbia, MD 21046 |
|-----------------------------------------|-----------------------------------------------------------------------------------|
| Contact Person:<br>Phone:<br>Facsimile: | Andrea Pilon Artman<br>410-953-8549<br>410-715-4511 |
| Date Prepared: | May 10, 2018 |
| Trade Name: | Cytal® Wound Particulate |
| Common Name: | Animal-Derived, Extracellular Matrix Wound Care Produc |
| Classification Name: | Dressing, Wound, Collagen |
| Regulatory Class: | Unclassified |
| Product Code: | KGN |
| Predicate Device: | ACell Inc. Cytal® Wound Matrix (K152721) |
| Reference Devices: | ACell Inc. MicroMatrix® (K172399)<br>Cook Biotech Inc. Cook® ECM Powder (K152033) |
#### Device Description
Cytal® Wound Particulate is composed of a resorbable, porcine-derived, extracellular matrix scaffold containing epithelial basement membrane, specifically known as Urinary Bladder Matrix (UBM). The devices are supplied as a dry, absorbent, white to off-white particulate sized <1000pm in multi-layer configurations (2 to 8 layers). The particulate is packaged in an amber glass vial with butyl stopper and crimp sealed. The device vial is then packaged in a peel-open outer pouch that is terminally sterilized using electron beam irradiation. Cytal® Wound Particulate can be applied to a wound either in the dry state or pre-hydrated with sterile saline, and can be used in conjunction with other sheet based extracellular matrix derived scaffolds indicated for wound management. In vitro and in vivo studies suggest that the product will be sloughed from the normal wound healing process or will be incorporated (remodeled) into the wound bed via enzymatic degradation, cellular infiltration, capillary growth, and/or integration by the surrounding host tissue. The device is intended for one time use.
#### Intended Use/Indications for Use
Cytal® Wound Particulate is intended for the management of wounds including: partial and fullthickness wounds, pressure ulcers, venous ulcers, chronic vascular ulcers, tunnel/undermined wounds, surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, second-degree burns, and skin tears), and draining wounds. The device is intended for one-time use.
{4}------------------------------------------------
#### Summary of Technological Characteristics
The technological characteristics of the Cytal® Wound Particulate are substantially equivalent to the cleared Cytal® predicate, as both are comprised of the same animal tissue-derived, collagen extracellular matrix (ECM) scaffold material. The devices are manufactured using the same processes except for the addition of a manufacturing process to reduce the size of the wound dressing from a layered sheet configuration to a particulate. Both the subject and predicate devices are terminally sterilized. Updated packaging has been implemented to account for the reduction in device size.
The minor differences between the Cytal® Wound Particulate and the identified predicate do not raise different questions of safety or efficacy, and performance testing demonstrates that the device has comparable performance to the predicate.
#### Performance Data
The following testing was performed to address the size change and to demonstrate substantial equivalence to the predicate device:
- Particle Size Characterization ●
- Dispensing (Deployability) ●
- Sterilization Validation
- Endotoxin
In all instances, the Cytal® Wound Particulate functioned as intended and the results observed were as expected.
### Conclusions
The Cytal® Wound Particulate is as safe and effective as the Cytal® Wound Matrix. The Cytal® Wound Particulate has the same intended uses, indications, and principles of operation as its predicate device. The minor technological differences between the Cytal® Wound Particulate and its predicate device do not raise different questions of safety or effectiveness. Performance data demonstrate that the Cytal® Wound Particulate is as safe and effective as the predicate device.