The Televere Podiatry X-Ray System HF is intended for use by qualified clinicians for the x-ray of hands and feet. Not for mammography. Not fluoroscopy.
Device Story
Televere Podiatry X-Ray System HF is a stationary X-ray system for podiatric imaging. It integrates a high-frequency X-ray generator (X-CEL HF 718BD) with Tigerview digital imaging software running on a standard PC. The system replaces the predicate's manual touchscreen exposure interface with a software-based interface, allowing clinicians to control exposure settings, acquire, display, edit (resize, contrast, crop), store, print, and distribute images within a PACS environment. Used in clinical settings by qualified clinicians, the device facilitates diagnostic imaging of hands and feet. The software interface provides 37 exposure presets for various anatomical regions (ankle, foot, calcaneus, etc.), improving workflow over the manual predicate. The system is compatible with flat panel detectors featuring AED functionality. The output is a digital radiographic image used by physicians to support clinical decision-making and diagnosis.
Clinical Evidence
No clinical trials were required. Substantial equivalence was demonstrated through bench testing, laboratory testing to recognized standards (e.g., 21 CFR 1020.30/31, IEC 60601 series), and inspection of sample clinical images to confirm image quality.
Technological Characteristics
Stationary X-ray system; high-frequency (HF) generator; stationary anode tube (1.0 mm focal spot); 3.2 mm Al filtration; 10 mA fixed current; 50-90 kV range. PC-based software interface for exposure control and PACS integration. Connectivity via Network Adaptor, USB, PCI, parallel, memory card, S-video, or SCSI. Compliant with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and IEC 62304 (software lifecycle).
Indications for Use
Indicated for use by qualified clinicians for x-ray imaging of hands and feet. Not for mammography or fluoroscopy.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is in blue and includes the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 20, 2018
Televere Systems, Inc. % Mr. Robert Bakin Regulatory Consultant Technology and Business Law Advisors, LLC 1244 Capuchino Avenue BURLINGAME CA 94010
Re: K180765
Trade/Device Name: Televere Podiatry X-Ray System HF Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: March 21, 2018 Received: March 23, 2018
Dear Mr. Bakin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
## Page 2 - Mr. Robert Bakin
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn
(http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) -n/a K180765
Device Name
Televere Podiatry X-Ray System HF
Indications for Use (Describe)
The Televere Podiatry X-Ray System HF is intended for use by qualified clinicians for the x-ray of hands and feet. Not for mammography. Not fluoroscopy. 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for TELEVERE SYSTEMS. The word "TELEVERE" is in large, orange, sans-serif font. Below it, in a smaller, gray, sans-serif font, is the word "SYSTEMS". The logo is simple and modern.
SQUARE STORE SHARE
510(k) Summary:
Televere Systems, LLC 1160 Cavalier Road Arnold, MD 21012 Email: rquaal@tigerview.com Establishment Registration Number: 2954358
CONTACT: Robert E. Bakin Email: rbakin@tblawadvisors.com Technology & Business Law Advisors, LLC 1244 Capuchino Avenue, Burlingame, CA 94010 Phone: 571-215-3507 Prepared March 20, 2018
1. Identification of the Device: Proprietary-Trade Name: Televere Podiatry X-Ray System HF Classification Name:Stationary X-ray System Product Code: KPR Common/UsualName:Stationary X-Ray System Device Class/RegulationNumber:Class Il per regulation 21 C.F.R. §892.1680
2. Equivalent legally marketed device: Model HF 718BD X-Ray System (K160857), X-CEL X-Ray Corp. Classification Name: Stationary X-ray System Product Code: KPR Common/UsualName:Stationary X-Ray System Device Class/RegulationNumber:Class Il per regulation 21 C.F.R. §892.1680
3. Associated legally marketed devices: Tiger View Professional (PACS Software) K061035, Televere Systems, LLC. This is the software used withthisdevice.
Classification Name: System, Image Processing, Radiological, Code LLZ Common/Usual Name: Picture Archiving and Communications System Device Class/Regulation Number: Class II per regulation 21 C.F.R. §892.2050
4. Indications for Use: The Televere Podiatry X-Ray System HF is intended for use by qualified clinicians for the x-ray of hands and feet. Not for mammography. Not fluoroscopy.
5. Description of the Device: The Televere Podiatry X-Ray System HF consists of a combination of previously cleared Tigerview digital imaging software (K061035), a high frequency X-Ray generator (X-CEL MODEL HF 718BD, K160857), a PC-based computer and a power supply. Tigerview software provides an exposure control upgrade of the previously cleared HF x-ray generator by replacing the manual touchscreen of the Predicate Device (K160857) with the Tigerview computer software interface. Tigerview software provides basic image adjustment features as well as x-ray exposure control settings. An image adjustment system allows the physician to acquire, display, edit (e.g., resize, adjust contrast, crop, etc.), review, store, print, and distribute medical images within a Picture Archiving and Communication System (PACS) environment. Tigerview software runs on standard PC-compatible computers and is compatible with capture devices which attach to the computer using a Network Adaptor, USB port, PCI slot, parallel port, memory card, S-video port on a video capture card, or SCSI card.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for TELEVERE SYSTEMS. The word "TELEVERE" is written in a bold, sans-serif font in orange color. Below "TELEVERE", the word "SYSTEMS" is written in a smaller, sans-serif font in gray color. The logo appears to be for a technology or engineering company.
SQUARE STORE VIEW SHARE
The previously cleared x-ray generator (K160857) contained minimal software functionalities and a manual x-ray exposure interface. The Proposed Device seeks to upgrade the manually entered x-ray exposure touchscreen of the Predicte Device with the Tigerview software computer interface. The combination of the digital imaging software and the x-ray generator does not affect the safety or efficacy of either component device alone, or in combination.
#### Integration-level requirements/restrictions:
The proposed device is compatible with flat panel detectors having AED functionality.
6. Safety and Effectiveness, Comparison to predicate device. The results of clinical image inspection, bench, and laboratory test results demonstrate that the new device is as safe and effective as the predicate device. Clinical images highlight equal or better image quality as compared to the predicate.
#### 7. Substantial Equivalence Chart
| | Predicate Device | Device Seeking Clearance |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | X-CEL X-Ray Model HF 718BD X-Ray System | Televere Podiatry X-Ray System HF |
| Indications for Use | The HF 718B X-Ray System is intended for<br>use by qualified clinicians for the x-ray of<br>hands and feet. | The Televere Podiatry X-Ray<br>System HF is intended for use by<br>qualified clinicians for the x-ray<br>of hands and feet. Not for<br>mammography. Not fluoroscopy. |
| Configuration | X-ray generator and Software only. No<br>digital flat panel. | SAME |
| X-Ray Tube | Stationary anode<br>Focal spot 1.0 mm NEMA<br>Filtration 3.2 mmAl minimum<br>X-ray tube current 10 mA fixed<br>Kilovoltage to x-ray tube up to 90 kV<br>Tube shielding 1 mm pb equivalent<br>minimum | SAME |
| Waveform | High Frequency (HF) | SAME |
| Exposure Control | Manual Touchscreen | Tigerview Software |
| Power<br>Requirements | 105-130 volt AC @ 58-62Hz. Single phase.<br>Less than 20 Amperes. | SAME |
| Exposure Factors | Exposure times from 50 to 500 mS<br>Adjustable kilo-voltage from 50 to 90 kV<br>No automatic exposure control or mAs<br>settings | SAME |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for TELEVERE SYSTEMS. The word "TELEVERE" is in large, orange, sans-serif font. Below that, in a smaller, gray, sans-serif font, is the word "SYSTEMS."
| Exposure Presets | 14 - For repeatedly used techniques. | 37 total (Ankle 10, Foot 9, Calcaneus 9, Other 9) |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
| Exposure switch | Deadman | SAME |
| Performance<br>accuracy | 21 CFR 1020.31 compliant<br>kVp accuracy at 50-90 kV = +/- 8%<br>mA accuracy = +/- 1%<br>Timer accuracy = +/- 5%.<br>Reproducibility = 0.002cv<br>Leakage Radiation = 34mR | SAME |
| X-ray Beam<br>Quality Metrics | "kVp" "HVL" "mR"<br>50 1.71 6.38<br>60 2.40 8.37<br>70 2.69 14.49<br>85 3.05 19.95<br>90 3.31 22.87 | SAME |
| Electrical Safety &<br>EMC | Electrical Safety per IEC 60601-1<br>and EMC per IEC 60601-1-2. | SAME |
#### 7. Bench Testing Conducted:
- · 21 CFR 1020.30 Diagnostic x-ray systems and their major components.
- 21 CFR 1020.31 Radiographic equipment
- 60601-1 3rd Edition Medical electrical equipment Part 1: General requirements for safety .
- 60601-1-2 3rd Edition Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic compatibility – Requirements and tests
- . 60601-1-6 Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance - Collateral Standard: Usability
- · 62304:2006 Medical device software Software life cycle processes
- 60601-2-54:2009 Ed.1.0 Medical electrical equipment Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy
- . 60601-3:2008-Ed. 2.0 Medical electrical equipment – Part 1-3: General requirements for basic safety and essential performance – Collateral Standard: Radiation protection in diagnostic X-ray equipment
- EC TR 60878 Graphical symbols for electrical equipment in medical practice
- · EN ISO 14971 Application of risk management to medical devices
- ISO 15223-1 Medical devices Symbols to be used with medical device labels, labeling, and information to be supplied.
8. Summary of ClinicalTesting: Sample clinical images from the new device were also provided however they were not necessary to demonstrate substantial equivalence.
9. Conclusion: After analyzing bench, clinicalimage, and externallaboratory testing to applicable standards, it is the conclusion of Televere that the Televere Podiatry X-Ray System HF is as safe and effective as the predicate device K170975, has few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate device.
3.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.