K180703 · Vita Zahnfabrik H.Rauter GmbH Co. · EIH · Aug 1, 2018 · Dental
Device Facts
Record ID
K180703
Device Name
VITA YZ ST and VITA YZ XT
Applicant
Vita Zahnfabrik H.Rauter GmbH Co.
Product Code
EIH · Dental
Decision Date
Aug 1, 2018
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.6660
Device Class
Class 2
Indications for Use
VITA YZ® ST and VITA YZ® XT materials are blanks used for fabricating dental restorations. VITA YZ ST is indicated for fully anatomical anterior and posterior crowns, multi-unit fully anatomical anterior and posterior bridges, fully anatomical crowns and multi-unit bridges on direct screw-retained restorations, and single-tooth and multi-unit bridge frameworks of anterior and posterior crowns and bridges. VITA YZ XT is indicated for fully anatomical anterior and posterior crowns, fully anatomical 3-unit anterior and posterior bridges, and anterior and posterior single-tooth and 3-unit bridge frameworks.
Device Story
VITA YZ ST and VITA YZ XT are pre-sintered yttrium-stabilized zirconia (Y-TZP) blanks; used as raw material for dental CAD/CAM milling systems. Dental technicians or clinicians mill blanks into anatomical crowns, bridges, or frameworks; restorations are subsequently sintered to achieve final physical properties. Devices provide high-translucency zirconia to match natural dentition. Output is a custom-milled dental restoration; used to restore tooth function and aesthetics. No software or algorithm is integral to the material itself; processing relies on external CAD/CAM hardware/software.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including ISO 6872:2015 (Dentistry, Ceramic materials), EN ISO 9693-2:2016 (Compatibility Testing), and biocompatibility testing per ISO 10993-1:2009 and ISO 7405:2008.
Technological Characteristics
Pre-sintered yttrium-stabilized tetragonal zirconia polycrystal (Y-TZP) blanks. Available in disc and block configurations. Compliant with ISO 6872:2015 and EN ISO 9693-2:2016. Flexural strength: VITA YZ ST (934 ± 104 MPa), VITA YZ XT (678 ± 62 MPa).
Indications for Use
Indicated for patients requiring dental restorations including fully anatomical anterior/posterior crowns, multi-unit bridges, and frameworks. VITA YZ ST is indicated for anterior/posterior crowns, multi-unit bridges, screw-retained restorations, and frameworks. VITA YZ XT is indicated for anterior/posterior crowns, 3-unit bridges, and frameworks.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and below that, the word "ADMINISTRATION" in a smaller font, also in blue.
August 1, 2018
VITA Zahnfabrik H.Rauter GmbH Co. % Nevine Erian Director, Regulatory Affairs & Compliance (Consultant) VITA North America 22705 Savi Ranch Parkway, Suite 100 Yorba Linda, California 92887
#### Re: K180703
Trade/Device Name: VITA YZ ST and VITA YZ XT Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: May 30, 2018 Received: May 31, 2018
#### Dear Nevine Erian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
## Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K180703
Device Name
VITA YZ® ST and VITA YZ® XT
Indications for Use (Describe)
VITA YZ® ST and VITA YZ® XT materials are blanks used for fabricating dental restorations.
VITA YZ® ST is indicated for fully anatomical anterior crowns, multi-unit fully anatomical anterior and posterior bridges, fully anatomical crowns and multi-unit bridges on direct screw-retained restorations, and single-tooth and multi-unit bridge frameworks of anterior and posterior crowns and bridges.
VITA YZ® XT is indicated for fully anatomical anterior and posterior crowns, fully anatomical 3-unit anterior and posterior bridges, and anterior and posterior single-tooth and 3-unit bridge frameworks.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
K180703
| Submitter | VITA Zahnfabrik H.Rauter GmbH Co.<br>Spitelgasse 3<br>Bad Sackingen, D-79713<br>Germany<br>Establishment Reg. No. 1000625496 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact | Bernd Walker<br>Head of Regulatory Affairs and Quality Systems<br>Phone (+49) 7761 562-361<br>Fax (+49) 7761 562-384<br>B.Walker@vita-zahnfabrik.com |
| Official Correspondent | Nevine Erian<br>Director, Regulatory Affairs & Compliance (Consultant)<br>VITA North America, Inc.<br>22705 Savi Ranch Parkway, Suite 100<br>Yorba Linda, CA 92887<br>Establishment Reg. No. 2082832<br>Phone (949) 370-7155<br>Fax (714) 221-6759<br>E-mail: nerian@vitanorthamerica.com |
| Date Prepared | May 29, 2018 |
- Trade/Device Name VITA YZ® ST and VITA YZ® XT 트
- Common Name Zirconia Based , CAD/CAM Material ■
- 트 Classification Name Porcelain powder for clinical use
- . Regulation Number 21 CFR 872.6660
- . Product Code EIH
#### Predicate Devices
VITA In–CERAM YZ Cubes for CEREC® (VITA Zahnfabrik GmbH) – K022996 – Primary Predicate
{4}------------------------------------------------
priti® multidisc Zr02 High Translucent ( pritidenta GmbH) – K161025 – Reference Predicate
#### Device Description
VITA YZ ST and VITA YZ XT discs are pre-sintered zirconia materials, partially stabilized with yttrium oxide (Y-TZP, yttria stabilized tetragonal zirconia polycrystal). VITA YZ ST is made of Super Translucent Zirconia and VITA YZ XT is made of Extra Translucent Zirconia. They offer higher translucency to better match natural dentition.
#### Statement of Intended Use
VITA YZ® ST and VITA YZ® XT materials are blanks used for fabricating dental restorations.
VITA YZ ST is indicated for fully anatomical anterior and posterior crowns, multi-unit fully anatomical anterior and posterior bridges, fully anatomical crowns and multi-unit bridges on direct screw-retained restorations, and single-tooth and multi-unit bridge frameworks of anterior and posterior crowns and bridges.
VITA YZ XT is indicated for fully anatomical anterior and posterior crowns, fully anatomical 3-unit anterior and posterior bridges, and anterior and posterior single-tooth and 3-unit bridge frameworks.
### Material Composition
VITA YZ ST and YZ XT are formulated with yttrium-stabilized zirconia which is identical to the predicate device VITA YZ T (KO22996). VITA YZ ST and VITA YZ XT have higher amounts of yttrium, for increased translucency, compared to VITA YZ T.
#### Technological Characteristics
VITA YZ products are offered in disc and block configurations, in different shades. Blocks have a mandrel attachment, to permit securing into a CAD/CAM machine for milling into a restoration.
#### Non-Clinical Performance Testing
VITA YZ ST and VITA YZ XT were tested and met the requirements of the following standards:
- . ISO 6872:2015 – Dentistry, Ceramic materials
- EN ISO 9693-2:2016 - Dentistry, Compatibility Testing, Ceramic-ceramic systems
{5}------------------------------------------------
Bench test results allowed us to conclude that VITA YZ ST and VITA YZ XT meet their intended use.
#### Biocompatibility
VITA YZ ST and VITA YZ XT meet biocompatibility requirements of the following standards:
- i ISO 10933-1:2009 – Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process
- . ISO 7405:2008 Dentistry - Evaluation of Biocompatibility of Medical Devices Used in Dentistry
#### Clinical Performance Data
Not applicable. No human clinical testing was performed to support the substantial equivalence of VITA YZ ST and VITA YZ XT.
#### Substantial Equivalence
#### Material
VITA YZ ST and YZ XT are similar in chemical composition as the predicate devices.
### Physical Properties
VITA YZ ST and YZ XT have similar physical properties as the predicate devices.
#### Technical Comparison of VITA YZ ST and YZ XT to Predicate Devices
| Technical Attribute | VITA YZ ST and XT | VITA IN-<br>CERAM YZ<br>(Primary<br>Predicate) | priti® multidisc<br>Zr02 HT<br>(Reference<br>Predicate) |
|--------------------------------------------------------------------------------------------|----------------------|------------------------------------------------|---------------------------------------------------------|
| Indications | | | |
| Fully anatomical anterior and<br>posterior crowns | ST – Yes<br>XT – Yes | Yes | Yes |
| Multi-unit fully anatomical<br>anterior and posterior bridges | ST – Yes<br>XT – Yes | Yes | Yes |
| Fully anatomical crowns and<br>multi-unit bridges on direct<br>screw-retained restorations | ST – Yes<br>XT – No | Yes | Yes |
{6}------------------------------------------------
| Technical Attribute | VITA YZ ST and XT | VITA IN-<br>CERAM YZ<br>(Primary<br>Predicate) | priti® multidisc<br>Zr02 HT<br>(Reference<br>Predicate) |
|------------------------------------------------------------------------------------------------------|-----------------------------------------------------|------------------------------------------------|---------------------------------------------------------|
| Single-tooth and multi-unit<br>bridge frameworks of anterior<br>and posterior crowns and<br>bridges. | ST - Yes<br>XT - Yes | Yes | Yes |
| Fully anatomical 3-unit<br>anterior and posterior bridges | ST - Yes<br>XT - Yes | Yes | Yes |
| Anterior and posterior single-<br>tooth and 3-unit bridge<br>frameworks | ST - Yes<br>XT - Yes | Yes | Yes |
| CAD/CAM Material | Yes | Yes | Yes |
| Physical Configuration | Disc Form | Disc & Block<br>Form | Disc & Block Form |
| Shades | Different shades | Different<br>shades | Different shades |
| Meets ISO 6872 Req'ts | Yes | Yes | Yes |
| Flexural Strength | VITA YZ ST 934 ± 104 MPa<br>VITA YZ XT 678 ± 62 MPa | 1165 ± 56 MPa | > 650 MPa |
### Conclusion
Information provided in this application demonstrates that VITA YZ ST and VITA YZ XT are substantially equivalent in safety and effectiveness to the predicate devices. VITA YZ ST & VITA YZ XT have similar indications for use, similar material composition, similar physical properties, and similar technological characteristics as VITA In-CERAM YZ Cubes for CEREC® and priti® multidisc Zr02 HT. The differences between VITA YZ ST & VITA YZ XT and the predicate devices do not impact safety and effectiveness, as VITA YZ ST & VITA YZ XT share same indications as the predicate devices, meet the same biocompatibility, and ISO 6872 test criteria.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.