CoaguChek XS Plus System

K180684 · Roche Diagnostics · GJS · Apr 19, 2018 · Hematology

Device Facts

Record IDK180684
Device NameCoaguChek XS Plus System
ApplicantRoche Diagnostics
Product CodeGJS · Hematology
Decision DateApr 19, 2018
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 864.7750
Device ClassClass 2

Indications for Use

The CoaguChek XS Plus system is intended for use by professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy. The system uses fresh capillary or non-anticoagulated venous whole blood.

Device Story

The CoaguChek XS Plus System is a point-of-care diagnostic device used by professional healthcare providers to monitor warfarin therapy. It measures prothrombin time using fresh capillary or non-anticoagulated venous whole blood samples. The system provides quantitative results to assist clinicians in adjusting anticoagulant therapy. This submission specifically validates the use of Super Sani-Cloth wipes for cleaning and disinfecting the device housing and components, replacing previously cleared sodium hypochlorite solutions. The device remains unchanged in its technological, material, and performance characteristics since its original clearance.

Clinical Evidence

Bench testing only; risk analysis performed per ISO 14971 to assess impact of new cleaning/disinfection protocol on device components.

Technological Characteristics

The device is a prothrombin time test system (21 CFR 864.7750). No changes to materials, sensing technology, or energy sources were made. The system is a point-of-care meter. Cleaning/disinfection validation was performed for Super Sani-Cloth wipes.

Indications for Use

Indicated for professional healthcare providers to perform quantitative prothrombin time testing for monitoring warfarin therapy using fresh capillary or non-anticoagulated venous whole blood.

Regulatory Classification

Identification

A prothrombin time test is a device used as a general screening procedure for the detection of possible clotting factor deficiencies in the extrinsic coagulation pathway, which involves the reaction between coagulation factors III and VII, and to monitor patients receiving coumarin therapy (the administration of one of the coumarin anticoagulants in the treatment of venous thrombosis or pulmonary embolism).

Predicate Devices

Submission Summary (Full Text)

{0} SPECIAL 510(K): DEVICE MODIFICATION OIR DECISION MEMORANDUM 510(k) Number: K180684 This 510(k) submission contains information/data on modifications made to the applicant’s own class II or class I devices requiring 510(k). The following items are present and acceptable: 1. The name and 510(k) number of the applicant’s previously cleared device: CoaguChek XS Plus System, K071041. 2. Applicant’s statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials. 3. A description of the device MODIFICATION, including clearly labeled diagrams, engineering drawings, photographs, user’s and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for new cleaning and disinfection instructions in the labeling to include Super Sani-Cloth Germicidal Disposable Wipes. 4. Comparison Information (similarities and differences) to applicant’s legally marketed predicate device including, labeling, intended use, target population, material, technology, performance, and design changes. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method used to assess the impact of the modification on the device and its components, and the results of the analysis. b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied. Sponsor followed ISO 14971 Second edition (Medical devices - Application of risk management to medical devices) to evaluate risk specific to cleaning and disinfection of the CoaguChek XS Plus System. The scope of the risk analysis covers disinfectant seeping into the device which may affect heater of the device, battery contact, temperature sensors and strip sensors. It was concluded that the proposed change does not pose any additional risks and there is no impact on the safety or effectiveness of the product. The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the applicant’s description of the modifications and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The applicant has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their pre-amendment) device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...