K180767 · Life Science Technology, Inc. · FLL · Jan 22, 2019 · General Hospital
Device Facts
Record ID
K180767
Device Name
FEMON Banana Thermometer
Applicant
Life Science Technology, Inc.
Product Code
FLL · General Hospital
Decision Date
Jan 22, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
The FEMON Banana Thermometer is a battery operated electronic device with intended use of measuring human body temperature. This device is intended for armpit temperature measurement for patients from ages 29 days and older.
Device Story
FEMON Banana Thermometer is a wearable, battery-operated electronic thermometer; consists of a body patch applied directly to the patient's armpit skin. Device measures body temperature and transmits data wirelessly via Bluetooth (2.4G BLE) to a mobile application on iOS or Android smart devices. Intended for home use; allows continuous monitoring. User selects from four measurement interval options (e.g., intensive monitoring every second for 20 seconds, or default 5-second intervals, 30-second, or 60-second intervals) to balance battery life and monitoring needs. Healthcare providers or patients view temperature data on the mobile app interface to track body temperature trends. Benefits include non-invasive, continuous temperature monitoring for patients 29 days and older.
Clinical Evidence
Bench testing only. Performance supported by clinical accuracy testing per ISO 80601-2-56, EMC and electrical safety testing (IEC 60601-1, IEC 60601-1-2), software verification and validation, and biocompatibility testing (cytotoxicity, sensitization, irritation) per ISO 10993.
Technological Characteristics
Wearable patch form factor; 3.0V DC lithium button cell power; Bluetooth 2.4G BLE connectivity. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-6 (usability), ISO 80601-2-56 (thermometer performance), ISO 10993-1/5/10 (biocompatibility). Measurement range 35-42°C; accuracy ±0.1°C. Software-based data transmission to mobile app.
Indications for Use
Indicated for armpit temperature measurement in patients aged 29 days and older.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
January 22, 2019
Life Science Technology Inc. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA Inc. 690 Roosevelt Irvine, California 92620
Re: K180767
Trade/Device Name: FEMON Banana Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: December 19, 2018 Received: December 21, 2018
Dear Priscilla Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Sapana Patel -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180767
Device Name FEMON Banana Thermometer
#### Indications for Use (Describe)
The FEMON Banana Thermometer is a battery operated electronic device with intended use of measuring human body temperature. This device is intended for armpit temperature measurement from ages 29 days and older.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
(K180767)
This summary of 510(k) is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: January 22, 2019
#### 1. 510K Applicant / Submitter:
Life Science Technology Inc. Suite B-1207, Gangseo Hangang Xi Tower, Yangcheon-ro 401, Gangseo-gu, Seoul, South Korea, 07528 Tel: +82-70-8620-8021
#### 2. Submission Contact Person
LK Consulting Group USA, Inc. 690 Roosevelt, Irvine CA 92620 Priscilla Juhee Chung Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com
#### 3. Device
- Proprietary Name: FEMON BananaThermometer .
- . Common Name: Electronic Thermometer
- Classification: Class II (21 CFR 880.2910), Clinical Electronic Thermometer .
- Product Code: FLL .
#### 4. Predicate Device
- . Primary Predicate Device: Fever Scout by VivaLink (K162137)
#### 5. Description:
The FEMON Banana Thermometer is a thermometer composed of a body patch to be directly attached to the patient's skin on armpit. The temperature information is then to be retrieved from the patch to the mobile phone application via Bluetooth.
{4}------------------------------------------------
## 8. Indications for Use
The FEMON Banana Thermometer is a battery operated electronic device with intended use of measuring human body temperature. This device is intended for armpit temperature measurement for patients from ages 29 days and older.
## 9. Substantial Equivalence Discussion:
The subject device, FEMON Banana Thermometer, has same intended use as the predicate device, and employs similar technology. The differences in the indications for use statement do not raise any different questions of safety or effectiveness.
The differences in technology do not raise different questions of safety and effectiveness. Like the predicate, the subject device functions to transmit continuous measurements of body temperature wirelessly to a mobile app. The device utilizes an adhesive that was evaluated for biocompatibility.
The technical specifications such as display device, working voltage, signal transmission, and receiver between the subject device and the predicate device are similar. The performance specifications such as measurement and accuracy range are similar as well.
The difference is that the subject device has 4 measurement interval options, whereas, the predicate device has only one option. The interval options of the subject device are offered to save battery life and also to choose intensive monitoring. This difference does not raise different questions in safety and performance because the user can choose the interval option for each situation.
| | Subject Device | Primary Predicate Device |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | FEMON<br>Banana Thermometer | Fever Scout |
| Manufacturer | LIFE SCIENCE<br>TECHNOLOGY Inc. | VivaLink |
| 510(k) Number | K180767 | K162137 |
| Product Code | FLL | FLL |
| Indications for<br>Use | The FEMON Banana Thermometer<br>is a battery operated electronic<br>device with intended use of<br>measuring human body temperature.<br>This device is intended for armpit<br>temperature measurement for<br>patients from ages 29 days and<br>older. | The wireless Fever Scout<br>Continuous<br>Monitoring thermometer is a<br>noninvasive and re-usable<br>electronic device for home use.<br>This product is intended for non-<br>urgent ambulatory continuous armpit<br>body temperature<br>monitoring from ages 29 days and<br>older. |
| Display Use<br>Specification | iOS device display<br>and Android device<br>display | iOS device display |
| Working Voltage | 3.0V DC | 3.0V DC |
| Battery | 3.0v Lithium Button Cell | MS Lithium Rechargeable Battery<br>3.0V |
| Reusability | Yes | Yes |
| Measurement<br>Range (°C) | 35~ 42°C | 35~ 42°C |
| Accuracy (°C) | ±0.1°C: 35~42°C | ±0.1°C: 37 ~39°C.<br>±0.2°C: 35 ~ 37°C and 39<br>~42°C |
| Temperature<br>Unit | °C or °F | °C or °F |
| Signal<br>Transmission | Wireless 2.4G<br>Bluetooth BLE | Wireless 2.4G<br>Bluetooth BLE |
| Receiver | Wireless 2.4G Bluetooth BLE<br>enabled smart devices running<br>Apple operating system iOS 9+ or<br>Android operating system 4.4+ | Wireless 2.4G Bluetooth BLE<br>enabled smart devices running Apple<br>operating system<br>iPhone 5S+ or newer. |
| Operating<br>Condition | 5~40°C<br>15-90% RH | 10~40°C<br>15-85% RH |
| Anatomical<br>Application | Armpit peel-and-<br>stick<br>contact thermometer<br>sensor | Armpit peel-and-<br>stick<br>contact thermometer<br>sensor |
| Temperature<br>Measurement<br>Interval | • Fever Check option for intensive<br>monitoring: continuous<br>measurement every second for 20<br>secs.<br>• Continuous transmitter measures<br>body temperature every 5<br>seconds (default setting). User can<br>change the setting to 30 seconds<br>or 60 seconds. | Continuous transmitter<br>measures body<br>temperature every 15<br>seconds |
| Standard<br>Conformance | IEC 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-1-6,<br>ISO 80601-2-56,<br>ISO 10993-1,<br>ISO 10993-5,<br>ISO 10993-10 | IEC 60601-1,<br>IEC 60601-1-2,<br>ISO 10993-1,<br>ISO 10993-5,<br>ISO 10993-10 |
{5}------------------------------------------------
{6}------------------------------------------------
## 10. Performance Tests
The following performance testing was included to support a determination of substantial equivalence of the device to the predicate:
- Clinical Accuracy Performance Testing (per ISO80601-2-56) ●
- . The EMC and Electrical Safety Testing (per IEC 60601-1 and IEC 60601-1-2
- Software Verification and Validation Testing in according with FDA Guidance for ● Software in Medical Devices
- Biocompatibility Tests (Cytotoxicity, Sensitization, Irritation) per ISO 10993. ●
The test results of the above tests performed on the subject device supported that it is substantially equivalent to the predicate device.
#### 11. Conclusions:
Based on the information provided in this premarket notification, Life Science Technology Inc. concludes that the FEMON Banana Thermometer is substantially equivalent to the predicate device as described herein in.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.