K180769 · Interrad Medical, Inc. · OKC · Jun 26, 2018 · General Hospital
Device Facts
Record ID
K180769
Device Name
SecurAcath
Applicant
Interrad Medical, Inc.
Product Code
OKC · General Hospital
Decision Date
Jun 26, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5970
Device Class
Class 2
Indications for Use
The SecurAcath device is indicated for short or long term securement of percutaneous indwelling catheters for intravenous use to the access site by means of a subcutaneous anchor.
Device Story
SecurAcath is a single-use, sterile, subcutaneous anchoring system for indwelling catheters. It consists of a base assembly, a cover, and a blunt nitinol anchor. The device is deployed by inserting the nitinol anchor into the subcutaneous space at the catheter access site and clamping the catheter shaft between the base and cover. This mechanism secures the catheter to the skin, reducing migration and accidental pull-out without impeding fluid flow. It is used in clinical settings to manage percutaneous catheters. The device is an accessory to the catheter; it does not process data or provide clinical outputs. Benefits include improved catheter stability and reduced risk of dislodgement.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including dimensional verification, joint tensile strength, base/cover interaction, hinge performance, and catheter securement performance testing. Risk management was conducted per ISO 14971:2012.
Technological Characteristics
Single-use, sterile subcutaneous securement device. Materials are identical to the predicate. Features a blunt nitinol anchor for subcutaneous deployment and a mechanical base/cover clamp. Sterilization via EtO (ISO 11135) to SAL 10^-6. Compatible with 3F-12F catheters. No software or electronic components.
Indications for Use
Indicated for short or long-term securement of percutaneous indwelling intravenous catheters to the access site via a subcutaneous anchor.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 26, 2018
Interrad Medical Inc % Denise Lenz Regulatory Consultant Libra Medical, Inc 8401 73rd Ave North, Suite 63 Brooklyn Park, Minnesota 55428
Re: K180769
Trade/Device Name: SecurAcath Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: Class II Product Code: OKC, KMK Dated: May 21, 2018 Received: May 29, 2018
Dear Denise Lenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180769
Device Name SecurAcath
Indications for Use (Describe)
The SecurAcath device is indicated for short or long term securement of percutaneous indwelling catheters for intravenous use to the access site by means of a subcutaneous anchor.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K180769 510(K) SUMMARY
| Manufacturer's Name: | Interrad, Inc<br>181 Cheshire Lane, Suite 100<br>Plymouth, MN 55441 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Corresponding Official: | Denise Lenz<br>Regulatory Consultant, Libra Medical, Inc.<br>8401 73rd Ave North, Suite 63<br>Brooklyn Park, MN 55428 |
| Telephone Number: | 612-965-3445 |
| Email: | dlenz@libramed.com |
| Preparation Date: | June 25, 2018 |
| Device Trade Name: | SecurAcath |
| Device Common or Usual Name: | Implanted subcutaneous securement catheter |
| Regulation Name: | Percutaneous, implanted, long-term intravascular catheter. |
| Regulation Number: | 21 CFR 880.5970 |
| Product Code: | OKC, KMK |
| Device Class: | Class II |
| Classification Panel: | General Hospital |
| Primary Predicate Device: | K120935; SecurAcath |
#### Purpose of 510(k)
The purpose of this 510(k) is to add the 10F and 12F sized SecurAcath devices to the SecurAcath family to secure 10F and 12F catheters and to update the product specification testing requirements for device fatigue and reliability testing.
## Indications for Use
The SecurAcath device is indicated for short or long term securement of percutaneous indwelling catheters for intravenous use to the access site by means of a subcutaneous anchor.
{4}------------------------------------------------
#### Device Description
The SecurAcath device is a standalone subcutaneous anchoring securement system used to secure the catheter to the access site. The securement is achieved by means of a blunt nitinol anchor deployed into the subcutaneous space at the catheter access site, and the clamping of the catheter shaft between the Base Assembly and Cover of the device.
The intended use of the SecurAcath device is to secure catheters by means of a subcutaneous anchor.
#### Technological Characteristics Comparison
The SecurAcath is a single use, sterile device for securing indwelling catheters. The device is a stand-alone accessory to percutaneous indwelling catheters. The securement to the catheter access site is achieved by means of a blunt nitinol Anchor deployed into the subcutaneous space at the catheter access site. The securement of the catheved by the clamping of the catheter shaft between the Base Assembly and Cover of the device. This reduces catheter migration and accidental pull-out while not significantly affecting fluid flow.
The changes between the subject device and predicate device is the addition of the 10F and 12F sized SecurAcath devices to the SecurAcath family to secure 10F and 12F catheters. Design changes for the 10F and 12F includes the removal of alignment fingers and the shaft lock shape change to oval.
| Characteristic | Predicate Device:<br>SecurAcath (K120935) | Subject Device:<br>SecurAcath (K180769) | Comparison |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indications for<br>Use | The SecurAcath Device is<br>indicated for short or long<br>term securement of<br>percutaneous indwelling<br>catheters for intravenous<br>use to the access site by<br>means of a subcutaneous<br>anchor. | The SecurAcath Device is<br>indicated for short or long<br>term securement of<br>percutaneous indwelling<br>catheters for intravenous<br>use to the access site by<br>means of a subcutaneous<br>anchor. | Same |
| Regulation<br>Number | 21 CFR 880.5970 | 21 CFR 880.5970 | Same |
| Product Code | OKC, KMK | OKC, KMK | Same |
| Device<br>Configuration | Single use, sterile device | Single use, sterile device | Same |
| Catheter<br>compatibility | 3F, 4F, 5F, 6F, 7F, 8F | 3F, 4F, 5F, 6F, 7F,<br>8F, 10F, 12F | Different |
| Biocompatibility | Materials testing in<br>accordance with ISO<br>10993-1 | Identical materials used in<br>new device configurations | Same |
| Sterility | EtO and SAL of 10-6<br>In accordance with ISO<br>11135 | EtO and SAL of 10-6<br>In accordance with ISO<br>11135 | Same |
{5}------------------------------------------------
#### Substantial Equivalence Discussion
The indications for use and intended use of the predicate 510(k) K120935, are equivalent to the subject 510(k) device.
The differences between the subject device and predicate device are the following:
- The additional sizes of 10F and 12F compatibility to the device ●
- Update the product specification testing requirements to the device fatigue and reliability . testing.
The additional sizes (10F, 12F) for SecurAcath have the same principles of operation, same technological characteristics, incorporate the same basic design, incorporate the same materials, have the same shelf life, and are packaged and sterilized using the same materials and processes as the previously cleared predicate device. The changes have been verified/validated through performance testing and risk management activities.
Based on the aforementioned modifications to the subject device, the subject device does not raise different types of safety and effectiveness questions when compared to the predicate device.
#### Performance Testing Summary
Based on a risk analysis of the changes, the following testing was conducted to demonstrate substantial equivalence to the predicate device:
Performance testing, includes dimensional verification, functional tests and securement reliability. Testing was performed to demonstrate that the device meets product specifications and is able to secure catheters to access sites. The device uses the same material as its predicate device and meets the same specifications as its predicate devices. Test results demonstrate that the device functions as intended. The following tests were performed:
- . Dimensional
- Joint Tensile Strength ●
- . Base & Cover Interaction
- Hinge performance ●
- . Catheter Securement Performance
- Catheter Interaction ●
A risk analysis was completed in accordance with ISO 14971:2012, Medical Devices -Applications of Risk Management to Medical Devices.
In all instances. the subject device functioned as intended and demonstrated equivalent performance to the predicate device.
## CONCLUSION
The modifications to the device do not raise different questions of safety and effectiveness and are supported by risk management activities. The SecurAcath is substantially equivalent to the Interrad Medical SecurAcath device, cleared under K120935.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.