The Uroview FD II is a solid state detector fluoroscopic X-ray system, primarily for urological applications (functional xray-diagnostics, endourology and minimal invasive urology/surgery). The system, which includes a radiologic/urologic treatment table, may be used for urological treatment, planning and diagnostic procedures including but not limited to: - Querying and retrieving patient information and /or images from other modalities. - X-ray examination of the urogenital area (e.g. cyctoscopy, kidney, bladder, ureter, urethra) including KUB, IVP, vasovesiculography, reflux-cystogram, cystourethrogram, and micturation cystourethrogram combined with uroflow measurements. - Ultrasound examinations (in conjunction with a stand-alone ultrasound system) of the kidney, bladder, prostate, scrotum. - Endourological interventions (e.g. of the urethra, prostate, bladder, sphincter, ostium, kidney and ureter, catheter placement, penile implant placement, transurethral resection of prostate or bladder, alternative treatment of the BPH and brachytherapy). - Percutaneous interventions (e.g. PCN nephrolithotomy, resection, percutaneous nephrostomy). - Laparoscopy (e.g. cholecystectomy, lymph node dissection, abdominal testis detection/correction. varicocele). - Application of fistula (kidney/bladder). - Simple procedures (e.g. urethra, testis, phimonis). - Introcorporeal shock wave lithotripsy. - Uroflow/urodynamics. - Pediatric radiological and therapeutic applications.
Device Story
Uroview FD II is a solid-state detector fluoroscopic X-ray system for urological diagnostics and interventions. It utilizes a cantilevered, isocentric tilting table, high-frequency generator, rotating anode X-ray tube, and dynamic flat panel detector. The system captures X-ray images, replacing traditional film/CR and image intensifier systems. Operated by clinicians in a clinical setting, it supports continuous/pulsed fluoroscopy and radiography. The system includes an automatic exposure control (AEC) with a measuring chamber to ensure consistent image quality. Output is displayed on video monitors for real-time procedural guidance, planning, and diagnostic review. It integrates with external modalities via DICOM 3.0. The device benefits patients by providing high-resolution imaging for minimally invasive urological procedures, reducing the need for more invasive surgeries.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical performance testing, including biocompatibility (ISO 10993), electrical safety (IEC 60601-1), electromagnetic compatibility, and software verification/validation.
Technological Characteristics
Solid-state flat panel detector (PaxScan 4343CB); high-frequency microprocessor-controlled generator (80 kW); rotating anode X-ray tube (RAD-60); automatic exposure control (SSMC601); moveable anti-scatter grid. Connectivity via DICOM 3.0. Complies with IEC 60601-1, 60601-1-2, 60601-2-43, 60601-2-54. Biocompatible per ISO 10993-1.
Indications for Use
Indicated for urological diagnostic, planning, and treatment procedures, including endourology, minimally invasive urology/surgery, and pediatric applications. Patient population includes those requiring urogenital X-ray examinations, ultrasound, endourological/percutaneous interventions, and lithotripsy.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
April 12, 2018
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Pausch Medical GmbH % Oliver Eikenberg, Ph.D. Senior Consultant, QA & RA EMERGO Global Consulting LLC 2500 Bee Cave Road, Building 1, Suite 300 AUSTIN TX 78746
Re: K180651
Trade/Device Name: Uroview FD II Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA Dated: February 23, 2018 Received: March 13, 2018
Dear Dr. Eikenberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Robert Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K180651
Device Name Uroview FD II
#### Indications for Use (Describe)
The Uroview FD II is a solid state detector fluoroscopic X-ray system, primarily for urological applications (functional xray-diagnostics, endourology and minimal invasive urology/surgery). The system, which includes a radiologic/urologic treatment table, may be used for urological treatment, planning and diagnostic procedures including but not limited to:
- · Querying and retrieving patient information and /or images from other modalities.
- · X-ray examination of the urogenital area (e.g. cyctoscopy, kidney, bladder, ureter, urethra) including KUB, IVP, vasovesiculography, reflux-cystogram, cystourethrogram, and micturation cystourethrogram combined with uroflow measurements.
- · Ultrasound examinations (in conjunction with a stand-alone ultrasound system) of the kidney, bladder, prostate, scrotum.
- · Endourological interventions (e.g. of the urethra, prostate, bladder, sphincter, ostium, kidney and ureter, catheter placement, penile implant placement, transurethral resection of prostate or bladder, alternative treatment of the BPH and brachytherapy).
- · Percutaneous interventions (e.g. PCN nephrolithotomy, resection, percutaneous nephrostomy).
- · Laparoscopy (e.g. cholecystectomy, lymph node dissection, abdominal testis detection/correction. varicocele).
- · Application of fistula (kidney/bladder).
- · Simple procedures (e.g. urethra, testis, phimonis).
- · Introcorporeal shock wave lithotripsy.
- · Uroflow/urodynamics.
- · Pediatric radiological and therapeutic applications.
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5 – 510(k) Summary
# Uroview FD II
K 180651
#### 1. Submission Sponsor
Pausch Medical GmbH
Graf-Zeppelin-Straße 1
Erlangen
Bavaria, D-91056
Germany
Office Ph: +49 (0)9131 / 99 92 - 0
Contact: Christian Stoian
Title: CEO
## 2. Submission Correspondent
Emergo Global Consulting, LLC 816 Congress Avenue, Suite 1400 Austin, TX 78701 Office Ph: (512) 327.9997 Contact: Dr. Oliver Eikenberg, Senior Consultant, RA/QA Email: project.management@emergogroup.com
## 3. Date Prepared
February, 23 2018
#### 4. Device Identification
| Trade/Proprietary Name: | Fluoroscopic, Image Intensified X-ray System |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | X-ray System |
| Classification Name: | Image-intensified fluoroscopic x-ray system |
| Regulation Number: | 892.1650 |
| Product Code: | JAA |
| Device Class: | Class II |
| Classification Panel: | Radiology |
| Guidance: | FORM FDA 3626 (1/14), A Guide for the Submission of Initial Reports on<br>Diagnostic X-Ray Systems and Their Major Components |
{4}------------------------------------------------
## 5. Legally Marketed Predicate Device
Uroview FD (K161019) from Pausch Medical GmbH, Germany
#### 6. Device Description
The Uroview FD II is a solid state detector fluoroscopic, X-ray system, primarily for urological applications (functional x-ray diagnostics, endourology and minimal invasive urology/surgery). The basic unit is a cantilevered, continuously adjustable, isocentric tilting patient the Uroview FD Skeleton, which can be raised and tilted to provide convenient access for the patient as well as optimum and ergonomic operating conditions for the user. The Uroview FD II X-ray system is a completely mounted system and is equipped with a digital imaging system with a dynamic flat detector, designed to replace traditional spot film devices using screen-film cassettes or "CR" plates and fluoroscopy with image intensifier CCD cameras. The system is also equipped with a generator and automatic, multilayer, square field collimation system intended for installation on stationary X-ray equipment. A rotating anode X-ray tube is mounted. The measuring chamber is placed between the patient and the detector in order to detect the actual dose value for the automatic exposure control (AEC), to provide consistent x-ray film appearance and to guarantee error-free images even at low kV values. Attenuation factor is low and X-ray scattering is reduced to minimum by using a moveable grid.
The Uroview FD II X-ray system includes the following major components:
- . Uroview FD Skeleton (urological table incl. tilting table)
- High Frequency RF X-ray generator
- X-ray tube incl. housing
- Collimator
- . Measuring chamber
- . Grid
- Dynamic flat panel detector
- Digital imaging workstation
- . Video monitors
- Accessories
The Uroview FD II is considered as an update to the model Uroview FD with some new components as part of the continuous X-ray system evolution. Major X-ray components are identical to the Uroview FD and new components use FDA-cleared or FDA-registered X-ray components. Therefore the Uroview FD II is considered similar by properties and technological characteristics.
The Uroview FD II system is designed to meet the requirements in accordance with relevant sections of 21 CFR 1020.30-1020.33.
{5}------------------------------------------------
## 7. Indication for Use Statement
The Uroview FD II is a solid state detector fluoroscopic X-ray system, primarily for urological applications (functional x-ray diagnostics, endourology and minimal invasive urology/surgery). The system, which includes a radiologic/urologic treatment table, may be used for urological treatment, planning and diagnostic procedures, including but not limited to:
- Querying and retrieving patient information and/or images from other modalities .
- . X-ray examinations of the urogenital area (e.g. cystoscopy, kidney, bladder, ureter, urethra) including KUB, IVP, vasovesiculography, reflux-cystogram, cystourethrogram, and micturation cystourethrogram combined with uroflow measurements.
- Ultrasound examinations (in conjunction with a stand-alone ultrasound system) of the kidney, bladder, prostate, scrotum.
- . Endourological interventions (e.g. of the urethra, prostate, bladder, sphincter, ostium, kidney and ureter, catheter placement, penile implant placement, transurethral resection of prostate or bladder, alternative treatment of the BPH and brachytherapy).
- Percutaneous interventions (e.g. PCN nephrolithotomy, resection, percutaneous nephrostomy) ●
- . Laparoscopy (e.g. cholecystectomy, nephrectomy, lymph node dissection, abdominal testis detection/correction, varicocele)
- . Application of fistula (kidney/bladder)
- Simple procedures (e.g. urethra, testis, phimosis)
- Intracorporeal shock wave lithotripsy ●
- . Uroflow/urodynamics
- Pediatric radiological and therapeutic applications
## 8. Substantial Equivalence Discussion
The update of the model Uroview FD to the new X-ray system Uroview FD II is identical by technological characteristics, design and mode of operation for the components "Tube, Measuring Chamber, Anti Scatter Grid".
The Uroview FD II further shares optimized component features or equivalent X-ray components to that of the predicate Uroview FD for Urological Table Uroview FD Skeleton, Generator, Digital Imaging System and Monitor beam- limiting device flat detector and digital imaging processing. The changes refer to design/size variations, image resolution, workstation functionalities and use of FDA-registered or FDA-cleared commercially available X-ray components, which are specifically designed for use in X-ray systems and therefore include standardized values and settings for electrical, mechanical and radiation requirements. As such the identified differences are considered to be optimized system variations, which are adequately controlled by testing of this X-ray system. The modifications of the Uroview FD II compared to the previous model Uroview FD are within the controls and predetermined specifications and are supported with verification and validation testing. The following table compares the Uroview FD II to the predicate device with respect to indications for use, principles of operation, technological characteristics, and performance testing.
The comparison chart reveals that the technological characteristics and as such the functions performed by the Uroview FD II offer substantially the same data and technology as the Uroview FD. The subject device is as safe and effective as its predicate device without raising any new safety and/or effectiveness concerns.
{6}------------------------------------------------
| | Proposed Device<br>Uroview FD II | Predicate Device<br>Uroview FD | Similarities /<br>Differences |
|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Pausch Medical GmbH | Pausch Medical GmbH | N/A |
| Product/Trade Name: | Uroview FD II | Uroview FD | N/A |
| 510(k): | Pending | K161019 | N/A |
| Establishment<br>Registration No. | 9610903 | 9610903 | N/A |
| Product Code: | JAA | JAA | Same |
| Regulation Number | 21 CFR 892.1650 | 21 CFR 892.1650 | Same |
| Device Classification | Solid State x-ray imager<br>(flat panel/digital imager) | Solid State x-ray imager<br>(flat panel/digital imager) | Same<br>classification |
| Class | II | II | Same |
| Indications for use | The Uroview FD II is a solid state detector fluoroscopic X-ray system, primarily for<br>urological applications (functional x-ray-diagnostics, endourology and minimal<br>invasive urology/surgery). The system, which includes a radiologic/urologic treatment<br>table, may be used for urological, treatment, planning and diagnostic procedures<br>including but not limited to:<br>• Querying and retrieving patient information and/or images from other modalities.<br>• X-ray examinations of the urogenital area (e.g. cyctoscopy, kidney, bladder, ureter,<br>urethra) including KUB, IVP, vasovesiculography, reflux-cystogram,<br>cystourethrogram, and micturation cystourethrogram combined with uroflow<br>measurements.<br>• Ultrasound examinations (in conjunction with a stand-alone ultrasound system) of<br>the kidney, bladder, prostate, scrotum.<br>• Endourological interventions (e.g. of the urethra, prostate, bladder, sphincter,<br>ostium, kidney and ureter, catheter placement, penile implant placement,<br>transurethral resection of prostate or bladder, alternative treatment of the BPH<br>and brachytherapy)<br>• Percutaneous interventions (e.g. PCN nephrolithotomy, resection, percutaneous<br>nephrostomy).<br>• Laparoscopy (e.g. cholecystectomy, nephrectomy, lymph node dissection,<br>abdominal testis detection/correction, varicocele).<br>• Application of fistula (kidney/bladder).<br>• Simple procedures (e.g. urethra, testis, phimosis).<br>• Intracorporeal shock wave lithotripsy.<br>• Uroflow/urodynamics.<br>• Pediatric radiological and therapeutic applications | | Same<br>The intended<br>use definition<br>is identical<br>except for the<br>device name<br>Uroview FD<br>versus<br>Uroview FD II |
| Complies with<br>Applicable Voluntary<br>EC, EMC-Standards | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-3<br>IEC 60601-1-6<br>IEC 60601-2-28<br>IEC 60601-2-43<br>IEC 60601-2-54 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-3<br>IEC 60601-1-6<br>IEC 60601-2-28<br>IEC 60601-2-43<br>IEC 60601-2-54 | Same |
| Complies with ISO 10993-1<br>for Biocompatibility | yes | yes | Same |
| Sterile: | N/A | N/A | N/A |
| Shelf Life (usable lifetime) | Ten years | Ten years | Same |
| | Proposed Device | Predicate Device | Similarities / |
| | Uroview FD II | Uroview FD | Differences |
| Overview of Component Specifications | | | |
| Major Components | 1. Uroview FD Skeleton | 1. Uroview FD Skeleton | Same |
| Basic Unit | Basic Unit (urological table) | (basic unit, urological table) | |
| | 2. X-ray Generator | 2. X-ray Generator | |
| | 3. X-ray tube and housing | 3. X-ray tube and housing | |
| | 4. Collimator | 4. Collimator | |
| | 5. Measuring chamber | 5. Measuring chamber | |
| | 6. Digital image workstation | 6. Digital image workstation | |
| | 7. Detector | 7. Detector | |
| Basic Unit | Uroview FD Skeleton | Uroview FD Skeleton | Similar |
| (urological table) | | | Newly designed<br>with similar<br>functionality |
| Basic Unit System | Optional right-handed or left-<br>handed version | Optional right-handed or left-<br>handed version | Same |
| Table height | Continuously adjustable by<br>motor drive: 64 to 116.5 cm<br>(25.2 to 45.9"), floating | Continuously adjustable by motor<br>drive: 64 to 116.5 cm<br>(25.2 to 45.9"), floating | Same |
| Tabletop extensions | 78 cm (30.7") x 75.9 cm (29.9") | 78 cm (30.7") x 75.9 cm (29.9") | Same |
| Tabletop-detector distance | 6.4 cm (2.5") | 6.4 cm (2.5") | Same |
| Source-detector distance | 115 cm (45.3") | 115 cm (45.3") | Same |
| Manual / Motorized<br>movement park position to<br>exposure position | 27 cm (10.6")<br>park position 27 cm (10.6") | 27 cm (10.6") | Similar,<br>additional park<br>position function |
| Tilt range | Motorized tilt: ± 88 °<br>Isocentric tilt: ± 20 °…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.