K180735 · Karl Storz SE & CO. KG · HIG · Oct 31, 2018 · Obstetrics/Gynecology
Device Facts
Record ID
K180735
Device Name
Endomat Select
Applicant
Karl Storz SE & CO. KG
Product Code
HIG · Obstetrics/Gynecology
Decision Date
Oct 31, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1700
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Suction/irrigation pumps and their accessories are used for the introduction of irrigation fluids into organs, joints and operating fields as well as the suctioning off of irrigation fluids and bodily fluids, secretions, tissue and gas during diagnostic or therapeutic interventions.
Device Story
Endomat Select (UP210) is a multi-functional, pressure-controlled, combined irrigation and suction pump for endoscopic/surgical procedures. Device inputs include user settings via LCD touch screen and encoded tubing sets that automatically configure operating modes. It utilizes a microprocessor-controlled peristaltic pump to deliver or suction fluids. Used in ORs by surgeons/staff; integrates with other Karl Storz devices via Storz Control Bus (SCB). Software monitors pressure; stops actuator and reverses pump direction if pressure exceeds limits; provides high-pressure alarms. User must confirm pressures >100mmHg. Benefits include maintained visualization, controlled distention, and automated overpressure protection, reducing risk of tissue damage.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including verification of pressure control, flow limitation, flow control, pressure limitation, suction control, and overpressure safety alarms/relief mechanisms.
Technological Characteristics
Microprocessor-controlled peristaltic pump; LCD touch screen interface; dual pressure sensors at pump head; SCB connectivity. Software-based overpressure protection (pump reversal). Tubing sets are encoded for automatic mode selection. Electrical safety per IEC 60601-1, EMC per IEC 60601-1-2, alarm systems per IEC 60601-1-8.
Indications for Use
Indicated for: infusion of sterile irrigant solutions into ureter/upper urinary tract and suctioning of fluids/tissue/gas during urological endoscopic procedures; liquid distention of uterus for diagnostic/operative hysteroscopy; infusion of sterile irrigant solutions into organs/operating fields during laparoscopic and open general surgery; sustained liquid irrigation/distention of joint/intra-articular spaces during arthroscopic surgery; infusion of sterile irrigant solutions for lens cleaning during endoscopically assisted Functional Endoscopic Sinus Surgery and transnasal pituitary gland surgery.
Regulatory Classification
Identification
A hysteroscopic insufflator is a device designed to distend the uterus by filling the uterine cavity with a liquid or gas to facilitate viewing with a hysteroscope.
Predicate Devices
World of Medicine AQUILEX FLUID CONTROL SYSTEM H112 (K112624)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 31, 2018
Karl Storz SE & Co. KG Nina Dusper Regulatory Affairs Manager Dr.-Karl-Storz-StraBe 34 Tuttlingen, 78532 Germany
Re: K180735
> Trade/Device Name: Endomat Select (Model Number: UP210) Regulation Number: 21 CFR§ 884.1700 Regulation Name: Hysteroscopic Insufflator Regulatory Class: II Product Code: HIG. BTA. LJH. OCX. HRX. GWG. EOB Dated: July 30, 2018 Received: August 2, 2018
Dear Nina Dusper:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Sharon M. Andrews -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K180735
Device Name
Endomat Select (Model Number: UP210)
Indications for Use (Describe)
Endomat Select is intended to:
· provide the infusion of the sterile irrigant solutions into the ureter and upper urinary tract, as well as to suction off the irrigations fluids, bodily fluids, secretions, tissue and gas during diagnostic and operative endoscopic urological procedures
· provide liquid distention of the uterus for diagnostic and operative hysteroscopy
· provide the infusion of the sterile irrigant solutions into organs and operating fields during diagnostic and operative procedures in laparoscopic and open general surgery
· provide sustained liquid irrigation of joint or intra-articular spaces during all phases of arthroscopic surgery
· provide the infusion of the sterile irrigant solutions in order to enable the Lens Cleaning during endoscopically assisted Functional Endoscopic Sinus Surgery and endoscopically assisted transnasal pituitary gland surgery
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller letters.
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KARL STORZ`s knowledge.
| Applicant: | KARL STORZ SE & Co. KG<br>Dr.-Karl-Storz-Straße 34<br>78532 Tuttlingen<br>Baden-Württemberg, GERMANY |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Nina Dusper<br>Regulatory Affairs Manager<br>+49 (0) 7461 708-8399<br>+49 (0) 7461 708-105 |
| Date of Preparation: | 1st February 2018 |
| Type of 510(k) Submission: | Traditional |
| Device Identification: | Trade Name: Endomat Select (Model Number: UP210)<br>Common Name: Suction Irrigation Pump |
| Regulation: | 21 CFR 884.1700 (Hysteroscopic Insufflator) |
| Class: | II |
| Product Code: | HIG: Hysteroscopic Insufflator |
| Additional Product Codes: | BTA: Pump, Portable, Aspiration (Manual or Powered)<br>LJH: System, Irrigation, Urological<br>OCX: Endoscopic Irrigation/Suction System<br>HRX: Arthroscope |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font.
| GWG: Endoscope, neurological | |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EOB: Nasopharyngoscope (flexible or rigid) | |
| Predicate Device(s): | World of Medicine AQUILEX FLUID CONTROL SYSTEM<br>H112 (K112624),<br>Applied Medical EPIX (K122619),<br>World of Medicine Uro Pro (K042457),<br>Karl Storz Endomat LC (K031457),<br>Atherex Dual Wave Arthroscopy Pump (K083707),<br>Karl Storz ClearVision (K072410),<br>Karl Storz Clear Vision (K013838)<br>**The above predicates have not been subject to any recall** |
| Device Description: | The Endomat Select is a multi-functional, pressure-<br>controlled, combined irrigation and suction pump. It can be<br>used for irrigation and, where appropriate dilation, during<br>Hysteroscopic (HYS), Arthroscopic (ART), Urological<br>(URO) and General Surgical or Laparoscopic (SURG)<br>interventions. The device can be used for suction during<br>Urological (URO) interventions. In addition the device can<br>function in an oscillatory mode (ENT/Neuro) providing a<br>fluid means to maintain lens clarity during use in Transnasal<br>procedures. The device has a modern LCD display with<br>touch screen user interface. The tubing sets are encoded so<br>that they can be identified by the device, which selects the<br>appropriate operating modes based on the tubing set<br>connected. The device can also be connected to other Karl<br>Storz devices using the proprietary Storz Control Bus (SCB)<br>[K994348].<br>The device protects the patient from overpressure via<br>software means. The software controlled pressure<br>measurement system can stop the actuator (roller pump)<br>should the pressure rise beyond the pre-determined limit and |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller letters.
will immediately reverse the pump roller direction to depressurize the fluid system if required. Additionally the device has high pressure alarms to alert the operating room staff of a high pressure situation. The device also requires positive action from the user to increase pressures above 100mmHg; the user is prompted to confirm their choice on the user interface.
The accessories for this device are provided separately. All of the accessories for use with this device are already marketed in the U.S. and are either Class I, Class II 510(k) exempt or Class II previously cleared devices. The generic tubing sets for use with this device are single use, sterile Class I accessories.
| Accessory description | Part number | Marketing Status |
|-----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Suction bottle, 5 1, sterilizable | 20300050 | Class II<br>510(k) Exempt |
| Cap for suction bottle 20 3000<br>50, 5 1 sterilizable | 20300034 | Class II<br>510(k) Exempt |
| Metal filter | 20300038 | Class II<br>510(k) Exempt |
| Bottle stand for suction bottle,<br>5 I | 20300032 | Class II<br>510(k) Exempt |
| Bottle stand holder | 20300033 | Class II<br>510(k) Exempt |
| Suction bottle, 0,5 I, sterilizable | 20300051 | Class II<br>510(k) Exempt |
| Cap, for suction bottle | 20300039 | Class II<br>510(k) Exempt |
| Bottle stand, for suction bottle | 20300231 | Class II<br>510(k) Exempt |
| Control cable | 20701070 | Class II<br>K973251 |
| Single pedal footswitch | 20014130 | Class II<br>K072410 |
| SCB connecting cord, length 100<br>cm | 20090170 | Class II<br>K994348 |
| Tubing Set, Irrigation, PC, for<br>single use, sterile, package of 10 | 031523-10 | Class I |
| Tubing Set, Suction, BS, for | 031647-10 | Class I |
| | single use, sterile, package of 10 | |
| Tubing Set, Suction, OS, for single use, sterile, package of 10 | 030647-10 | Class I |
| Tubing Set, Irrigation, FC, for single use, sterile, package of 10 | 031524-10 | Class I |
| Tubing Set, Irrigation, CV, for single use, sterile, package of 10 | 031529-10 | Class I |
| Intended Use: | Suction/irrigation pumps and their accessories are used for the introduction of irrigation fluids into organs, joints and operating fields as well as the suctioning off of irrigation fluids and bodily fluids, secretions, tissue and gas during diagnostic or therapeutic interventions. | |
| Indications for Use | Endomat Select is intended to:<br>• provide the infusion of the sterile irrigant solutions into the ureter and upper urinary tract, as well as to suction off the irrigations fluids, bodily fluids, secretions, tissue and gas during diagnostic and operative endoscopic urological procedures<br>• provide liquid distention of the uterus for diagnostic and operative hysteroscopy<br>• provide the infusion of the sterile irrigant solutions into organs and operating fields during diagnostic and operative procedures in laparoscopic and open general surgery<br>• provide sustained liquid irrigation and distention of joint or intra-articular spaces during all phases of arthroscopic surgery<br>• provide the infusion of the sterile irrigant solutions in order to enable the Lens Cleaning during endoscopically assisted Functional Endoscopic Sinus Surgery and endoscopically | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Karl Storz Endoskope. The word "STORZ" is in large, bold, blue letters. Below that, the words "KARL STORZ - ENDOSKOPE" are in smaller, bold, blue letters.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for Karl Storz Endoskope. The word "STORZ" is in large, bold, blue letters on the top line. The words "KARL STORZ - ENDOSKOPE" are in smaller, bold, blue letters on the bottom line.
| | | assisted transnasal pituitary gland surgery | |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | | ** The above Indications for Use represent a combination of the<br>indications for the cited predicate devices and therefore there is no new<br>intended use. ** | |
| Technological<br>Characteristics: | The proposed subject device is a multifunctional design that<br>incorporates the functionality of multiple predicate devices.<br>The technological differences between each of the subject<br>device's functions and the respective predicate devices are<br>summarized in the Table 1.1. below, and they do not raise<br>any different issues of safety and effectiveness. | | |
| Non-Clinical<br>Performance Data: | The Endomat Select follows the FDA recognized consensus<br>standards and is tested according to the following standards<br>and FDA Guidances:<br>Electrical Safety and EMC IEC 60601-1 IEC 60601-1-2 IEC 60601-1-8 Software Verification and Validation Testing Guidance for the Content of Premarket<br>Submissions for Software Contained in<br>Medical Devices issued on May 11, 2005 Level of Concern: Major Summative Usability Testing IEC 62366 Performance Testing Hysteroscopic and Laparoscopic Insufflators:<br>Submission Guidance For A 510(K) issued on<br>August 1, 1995 for Product Code HIG Performance tests of pressure controlled<br>irrigation mode | | |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller letters.
| Performance tests of intrauterine pressure | |
|---------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Additional bench testing was performed to ensure the device | |
| met its design specifications. | |
| Verification Test | Conclusion |
| Performance:<br>Pressure Control | Pass |
| Performance:<br>Flow Limitation | Pass |
| Performance:<br>Flow Control | Pass |
| Performance:<br>Pressure Limitation | Pass |
| Performance:<br>Suction Control | Pass |
| Safety:<br>Overpressure Warning (in pressure relevant<br>GYN and URO interventions) | Pass |
| Safety:<br>Overpressure Warning (in pressure relevant<br>ARTHRO interventions) | Pass |
| Safety:<br>Overpressure Alarm<br>(in pressure relevant GYN and URO<br>interventions) | Pass |
| Safety:<br>Pressure Relief<br>(in pressure relevant GYN and URO<br>interventions) | Pass |
| The bench testing performed verified and validated that the<br>Endomat Select has met all its design specification and is | |
| substantially equivalent to its predicate devices. | |
| Clinical Performance Data: | Clinical testing was not required to demonstrate the<br>substantial equivalence to the predicate devices. Non-clinical<br>bench testing was sufficient to assess safety and effectiveness<br>and to establish the substantial equivalence. |
| Conclusion: | The non-clinical testing demonstrates that the device is as<br>safe, as effective and performs as well as or better than the<br>legally marketed predicate devices. |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller letters.
## Table 1.1. Comparison of Technological Characteristics
| | Subject Device | Predicate Devices by application | | | | | | |
|--------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Karl Storz | Applied Medical | World of Medicine | World of Medicine | Karl Storz | Artherex | Karl Storz | Karl Storz |
| | Endomat Select | EPIX | AQUILEX FLUID<br>CONTROL SYSTEM<br>H112 | Uro Pro | Endomat LC | Dual Wave Arthroscopy<br>Pump | Clearvision | Clearvision<br>Lens Irrigation System |
| Application | Multi | Surgery<br>( Laparoscopy ) | Hysteroscopic | Urological | Urological Suction | Arthroscopic | Lens Cleaning | Lens Cleaning |
| | | | | | | | | |
| Design | Microprocessor<br>controlled, pressure<br>monitoring peristaltic<br>Pump for irrigation or<br>suction. | Microprocessor<br>controlled, peristaltic<br>pump for irrigation<br>with conduit for<br>external suction. | Microprocessor<br>controlled, pressure<br>monitoring peristaltic<br>pump for irrigation<br>and/or suction. | Microprocessor controlled,<br>pressure monitoring<br>peristaltic pump for<br>irrigation or suction. | Microprocessor<br>controlled, peristaltic<br>pump for suction. | Microprocessor<br>controlled, pressure<br>monitoring peristaltic<br>pump for irrigation and<br>suction. | Microprocessor<br>controlled, oscillatory<br>peristaltic pump | Microprocessor<br>controlled, oscillatory<br>peristaltic pump |
| Intended Use | Suction/irrigation<br>pumps and their<br>accessories are used<br>for the introduction of<br>irrigation fluids into<br>organs, joints and<br>operating fields as wel<br>as the suctioning off of<br>irrigation fluids and<br>bodily fluids,<br>secretions, tissue and<br>gas during diagnostic<br>or therapeutic<br>interventions. | The Epix® Suction<br>Irrigation System is<br>intended as a general<br>purpose suction<br>and/or irrigation<br>device for use in<br>laparoscopic and open<br>general surgery,<br>laparoscopic and oper<br>gynecological surgery,<br>laparoscopic and open<br>urologic surgery. This<br>device delivers sterile<br>irrigant solution and<br>serves as a conduit for<br>suction. | Aquilex™ Fluid Control<br>System is intended to<br>provide liquid<br>distension of the<br>uterus during<br>diagnostic and<br>operative<br>hysteroscopy, and to<br>monitor the volume<br>differential between<br>the irrigation fluid<br>flowing into and out of<br>the uterus. | The URO Pro is an irrigation<br>roller pump intended for<br>the infusion of sterile<br>solutions into the ureter<br>and upper urinary tract<br>during diagnostic and/or<br>therapeutic endoscopic<br>urologic procedures. It is<br>indicated for the infusion of<br>sterile solutions through an<br>endoscope into the<br>urogenital tract to irrigate<br>the ureter or upper urinary<br>tract in a controlled<br>manner and to improve<br>visibility of the surgical<br>field during urologic<br>procedures. | Suction/irrigation pumps<br>and their accessories are<br>used for the introduction<br>of irrigation fluids into<br>organs, joints and<br>operating fields as well<br>as the suctioning off of<br>irrigation fluids and<br>bodily fluids, secretions,<br>tissue and gas during<br>diagnostic or therapeutic<br>interventions. | The AR-6480 is intended<br>to provide continuous<br>pulse-free flow that<br>reacts immediately to<br>changes in the intra-<br>articular pressure so that<br>joint distention can be<br>sustained even under<br>high shaver extraction<br>volumes or secondary<br>outflow. | Irrigation pumps and<br>their accessories are<br>used for the introduction<br>of irrigation fluids into<br>organs and operating<br>fields during endoscopic<br>interventions. The<br>clearvision is a lens<br>irrigation pump for the<br>fields of ENT and<br>Neurosurgery. | The KSEA Clearvision is<br>a lens irrogation system<br>for cleaning the lens<br>and maintaining clear<br>visualization without<br>removing the scope<br>from<br>the surgical site during<br>sinus surgery. |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image is a logo for Karl Storz Endoskope. The word "STORZ" is in large, bold, blue letters on the top line. The words "KARL STORZ - ENDOSKOPE" are in smaller, blue letters on the bottom line.
KARL STORZ Premarket Notification
Endomat Select
Summary of Safety and Effectiveness
| | HIG, BTA, LJH, OCX,<br>HRX, GWG, EOB | BTA | HIG | LJH | OCX | HRX | GWG | EOB |
|-----------------------------------------------|---------------------------------------------------|-----------------------------------------------------|-----------------------------------------------|------------------------------------------------|--------------------------------------------------|----------------------------------------------------------------------------|----------------------------|----------------------------|
| Procode | | | | | | | | |
| Pressure<br>measurement<br>location and | Yes, dual pressure<br>sensors at the pump<br>head | No pressure<br>measurement | Yes, single transducer<br>at the pump head | Yes, dual pressure sensors<br>at the pump head | No pressure<br>measurement | Yes, dual sensor | No pressure<br>measurement | No pressure<br>measurement |
| User<br>interface/Selection<br>of Application | LCD touch screen | Digital flowrate display<br>and membrane<br>buttons | Digital display and button | Digital display and<br>membrane buttons | Digital flowrate display<br>and membrane buttons | LCD touch screen | On/Off | On/Off |
| Overpressure<br>Safety | Yes, using software<br>and roll pump reversal | No | Yes, using software<br>and roll pump reversal | Yes, using software and roll<br>pump reversal | N/A | Software controlled<br>pump stops at<br>>300mmHg (IFU<br>recommends manual | N/A | N/A |
| Features | | | | | | | | |
| Indicates<br>Operating | Yes | No | Intrauterine | Yes | No | Yes | No | No |
| Alarm for over<br>Pressure | Yes | No | Yes | Yes | No | Yes | No | No |
| Volume difference<br>monitor | No | No | Yes | No | No | No | No | No |
| Indicates<br>Operating Fluid<br>Flow | Yes | Yes | No | Yes, nominal set point | No | No | No | No |
| Alarm for over<br>Fluid Flow | Yes | No | Yes | Yes | No | No | No | No |
| Indicates<br>Maximum Suction<br>Pressure | No | No | No | No | No | No | No | No |
| Alarm for over<br>Suction Pressure | No | No | No | No | No | No | No | No |
| Accessories | | | | | | | | |
| Foot Switch | Yes | No | No | No | Yes | Yes | Yes | Yes |
| Tubing Sets | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| External pressure<br>sensor (optional) | No | No | No | No | No | No | No | No |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for STORZ. The word "STORZ" is written in large, bold, blue letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE", also written in blue. The logo is simple and professional.
Summary of Safety and Effectiveness
| Performance | | | | | | | | |
|----------------------|---------------------------|-----------------|-------------|-------------|------------------------|------------------|-----------|-----------|
| Irrigation Pressure | | | | | | | | |
| Lap/Surgical | 100,300 and 500mmHg | Not Given | | | | | | |
| Hysteroscopic | 20-150mmHg | | 40-150 mmHg | | | | | |
| Urological | 20-150mmHg | | | 15-150 mmHg | | | | |
| Urological suction | N/A | | | | N/A | | | |
| Arthroscopy | 20-150mmHg | | | | | 10-120 mmHg | | |
| Clearvision | N/A | | | | | | N/A | |
| Clearvision | N/A | | | | | | | N/A |
| Irrigation Flow | | | | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.