The MINE is intended to be used by trained dentists and dental technicians as extra-oral x-ray source for producing diagnostic x-ray images using intra-oral image receptors or film. Its use is intended for both adult and pediatric subjects.
Device Story
Portable dental X-ray system; hand-held enclosure containing X-ray generator, controller, and beam-limiting device. Operates on 11.1V DC rechargeable Li-Polymer battery. High-frequency inverter converts DC to AC to drive X-ray tube. Produces X-ray beam to penetrate tooth/jaw area; images captured on external intra-oral receptors or film. Used in dental clinics by dentists or dental technicians. User interface allows selection of tooth region, patient type (adult/child), and receptor type (film/sensor). Provides diagnostic images for dental assessment; small radiation dose minimizes exposure.
Clinical Evidence
Bench testing only. Compliance with IEC/EN 60601-1, 60601-1-3, 60601-2-28, 60601-2-65, and EN/IEC 60601-1-2 (EMC) standards. Phantom images provided to demonstrate performance within an intraoral imaging chain. Compliance with 21 CFR 1020 federal performance standards confirmed.
Indicated for use by trained dentists and dental technicians as an extra-oral x-ray source for producing diagnostic images of teeth and jaws in adult and pediatric patients using intra-oral receptors or film.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font size below.
RHT Co., Ltd. % Ms. April Lee Consultant Withus Group, Inc. 106 Superior IRVINE CA 92620
Re: K180671
Trade/Device Name: MINE Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: EHD Dated: May 16, 2019 Received: May 17, 2019
Dear Ms. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
June 5, 2019
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180671
Device Name MINE
Indications for Use (Describe)
The MINE is intended to be used by trained dental technicians as extra-oral x-ray source for producing diagnostic x-ray images using intra-oral image receptors or film. Its use is intended for both adult and pediatric subjects.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
#### Submitter
RHT Co., Ltd. Joon-Ho, Oh 207, cheomdangwagiro, 208 bungil 50, Bukgu, Gwangju, 61012, South Korea Email: hi990@naver.com Phone: +82-62-431-7752 Fax: +82-62-434-7752
#### Official Correspondent
Withus Group Inc April Lee 106 Superior, Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: +1-909-274-9971 Fax: 1-909-460-8122
### Device Information
- K number: K180671 ●
- Trade Name: MINE
- Common Name: Portable X-ray System ●
- Classification Name: Extra-oral source x-ray system
- Product Code: EHD ●
- Panel: Dental ●
- Regulation Number: 872.1800 ●
- Device Class: Class II ●
- Date prepared: 06/04/2019
#### Predicate Device
- K number: K122124 ●
- Trade Name: EXARO
- Common Name: Portable X-ray System ●
- Classification Name: Extra-oral source x-ray system ●
- Product Code: EHD
- Panel: Dental ●
- Regulation Number: 872.1800 ●
- Device Class: Class II ●
#### Device Description
MINE, a portable dental X-ray system, operates on 11.1V DC supplied by a rechargeable LiPolymer battery pack, The X-ray tube head, controls and power source are assembled into a single hand-held enclosure. The package includes a battery charger.
The portable X-ray system. MINE, being composed of X-ray generator, controller, and beam limiting device is designed to diagnose tooth and jaw through generated and controlled X-ray. The operating principle of MINE starts from the generation of X-ray by high voltage electricity, which in turn penetrates tooth and jaw area after flowing through X-ray tube and produces X-ray images on X-ray receptors (i.e. chemical film or digital sensor)
This device contains a high frequency inverter that converts direct to alternating current, X-ray tube head, electrical protective devices, and other elements. The MINE produces sharp and clear images and prevents patients and dentists from radiation exposure with utilizing small dose of radiation.
{4}------------------------------------------------
| Characteristic | MINE | |
|-------------------|----------------------|------------|
| Anode current | 2mA | |
| Expose time | 0.01~1.3sec | |
| Battery | Battery part No. | BT17070001 |
| | Battery Current | 1500mA |
| | Battery Max. Current | 25A |
| Battery case size | (max) 52x62x22 | |
## Indications for use
The MINE is intended to be used by trained dentists and dental technicians as extra-oral x-ray source for producing diagnostic x-ray images using intra-oral image receptors or film. Its use is intended for both adult and pediatric subjects.
### Summary of Technological Characteristics
| | NEW | Predicate Device |
|--------------------------------|-----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Manufacturer | RHT Co., Ltd. | OSSTEM IMPLANT Co,. Ltd. |
| Model | MINE | EXARO |
| 510(k)<br>Number | K180671 | K122124 |
| Energy<br>Source | Rechargeable 11.1V, DC Lithium<br>Polymer Battery pack | Rechargeable 25.2V, DC Lithium<br>Polymer Battery pack |
| Expose Time | 0.01sec ~ 1.30sec | 0.01~2.0 seconds, 0.01 increments |
| Time Accuracy | ±(10%+1ms) | ±(10%+1ms) |
| Heat Capacity | 4300J (6 KHU) | 8.5 KHU |
| Power Output | 100W | 100W |
| mA | 2mA Fixed | 2mA Fixed |
| kVp | 60kV Fixed | 60kV Fixed |
| Focal Spot | 0.4mm | 0.8mm |
| Wave Form | Constant Potential (DC) | Constant Potential (DC) |
| Safety, EMC and<br>Performance | IEC/EN 60601-1, IEC/ EN 60601-1-3<br>and IEC/EN 60601-2-28 | IEC 60601-1, IEC 60601-1-2, IEC<br>60601-2-7, IEC 60601-2-28 |
| Source to skin<br>Distance | 20cm | 20cm |
| Cone Diameter | 6cm | 6cm |
| User Interface | Exposure time: up, down.<br>Selection buttons of parts of teeth,<br>adult and child, film and sensor with | Exposure time: up, down.<br>Selection buttons of parts of teeth,<br>adult and child, film and sensor |
{5}------------------------------------------------
| | display. 12channel save. | with display. |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Exposure switch | Control panel key | Control panel and remote controller |
| Tube head<br>Mounting | Yes | Yes |
| Principle of<br>Operation | X-ray generated by high voltage<br>electricity into X-ray tube, which<br>penetrates hand, tooth and jaw, and<br>makes X-ray images on receptor<br>(Chemical Film or Digital Sensor) | X-ray generated by high voltage<br>electricity into X-ray tube, which<br>penetrates hand, tooth and jaw, and<br>makes X-ray images on receptor<br>(Chemical Film or Digital Sensor) |
| Intended Use | intended to be used by trained dentists and dental technicians as extra-oral x-<br>ray source for producing diagnostic x-ray images using intra-oral image<br>receptors or film. Its use is intended for both adult and pediatric subjects. | |
The subject and predicate devices are similar in indications, design, technology, functions, source to skin distance and principle of operation.
The difference between the two devices is as follows.
- Heat Capacity: The heat capacity of the subject device is 4300J and predicate device's is 6000J. ● The subject device has Less Heat Capacity than the predicate device, but the efficiency is similar.
- Focal Spot: The focal spot of the subject device is 0.4mm and predicate device's is 0.8mm. ● Reducing focal spot size results in a sharper image quality than predicate device.
- . Energy Source: The Energy source of the subject device is 11.1V and predicate device's is 25.2V. Since this product is designed with low voltage, it can be used with low voltage battery.
Any differences do not raise different questions of safety and effectiveness than the predicate. Therefore, there is no difference between the subject and predicate with respect to the indications or technology.
# Non-clinical Testing
The compliance of MINE will satisfy the applicable requirements of the Underwriters Laboratories Standard for Safety-IEC/EN 60601-1. IEC/ EN 60601-1-3. IEC/EN 60601-2-28 and IEC 60601-2-65. All required documents and reports will be submitted to the appropriate oversight agency to establish compliance with the applicable requirements.
EMS test was performed for MINE in accordance with Standard EN/IEC 60601-1-2. All test results were complied with the requirements.
Phantom images were provided to demonstrate the overall performance of the device as part of a complete intraoral x-ray imaging chain.
The subject device, MINE was found to be compliant with all applicable federal performance standards under 21 CFR 1020.
# Conclusion
In reference to the Federal Food, Drug, and Cosmetic Act, 21 CFR Part 807 and the comparison information provided substantial equivalent chart above, the RHT Co.,Ltd. believes that the MINE is substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.